241 18 8MB
English Pages 12 Year 2023
COVER Page 1
EXAM Q1 Generator Elution Generator Elution: We Need: saline vial, empty 11ml vial for Tc 99m & lead shield, forceps, alcohol wipe. 1- Wipe the saline vial , then insert on its [lace on the generator 2- Wipe 11ml vial , insert it inside lead shield, then insert it on the generator. 3- After 2-3 min or after the bubbles on the saline vial stops, remove Tc99m vial and recap the generator needle. 4- Insert Tc99m vial in the dose calibrator, measure activity, print label return vial in the lead shield & stick the label on it. Elution Concentration: E.X: 38.8 mCi in 10 ml @ 9:00 am
=38.8/10= 3.8 mCi in 1 ml @ 9:00 am
Elution Quality Control: QC 1: Radio Nuclide Purity (99Mo Break Through (ASSAY)): 1234-
Insert Empty lead on Dose calibrator & measure. Insert Tc99m vial in lead, then Insert on Dose calibrator& measure. Insert Empty plastic dipper on Dose calibrator & measure. Insert Tc99m vial in plastic dipper, then Insert on Dose calibrator& measure.
• For QC2 & QC3: Draw 0.03ml of Elution in Insulin Syringe.
QC 2: Chemical Purity (ALUMINA Break Through): 1- From ALUMINA Break Through QC kit, Take one strip 2- Add a drop of standard AL3 on the strip 3- Add a drop of Tc99m elution on the strip AL3 should be < 10 Tc99m
QC 3: Radio Chemical Purity : Impurities Are: ( Free Tc99m, Hydrolyzed Tc02, Tc labelled to other compounds) 1- Add 0.1ml of Acetone on a vial 2- Add a drop of Tc99m elution on a RED strip bottom line 3- Put the strip on the Acetone Vial 4- Wait until it move to the top line 5- Cut the strip on the middle line & put the two strips on two tubes & measure them. Background
0.8
Top free TCO4-
8.8
8.8 - 0.8 = 8
Bottom TCO2-
1.2
1.2-0.8=0.4
=8/8.4*100=95.2 %
Generator Page 2
EXAM Q2 KIT Prepration Prepare HIDA for two patients, 1st pt will be injected @ 10:30 am, 2nd pt will be injected @ 12:00 am,Now the time is 9:00am? ANSWER: • Take vial from the fridge, check Expiry Date • Put it in shield and swap it.
Calculate PT Dose: Adult Dose for HIDA is (5 mCi)
A = Ao e-λt/T half
• For 1st pt we should draw now
• For 2nd pt we should draw now
5.9 + 7 = 12.9 + 10% = 13 mCi , Min activity in HIDA kit is 10 mCi & Max activity is 40 mCi, so we add about 20 mci to HIDA vial.
Calculate Elution concentration NOW: Elution Concentration is 3.8 mCi in 1 ml @ 9:00 am , NOW TIME is 9:30am Volume of activity that we should add to HIDA vial @ 9:30am =
KITS Page 3
• Draw 5.57 ml from elution & add to HIDA vial, draw same amount of air, discard the syrnge. • Print label • Shake and incubate for about 30 min
Qc for HIDA: 123456-
Draw 0.03ml from Tc 99m HIDA vial in Insulin Syringe. Add 0.1ml of H2o on a vial Add a drop on a blue strip bottom line Put the strip on the H2o Vial Wait until it move to the top line Cut the strip on the middle line & put the two strips on two tubes & measure them.
7891011-
Add 0.1ml of 20%Nacl on a vial Add a drop on a orange strip bottom line Put the strip on the 20%Nacl Vial Wait until it move to the top line Cut the strip on the middle line & put the two strips on two tubes & measure them.
Draw Dose for the two patients : Tc 99m HIDA activity 18.3 mCi in 5.6 ml @ 9:30 am , NOW TIME is 9:42am Tc 99m HIDA concentration = 18.3/5.6=3.2679 3.26 mCi/ml • For1 st pt we should draw now
•
• For 2nd pt we should draw now
• Draw the dose for the two pt, & put it in a shield
KITS Page 4
Dose Calibrator QC
A-ACCURACY: • • • •
To show that the Dose Calibrator give correct reading through entire scale. Done annually Should be within -+ 5% of the standard Different long half-life sources with different energies (Low Co57-MED Cs137, Ba133-High Co60) can be measured with the dose calibrator and their value is compared to the expected calculated value.
B-CONSTANCY (PRESICION): • • • •
To show that the Dose Calibrator give reproducible reading on all various isotope setting. Done daily Should be within -+ 5% Cs137 with T half of 30 years, is measured daily
C-LINERATY: • To show that the Dose Calibrator reading is linear from • Tc99m Is measured until it reach 30 , about 96 hr
D-GEOMETRY: • To show that the Dose Calibrator give correct reading regardless of size or geometry. • Done after repair or moving the DC • Done in two ways: 1. 25 Tc99m Is measured, then dilated in (1,2,3,4,5,6,7,8,910 ml Saline) & measured 2. 20 mCi Tc99m Is measured at different distance in DC ( bottom, 2, 4, 6 cm)
Dose Calibrator Page 5
Kits Table #
KIT
MANUFACTU RER
ACTIVITY ADDED TO VIEL mCi
VOLUME ml
STD
MIN
MAX
STD volume
MIN
MAX
1
DMSA
curium
100
1
6
2
MAG-3
curium
80
3
10
3
MDP/HDP
curium
500
3 to10
4
DTPA
curium
5
HIDA
Polatic
6
MIBI
curium
7
Tin Colloid
Polatic
2.7
27
Nano Colloid
Rotop
5
150
1
5
MAA
curium
2.5
100
2.5
10
8
9
300 10
40
2
300
50
Kits Table Page 6
10
5
INCUBATIO N
SHIELF LIFE
MINUTE
HOURS
15-10
ADULT PATIENT DOSE
PEDIATRIC DOSE
mCi
MIN mCi
0.5
NOTES
max localization 3-6 hr, shake 5 min
8hr
>95%
0.8-3.2 mci
8hr
>94%
1-5 mci, baby min .4 0.4 mci
Boil 10-Cool 10 min upright
8--20, baby min 1 mci
Delay at least 2 hr
1
15-30
8hr
renal 1-10 mci
30
5hr
3-5 mCi, baby min .5 0.5 mci
shake 2 min
PARA 13.5-18.9
Boil 10-Cool 10 min upright
1 to 3
10hr
5
QC
20
15
>94%
4hr
2.1
4-5.4mCi
ge 1 mci soilid in eggs
0.2-0.3 ml per injection
6 hr
>95%
Lymph-Sentinal 0.5-2.97 per injection site
8hr
>90%
1-5 mci
0.27
DO NOT VENT, 18 g needle,
Back cover Page 7