253 36 2MB
English Pages 375 Year 2002
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
Jose A. C. Broekaert Analytical Atomic Spectrometry with Flames and Plasmas
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
Analytical Atomic Spectrometry with Flames and Plasmas Valeur, B.
Molecular Fluorescence. Principles and Applications 2001. ISBN 3-527-29919-X
G unzler, H. and Williams, A.
Handbook of Analytical Techniques 2001. ISBN 3-527-30165-8
H ubschmann, H.-J.
Handbook of GC/MS 2001. ISBN 3-527-30170-4
Welz, B. and Sperling, M.
Atomic Absorption Spectrometry Third, Completely Revised Edition 1998. ISBN 3-527-28571-7
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
Jose A. C. Broekaert
Analytical Atomic Spectrometry with Flames and Plasmas
Weinheim ± New York ± Chichester ± Brisbane ± Singapore ± Toronto
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
Prof. Dr. Jose A. C. Broekaert UniversitaÈt Leipzig Institut fuÈr Analytische Chemie LinneÂstraûe 3 04103 Leipzig Germany
Typesetting Asco Typesetters, Hong Kong Printing betz-druck gmbH, D-64291 Darmstadt Bookbinding Wilhelm Osswald & Co., 67433 Neustadt ISBN
9 This book was carefully produced. Nevertheless, author and publisher do not warrant the information contained therein to be free of errors. Readers are advised to keep in mind that statements, data, illustrations, procedural details or other items may inadvertently be inaccurate. Library of Congress Card No.: applied for A catalogue record for this book is available from the British Library. Die Deutsche Bibliothek ± CIP Cataloguing-in-Publication-Data A catalogue record for this publication is available from Die Deutsche Bibliothek ( WILEY-VCH Verlag GmbH, D-69469 Weinheim (Federal Republic of Germany). 2002 All rights reserved (including those of translation in other languages). No part of this book may be reproduced in any form ± by photoprinting, micro®lm, or any other means ± nor transmitted or translated into machine language without written permission from the publishers. Registered names, trademarks, etc. used in this book, even when not speci®cally marked as such, are not to be considered unprotected by law. Printed in the Federal Republic of Germany. Printed on acid-free paper.
3-527-30146-1
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
To my wife Paula and our daughters Ilse, Sigrid and Carmen
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
vii
Contents Preface
xi
Introduction
1
1
Basic Principles
1.1 1.2 1.3 1.4 1.5 1.6 1.7
Atomic structure 4 Plasmas 8 Emission and absorption of radiation Ionization 18 Dissociation 23 Sources for atomic spectrometry 27 Analytical atomic spectrometry 31
2
Spectrometric Instrumentation
2.1 2.2 2.2.1 2.2.2 2.2.3 2.3 2.3.1 2.3.2 2.3.3 2.3.4 2.4
Figures of merit of an analytical method Optical spectrometers 51 Optical systems 52 Radiation detectors 61 Non-dispersive spectrometers 70 Mass spectrometers 72 Types of mass spectrometers 73 Ion detection 80 Ion extraction 82 Ion optics and transmission 84 Data acquisition and treatment 84
3
Sample Introduction Devices
3.1 3.2 3.3 3.4 3.4.1 3.4.2
Sample introduction by pneumatic nebulization 90 Ultrasonic nebulization 103 Hydride and other volatile species generation 105 Electrothermal vaporization 109 The volatilization process 109 Types of electrothermal devices 111
4
9
34 34
88
viii
Contents
3.4.3 3.4.4 3.5 3.5.1 3.5.2 3.5.3 3.5.4 3.6
Temperature programming 114 Analytical performance 115 Direct solids sampling 117 Thermal methods 117 Slurry atomization 120 Arc and spark ablation 124 Laser ablation 131 Cathodic sputtering 135
4
Atomic Absorption Spectrometry Principles 148
4.1 4.2 4.2.1 4.2.2 4.3 4.3.1 4.3.2 4.3.3 4.4 4.4.1 4.4.2 4.4.3 4.5 4.5.1 4.5.2 4.5.3 4.5.4 4.5.5 4.6 4.6.1 4.6.2 4.6.3 4.6.4 4.7 4.8 5
5.1 5.2 5.3 5.4 5.4.1 5.4.1.1
148
Atomic absorption spectrometers 150 Spectrometers 150 Primary radiation sources 152 Flame atomic absorption 158 Flames and burners 159 Nebulizers 161 Figures of merit 162 Electrothermal atomic absorption 164 Atomizers 165 Thermochemistry 168 Figures of merit 169 Special techniques 172 Hydride and cold-vapor techniques 172 Direct solids sampling 174 Indirect determinations 175 Flow injection analysis 175 Diode laser atomic absorption spectrometry 176 Background correction techniques 177 Correction for background absorption with the deuterium lamp technique 177 Background correction with the aid of the Zeeman eect 179 Smith±Hieftje technique 182 Coherent forward scattering 183 Fields of application 184 Outlook 191 Atomic Emission Spectrometry Principles 192
192
Atomic emission spectrometers 202 Flame emission 210 Arcs and sparks 210 Arc emission spectrometry 210 Arc characteristics 210
Contents
5.4.1.2 5.4.1.3 5.4.2 5.4.2.1 5.4.2.2 5.5 5.5.1 5.5.1.1 5.5.1.2 5.5.2 5.5.2.1 5.5.2.2 5.5.2.3 5.5.2.4 5.5.3 5.5.4 5.6 5.6.1 5.6.2 5.6.3 5.6.4 5.6.5 5.7
Dc arc spectrometry 211 Ac arc spectrometry 213 Spark emission spectrometry 213 Sparks 213 Analytical features 215 Plasma source AES 216 Dc plasma-jet AES 217 Types of plasma jets 217 Three-electrode plasma jet 218 Inductively coupled plasma AES 219 The inductively coupled plasma 219 Instrumentation 221 Analytical performance 223 Applications 232 Low-power high-frequency plasmas 233 Microwave plasmas 234 Glow discharge AES 241 Hollow cathodes for AES 242 Furnace emission spectrometry 243 Dc glow discharges with a ¯at cathode 244 Rf glow discharges 248 New developments 249 Laser sources 251
6
Plasma Mass Spectrometry
6.1 6.1.1 6.1.2 6.1.3 6.1.4 6.2 6.2.1 6.2.2 6.2.3
ICP mass spectrometry 255 Instrumentation 255 Analytical features 257 Applications 268 Outlook 272 Glow discharge mass spectrometry Instrumentation 277 Analytical performance 281 Analytical applications 282
7
Atomic Fluorescence Spectrometry Principles 290 Instrumentation 293 Analytical performance 295
7.1 7.2 7.3 8
8.1 8.2 8.3
254
275
290
Laser Enhanced Ionization Spectrometry Principles 297 Figues of merit 300 Analytical applications 301
297
ix
x
Contents
9
Sample Preparation for Atomic Spectrometry
9.1 9.2 9.3 9.3.1 9.3.2 9.3.3 9.3.4 9.4 9.5
Sample preparation in direct compact sample analysis Grinding, sieving and compaction of powders 302 Sample dissolution 304 Wet chemical methods 304 Fusion procedures 304 Microwave-assisted methods 305 Combustion techniques 305 Flow injection analysis 305 Leaching sample preparation methods 306
10
Comparison with Other Methods Power of detection 307 Analytical accuracy 309 Economic aspects 310
10.1 10.2 10.3
Literature Index
348
312
307
302 302
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
xi
Preface Spectrochemical analysis is a powerful instrumental principle for the determination of the chemical elements and their species in a variety of sample types of different size, at widely dierent concentration levels and with very diering cost performance ratios and time consumption. In addition, not only monoelement but also multielement determinations are possible with widely diering precision and accuracy using the various dierent methods. The basic principles of spectrochemical analysis are related to the atomic and molecular structure and also to gas discharge physics as well as to instrumentation and measurement sciences. Therefore, research into spectrochemical analysis requires knowledge of the aforementioned disciplines to enable innovative developments of new methodologies to be achieved in terms of the improvement of power of detection, accuracy and cost performance ratios, these being the driving forces in analytical innovation. The development of analytical procedures also requires the analytical chemist to have a knowledge of the theory and the principles of the above mentioned disciplines. It is the aim of this monograph to bring together the theory and principles of todays spectrochemical methods that make use of ¯ames and plasma sources. This should enable researchers to enter the ®eld of spectrochemical research, where innovation is through the use and development of new sources and the application of new types of spectrometers, and also to face challenges from emerging ®elds of application, which is as straightforward today as it was even in the time of Bunsen and Kirchho. This work should appeal both to chemists and physicists, the cooperation of whom is instrumental for progress to be made in this ®eld of analytical chemistry as well as to users from dierent areas of science, including the life sciences, material sciences, environmental sciences, geochemistry, chemical process technology, etc. This present work could also be viewed as a resume of the theoretical background, which manufacturers of instrumentation for atomic absorption spectrometry, arc, spark and glow discharge emission spectrometry as well as ICP emission spectrometry and plasma mass spectrometry with ICPs or glow discharges and laser based techniques can recommend to their interested users to make the most ecient use of these analytical methods in their respective ®elds of application. Also research associates entering the ®eld of atomic spectroscopy with ¯ames and plasmas should ®nd the necessary basics and references to further literature in this book.
xii
Preface
The work also describes a number of achievements of over thirty years of research performed at the University of Gent (Belgium), the Institute for Spectrochemistry and Applied Spectroscopy (ISAS), Dortmund, the University of Dortmund and the University of Leipzig, which have been made possible through many interactions and collaborations with experts in the ®eld, whom I thank thoroughly. A great deal of knowledge gained from my teachers and in interaction with prominent senior researchers in the ®eld worldwide and especially at the Council for Scienti®c and Industrial Research and the University of Stellenbosch (South-Africa), the Universitaire Instelling Antwerpen (UIA) (Belgium) and Indiana University, Bloomington (IN, USA) as well as results obtained while collaborating with colleagues and with students made this book possible, for which all of them are gratefully acknowledged and thanked. Leipzig, June 2001
Jose A. C. Broekaert
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
348
Index
Index
a a two-level system 290 AAS 307, 310 diode laser 148 AAS with diode laser 156 Abel inversion 29 ablation 124 arc 230 arc and spark 124 spark 230 laser 230 ablation device spark 127 ablation rate 125, 140, 141, 144, 251 abnormal characteristic 144 abnormal glow discharge 138 absolute intensity 197 absolute measurement 129 absorbance 164 absorbance signal 163 absorption 9, 11, 131 absorption coecient 13, 181 absorption path 159 absorption pro®le 149, 176 absorption signal 167 absorption transition probability 17 absorption volume 150, 166 abundance natural 266 accelerating voltage 76 accuracy 201, 282 acid concentration matching 185 acid digestion 289 adiabatic expansion 255 adsorption 186 adsorption±desorption process 167 aerosol 90 aerosol droplet 104 aerosol production eciency 104 AES 193, 202
dc plasma-jet 217 glow discharge 241 hollow cathode 242 inductively coupled plasma 219 AFS 310 laser-excited 295 afterglow 216 air operated CMP 235 airborne dust 190, 243 Al-based alloy 284 Al2 O3 132 Al2 O3 ®lled column 230 Al2 O3 powder 286 Al2 O3 122, see aluminium oxide aluminum 187, 246 ambipolar diusion 212, 224 ampli®cation 197 analysis isotopic dilution 266 analysis of microsample 161 analysis time 126 analyte atom 149 analyte introduction eciency 109 analyte line physical width 148 analyte volatilization 113, 169 analytical accuracy 309 analytical atomic spectrometry 31 analytical evaluation function 35, 197 analytical ®gures of merit 279 analytical line 149 analytical precision 125, 224 analytical sensitivity 169, 182 anion exchange 271 anode current 197 anode region 137 anomalous Zeeman eect 154 Antartic ice 295 APDTC complex 227 arc 2, 30, 31, 210
Index characteristic 210 disk stabilized 217 emission spectrometry 210 arc ablation 124 arc channel 125 arc temperature 211 arrangement Mattauch-Herzog 257 ashing step 172 associative ionization 242 Aston dark space 137 atmospheric particulate matter 286 atmospheric pressure discharge 126 atom line intensity 19 atom loss 167 atom number density 241 atom reservoir 141, 291, 294, 299 atom supply 167 atomic absorption 1 atomic absorption spectrometry 32, 109, 112, 148 electrothermal atomization 87 ¯ame 2 graphite furnace 2 atomic emission 14, 32 glow discharge 309 atomic emission spectrometry 192 arc 309 spark 309 atomic ¯uorescence 14, 33 atomic ¯uorescence spectrometry 290 atomic species 4 atomic spectra 202 atomic spectral line width 148 atomic spectrometry diode laser 281 atomic structure 4 atomic term 6 atomic vapor 150 atomization step 169 atomization temperature 166 atomizer 150, 167 graphite furnace 165 cup 165 ®lament 165 autocollimation 207 axial viewing 227
b background 260 background absorption 149 background correction procedure 177
background correction 87, 151, 172, 204, 205, 222, 224 background intensity 258 background spectral interference 284 Balmer series 4 band emission 202, 215 barrier-layer eect 82 bastnaesite 195 beam-stop 74 Beer±Lambert law 14 Bessel box 256 binder 285, 303 binning 69 bio-medical risk assesment 3 biological material 112, 305 biological sample 123, 168, 266, 270 biological substance 268 black body radiation 17 blackening 62, 63 blank 47, 106, 200, 201 sample 200 ``blaze'' angle b 57 blood 270 blooming 70 boiling point 116, 229 Boissel switch 133 Boltzmann plot 26 Boltzmann's law 9 bond energy 139 box-car integrator 295, 299 brass 134, 252 Bremsstrahlung 18 broadband absorption 151 broadening Doppler 15, 16, 157, 170 Lorentzian 15, 16 natural 15, 16 pressure 15 resonance 15 Stark 15, 21 bulk analysis 246 bulk concentration major 282 minor 282 trace 282 burn-in time 141, 245 burner 159 air±acetylene 161 nitrous oxide 161 multislit 159 burning crater 124, 141, 245 magnetic ®eld 147 burning crater pro®le 143 burning gas 159
349
350
Index burning gas mixture 159 burning spot 127, 143 burning voltage 124, 125, 128, 136, 137, 144, 241
chemical 188 chemical hydride generation 107 chemical interference 163 chemical vapor deposition 115 w 2 test 48, 49 c chip C, H, N, O analyzer 216 microstructured system 222 calibration chromatography function 34 gas 230 standard addition 38 liquid 99, 230 calibration constant 198 chromium 271 calibration curve 149 Cisplatin 288 calibration function 35, 84 clinical analysis 232 calibration function relate 197 clinical chemistry 187 capacity 214 cluster ion 258, 276 carbide 110, 118 CN band 159, 212 carbon arc 211 coated glass 248 carbon deposit 227 Co-based alloy 284 carbon-rod atomizer 111 coecient of variation 36 carbothermal reduction 166 coherent forward scattering 183 cascade impactor 101 coincidence 224 catalyst 189 cold vapor 108 cathode cold-vapor technique 172, 173 ¯at 281 collector electrode 299 pin form 281 collision cathode current 197 the ®rst kind 138 cathode dark space 137 the second kind 138 cathode fall region 138 collisional decay 298 cathode fall 211 collisional±radiative model 242 cathode layer 137, 211 collisions of the ®rst kind 9 cathode plume 276 collisions of the second kind 9 cathodic sputtering 152, 175, 244 column chromatography 190 cavity combined analytical procedure 166 cathode 242 combustion 305 CCD 194, 204, 206, 209, 223, 251 compact ceramic 134 CCD detection 229 compact sample 123, 128 CCD detector 154, 200 compaction 303 CCD spectrometer 199, 213, 225 compromise condition 223 CCD 60, see charge coupled device concentric glass nebulizer 92 cement 188 concomitant 170, 173 ceramic 269 conductive sample 124 ceramic powder 121, 258 con®dence level 36 ceramic sample 286 contamination 186, 304 characteristic 62, 136, 137, 243, 245, 248, 251 continuous mode 133 abnormal 136 continuous sample aspiration 161 current±voltage 211 continuous sample nebulization 162 normal 136 continuous source 153, 183 characteristic concentration 162 continuous source AAS 153, 154 characteristic mass 171 continuum radiation 18, 172 charge coupled device 59, see CCD continuum source 151 charge injection device (CID) 68 cool plasma condition 262 charge transfer 136, 220 cooled hollow cathode 243 charge transfer device 67, see CTD copper arc 213 charge-coupled device (CCD) 68, 208 corona discharge 136
Index counter electrode 124, 127 coupling 7 crater diameter 251 crater pro®le 145 rf discharge 146 crater 133, 246 depth 134 diameter 133 critical concentration ratio 224 cross-contamination 130 cross-¯ow nebulizer 227 cross section 8, 297 crossed polarizers 183 crossed-dispersion 206 crossed-dispersion mode 59 cryocooling 285 cryopump 84 CsCl 213 CuaZn alloy 268 current-carrying plasma 235 current modulated 156 current±voltage characteristic 137, 141, 142 cuvette 165, 173 Czerny±Turner monochromator 199
d D2 -lamp technique 178 Daly±Multiplier 277 dark current 45, 65 data acquisition and treatment 84 dc arc 10 stabilized 212 dc arc spectrography 213 dc discharge 124 dc glow discharge with a ¯at cathode 244 dead time 81 Debye length 83 decay function 109 degassing 285 degeneracy 196 degree of dissociation 160 degree of ionization 20, 257 densitometer 62, 63, 254 departure from LTE 226 depression 164 depth-pro®le 248, 287 depth-pro®le analysis 246 depth-pro®ling analysis of steel 287 depth resolution 287 desolvate 104 desolvation 267 membrane 103 detecter noise 148
detection phase-sensitive 294 detection limit 182, 184, 201, 218, 223, 235, 237, 238, 243, 263, 277, 295, 296 ¯ame AAS 163, 169 furnace AAS 169 detection of the halogen 243 detector spectral response 13 determination sequential 202 simultaneous 202 deuterium lamp technique 177 diatomic molecule 26 diraction angle 207 diraction order 209 diusion 167, 242 diusion coecient 167 digestion under resistance heating 186 diode 66 diode AAS 149 diode array 70 diode-array 66 diode laser 154 diode laser atomic absorption spectrometry 176, see also diode laser AAS dipole 186 direct compact sample analysis 302 direct insertion probe 279 direct sample insertion 89, 228, 229 direct solids nebulizer 126 direct solids sampling 114, 117, 170, 174, 230, 268 discharge 2, 141 dielectric barrier (db) 281 electrodeless 235 hollow cathode 279, 295 restricted 136 dc 135 rf 135 single-electrode 235 spark 127 diuse spark 127 discharge gap 213 discharge lamp a ¯oating anode tube 141 ¯at cathode 141 discharge parameter 141 discharge under reduced pressure 11, 31, 135, 152, 297 discharges under reduced pressure 177, 294 discrete sampling 99, 161, 222 dispenser 165 dispersive element 52
351
352
Index displacement energy 139, 140 dissociation 23, 168 collision-induced 266 reaction-induced 266 dissociation energy 25, 160 dissociation equilibrium 168 dissociation of oxide 160 dissociation reaction 160 dissolution eciency 185 dissolution speed 185 dissolved metal 216 divergence 132, 133 Doppler pro®le 158 Doppler spectral width 15 Doppler-free spectroscopy 158, 301 double arc 109 double peak 116, 170, 171 doubly charged ion 276 drinking water 190 droplet 90, 126, 161 droplet diameter 91 droplet distribution 97 droplet size 100, 102 droplet size distribution 100 Druyvenstein distribution 9 dry solution residue 254, 287 drying stage 114, 173 dual-beam 150 dual-channel instrument 151 dynode 64
e Eagle mounting 61 easily ionized element 211, 224 Echelle grating 207 Echelle monochromator 218 dual 153 Echelle spectrograph 207 Echelle spectrometer 59, 153 8-hydroxyquinoline 270 Einstein coecient for spontaneous emission 11 Einstein transition probability 10 einzel lens 256 ``einzel'' lens 84 elastic collision 136 elctrically non-conducting material 131 electric ®eld 135 electrical discharge 254 electrically conducting solid 129 electrically-conductive sample 281 electrically-conducting sample 244, see also Grimm lamp electrically non-conducting powder 248, 285
electrochemical hydride generation 106, 238 electrodeless discharge lamp 152, 290, 293 electroerosion 89 electrolysis 106 electrolytic hydride generation 106, 107 electromagnet 181 electron charge 197 electron±ion recombination 242 electron impact 138, 220 electron microprobe 121 electron microprobe line scan 143 electron number 18 electron number density 220, 221 electron pressure 20, 211 electron-probe microanalysis 170 electron probe micrograph 243 electronic energy level 23 electrostatic analyzer 73, 277 electrothermal AAS 164 electrothermal atomic absorption 164 electrothermal evaporation 89, 117, 228, 233 electrothermal vaporization 267, 289 element analyte 226 reference 226 elemental mass spectrometry see ICP-MS inductively coupled plasma 3 spark source 3 elemental species 116 element-speci®c detection 157, 237, 251 emission 9 emission spectra 4 emission spectrometry furance 243 spark 213 emulsion 64 end-on 221 energy ion 262 energy distribution 277 energy focusing 75 energy level 13 enriched uranium 158 entrance collimator 52, 56 entrance slit 53 environmental 3 environmental monitoring 220 environmental sample 186 environmental work 270 environmentally-relevant sample 232 error of the ®rst kind 47 error of the second kind 47 ETV 123
Index evaporation 122 evaporation stage 114 excess noise 40 excitation 138, 192, 245 excitation condition 225 excited state 156 excreted oxide 216 exit collimator 56 extraction of ion 255
f 1=f noise 156 FANES (furance atomic non-resonance emission spectrometry) 243 FAPES 272, see also furnace atomic plasma emission source Faraday-cup 277 Faraday dark space 137 Faraday eect 183 FIA manifold 162 ®eld emission 136 ®eld ionization 298 ®gure of merit 34, 162, 300, 308 detection limit 199 ®lament 112 ®lament plasma 237 ®lament-type MIP 238 ®ller gas 276 ®ltering 40 ®nd use 272 ¯ame 1, 30, 159, 290, 299 air±acetylene 159 C2 H2 aN2 O 294 carbon rich 164 chemical 294 emission 210 high-temperature 160 hydrogen±air 159 nitrous oxide±acetylene 159, 210 propane±air 159 ¯ame AAS 150, 158 ¯ame atomic absorption 158 ¯ame temperature 160 ¯ashlamp 132 ¯at cathode 251 ¯at Grimm-type glow discharge 147 ¯icker noise 40 ¯oating restrictor 278 ¯ow cell 108, 229 ¯ow-cell 173 ¯ow-cell hydride generation 233, 239 ¯ow-cell type hydride generator 105 ¯ow injection 161, 176 ¯ow injection analysis 305
¯ow injection analysis (FIA) 99 ¯ow injection analysis 175, see also FIA ¯ow injection procedure 162 ¯ow-through cell 105 ¯uorescence non-resonance 290, 295 resonance 290 ¯uorescence volume 294 ¯uorination 305 ¯ux 131, 196 acid 304 alkali 304 food 123 formation function 109 Fourier transform 79 Fourier transform spectrometry 245 fractionated volatilization 135 free atom concentration 164, 165 free chemical energy 120 free sample aspiration 224 free-running 133 freon 118 frequency 79, 125 frequency doubling 133, 156 fresh water sample 228 fs-laser 252 full-widths at half maximum 16 fuming 264 fundamental noise 40 furnace 109, 111, 290 refractory metal 112 furnace AAS 123, 232 fusion 304
g GaAs 189, see also gallium arsenide galvanic detection 299 gas 289 purging 110 gas chromatography 190 gas chromatography coupled with ICP-MS 271 multi-capillary 271 gas ¯ow 91 aerosol carrier 258 carrier 262 injector 262 intermediate 220 nebulizer 222, 262, 265 outer 220 gas ¯ow dynamic 225 gas jet 141 gas±liquid separator 107 gas-phase reaction 276
353
354
Index gas sampling glow discharge 280 gas temperature 121, 138 Gauûian distribution 16 Gaussian function 36 GC-MPT-TOFMS 274, 275 GD dc microsecond-pulsed 279, see glow discharge gas sampling 280, see glow discharge rf-planar-magnetron 279, see glow discharge generator high-frequency 219 generation of hydride 108 geological 1 geological sample 232, 269 glass 188, 286 gliding spark 130 glow discharge 68, 125, 137, 175, 193, 275, 290 a planar cathode 144 gas-sampling 251, 252 jet-assisted 249 microwave assisted 249 radio-frequency 138 rf 145, 248 glow discharge mass spectrometry 255, 275, see GD-MS glow discharge OES detection limit 216 glow discharge source 44 glow emission 136 gold±platinum gauze 108 gold sponge 173 grain size 121, 123 graininess 141 graphite atomizer 113 graphite cup 228 graphite furnace 118, 228, 295 transversally heated 174 graphite furnace AAS 150 graphite furnace atomic absorption spectrometry 123 graphite furnace atomization 237 graphite furnace evaporation 113 graphite platform 165 graphite powder 118 graphite tube 111, 112 pretreatment 108 pyrolytic graphite coated 169 grating concave 59 echelle 59, 209 holographically ruled 56
mechanically ruled 56 plane 58 Grimm lamp 244 Grimm-type discharge 287 Grimm-type source 277 grinding 123, 124, 302 groove 207 ground state 132 ground state atom 148, 296
h Hû line 21 halfwidth 203 halogenated hydrocarbon 273 halogenation 125 hard-sphere 139 heat decomposition 122 latent heat 121 heat conductance 135 heated spray chamber 102 heating stage 114 high eciency nebulizer (HEN) 92 high-energy preburn 141, 245 high-frequency discharge 152 high-power nitrogen discharge 272 high-purity Ag powder 286 high-purity Ga 284 high-purity substance 189, 283 high-purity Ti 285 high resolution 87 high-resolution spectrometer 185 high-temperature superconductor 232, 284 Hittorf dark space 137 hollow cathode 251, 287 hollow cathode discharge 25 hollow cathode lamp 14, 152, 177, 290 hollow cathode source 293 hop 155 hot hollow cathode 243 hot-trapping 107, 173 hydraulic high pressure nebulization system 102 hydride 89, 105 hydride forming element 238 hydride generation 105, 229, 289 continuous-¯ow 252 electrochemical 230 hydride technique 150, 172 hydrogeological sample 269
i ICP atomic emission 87 ICP generator 221
Index ICP mass spectrometry 255, see ICP-MS ICP-AES 87, 232 sequential 226 simultaneous 226 ICP-MS 43, 255, 272, 307, 310 laser ablation 270 low-resolution 257 ICP-OES 41, 42, 194, 307 ICP-optical emission spectrometer 219 ignition pulse 125 illumination 52 image on the entrance slit 54 intermediate image 53 image 86, 203, 204 image dissector tube 67, 68 imaging mirror 52 impact excitation 241 impact ionization 241 impactor 286 impulse theory 139, 144 in-depth resolution 141 indirect determination 175 inductance 214 inductively coupled plasma 2, 193, see ICP inertion of the emulsion 63 instrument cost 310 instrumental drift 176 integrated absorption 167 integrator 65 interference 116, 170, 201, 224, 265 additive 85, 224 spectral 85 calcium-phosphate 163 chemical 172 ¯ame AAS 163 isobaric 265 matrix 269 multiplicative 85, 86 signal enhancement or depression 86 physical 170 spectral 202, 205, 258, 260, 261, 264, 265 interference noise 40 interferogram 71 internal standard 192, 212, 225, 265 internal standardization 85, 125, 135, 212, 225, 264 ion cluster 257, 265 doubly charged 265 doubly charged analyte 258 ®ller gas 276 polyatomic 260 ion current 277 ion detection 80
dual 82 Faraday collector 81 ion counting 81 photographic plate 80 ion energy 259 ion exchange 186 ion extraction 82 ion intensity 277 ion lens 255 ion line intensity 19 ion optics 83, 84 ion sampling 82 ion source 109 ionization 18, 192, 262 ionization buer 164 ionization energy 22, 136 ionization equilibrium 164, 224 ionization interference 256 irradiance 14 isobaric interference 276 isothermal atomizer 191 isothermal furnace 170 isothermic distillation 304 isotope ratio 266 isotope ratio measurement 267 isotopic abundance 258 isotopic analysis 158
j jet expansion 83 jet-assisted glow discharge 143 jet-enhanced sputtering 249
k kerosene 227 kinetic energy 136
l La- and Sr-compound 164 Lambert±Beers' law 184 laminar-¯ow clean bench 304 lamp deuterium 293 tungsten halogenide 293 Langmuir probe 138, 248 laser 2, 30, 31, 131 ablation 131 continous wave 298 Cu-vapor laser-pumped dye 294 dye 293 excimer 293 femtosecond 135
355
356
Index laser (cont.) gas 133 high-power 251 Nd:YAG 133, 134, 135, 230 pulsed dye 294, 299 rod 132 semiconductor 133 solid state 132 source 131 laser ablation 89, 134, 175, 268, 302 laser atomic ¯uorescence 156 laser light scattering 122 laser photoionization 301 laser plume 295 laser radiation 297 laser source 251 layered ceramic 248 leaching 306 least square linear regression 35 lens 52 level rotational 24 vibrational 24 life science 3, 187, 270 lifetime 15, 298 lifetime of the excited state 11 limit of detection 46 limit of determination 50, 226 limit of guarantee of purity 50 limits of detection 284 LINA spark 231 LINA-sparkTM 135 line atomic 18 atomic emission 225 hard 23, 223 intensity 194 ion 18 rotational 23, 25 soft 23, 223 line coincidence 193 line frequency 196 line pair 12 line pair selection 225 line pro®le 17, 74 line source 149 line spectrum 152 line width 157 line wing 202 linear dynamic range 37, 182, 226, 246, 296, 301 linear regression 84 linearity 182
linearization 163 line-to-background intensity ratio 249 line-to-background ratio 251 liquid chromatography particle beam 280 local thermal equilibrium 20, 28, 220 local thermodynamic equilibrium 138 longitudinal ®eld 179, 180 loop 118, 162 Lorentzian distribution 16 low pressure ICPs 280 low pressure MIP 280 low-pressure discharge 30, 88 low-pressure ICP 273
m MacSimion 84, 256 magnetic ®eld 126, 144, 212, 250 magneto-optical eect 183 magnetron 234 Marsch method 107 mass analyzer 75 mass resolution 76, 265 mass spectra 258 mass spectrometer 72, 84 double-focussing 277 dual-channel 257 ion cyclotron resonance analyzer 79 ion trap 78 multiple-collector magnetic mass analyzer 80 quadrupole 258, 277 quadrupole mass ®lter 3 rf-GD-TOF 279 sector ®eld 254 simultaneous 80 mass spectrometry 254 elemental 254 glow discharge 309 isotope dilution 267 resonance ionization 301 thermionic 72 time-of-¯ight 272 mass spectrum 280 matching unit 221 Mathieu equation 74 matrix destruction 114, 119 matrix eect 85, 129, 218, 301, see also interference matrix interference 215, 309 matrix modi®cation 115, 169 matrix modi®er 114, 115 matrix-free determination 107, 115
Index Maxwell distribution 8 Maxwell velocity distribution 10 mean droplet diameter 222 medical sample 270 medium voltage spark 127 medium voltage spark OES 247 Meinhard nebulizer 92 melting point 122 melting with ¯uxes 186 memory-eect 104 mercury 108, 190 mercury cold vapor technique 108 metal 246, 269 metal alloy 284 metal argide 282, 284 metal halogenide 160 metal sample 302 metalloprotein 270 metallurgical laboratory 215 metals analysis 232 metals industry 2, 187 metastable argon 237 metastable helium 237 metastable level 220 metastable state 156 meteorite 285 MIBK 227 Michelson interferometer 70 microchannel-plate 66 microdensitometer 81 microdistributional analysis 134 microsample 112, 117 microstrip line 240 microstructured resonant cavity 240 microtechnique 105 microwave assisted digestion 185 microwave induced plasma 157 microwave induced plasma atomic emission detector 273 microwave-assisted treatment 305 microwave plasma 2, 88, 234, 255 capacitively coupled 236 microwave plasma torch 239 mill 186 mini-torch 219 MIP 42 MIP torch 238 modelling 241 modi®er 119 Mohs hardness scale 302 molecular band 25, 178, 260 molecular gas 127, 272 molecular species 23, 113, 157, 171
molybdenum 187 momentum orbital impulse 6 spin 6 monochromatic radiation 14 monochromator 149 Czerny±Turner 150, 203 Ebert 150, 203 monocrystal 139 Monte Carlo model 241 Monte Carlo simulation 143 mounting Echelle 60 Paschen±Runge 60 MS ICP-TOF 279 quadrupole ion-trap 280 MSP 240 temperature 240 excitation 240 gas 240 MSP miniaturized MIP 240 multi-CCD system 205 multielement capacity 218, 226 multielement determination 184, 226 multielement method 184 multilayer 287, 288 multiply charged Ar specie 284
n Na®on membrane 106 nascent hydrogen 105 natural width 15 Nd:YAG laser 251, 268 nebulization direct injection 222 high pressure 102, 222, 228 jet-impact 101 pneumatic 222 thermospray 228 ultrasonic 103, 228, 267 nebulization chamber 90, 96, 101, see spray chamber nebulization eect 100, 164, 224, 265 nebulization eciency 92, 102 nebulizer Babington 94, 121, 227 Babington-type 222 concentric 91, 161 cross-¯ow 94, 161, 222, 227 direct injection 96 eciency 91 fritted-disk 96, 222
357
358
Index nebulizer (cont.) grid 227 high-eciency 222, 227 jet-impact 222 LeÂgeÁre 227, 239 maximum dissolved solid 227 Meinhard 222 microconcentric high-eciency 271 microultrasonic 228 oscillating 97, 222 pneumatic 90 nebulizer gas ¯ow 223 negative glow 137, 243, 245 negative ion 264 Nier±Johnson geometry 277 niobium 187 noble metal 188 noise 39, 163, 264, 300 ampli®er 47, 200 background 46 dark current 200 detector 201, 225 excess 40 1=f 41, 42, 264 ¯icker 40, 47, 200, 224, 246 fundamental 40, 41 interference 40 nebulizer 224 non-fundamental 40 photon 200 pixel-to-pixel 201 random 40 shot 45 type 39 white 40, 246, 264 noise power spectra 40, see noise spectrum noise spectra 225 noise spectrum 40 non-absorbed radiation 149 non-conducting sample 138 non-conductive sample 286 non-element speci®c absorption 168 non-fundamental noise 40 non-LTE 243 non-LTE plasma 30, 241 non-metal 156 non-resonant AFS 293 non-resonant ¯uorescence 292 non-resonant line 250 non-resonant transition 298 non-speci®c absorption 173 norm temperature 211 normal distribution 49 nucleation 116
Nuclepore ®lter 95, 121, 122, 128, 303 Nukuyama±Tanasawa 100 number density analyte 196 electron 196
o observation axial 223 radial 223 observation height 223 observation zone 218 OES using a Grimm-type glow discharge 247 o-axis 76, 153 oil analysis 189, 216 on-line matrix removal 176 on-line preconcentration 99, 176 on-line trace matrix separation 162 operating stability 93 optical aberration 198 optical beam 165 optical conductance 59 optical emission spectrometry 112 optical spectra 8, 112 optical spectrometer 84 optical spectrometry 112 optical transmittance 196 optimization maximum 262 optimization study 262 optimization 163, 261 optogalvanic eect 297 ore 189 organic matrices 168 organolead compound 172 organolead 190, 271 organotin 190 oscillator strength 13, 299 over-ionization 220 over-population 220 oxyanions 163, 164
p P-branch 24 paint analysis 189 particle diameter 128 particle size distribution 104, 122 particle size 121, 123, 303 partition function 9, 20, 226 Paschen curve 244 Paschen series 4 Paschen±Runge 203, 206 peak area 116
Index peak height 116 peak skew 273 pellet 125, 248, 285 Peltier cooling element 235 Peltier element 102 penetration rate 139, 287 Penning eect 220 Penning ionization 138, 242, 275, 276 peristaltic pump 94 permanent magnet 181 Pfund series 4 pharmaceutical 189 photocurrent 64 photoelectric measurement 47, 198 photographic detection 2 photographic emulsion 201 photomultiplier 64, 150 solar-blind 59, 65, 210 photon 196 photon ¯ux 197 photon stop 84 photoplate 59 physical interference 164 physical line width 149 physical width 15, 198, 223 p-component 179, 180 pin-shaped electrode 287 pin-shaped sample 277 Planck's constant 196 Planck's law 8, 11 plasma 8, 28, 290 current-carrying 217 current-free 217 high-frequency 233 low-power 233 inductively coupled 219 laser-produced 294 microwave 235 capacitively coupled 235 microwave induced 235 stabilized capacitively coupled 233 toroidal 219 transferred 217 plasma discharge high-frequency 30 microwave 30 plasma jet 31 three-electrode 218 plasma mass spectrometry 129, 254 plasma parameter 248 plasma source AES 216 plasma spectrometry 89 plasma temperature 21 plasma tomography 29
plasmaarray spectrometer 68 platform 115 platform technique 113 plume 251 pneumatic nebulization 89, 222 pneumatic nebulizer 161 optimization 100 point to plane 127 Poiseuille law 100 Poisson distribution 45 polarizer 180, 182 polyatomic compound 160 population inversion 131, 132 positive column 137 powder sampling syringe 174 power 223, 263 power detection 215 power of detection 46, 99, 109, 115, 148, 156, 162, 163, 169, 220, 223, 246, 263, 268, 307 absolute 134 practical resolution 59 precision analysis 125 precision 35, 104, 266 preconcentration 107 predisperser 56, 153 prespark 215 primary combustion zone 160 primary radiation source 152 primary source 148, 149, 290, 291, 293 prism 1 probability 199 pro®le absorption 148, 292 spectral 292 pro®le function 17 pro®le of spectral line 16 pro®le of the line 76 protein fraction 270 Pt powder 284 PTFE vessel 185 pulse dierential height analysis 215 pulse length 294 pulse width 46 pulsed laser 299 pulsed mode 133 pulverization 123 pumping eciency 132 pumping process 291 pyrolytic graphite 112 pyrolytic graphite coated graphite tube 170 pyrolytically coated 110 pyrolytically coated graphite 172
359
360
Index
q Q-branch 24 Q-switched 133 quadrupole mass spectra 282 quadrupole 256 qualitative analysis 86, 193 quantitative analysis 192 quantitative atomic emission spectrometry 194 quantum eciency 64, 65, 197 quantum number magnetic 5 main 5 orbital 5 rotational 24 spin 6 quarternary ammonium salt 168 quartz ®ber optics 54 quasi-simultaneous measurement of line and background absorption 177 quaternary ammonium salt 114 quenching 291
r R-branch 24 radial viewing 227 radiance 132 radiant density 179, 198, 290, 292, 293 radiant ¯ux 51, 151 radiant intensity 64 radiation ¯uorescence 294 radiation density 11, 150 radiation detector photoelectric detection device 61 photographic emulsion 61 photomultiplier 61 radiation source 10, 17, 51, 109, 192, 198 radiation trapping 220 radiative 242 radiative decay 242 radiative recombination 220 radical 171 radioisotope 286 radiotracer 169 Raman shifting 299 random noise 40 rare earth element 283 raw material 187 Rayleigh scattering 172 reciprocal linear dispersion 170 recombination 136, 138, 165, 241 recombination continuum 18
recombination process 110 reduced pressure discharge 2, see glow discharge reducing ¯ame 159 reduction 112 reference element 85, 125, 151 reference signal 197 re¯ectivity 196 re¯ectron 256 refractive index 154 refractory 110 refractory carbide 112 refractory element 164 refractory powder 229, 232, 304 refractory sample 186 relative method 197 relative method of analysis 170 relative sensitivity factor 281 relative standard deviation 195 remote sampling 125 removal 100 repeatability 36 repeller 76 residence time 109, 167, 258 residual gas impurity 276 resistance 214 resolution 78, 79, 277 resolving power 148 resonance ¯uorescence 292 resonance line 156, 163, 183 resonance radiation 148 resonant transition 298 resonator 132, 234, 236 response time 291 restrictor tube con®guration 142 Reticon 66 reversed-phase chromatography 271 rf coil 256 rf discharge 138 rf glow discharge 144 rf shielding 278 rf-GD-MS 283, 286 rf-powered glow discharge 278 rinsing time 222 R-L-C-circuit 221 rotating arc 118 rotation mill 302 rotation±vibration band spectra 210 rotation±vibration hyper®ne structure 178 rotational energy 24 rotational hyper®ne structure 23 RSF 284, 286 Russell and Saunders 7
Index Rydberg constant 4 Rydberg state 298, 301
s Saha constant 19 Saha equation 19 Saha-Eggert equation 257 sample 88 ¯at 278 inhomogeneity 131 pin 278 sample decomposition 185 sample decomposition step 107 sample dissolution 302, 304 sample feeding 94 sample homogeneity 115 sample inhomogeneity 284 sample introduction 88 pneumatic nebulization 90 sample preparation 302 sample solution consumption 161 sample uptake 222 sample volatilization 88, 126, 246 sampler 83, 255 sampler clogging 264 sampling boat 229 sampling depth 134 sampling eciency 161 saturation 291, 293 Sauter diameter 100 scanning electron micrograph 129 scanning electron microscopy 96 scanning monochromator 193 scattered radiation 183 scattering 183 scintillator 70 seawater analysis 270 secondary cathode 285, 286 secondary electron emission 136 sector ®eld 256 high-resolution 258 sediment 286 Seidel 63 selection rule 6 selective volatilization 118, 130 self biasing 138 self-absorption 17, 153 self-aspirating 92 self-reversal 17, 126, 153, 182, 193, 227, 246, 250, 296 self-sputtering 140 semi-conductor 189, 281 semiconductor-grade material 283
semi-quantitative analysis 254 semi-transparent mirror 151 sensitivity 85, 86, 184, 300 sequential 222 serum 187 Seya±Namioka mounting 61 sheathing gas 218 shock wave 83 SiC 123, 285, 303, see silicon carbide side-on 221 side-on observation 29 sieve 303 sieving 302 sifter 125 s-component 179, 180 signal depression 84, 261 signal enhancement 84, 261 signal generation 88 signal-to-background 116 signal-to-background ratio 112 signal-to-noise 66 signal-to-noise ratio 44, 47, 59, 69, 293 silicon±boron±carbo±nitride 225 Silsbee focussing 140 Simplex optimization 223 simultaneous 222 simultaneous detection 75 simultaneous emission spectrometer 194 single beam 150 single-channel instrument 151 SIT vidicon 204 Skewedness and excess test 48 skimmer 83, 255 cone 279 potential 279 slag 189 slit entrance 196 exit 196 slope 37 slurry 95, 114, 120 slurry atomization 120, 174 slurry nebulization 95, 268 slurry sampling 188 small sample 99 Smith±Hieftje technique 182 SNR value 101, see signal-to-noise ratio soft plasma 272 soil 285, 286 solid sample 211 solid state detector photodiode array 67 SIT vidicon system 67
361
362
Index solid state detector (cont.) vidicon 67 solid-phase extraction 162 solids ablation 251 solid-state detector 206 charge injection device 66 charge-coupled device 66 solution 184 solvent residue 172 source 27, 30 continuous 292 pulsed 292 space angle 196 space charge positive 136 space charge eect 265 spark 2, 30, 31, 126, 210 ac 213 channel 213 condensed 127 critically damped 215 detection limit 216 emission spectrometry 213 generator 214 high-voltage 126, 128, 215 medium voltage 215 oscillating 215 overcritically damped 215 repetition rate 127 single 215 unidirectional 215 spark ablated aerosol 239 spark ablation 88, 128 spark ablation coupled with ICP-OES 231 spark-ablation ICP-MS 282 spark chamber 128, 129 miniaturized 127 spark channel 128 spark emission spectrometry 88, 131 spark frequency 214 spark gap 127 spark source mass spectrography 254 spark source mass spectrometry 254, 284 spark train 127 sparking chamber 127 spatially isothermal graphite furnace 165 spatially-resolved measurement 221 speciation 190, 271, 289 species density 143 spectra 1, 195 display 86 ICP atomic emission 204 ICP mass 261
high-resolution 261 ICP-MS 258 spectral background 203, 220 spectral bandwidth 148, 149, 163, 177 spectral dispersion angular 57 reciprocal linear 57 spectral interference 144, 163, 193, 282, 296, 301, 309 spectral line 12, 193 spectral line table 193 spectral line width 155 spectral radiance 152 spectral radiance t 59 spectral range 154, 209 spectral resolution 194 spectral scan 87, 202 spectral slit width 57, 198 spectral stripping 194 spectral term 5 spectrochemical analysis 192 spectrograph 2 spectrometer atomic absorption 150 atomic emission 202 CCD 240 Czerny±Turner mounting 58 dispersive spectral apparatus 55 Ebert mounting 59 Echelle 206 Fastie±Ebert mounting 59 Fourier transform 70 grating 52 Hadamard transformation 70 ICP atomic emission 258 mass 73, 74, 76 magnetic mass analyzer 74 quadrupole 73 time-of-¯ight 76 multichannel 151 non-dispersive 70 non-dispersive spectral apparatus 55 optical 51 optical mounting 58 Paschen±Runge 61, 223 sequential 203 simultaneous 203 two-channel 226 spectrometry ac arc 213 dc arc 211 laser enhanced ionization 297 spectroscope 192 spray chamber 91
Index cyclone 91 single pass 98 spray chamber 90, see nebulization chamber spray chamber±burner assembly 161 sputter pro®le 250 sputtering 126, 245, 275 cathodic 135 sputtering eciency 144 sputtering equilibrium 141 sputtering rate 139 sputtering yield 139, 140 stabilized temperature platform furnace 167 stack gas 235 standard addition 86, 172, 260 standard deviation 36, 199, 200 statistical weight 19, 22 steel 187, 269, 282, 284, 287 brass-coated 287 high alloy NiaCr 131 low-alloyed 252 steel analysis 282 steel mill 303 stigmatic spectral apparatus 58 stimulated emission 131 stoichiometry 189 stray-light measurement 100 stray-radiation 52, 202, 294 strip line 234 structured background 181 student table 37 superalloy 282 superconductor PbaBiaSraCaaCuaO 225 superconductor material 188 surface tension 170 surfactant 170 surfatron 236 suspension 120 switch electro-acoustic 133 opto-acoustic 133
t tandem source 33 tandem source concept 88 tantalum 187 temperature electron 27, 193, 221, 245, 276 excitation 12, 27, 196, 220, 226 gas 27 gas kinetic 16 ionization 28 norm 22 plasma 223
rotation 220 rotational 25, 26, 27 spectroscopic measurement 13 volatilization 168 temperature pro®le 165 temperature program 169 temperature programming 114 temperature tuning 155 term scheme 7 theoretical resolving power 17 thermal equilibrium 10, 11 thermal evaporation 109, 211 thermal matrix removal 171 thermal method 117 thermal spray 89 thermal volatility 130 thermal volatilization 89, 129 thermally stable compound 172 thermally stable oxide 163, 295 thermionic diode 300 thermochemical aid 211 thermochemical behavior 117, 168 thermochemical decomposition 172 thermochemical modi®er 119, see matrix modi®er thermochemical reaction 113, 118 thermochemical reagent 114, 119, 168, 229 thermochemistry 168 thermospray 228 Thomson scattering 239, 241 three-body collision 242 three-body recombination 276 three-level system 292 time-of-¯ight 272 time-of-¯ight mass spectrometry 273 time-of-¯ight system 256 time-resolved absorption spectra 154 TM010 resonator 237 TOF-ICP-MS 267 TOF-MS plasma 272 torch Fassel 220 green®eld 220 toroidal argon discharge 237 toroidal MIP 238 toxic element 187 trace-matrix separation 123, 269 trace-O-mat 305 tracer experiment 266 trajectory 75 transfer of free atom 166 transformation equation 63 transformation function 62
363
364
Index transient 116 transient signal 76, 87, 110, 176, 267 transition non-resonant 294 resonant 294 transition probability 13, 24, 196, 290, 298 transmission 62, 84 transmittance 51 transport 167 transport eciency 169 transversally-heated furnace 165 transverse magnetic ®eld 179 trapping 107, 173, 238 hot 107 trapping hydride 238 TTA 162, see also thenoyltri¯uoroacetone tube furnace 109, 165 tunable diode laser 191 tunable diode laser source 176 tunable laser 292, 299 tuned laser 290 tungsten 187 tungsten coil 166 tungsten ®lament 228, 233 tungsten ®lament atomizer 166 tungsten furnace 165 tungsten trioxide 174 tuning range 155 turbomolecular pump 84, 255 turret 185 two-body collision 242 two-line method 21 two-photon spectroscopy 301
u U3 O8 213 ultrasonic nebulization 89, 123 ultrasonic stirring 123 uncertainty 38 undissociated molecule 171 unidirectional spark 127 urine 270, 288
vesicle mediation 108 vibrational ®ne structure 23 vidicon 59 viewing port 165 viscosity 170 viscosity drag force 121 Voigt eect 183 Voigt pro®le 16 volatile element 164 volatile species formation 108 volatile species generation 105 volatilization 128, 165 volatilization eect 109 volatilization interference 161 volatilization process 27, 109 volatilization temperature 119 VUV wavelength 216
w wandering eect 126 wash-out time 90 waste water analysis 270 water analysis 190 water loading 102 waveguide 234 wavelength gap 155 wavelength modulation laser AAS 157 WC 123 wet chemical dissolution 304 wetting agent 121 white noise 40 wire 130 wire cup 118 wire loop atomization 112 working coil 219
x xenon lamp 290 x-ray ¯uorescence 122 x-ray spectrometry 311 xylene 227
z v vapor 289 vapor cloud 109 vapor condensation 128 vapor sampling 289 vaporization electrothermal 109 vegetation 286 velocity of light 13 vesicle 174
Zeeman AAS 179 Zeeman eect 6, 154, 179 anomalous 179 longitudinal ®eld 180 normal 179 Zeeman splitting 180, 182 Zeeman-eect background correction 158 zero-background 183 ZrO2 162, 186, 194, see also zirconium dioxide
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
1
Introduction Atomic spectroscopy is the oldest instrumental elemental analysis principle, the origins of which go back to the work of Bunsen and Kirchho in the mid-19th century [1]. Their work showed how the optical radiation emitted from ¯ames is characteristic of the elements present in the ¯ame gases or introduced into the burning ¯ame by various means. It had also already been observed that the intensities of the element-speci®c features in the spectra, namely the atomic spectral lines, changed with the amount of elemental species present. Thus the basis for both qualitative and quantitative analysis with atomic emission spectrometry was discovered. These discoveries were made possible by the availability of dispersing media such as prisms, which allowed the radiation to be spectrally resolved and the line spectra of the elements to be produced. Around the same time it was found that radiation of the same wavelength as that of the emitted lines is absorbed by a cold vapor of the particular element. This discovery was along the same lines as the earlier discovery made by Fraunhofer, who found that in the spectra of solar radiation line-shaped dark gaps occurred. They were attributed to the absorption of radiation by species in the cooler regions around the sun. These observations are the basis for atomic absorption spectrometry, as it is used today. Flames proved to be suitable sources for determinations in liquids, and in the work of Bunsen and Kirchho estimations were already being made on the smallest of elemental amounts that would still produce an emission or absorption signal when brought into a ¯ame. From this there was already a link appearing between atomic spectroscopy and the determination of very small amounts of elements as being a basis for trace analysis. With industrial developments arose a large need for the direct chemical analysis of solids. This resulted from expansion of production processes, where raw materials are subjected to large-scale processes for the production of bulk materials, from which products of increased value, complying to very strict speci®cations, are manufactured. The search for appropriate raw materials became the basis for mining, which was then developed on a large scale. Geological prospecting with the inevitable analyses of large amounts of samples for many elements, often down to low concentration levels as in the case of the noble metals, took place. Also trading of raw materials developed, which intensi®ed the need for highly accurate characterization of ores and minerals, a development that today is becoming more strin-
2
Introduction
gent. Accordingly, analytical methodology that allows widely diverse materials to be characterized for many elements became necessary. Because of economic implications this information must frequently be obtained rapidly, which necessitates so-called multielement methods for the direct analysis of solids. Not only is there a need for the characterization of raw bulk materials but also the requirement for process controled industrial production introduced new demands. This was particularly the case in the metals industry, where production of steel became dependent on the speed with which the composition of the molten steel during converter processes could be controlled. After World War II this task was eciently dealt with by atomic spectrometry, where the development and knowledge gained about suitable electrical discharges for this task fostered the growth of atomic spectrometry. Indeed, arcs and sparks were soon shown to be of use for analyte ablation and excitation of solid materials. The arc thus became a standard tool for the semi-quantitative analysis of powdered samples whereas spark emission spectrometry became a decisive technique for the direct analysis of metal samples. Other reduced pressure discharges, as known from atomic physics, had been shown to be powerful radiation sources and the same developments could be observed as reliable laser sources become available. Both were found to oer special advantages particularly for materials characterization. The need for environmentally friendly production methods introduced new challenges for process control and fostered the development of atomic spectrometric methods with respect to the reliable determination of elements and their species in both solids, liquids and gaseous samples. The limitations stemming from the restrictive temperatures of ¯ames led to the development of high temperature plasma sources for atomic emission spectrometry. Thus, as a result of the successful development of high-frequency inductively coupled plasmas and microwave plasmas these sources are now used for routine work in practically all large analytical laboratories. Accordingly, atomic emission spectrometry has developed into a successful method for multielement analyses of liquids and solids as well as for determinations in gas ¯ows. This is due to the variety of sources that are available but also to the development of spectrometer design. The way started with the spectroscope, then came the spectrographs with photographic detection and the strongest development since photoelectric multichannel spectrometers and ¯exible sequential spectrometers, has recently been with array detectors becoming available. In time, the use of ¯ames as atom reservoirs for atomic absorption spectrometry was also transformed into an analytical methodology, as a result of the work of Walsh [2]. Flame atomic absorption spectrometry became a standard tool of the routine analytical laboratory. Because of the work of L'vov and of Massmann, the graphite furnace became popular as an atom reservoir for atomic absorption and gave rise to the widespread use of furnace atomic absorption spectrometry, as oered by many manufacturers and used in analytical laboratories, especially for extreme trace analysis. However, in atomic absorption spectrometry, which is essentially a single-element method, developments due to the multitude of atomic reservoirs and also of primary sources available, is far from the end of its development. Lasers will be shown to give new impetus to atomic absorption work and also to make
Introduction
atomic ¯uorescence feasable as an extreme trace analysis method. They will also give rise to new types of optical atomic spectrometry such as laser enhanced ionization spectrometry. The sources investigated and that are being used successfully for optical atomic spectrometry are also powerful sources for elemental mass spectrometry. This led to the development of classical spark source mass spectrometry moving through to what are today important plasma mass spectrometric methods, such as glow discharge and inductively coupled plasma mass spectrometry. The development of elemental mass spectrometry started with the work of Aston on the elemental mass separator. Here the ions of dierent elements are separated on the basis of their de¯ection in electrical and magnetic ®elds. This development lead to high-resolution but expensive instrumentation, with which highly sensitive determinations could be performed, as e.g. are necessary for high-purity materials that are required for the electronics industry. Towards the end of the 1970s, however, so-called quadrupole mass ®lters developed to such a high standard that they could replace conventional mass spectrometers for a number of tasks. The use of mass spectrometry instead of optical emission spectrometry enabled considerable gains in the power of detection to be made for the plasma sources developed around that time. The development of this ®eld is still proceeding fully, both with respect to the ion sources and the types of mass spectrometers used as well as with respect to detector technology. All the items mentioned will ®nally make an impact on the analytical performance of plasma mass spectrometry. Certainly the latter is considerably more advantageous than optical methods as a result of the possibility of detecting the various isotopes of particular elements. The dierent atomic spectrometric methods with ¯ames and plasmas have to be judged by comparing their analytical ®gures of merit with those of other methods for elemental analysis, a point which has to be seen through a critical eye with respect to the analytical problems to be solved. The scope for plasma spectrometry is still growing considerably, as it is no longer elemental determinations but the determination of the elements as present in dierent compounds that is becoming important, for problems associated with the design of new working materials, challenges in life sciences as well as for environmental and bio-medical risk assesment. This gives the area of interfacing atomic spectrometry with separation sciences a strong impetus, which needs to be treated in depth, both from the development of types of interfaces as well as from the point of view of reshaping existing and developing new sources of suitable size and cost±performance ratios.
3
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
4
1
Basic Principles 1.1
Atomic structure
The basic processes in optical atomic spectrometry involve the outer electrons of the atomic species and therefore its possibilities and limitations can be well understood from the theory of atomic structure itself. On the other hand, the availability of optical spectra was decisive in the development of the theory of atomic structure and even for the discovery of a series of elements. With the study of the relationship between the wavelengths of the chemical elements in the mid-19th century a fundament was obtained for the relationship between the atomic structure and the optical line emission spectra of the elements. In 1885 Balmer published that for a series of atomic lines of hydrogen a relationship between the wavelengths could be found and described as: l k n 2 =
n 2 ÿ 4
1
where n 2; 3; 4; . . . for the lines Ha ; Hb ; Hg etc. Eq. (1) can also be written in wavenumbers as: n 0 1=l R
1=2 2 ÿ 1=n 2
2
where n 0 is the wavenumber (in cmÿ1 ) and R the Rydberg constant (109 677 cmÿ1 ). The wavenumbers of all so-called series in the spectrum of hydrogen are given by: n 0 1=l R
1=n12 ÿ 1=n22
3
where n2 is a series of numbers > n1 and with n1 1; 2; 3; 4; . . . for the Lyman, Balmer, Paschen and Pfund series, respectively. Rydberg applied the formula of Balmer as: n 0 R Z 2
1=n12 ÿ n22
4
where Z is the eective charge of the atomic nucleus. This formula then also allows calculation of the wavelengths for other elements. The wavenumbers of the atomic
1.1 Atomic structure
spectral lines can thus be calculated from the dierence between two positive numbers, called terms, and the spectrum of an element accordingly contains a large number of spectral lines each of which is related by two spectral terms. The signi®cance of the spectral terms had already been re¯ected by Bohr's theory, where it is stated that the atom has a number of discrete energy levels related to the orbits of the electrons. These energy levels are the spectral terms. As long as an electron is in a de®ned orbit no electromagnetic energy is emitted but when a change in orbit occurs, another energy level is reached and the excess energy is emitted in the form of electromagnetic radiation. The wavelength is given according to Planck's law as: E h n h c=l
5
Here h 6:623 10ÿ27 erg s, n is the frequency in sÿ1 , c 3 10 10 cm/s is the velocity of light and l is the wavelength in cm. Accordingly: n 0 1=l E=h c E1 =
h c ÿ E2 =
h c T1 ÿ T2
6
T1 and T2 are the Bohr energy levels and the complexity of the emission spectra can be related to the complexity of the structure of the atomic energy levels. For an atom with a nucleus charge Z and one valence electron, the energy of this electron is given by: Eÿ
2 p Z 2 e4 m n2h2
7
m m M=
m M, with m being the mass of the electron and M the mass of the nucleus; n is the main quantum number
n 1; 2; 3; . . . and gives the order of the energy levels. Through the movement around the atomic nucleus, the electron has an orbital impulse moment L of which the absolute value is quantitized as: jLj h=
2p
p l
l 1
8
l is the orbital quantum number and has values of: 0; 1; . . . ;
n ÿ 1. The elliptical orbits can take on dierent orientations with respect to an external electric or magnetic ®eld and the projections on the direction of the ®eld also are quantitized and given by: L z h=
2pm l
9
L z is the component of the orbital momentum along the ®eld axis for a certain angle, m l Gl; G
l ÿ 1; . . . ; 0 is the magnetic quantum number and for each value of l it may have
2l 1 values.
5
6
1 Basic Principles
When a spectral line source is brought into a magnetic ®eld, the spectral lines start to display hyper®ne structures, which is known as the Zeeman eect. In order to explain these hyper®ne structures it is accepted that the electron rotates around its axis and has a spin momentum S for which: p jSj h=
2p S
S 1
10
The spin quantum number m s determines the angles possible between the axis of rotation and the external ®eld as: sz h=
2pm s
11
where m s G12. The orbital impulse momentum and the spin momentum are vectors and determine the total impulse momentum of the electron J as: J LS
with j Jj h=
2p
p j
j 1
12
j l G s and is the total quantum number. In the case of an external magnetic or electrical ®eld, the total impulse momentum also has a component along the ®eld, whose projections on the ®eld are quantized and given by: Jz h=
2p mj
with mj G j; G
j ÿ 1; . . . ; 0
13
This corresponds with possible 2 j 1 orientations. The atomic terms dier by their electron energies and can be characterized by the quantum numbers using the so-called term symbols: n m lj
14
Here l 0; 1; 2; . . . and the corresponding terms are given the symbols s (sharp), p (principal), d (diuse), f (fundamental), etc., originally relating to the nature of dierent types of spectral lines: n is the main quantum number, m is the multiplicity
m 2s G 1 and j is the total internal quantum number. The energy levels of each element can be given in a term scheme. In such a term scheme, also indicated are which transitions between energy levels are allowed and which ones are forbidden. This is re¯ected by the selection rules. According to these, only those transitions are allowed for which Dn has an integer value and at the same time Dl G1, D j 0 or G1 and Ds 0. The terms of an atom with one valence electron can easily be found, e.g., for Na
1s 2 2s 2 2p 6 3s 1 , in the ground level: 3 2 S1=2 [l 0 (s), m 2:1=2 1 2
s 1=2 and j 1=2
j jl G sj]. When the 3s electron goes to the 3p level, the term symbol for the excited level is: 3 2 P1=2; 3=2 [l 1 (p), m 2:1=2 1 2 as s 1=2 and j 1=2; 3=2]. The terms have a multiplicity of 2 and accordingly the lines have a doublet structure.
1.1 Atomic structure Fig. 1. Atomic energy level diagram for the sodium atom. (Reprinted with permission from Ref. [3].)
The term schemes of the elements are well documented in the work of Grotrian [3]. For the case of the Na atom the term scheme is represented in Fig. 1. When atoms have more than one valence electron, the term schemes become more complex as a coupling between the impulse and orbital momentums of the individual electrons occurs. According to Russell and Saunders
L ÿ S a coupling applies, where the orbital moments of all electrons have to be coupled to a total orbital momentum, as with the spin momentum. This coupling applies for elements with Z below 20, where it is accepted that the spin±orbital interactions are much lower than the spin±spin and the orbital±orbital interactions. The fact that none of the electrons in an atom can have the same set of quantum numbers is known as the Pauli rule. The total quantum number L is obtained as L Sl, S Ss and J L ÿ S; . . . ; L S. The term symbol accordingly becomes: M
LJ
15
For the case of magnesium
1s 2 2s 2 2p 6 3s 2 , the ground level is 3 1 S0 (as there are two 3s electrons which must have antiparallel spins L 0 as l1 0 and l2 0, S 0 as s1 1=2 and s2 ÿ1=2, and J 0 as both L and S 0). The excited level
1s 2 2s 2 2p 6 3s3p is characterized by the terms: 3 1 P1 (L 1 as l1 0 and l2 1, S 0 as s1 1=2 and s2 ÿ1=2, and J jL G 1j 1) but also 3 3 P2 , 3 3 P1 and 3 3 P0 (as for the spins s1 1=2 and s2 1=2, S 1, and further J 0; 1; 2 parallel). Here singlet
m 1 and triplet
m 3 terms are present in the term scheme. Also a j j coupling is possible, when the interaction between spin and orbital momentum of the individual electrons is decisive. With a number of electrons the coupling becomes more complex and leads to a high number of terms and accordingly line-rich atomic spectra. Also not only neu-
7
8
1 Basic Principles
tral atoms but also ions with dierent levels of ionization have term schemes, making the optical spectra very line rich. Indeed, for 90 elements between 200 and 400 nm more than 200 000 atomic lines have been listed, and many others are missing from the tables. From Planck's law, as given by Eq. (5), the relationship between the optical atomic spectra of the elements and energy level transitions of the valence electrons can be understood. Indeed, the wavelength corresponding to a transition over an energy dierence of 1 eV according to Planck's law corresponds to a wavelength of: 1 eV 1:6 10ÿ12 erg 6:62 10ÿ27 erg s 3 10 10 cm=s 1=l
cm or 1240 nm. Accordingly, the optical wavelength range of 200±800 nm corresponds to energies of 2±7 eV, this being the range involved in transitions of the valence electrons.
1.2
Plasmas
Partially ionized gases are usually denoted as plasmas [4]. They contain molecules, radicals and atoms but also ions and free electrons and result from the coupling of energy with matter in the gaseous state. As has been previously stated for atoms, radicals, molecules and ions also present in the plasma can be in their ground states and in excited states and radiation can be emitted or absorbed when transitions from one state to another occur. The wavelength of the radiation can be obtained from Planck's law whereas the intensities of the discrete lines depend on the number densities of the species and the states involved. Transfer of energy for the dierent species in a plasma results from the nonradiative as well as from the radiative processes taking place. Non-radiative processes involve collisions and radiative processes involve emission, absorption and ¯uorescence of radiation. The eciency of collision processes is described by the cross section s
v. This re¯ects the loss in impulse a particle with mass m and velocity v undergoes when it collides with a particle with mass M. It can be given by:
s
v 2p
p 0
p
v; y
1 ÿ cos y sin y dy
16
This expression shows that apart from loss of momentum a change in direction may also result from collisions. The mean collision cross section is denoted as: hs
vi. A collision frequency is described as hs
v vi and a mean collision frequency as hs
v vi=hvi. Apart from the cross section, however, the velocity distribution for a given species is important for describing the energy transfer in a plasma. In the case of a Maxwell distribution the velocity distribution is given by: p p 0 dn=n 2=
p u 0 eÿu du 0
17
1.3 Emission and absorption of radiation
In the case of a so-called Druyvenstein distribution: dn=n 1:039
p u 0 exp
ÿ0:548 u 02 du 0
18
u 0 E=kT, where E is the mean energy of the particles and T the absolute temperature. For a plasma contained in a closed system and which is in so-called thermal equilibrium, the population of the excited levels for every type of particle is given by Boltzmann's law: nq =n0 gq =g0 exp
ÿEq =kT
19
nq is the number density of particles in the excited state, n0 is the number density of particles in the ground state, gq and g0 are the statistical weights of the corresponding levels, Eq is the excitation energy of the state q, k is Boltzmann's constant (1:38 10ÿ16 erg K) and T is the absolute temperature. In Eq. (19) a relationship is formulated between the temperature and the atom number densities in a single excited state and in the ground state, respectively. As the latter is not constant, the Boltzmann equation can be better formulated as a function of the total number of particles n distributed over all states. Then nq =n gq exp
ÿEq =kT=Sm gm exp
ÿEm =kT
20
as n Sm n m . The sum Zm Sm gm exp
ÿEm =kT is the partition function. This is a function of the temperature and the coecients of this function for a large number of neutral and ionized species are listed in the literature (see e.g. Ref. [5]). When Eq is expressed in eV, Eq. (20) can be written as: log naq log na log nq ÿ
5040=T Vq ÿ log Z
21
1.3
Emission and absorption of radiation
In a steady-state plasma the number of particles leaving an energy level per unit of time equals the number returning to this level [6]. In order to characterize such an equilibrium, all processes which can lead to excitation as well as to de-excitation have to be considered. The most important energy exchange processes in a plasma are as follows.
. (1a) Collisions where atoms are excited to a higher level by collision with energetic neutrals (collisions of the ®rst kind).
. (1b) Collisions where excited neutrals loose energy through collisions without emission of radiation (collisions of the second kind).
. (2a) Excitation by collision with electrons.
9
10
1 Basic Principles
. (2b) De-excitation where energy is transferred to electrons. . (3a) Excitation of atoms by the absorption of radiation. . (3b) De-excitation of atoms by spontaneous or stimulated emission. When n is the number density of the ®rst type of particles and N the one of a second species that is present in excess
n f N, the following equilibria can be considered: a N n0 b N nq
22
a e ne n 0 b e n e nq
23
B 0 rn n0
A B rn nq
24
ne is the electron number density, A, B and B 0 are the Einstein transition probabilities for spontaneous emission, stimulated emission and absorption and ae ; a; b e and b are functions of the cross sections for the respective processes as well as of the velocity distribution of the particles involved. rn is the radiation density (frequency n). When the system is in so-called thermodynamic equilibrium, the neutrals and the electrons have the same Maxwell velocity distribution and at a temperature T we have: nq =n0 a=b ae =b e B 0 =
A=rn B gq =g0 exp
ÿEq =kT
25
Thus each process is in equilibrium with the inverse process and the Boltzmann distribution of each state is maintained by collisions of the ®rst and the second kind, including the ones with electrons, and there are no losses of energy through the emission of radiation or any absorption of radiation from an external source. In a real radiation source this perfect equilibrium cannot exist and there are losses of energy as a result of the emission and absorption of radiation, which also have to be considered. However, as long as both only slightly aect the energy balance, the system is in so-called local thermal equilibrium and: a N n0 ae ne n0 B 0 rn n0 b N nq b e ne nq
A B rn nq
26
from which nq =n0 can be calculated as: nq =n0
a N ae ne B 0 ru =b N b e ne
A B ru
27
The population of the excited states is determined by the excitation processes in the radiation source, as re¯ected by the coecients in Eq. (26). In the case of a dc arc for instance a N g ae ne B 0 rn and b N g b e ne
A B rn . This leads to:
1.3 Emission and absorption of radiation
nq =n0 a=b gq =g0 exp
ÿEq =kT
28
As the radiation density is low, it can be accepted that the dc arc plasma is in thermal equilibrium. The excited states are produced predominantly and decay through collisions with neutrals. The simpli®cation which leads to Eq. (26) does not apply to discharges under reduced pressure, where collisions with electrons are very important as are the radiation processes. Moreover, the velocity distributions are described by the Druyvenstein equation. These sources are not in thermal equilibrium. Excited states are prone to decay because of their high energy and therefore mainly have short lifetimes. The decay can occur by collisions with surrounding particles (molecules, atoms, electrons or ions) or by emission of electromagnetic radiation. In the latter case the wavelength is given by Planck's law. When the levels q and p are involved, the number of spontaneous transitions per unit of time is given by: ÿdNq =dt Aqp Nq
29
where Aqp is the Einstein coecient for spontaneous emission (in sÿ1 ). When considering an optically thin sytem with atoms in the excited state q, which decay spontaneously to a level p under the emission of radiation, the number of transitions per unit time at each moment is proportional to the number of atoms in the state q. When several transitions can start from level q
q ! p1 ; q ! p2 ; . . . ; q ! pn Eq. (29) becomes: ÿdNq =dt Nq Sp Aqp Nq nq
30
Here nq is the inverse value of the mean lifetime of the excited state q. For levels in which a decay by an allowed radiative transition can take place the lifetime is of the order of 10ÿ8 s. When no radiative transitions are allowed we have metastable levels (e.g. Ar 11.5 and 11.7 eV), which only can decay by collisions. Therefore, such levels in the case of low pressure discharges may have very long lifetimes (up to 10ÿ1 s). In the case of the absorption of electromagnetic radiation with a frequency nqp and a radiation density rn, the number density of Nq increases as: dNq =dt Bqp Nq nr
31
For the case of stimulated emission, atoms in the excited state q only decay when they receive radiation of the wavelength lqp and ÿdNq dt Bqp Nq rn
32
For the case of thermal equilibrium: gq Bqp gp Bpq
33
11
12
1 Basic Principles
and: A
8phn 3 =c 3 Bqp
8phn 3 =c 3
gq =gp Bqp
34
where gp and gq are the degeneratives of the respective levels with g 2J 1. The intensity
Iqp of an emitted spectral line is proportional to the number density of atoms in the state q: Iqp Aqp naq h nqp
35
or after substitution of naq , or nq for atomic species, according to Eq. (20): Iqp Aqp h nqp na
gq =Za exp
ÿEq =kT
36
When multiplying with d=
4p, where d is the depth of the source (in cm), one obtains the absolute intensity. T is the excitation temperature, which can be determined from the intensity ratio for two lines (a and b) of the same ionization stage of an element as: T 5040
Va ÿ Vb =flog
gAa =
gAb ÿ log
la =lb ÿ log
Ia =Ib g
37
In order to determine the excitation temperature with a high precision, the thermometric species should have a high degree of ionization, otherwise the temperature or the geometry of the discharge change when the substance is brought into the source. Furthermore, the dierence between Va and Vb must be large. Indeed, the error of the determination can be obtained by dierentiating Eq. (37): dT=T T=5040
Va ÿ Vb 0:434 dI=I
38
gAa =
gAb must be large when
Va ÿ Vb is large and Ia =Ib should not be particularly small or large. In addition the transition probabilities must be accurately known. Indeed, the error of the determination of the temperature strongly depends on the accuracy of
gAa =
gAb , as by dierentiating Eq. (37) with respect to log
gAa =
gAb one obtains: dT T 2 =5040
Va ÿ Vb d
gAa =
gAb
39
Often the line pair Zn 307.206/Zn 307.59 is used, for which: Va 8:08 eV and Vb 4:01 eV and
gAa =
gAb 380 and: T 20510=2:58 logI307:6
nm =I307:2 nm
40
This line pair is very suitable because the ionization of zinc is low as a result of its relatively high ionization energy, the wavelengths are close to each other, which
1.3 Emission and absorption of radiation
minimizes errors introduced by changes in the spectral response of the detector, and the ratio of the gA values is well known. The excitation temperature can also be determined from the slope of the plot lnIqp =
gq Aqp nqp or lnIqp l=
gAqp versus Eq , which is ÿ1=k T as: lnIqp =
gq Aqp nqp ln
h n=Z ÿ Eq =
k T
41
The l=gA values for a large number of elements (such as argon, helium, iron) and their lines have been compiled [6]. The determination of excitation temperatures in spatially inhomogeneous plasmas has been treated extensively by Boumans [7] and is described later (see Ref. [8]).
. Example For the case of a 2 kW inductively coupled plasma the four iron lines Fe I 381.58 nm, Fe I 383.04 nm, Fe I 382.44 nm and Fe I 382.58 nm have relative intensities of 5, 10, 2.3 and 7.4 a.u., respectively. When using the transition probability products gA of 66, 36, 1.26 and 26 (see Ref. [9]) as well as the excitations energies of 38175, 33096, 26140 and 33507 cmÿ1 , respectively, it can be calculated that the excitation temperature should be 5000 K. Limitations to the spectroscopic measurement of the temperatures from line intensities lie in possible deviations from ideal thermodynamic behavior in real radiation sources, but also in the poor accuracy of transition probabilities. They can be calculated from quantum mechanics, and have been determined and compiled by Corliss and Bozman at NIST [10] from measurements using a copper dc arc. These tables contain line energy levels, transition probabilities and the so-called oscillator strengths for ca. 25000 lines between 200 and 900 nm for 112 spectra of 70 elements. Between the oscillator strength f qp (being 0.01±0.1 for non-resonance and nearer to 1 for resonance lines) there is the relationship [11]: f qp
gq =gp Aqp
m c 3 =
8 p 2 e 2 n 2
42
and f qp 1:499 10ÿ16 l 2 gq =gp Aqp
43
As g is known to be 2J 1, oscillator strengths can be converted into transition probabilities and vice versa. According to the classical dispersion theory, the relationship between the absorption and the number density of the absorbing atoms is given by:
Kn dn
pe 2 =mc Nn f
44
Kn is the absorption coecient at a frequeny n; m is the mass and e the charge of the electron, c is the velocity of light, Nn is the density of atoms with a line at a
13
14
1 Basic Principles
frequency between n and n dn and is almost equal to N; f is the oscillator strength. This relationship applies purely to monochromatic radiation. As the widths of the absorption lines in most atom reservoirs are of the order 1±5 pm, the use of a primary source which emits very narrow lines would be advantageous. Indeed, when using a continuous source one would need a spectral apparatus with a practical resolving power of at least 500 000 to reach the theoretically achievable values of Kn and this certainly would lead to detector noise limitations as a result of the low irradiances. Therefore, it is more advantageous to use sources which emit relatively few narrow lines and to use a low-resolution monochromator which just isolates the spectral lines in the spectra. The relationship between the absorption A and the concentration of the absorbing atoms in an atom reservoir is given by the Beer±Lambert law. When I0 is the intensity of the incident radiation, l the length of the atom reservoir and I the intensity of the sorting radiation, the change in intensity dI resulting from the absorption within the absorption path length dl is given by: ÿdI k I0 c dI
45
or
I
I0
dI=I k c
l
0
dl
46
A log
I0 =I log
1=T, in which A is the absorbance and T the transmission. Accordingly, Eq. (46) becomes: Akcl
47
The absorbances are additive. The Lambert±Beer law, however, is only valid within a restricted concentration range. This is due to the fact that not all radiation reaching the detector has been absorbed to the same extent by the analyte atoms. Normally the calibration curve at high concentrations levels o asymptotically to the signal for the non-absorbed radiation. The latter consists of contributions from non-absorbed lines of the cathode material in the hollow cathode lamp or of lines of the ®ller gas within the spectral bandwidth of the monochromator. The calibration curve is only linear in the concentration range where the ratio of the widths of the emission to the absorption line is < 1=5. Also incomplete dissociation of analyte molecules leads to a curvature towards the concentration axis, and this incomplete dissociation then becomes limiting at high concentrations. A decrease in the ionization at higher analyte concentrations, however, may cause curvature away from the concentration axis. For all these reasons, in atomic absorption spectrometric measurements deviations from linearity are common and the linear dynamic range is much smaller than in atomic emission or atomic ¯uorescence.
1.3 Emission and absorption of radiation
Line broadening
Atomic spectral lines have a physical width as a result of several broadening mechanisms [12]. The natural width of a spectral line is due to the ®nite lifetime of an excited state
t. The corresponding halfwidth in terms of the frequency is given by: DnN 1=
2pt
48
For lines corresponding to transitions that are allowed according to the selection rules, lifetimes are of the order of 10ÿ8 s and, accordingly, for most spectral lines this results in a natural broadening with a halfwidth of the order of 10ÿ2 pm. The Doppler spectral width results from the movement of the emitting atoms and their velocity component in the viewing direction. The respective halfwidth is given by: DnD 2
p p
ln 2=c n0
2 R T=M
49
where c is the velocity of light, n0 is the frequency of the line maximum, R is the gas constant and M the atomic mass. The Doppler broadening thus strongly depends on the temperature. Accordingly, it is also often denoted as temperature broadening and re¯ects the kinetic energy of the radiating species (atoms, ions or molecules). The relevant temperature is denoted as the gas temperature or Doppler temperature. The measurement of the Doppler broadening thus allows the determination of the gas temperatures in spectroscopic sources (see line pro®les). For the light elements, the Doppler broadening is larger than it is for analytical lines with shorter wavelengths. For the Ca 422.6 nm line in the case of a hollow cathode discharge at a few mbar pressure, the Doppler broadening at 300 K for instance is 0.8 pm whereas at 2000 K it is 2 pm [13]. The Lorentzian broadening or pressure broadening results from the interaction between the emitting atoms of the element considered and atoms of other elements. The halfwidth is given by: DnL
2=p sL2 N
p 2 p R T
1=M1 1=M2
50
M1 and M2 are the atomic masses, N is the concentration of the foreign atoms and sL is their cross section. The pressure broadening is low in the case of discharges under reduced pressure. For the case of the Ca 422.6 nm line this type of broadening at a temperature of 300 K and a pressure of 9 mbar is only 0.02 pm. At atmospheric pressure, however, this type of line broadening is the predominant one. Furthermore, isotopic structure and hyper®ne structure, and also resonance broadening, resulting from the interaction between radiating and non-radiating atoms of the same species, and Stark broadening, resulting from interaction with electrical ®elds, contribute to the physical widths of the spectral lines.
15
16
1 Basic Principles
The natural broadening and the Lorentzian broadening have a Lorentzian distribution, which is given by: I
n I0 =f1 2
n ÿ n0 =Dn 2 g
51
The Doppler broadening has a Gauûian distribution, described by: I
n I0 exp ÿf2
n ÿ n0 =2DnD
p
ln 2g
52
The combination of both types of pro®le functions (which is normally due to the predominance of both pressure and temperature broadening) results in a so-called Voigt pro®le, which can be described by: V
a; o a=p
y ÿy
fexp
ÿ y 2 d y=a 2
o ÿ y 2 g
o 2
n ÿ n0 =DnD
53
p p
ln 2 and a DnD
ln 2
when the contribution of the natural width is neglected. From the physical line widths of the spectral lines, which in most cases are between 1 and 20 pm, and the information on the pro®le functions discussed above, the contributions of the different processes can be calculated by deconvolution methods. However, the experimentally measured full-widths at half maximum (FWHM) must be corrected for the bandwidth of the spectral apparatus used. In the determination of absorption pro®les with the aid of very narrow bandwidth tunable laser sources, the spectral bandwidth of the probing beam can often be neglected. Equation (53) can e.g. be solved by a mathematical progression as: V
a; o a=constant
a 2 o 2 f1 3
a 2 ÿ o 2 =2
a 2 o 2 2 3
a 4 10a 2 o 2 5o 4 =4
a 2 o 2 4 g
54
From this solution a and o and accordingly DnL and DnD can be determined and from the latter the gas kinetic temperature. This is a measure of the kinetic energy of atoms and ions in the plasma. Owing to the line broadening mechanisms, the physical widths of spectral lines in most radiation sources used in optical atomic spectrometry are between 1 and 20 pm. This applies both for atomic emission and atomic absorption line pro®les. In reality the spectral bandwidth of dispersive spectrometers is much larger than the physical widths of the atomic spectral lines. The pro®les of spectral lines as are obtained in plasma sources at atmospheric pressure are illustrated by the high-resolution records of a number of rare earth spectral lines obtained in the case of an inductively coupled plasma source (ICP) (Fig. 2) [14].
1.3 Emission and absorption of radiation
Fig. 2. Line profiles of some rare earth element atomic emission lines in inductively coupled plasma atomic emission spectrometry (photographic measurements obtained with a high-resolution grating spectrograph):
theoretical resolving power: 460 000. (1): Ho II 345.6 nm, (2): Er II 369.2 nm, (3): Yb II 369.4 nm, (4): Y II 371.0 nm, (5): Tm II 376.1 nm, (6): Eu II 381.9 nm and (7): La II 398.8 nm. (Reprinted with permission from Ref. [14].)
Self-absorption
The radiation emitted in a radiation source is absorbed by ground state atoms of the same species. This phenomenon is known as self-absorption (for an explanation, see e.g., Ref. [15]). As the chance that an absorbed photon is re-emitted is < 1, this causes the observed radiation to be smaller than the emitted radiation. When I0 is the intensity emitted at the wavelength of the line maximum and PE
n is the pro®le function, the intensity distribution for a line emitted by a radiation source over the pro®le of a line equals I0 PE
n and the intensity observed after the radiation has passed through a layer with a number density of absorbing atoms of nA is: I
n I0 PE
n expÿr PA
n=PA
n0
55
n0 is the frequency at the line center, PA
n is the absorption pro®le function, PA
n0 is its value at the line center and p is an absorption parameter: p A B PA
n nA
56
p increases with the absorption transition probability and thus is larger for resonance lines which stem from transitions ending at the ground level. It is also larger in sources with a high analyte number density nA. As the absorption is maximum in the center of the line, self-absorption always leads to ¯atter line pro®les. Selfabsorption increases with the analyte number densities in the source and with the number densities of emitting and absorbing analyte atoms the intensity of a line goes to that given by Planck's law for black body radiation. When a minimum occurs in the absorption pro®le, then the line undergoes so-called self-reversal, when the absorption parameter is then >1. Self-reversal only occurs when there is a strong temperature gradient in the radiation source and when the analyte number densities both in the hot as well as in the cooler zones of the plasma are high.
17
18
1 Basic Principles
Continuum radiation
Apart from the atomic and ion lines of the species present in a plasma source an emission spectrum has a continuum on which the emission lines are superimposed. This extends over the whole spectrum. It is due to the interactions between free electrons (``Bremsstrahlung'') and to the interaction of free and bound electrons (``recombination continuum''). The former is particularly important in the UV spectral region, whereas the latter is important at longer wavelengths. The spectral intensity distribution for the continuum radiation is given by: I
n dn K ne nr r 2 =
Te 1=2 exp
ÿhn=
kTe dn
freeafree K 0 1=
j 3 ne nZ
Z 4 =
Te3=2 expÿ
Uj ÿ hn=kTe
freeabound
57
K and K 0 are constants, ne is the electron number density, nr is the number density of the ions with a charge of r times the elementary charge, Te is the electron temperature, and is a measure of the kinetic energy of the electrons in the plasma. n is the frequency, h and k are the Planck and Boltzmann constants, respectively, Uj is the ionization energy from the term with quantum number j, nZ is the number density of the atoms with atomic number Z. The intensity per unit of wavelength (radiant density Bl ) is obtained by multiplying with 4pc=l 2 . Accordingly, for a hydrogen plasma the intensity of the ``Bremsstrahlung'' is given by: Bl
2:04 10ÿ32 ne2 =
l 2 Te1=2 expÿ
1:44 10 8 =
lTe
58
At complete ionization of the hydrogen (e.g. when added to a plasma with another gas as the main constituent) ne p=
2 k Te has a maximum at a wavelength of l
7:2 10 7 =Te or at a ®xed wavelength, the maximum intensity is found at a temperature Te
5:76 10 7 =l. Thus, the electron temperature can be determined from the wavelength dependence of the continuum intensity. As Te is the electron temperature, absolute measurements of the background continuum emission in a plasma, e.g. for the case of hydrogen, allow determination of the electron temperature in a plasma, irrespective of whether it is in local thermal equilibrium or not. Similar methods also make use of the recombination continuum and of the ratio of the ``Bremsstrahlung'' and the recombination continuum.
1.4
Ionization
Provided that sucient energy is transferred to a plasma, atoms can also be ionized. This depends on the temperature of the plasma and also on the ionization energy of the elements. As these ions have discrete energy levels between which
1.4 Ionization
transitions are possible, just as atoms do, ionic spectra also will be important when considering the emission of radiation by a plasma. The ionization of atoms (a) of the element j into ions (i) is an equilibrium: naj T nij ne
59
and the equilibrium constant Snj
T, known as the Saha constant, is given by: Snj
T
nij ne =naj
60
The degree of ionization aj for an element j is given by: aj nij =nj nij =
naj nij
61
naj and nij are the concentrations of the atoms and the ions and can be expressed as a function of the total number of atoms nj by: naj
1 ÿ aj nj
and nij aj nj
62
Accordingly, using the notation given in Ref. [7] the intensity of an atom line can be written as: Iqp Aqp h nqp gq =Zaj
1 ÿ aj nj exp
ÿEq =kT
63
and the intensity of an ion line is given by: Iqp A qp h nqp gq =Zij aj nj exp
ÿEq =kT
64
These expressions for the intensities contain three factors that depend on the temperature, namely the degree of ionization, the Boltzmann factors and the partition functions. In particular, aj can be written as a function of the electron number density and the Saha function as: aj =
1 ÿ aj ne Snj
T
65
However, the latter is also given by the well-known Saha equation. With the aid of wave mechanics and through dierentiation of the Boltzmann equation, the Saha function in terms of the partial pressures can also be expressed as: Spj
T
pij pe =
paj
2pm 3=2
kT 5=2 =h 3 2Zij =Zaj exp
ÿEij =kT
66
The factor of 2 is the statistical weight of the free electron (2 spin orientations),
19
20
1 Basic Principles
k 1:38 10ÿ16 erg/K, m 9:11 10ÿ28 g, h 6:67 10ÿ27 erg s and 1 eV 1:6 10ÿ12 erg. This leads to the expression:
Pij pe = paj Spj
T 6:58 10ÿ7 T 3=2 Zij =Zaj exp10
ÿ5040Vij =T
67
or in the logarithmic form: log Spj 5=2 log T ÿ 5040=T Vij log
Zij =
Zaj ÿ 6:18
68
where Vij is the ionization energy (in eV). The Saha equation is only valid for a plasma which is in local thermal equilibrium, where the temperature in the equation is then the ionization temperature. When this condition is not full®lled, the equilibrium between the dierent states of ionization is given by the so-called Corona equation [16]. Accordingly, the degree of ionization in a plasma can be determined from the intensity relationship between an atom and an ion line of the same element as: log
aj =
1 ÿ aj log
Iqp =Iqp ÿ log
gq A qp nqp =
gq Aqp nqp
5040=T
V ÿ Vq log
Zij =Zaj
69
This method can again only be applied for a plasma in local thermal equilibrium, the temperature of which is known. The partition functions Zaj and Zij for the atom and ion species, respectively, are again a function of the temperature and the coecients of these functions have been calculated for many elements [5]. Furthermore, the accuracy of the gA values and of the temperatures is important for the accuracy of the determination of the degree of ionization. One often uses the line pairs Mg II 279.6 nm/Mg I 278.0 nm and Mg II 279.6 nm/Mg I 285.2 nm for determinations of the degree of ionization of an element in a plasma. Once aj is known, the electron pressure can also be determined. Indeed, from logaj =
1 ÿ aj logSpj
T= pe
70
one can calculate log pe ÿlogaj =
1 ÿ aj log Spj
T
71
When taking into account Eqs. (68) and (69), this results in: log pe ÿlog
Iqp =
Iqp log
gq A qp nqp =
gq Aqp nqp
ÿ
5040=T
Vij Vq ÿ Vq 5=2
log T ÿ 6:18
72
By dierentiating Eq. (72) with respect to temperature it can be found that the error
1.4 Ionization
in the determination of the electron pressure as a result of errors in the determination of the temperature is given by: d
log pe
0:434 5=2
5040=T
Vij Vq ÿ Vq
dT=T
73
Eq. (72) also shows that the intensity ratio of the atom and ion lines of an element will change considerably with the electron pressure in the plasma. Elements with a low ionization energy such as Na will thus have a strong in¯uence on the intensity ratios of the atom and ion lines of other elements. This is analytically very important as it is the cause of the so-called ionization interferences, found in classical dc arc emission spectrometry but also in atomic absorption and plasma optical emission as well as in mass spectrometry. When the plasma is not in local thermal equilibrium (LTE), the electron number densities cannot be determined on the basis of the Saha equation. Irrespective of the plasma being in local thermal equilibrium or not, the electron number density can be derived directly from the Stark broadening of the Hû line or of a suitable argon line. This contribution to broadening is a result of the electrical ®eld of the quasi-static ions on one side and the mobile electrons on the other side. As described in Ref. [17] it can be written as: dl 21 1:75a
1 ÿ 0:75ro
74
where r is the ratio of the distance between the ions
rm and the Debye path length
rD , o is the broadening due to the interaction of the electrons
Ane and a 1=4 is the contribution of the interaction with the quasi-static ions
Ane . rm
4p ne =3ÿ1=3
75
rD
k T=
4p e 2 ne 1=2
76
dl can thus be calculated as a function of ne . Accordingly, from the widths of the Ar I 549.59 or the Ar I 565.07 nm lines, which are due mainly to Stark broadening, ne can be determined directly and is independent of the existence of LTE. Thus the temperature can also be determined when combined with measurement of the intensities of an atom and an ion line of the same element. Indeed, log ni =na ÿlog ne 3=2 log T ÿ
5040=T Vij log
Zij =Zaj 15:684
77
which can be combined with Eq. (72). Because with the ``two-line method'' using lines of the same ionization level for the determination of temperatures, it is dicult to ful®l all conditions necessary to obtain highly accurate values [see Eqs. (38) and (39)], a method was developed that enables the plasma temperature to be determined from intensities of lines belonging to dierent ionization levels. When Ii is the intensity of an ion line and I the intensity of an atom line (in general both lines have to belong to two adjacent
21
22
1 Basic Principles
ionization levels), one can write: Ii =I 2
Ai gi li =Agl
2pmkT 3=2 =
h 3
1=ne
T 3=2 expÿ
Ei ÿ E Ei ÿ DEi =
kT
78
DEi is a correction for the ionization energy of the lowest level. The plasma temperatures can also be determined from the measurements of absolute line intensities. A survey of all methods used and discussed in the various chapters is given in Refs. [7, 8, 12]. Norm temperatures
From Eqs. (63) and (64), which give the intensity of a line, and from the Saha equation [Eq. (68)], it can be understood that for each spectral line emitted by a plasma source there is a temperature where its emission intensity is maximum. This is the so-called norm temperature. In a ®rst approximation [18], it can be written as: Tn
0:95Vij 10 3 =1 ÿ 0:33 a 0:37 log
Vij =10 ÿ 0:14 log Pe
79
where Vij is the ionization energy, a Va =Vij and Ve is the excitation energy. Pe is the electron pressure (in atm; 1 atm A 101 kPa) and is a function of T and ne : Pe 1:263 10ÿ12 ne T
80
In the case where we have to consider the norm temperature for a line of an element which is only present as an impurity in a plasma, e.g., one formed in a noble gas, the dilution in the plasma (a) also has to be considered. For a system with more components Pe is given by: Pe
2i 1=
2i 3
gi =gi1 4a=
1 a 2 P
81
gi and gi1 are the statistical weights of the ions with charge i and i 1, respectively. Accordingly, as a result of the dilution the change in the norm temperature
Tn at a dilution of a will be given by: DTn =Tn 0:14 log4a=
1 a 2 < 1
82
At a dilution of 0.1 the change in norm temperature will thus be ÿ7:2%. In a source such as the inductively coupled plasma the analyte dilution can be very high [of the order of 10 8 (1 mL/min of a 1±10 mg/mL solution for an element with a mass of 40, which is nebulized with an eciency of 1% into an argon ¯ow of
1.5 Dissociation
10 L/min). In an ICP the norm temperatures for lanthanum atom lines such as La I 418.7 nm (Vij 5:61 eV and Va 2:96 eV), will thus change from ca. 5000 K for a pure lanthanum plasma to 2830 K as a result of the large dilution. However, for atom lines of elements with relatively low ionization energies, although the change in the norm temperatures as a result of the analyte dilution is high, it is much less than for ion lines. For the La II 412.3 nm line (second ionization energy: 11.43 eV and Va : 3.82 eV) Tn is 9040 K. From what is known about the norm temperatures, it becomes clear which types of lines will be optimally excited in a plasma of a given temperature, electron pressure and gas composition, and the norm temperatures thus give important indications for line selection in a source of a given temperature. Atom lines often have their norm temperatures below 4000 K, especially when the analyte dilution in the plasma is high, whereas ion lines often reach 10 000 K. Both types of lines are often denoted as ``soft'' and ``hard'' lines, respectively.
1.5
Dissociation
The dissociation of molecular plasma gases or analyte molecules which are brought into the radiation source is an equilibrium reaction. Accordingly, thermally stable radicals in particular or molecules are always present in a radiation source. They emit molecular bands which occur along with the atomic and ionic line spectra in the emission spectrum. Radicals and molecules may also give rise to the formation of cluster ions, the signals of which will be present in the mass spectra. The main species stemming from the plasma gases are: CN, NH, NO, OH and N2 (or N 2 ). From the analytes, predominantly thermally stable oxides remain (e.g., AlO , TiO , YO , etc.). A thorough treatment of molecular spectra is available in many classical textbooks (see e.g., Refs. [19, 20]). Molecules or radicals have dierent electronic energy levels
1 S; 2 S; 2 P; . . ., which have a vibrational ®ne structure
n 0; 1; 2; 3; . . . and the latter again have a rotational hyper®ne structure
J 0; 1; 2; 3; . . .. The total energy of a state is then given by: Ei Eel Evibr Erot
83
Eel is of the order of 1±10 eV, the energy dierence between two vibrational levels of the same electronic state is of the order of 0.25 eV and for the case of two rotational levels of a vibrational band the energy dierence is of the order of only 0.005 eV. Through a transition between two rotational levels a rotational line is emitted. When the rotational levels considered belong to the same electronic level, the wavelength of the radiation emitted will be in the infrared region. When they belong to dierent electronic levels, their wavelengths will be in the UV or in the visible region. Transitions are characterized by the three quantum numbers of the states
23
24
1 Basic Principles
involved, namely: n 0 ; n 0 ; j 0 and n 00 ; n 00 ; j 00 . All lines which originate from transitions between rotational levels belonging to dierent vibrational levels of two electronic states form the band: n 0 ; n 0 ! n 00 ; n 00 . For these band spectra the selection rule is D j j 0 ÿ j 00 G1; 0. Transitions for which j 00 j 0 1 then give rise to the P-branch, j 00 j 0 ÿ 1 to the R-branch and j 0 j 00 to the Q-branch of the band. The line corresponding with j 0 j 00 0 is the zero line of the band. When n 0 n 00 0, it is also the zero line of the system. The dierence between the wavenumber of a rotation line (in cmÿ1 ) and the wavenumber of the zero line in the case of the P and the R branch is a function of the rotation quantum number j and the rotational constant Bn for which: Ej =
hc Bn j
j 1
84
The functional relationship is quadratic and is known as the Fortrat parabola. For the CN radical and the N 2 molecular ion, the transitions giving rise to band emission between 370 and 400 nm, together with the rotational line pattern are represented in Fig. 3 [21]. For the violet system of the CN band, there is no Qbranch and the lowest j in the R-branch is j 1. Molecular and radical band spectra thus consist of electronic series, which in their turn consist of various vibrational bands, which again consist of rotational lines, many often only partially resolved. As in the case of atomic spectral lines, the intensity of a rotation line can be written as: Inm Nm Anm h nnm 1=2p
85
where Nm is the population of the excited level and nnm the frequency of the emitted radiation. The transition probability for dipole radiation is: 3 Anm
64 p 4 nnm =3k 1=
gm SjRni mk j 2
86
i and k are degenerate levels of the upper (m) and the lower state (n). Rni mk is a matrix element of the electrical dipole moment and gm is the statistical weight of the upper state. Nm is given by the Boltzmann equation: Nm N
gm =Z
T exp
ÿEr =kT
87
where Er is the rotational energy of the excited electronic and vibrational level and is given by: Er h c Bn 0 J 0
J 0 1
88
Bn 0 is the rotational constant and J 0 is the rotational quantum number of the upper state (m). For a 2 Sg ± 2 Su transition the term SjRni mk j 2 J 0 J 00 1 where J 0 and J 00 are the rotational quantum numbers of the upper and the lower states, respec-
1.5 Dissociation
tively. Accordingly: 4 Inm
16 p 3 c N nnm =3Z
T
J 0 J 00 1
exp
ÿh c Bn 0 J 0
J 0 1=kT
89
or 4 lnInm =
J 0 J 00 1 ln16 p 3 c N nnm =3Z
T
ÿ h c Bn 0 J 0
J 0 1=kT
90
By plotting lnInm =
J 0 J 0 1 versus J 0
J 0 1 for a series of rotational lines a so-called rotational temperature can be determined. It characterizes the kinetic energy of the molecules and radicals, by which the band spectra are emitted. It is also a good approximation of the temperature re¯ecting the kinetic energy of the neutrals and ions in a plasma. For the case of a hollow cathode discharge the Boltzmann plot and the temperatures as measured from CN and N 2 band hyper®ne structures are given in Fig. 3. Spectral lines of molecular bands emitted by molecules and radicals present in a plasma often interfere with the atomic spectral lines in atomic emission spectrometry. However, in atomic absorption spectrometry the absorption by molecular bands stemming from undissociated molecules in the atom reservoir also lead to systematic errors and require correction. Furthermore, in mass spectrometry molecular fragments give rise to signals, which can also interfere with the signals from the ionized analyte atom. Therefore, it is important to study the dissociation of molecular species in the high-temperature sources used as radiation sources, atom reservoirs or ion sources. In dierent plasma sources a series of band-emitting species stem from the working gas. In this respect, for instance, N2 , N 2 , CN, OH and NH band emission has to be mentioned. However, undissociated sample and analyte species are also present in the plasma. In particular, thermally stable molecules such as AlO, LaO, BaO, AlF, CaF2 and MgO may be present in atomic spectrometric sources. It is important to understand their dissociation as a function of the plasma temperature and the plasma composition. This dependence can be described by a dissociation equation, which is similar to the Saha equation: Kn
2p=h 2
mX mY =mXY
kT 3=2
ZX ZY =ZXY exp
ÿEd =kT
91
where: Kn
nX nY =nXY
92
Z represents the partition functions for the dierent atomic and molecular species, MX , MY and MXY are the respective masses and Ed is the dissociation energy. This can be rewritten as:
25
26
1 Basic Principles
Fig. 3. Determination of rotational
temperatures in a transitional hollow cathode. Rotational lines used are of the R-branch of 2 S± 2 S (0,0) CN 388.3 nm and 2 S± 2 S (0,0) N 2 391.4 nm band (positive m J 0 values belong
to the R-branch) (a); Boltzmann plots for a graphite hollow cathode (fi 4:76 mm) operated at 80 mA and 330 Pa argon (b). (Reprinted with permission from Ref. [21].)
log Kn 20:432 3=2 log
MX MY =MXY log
ZX ZY =ZXY ÿ log g 1=2 log T log B log
1 ÿ 10ÿ0:625o=T ÿ
5040=T Vd
93
For most diatomic molecules B and o (in cmÿ1 ) have been listed in tables (see e.g. Ref. [19]). Vd then is given in eV and for a diatomic molecule: Z kT=
h c B g exp
ÿEd =kT=1 ÿ exp
h c o=kT with g 1
1 S; 2
2 S; 1 p; 2 p1=2 ; 2 p3=2 ; . . .; 3
3 S; . . ..
94
1.6 Sources for atomic spectrometry
Thus for a metal oxide (XY), from nXY =nX NY =Kn the ratio of the number densities for the metal oxide and the metal atoms
nX as well as the degree of dissociation can be calculated when the plasma temperature, the partial pressure of oxygen in the plasma
pY and the dissociation constant are known. For refractory oxides of relevance in dc arc analysis, these data are listed, for example, in Ref. [7].
1.6
Sources for atomic spectrometry
In atomic spectrometry the sample material is brought into a high-temperature source (plasma, ¯ame, etc.) with the aid of a sampling device. The sample, which may be a liquid, a gas mixture or a solid must be transformed into a vapor or an aerosol. This involves sample nebulization or various volatilization processes (e.g., by thermal evaporation or sputtering). It is advantageous to supply as much energy as possible for this process. The volatilization processes, the principle of which is to lead to a physical or chemical equilibrium, will possibly result in complete atomization, irrespective of the state of aggregate, of the eventual solid state structure or of the chemical composition of the sample. This is very important both for obtaining the highest sensitivity as well as for keeping the matrix interferences involved in the analyses at the lowest possible level. The eectiviness of the volatilization processes involved, the plasma temperatures describing the kinetic energy distributions of the various plasma components as well as their number densities will all in¯uence the atomization of the sample in the source. The rotational temperatures are relevant to all processes in which molecules, radicals and their dissociation products are involved. They can be obtained from the intensity distribution for the rotational lines in the rotation±vibration spectra, as described by Eqs. (83±90). The molecules OH, CN etc. have often been used to measure temperature (see e.g. Refs. [21±23]). The gas temperature is determined by the kinetic energy of the neutral atoms and the ions. It can be determined from the Doppler broadening of the spectral lines, as described by Eq. (49). However, to achieve this contributions of Doppler broadening and temperature broadening have to be separated, which involves the use of complicated deconvolution procedures as e.g. shown for the case of glow discharges in Ref. [24]. Whereas the rotational and the gas temperature are particularly relevant to the evaporation processes in the plasma, the electron temperature, being a measure of the kinetic energy of the electrons, is relevant to the study of excitation and ionization by collisions with electrons. This is an important process for generation of the analyte signal both in optical atomic emission and in mass spectrometry. The electron temperature can be determined from the intensity of the recombination continuum or of the ``Bremsstrahlung'', as described by Eq. (57). The excitation temperature describes the population of the excited levels of atoms and ions. Therefore it is important in studies on the dependence of analyte line intensities on various plasma conditions in analytical emission spectrometry.
27
28
1 Basic Principles Tab. 1.
Temperatures (K) for sources in atomic spectrometry.
Source
Trotational
Texcitation
Telectron
Tionization
Arc discharges Spark Inductively coupled plasma Microwave plasmas Glow discharges
5000 Ð 4800 2000 600
5000 20 000 5000 4000 20 000
5500 20 000 6000 6000 30 000
5000 2000 6000 6000 30 000
LTE LTE (transient) ALTE departures from LTE non-LTE
It can be determined from the intensity ratio of two atomic emission lines of the same element and ionization state [see Eq. (42)] or from plots of the appropriate function for various atomic emission lines versus their excitation energies. The ionization temperature is relevant for all phenomena involving equilibria between analyte atoms, ions and free electrons in plasmas. In the case of thermal equilibrium, it occurs in the Saha equation [Eqs. (66, 68)] and can be determined from the intensity ratio of an ion and an atom line of the same element. In all other cases ionization temperatures can be determined from the ne value obtained from Stark broadening [see Eqs. (74, 77)]. The dierent temperatures for the most important sources in atomic spectrometry are listed in Table 1. In a plasma, which is at least in local thermal equilibrium, all the temperatures discussed are equal. In addition, this implies that the velocity distribution of all types of particles in the plasma (molecules, atoms, ions and electrons) at any energy level can be described by the Maxwell equation [Eq. (17)]. For all species the population of the dierent levels is then given by the Boltzmann equation [Eq. (19)]. Furthermore, the ionization of atoms, molecules and radicals can be described by the Saha equation [Eqs. (66, 68)] and the related equations for the chemical equilibrium. Finally, the radiation density in the source conforms to Planck's law and both the exchange of kinetic energy between the particles as well as the electromagnetic radiation exchange are in equilibrium with each other. The real plasma sources used in atomic spectrometry are at best in so-called local thermal equilibrium. However, contrary to the case of thermal equilibrium all processes between the particles do not involve emission or absorption of electromagnetic radiation, as the plasma cannot be considered as a completely closed system. Moreover, real plasma sources are extremely inhomogeneous with respect to temperature and species number density distributions. Accordingly, the above mentioned equilibria only occur within small volume elements of the sources, where gradients can be neglected. Many plasmas, however, have a cylindrical symmetry and can be observed laterally. The observer then integrates the information provided for many volume elements along the observation direction and: I
x 2
y 0
I
x; y d y
95
where x and y are the coordinates of the volume element and integration of an
1.6 Sources for atomic spectrometry Fig. 4. Abel inversion procedure used for the determination of radial distributions in plasma sources. x: lateral position, y: direction of observation, r: radial distance.
intensity I is performed along the x-axis. When considering I
r, with r the radial distance away from the plasma center (Fig. 4), this becomes: I
x 2
y x
I
r r dr=
p
r 2 ÿ x 2
96
From these integral values, however, the values at a well-de®ned radial distance away from the plasma center can be calculated. This necessitates side-on observation along several parallel axes, which are equidistant with respect to each other and an Abel-inversion. Indeed, when I 0
x is the ®rst derivative of the function I
x, describing the variation of a measured value as a function of the lateral position (x) during side-on observation (Fig. 4), the radial values at a distance r are given by: I
r ÿ
1=p
y r
p I 0
x dx=
r 2 ÿ x 2
97
This integral can only be solved numerically. One has to write each laterally measured intensity as a sum of the dierent volume elements with a given coecient denoting their contribution. For certain numbers of lateral observation positions
3; 10; . . . these Abel coecients are listed in the literature [25]. By inversion of the corresponding matrix the radial values are then obtained. This allows determination of the radial distributions of emissivities, absorbances, temperatures or species number densities in plasmas. By repeating this procedure at dierent heights in a plasma one can perform plasma tomography. Similar results, which do not require the assumption of cylindrical symmetry in the source, can now be obtained by imaging spectrometry using two-dimensional detectors (for a discussion, see e.g. Ref. [26]). In real plasmas departures from thermal equilibrium often occur. For the extreme case, as encountered in plasmas under reduced pressure, the emission or the
29
30
1 Basic Principles Fig. 5. Sources for atomic spectrometry (Reprinted with permission from Ref. [28].) (A): arc; (B): spark; (C): flame; (D): plasma sources; (E): lowpressure discharges; (F): furnace; (G): laser.
absorption of radiation becomes so important that there is no longer a clear relationship between the mean kinetic energies of the species and the excitation temperatures. The latter then lose the physical meaning of a temperature. The absence of local thermal equilibrium in these so-called non-LTE plasmas relates to the existence of high ®eld gradients or ac ®elds, by which only the light electrons and not the heavy atoms and ions can follow the ®eld changes and take up the dissipated energies fully. Accordingly, the mean kinetic energy of the electrons and thus also the electron temperature will be much higher than the gas kinetic temperature. For a plasma, which is not in local thermal equilibrium, the equilibrium between species at dierent ionization levels at dierent temperatures will be given by the so-called Corona equation, as is, for example, derived for the sun by astrophysicists [27]. Sources for atomic spectrometry include ¯ames, arcs, sparks, low-pressure discharges, lasers as well as dc, high-frequency and microwave plasma discharges at reduced and atmospheric pressure (Fig. 5) [28]. They can be characterized as listed in Table 2. Flames are in thermal equilibrium. Their temperatures, however, at the highest are 2800 K. As this is far below the norm temperature of most elemental lines, ¯ames only have limited importance for atomic emission spectrometry, but they are excellent atom reservoirs for atomic absorption and atomic ¯uorescence spectrometry as well as for laser enhanced ionization work. Arcs and sparks are
1.7 Analytical atomic spectrometry Tab. 2. Use of sources in analytical atomic spectrometry.
Source
Chemical ¯ame Arc Spark Electrically-heated furnace Arc jet plasma Inductively coupled plasma Microwave plasma Glow discharge Laser plasma Exploding wire/foil
Optical spectrometry
Mass spectrometry
emission
absorption
fluorescence
well known as sources for atomic emission spectrometry. The high temperatures obtained in spark sources anticipate that ion lines in particular will be excited, the norm temperatures of which are often beyond 10 000 K, whereas in arc sources atom lines will be predominant. In plasma sources under reduced pressure the gas kinetic temperatures are low. Accordingly, their atomization capacity will be limited. When these sources are used as both atomic emission sources and as primary sources in atomic absorption the line widths will be very narrow. Moreover, particularly when gases with high ionization energies such as helium are used, lines with high excitation energies, such as the lines of the halogens, can also be excited. The discharges under reduced pressure are valuable ion sources for mass spectrometry as analyte ionization takes place, and because of the low pressure the coupling with a mass spectrometer operated at 10ÿ5 mbar becomes easier. So-called plasma jet and plasma sources at atmospheric pressure are of particular use for the emission spectrometric analysis of solutions. Their gas kinetic temperatures are high enough so as to achieve a complete dissociation of thermally stable oxides and both atom lines as well as ion lines occur in the spectra. As re¯ected by the fairly high ionization temperatures, they are powerful ion sources for mass spectrometry and plasma mass spectrometry is now one of the most sensitive methods of atomic spectrometry. Lasers are very suitable devices for material ablation in the case of solids. Owing to their high analyte number densities, the plasmas are subjected to high self-absorption thus it is more appropriate to use them as sources for material volatilization only and to lead the ablated material into a second source.
1.7
Analytical atomic spectrometry
Analytical atomic spectrometry nowadays includes the use of ¯ames and plasma discharges for optical and mass spectrometry. The sources are then used directly as
31
32
1 Basic Principles Fig. 6. Term schemes for (a): atomic emission, (b): absorption and (c): fluorescence spectrometry (resonant: hn and non-resonant: hn 0 ).
emission sources or atom reservoirs for atomic absorption or atomic ¯uorescence or they are used for ion production. In optical atomic spectrometry, atomic emission, absorption and ¯uorescence all have their speci®c possibilities and analytical features. The type of information obtained is clear from the transitions involved (Fig. 6). In atomic emission, thermal or electrical energy is used to bring the analyte species into an excited state, from which they return to their ground state through emission of radiation characteristic of all species that are present and that were suciently excited. Thus from the principle of atomic emission spectrometry it is a clearly multielement method. The number of elements that can be determined simultaneously is only limited by the availability of suciently sensitive interferencefree spectral lines. In the case of atomic emission spectrometry the selectivity is achieved by the isolation of the spectral lines with the aid of the exit slit of the spectrometer. This puts high demands on the optical quality of the spectral apparatus used and certainly requires a spectrometer with high spectral resolution. Moreover, the lines are superimposed on a spectral background, which is partly structured as the result of the presence of radicals and molecular ions emitting band spectra. Broad wings arising from neighboring spectral lines, e.g. of matrix constituents, may also occur and ®nally there will be a continuum resulting from the interaction of free and bound electrons (see Section 1.3, continuum radiation). The intensity of the spectral background on which the line to be measured is superimposed, may thus dier considerably from one sample to another. It must be subtracted from the total line intensities, as only the radiation emitted by the analyte atoms is relevant for the calibration. This can be done by estimating the spectral background intensity ``under'' the analytical line for the background intensities at the wavelength of the analytical line using a blank sample. It is often safer to estimate it from the spectral background intensities close to the analytical line (e.g. on each side) in the spectrum of the sample itself. The latter is certainly the case when the line is positioned on a wing of a band or a broad matrix spectral line, providing dierent spectral background intensities on either side of the analytical line. The means to correct for the spectral background must be available in every atomic emission spectrometer used for trace analysis. In atomic absorption spectrometry we need a primary source delivering monochromatic radiation of which the wavelength agrees with that of a resonance line of the elment to be determined. The spectral width must be narrow with respect to the absorption pro®le of the analyte line. From this point of view atomic absorption
1.7 Analytical atomic spectrometry
is a single-elemental method, of which the dynamic range is usually much lower than in atomic emission spectrometry. It is, at a ®rst approximation, a zero-background method, when neglecting absorption due to radiation scattering and molecular absorption. As the ®nal spectral selectivity is realized by the primary source, the spectral apparatus only has to enable line isolation in a spectrum of the element to be determined, where the spectral lines are very narrow. Accordingly, the demands on the spectral resolving power of the spectrometer are much lower than in atomic emission spectrometry. In atomic ¯uorescence, the excitation can be performed both with white as well as with monochromatic sources, which consequently aects the ¯uorescence intensities obtainable and the freedom from stray radiation limitations. The latter are particularly low with monochromatic primary sources and when using ¯uorescence lines with wavelengths diering from that of the exciting radiation. Generally, in atomic ¯uorescence the linear dynamic range is higher than in atomic absorption and spectral interference as well as background interferences are just as low. In the case of atomic absorption and atomic ¯uorescence the selectivity is thus already partly realized by the radiation source delivering the primary radiation, which in most cases is a line source (hollow cathode lamp, laser, etc.). Therefore, the spectral bandpass of the monochromator is not as critical as it is in atomic emission work. This is especially true for laser based methods, where in some cases of atomic ¯uorescence a ®lter is sucient, or for laser induced ionization spectrometry where no spectral isolation is required at all. For glow discharges and inductively coupled high-frequency plasmas ion generation takes place in the plasmas. In the ®rst case mass spectrometry can be performed directly on solids and in the second case on liquids or solids after sample dissolution. In the various atomic spectrometric methods, real samples have to be delivered in the appropriate form to the plasma source. Therefore, in the treatment of the respective methods extensive attention will be given to the techniques for sample introduction. In an atomic spectrometric source both the atomic vapor production as well as signal generation processes take place. The ®rst processes require high energy so as to achieve complete atomization as already discussed, whereas the signal generation processes in many cases would pro®t from a discrete excitation which makes, use for example, of the selective excitation of the terms involved. Therefore, in a number of cases the use of so-called tandem sources, where the analyte vapor generation and the signal generation take place in a dierent source (for a discussion, see Ref. [29]), may oer advantages with respect to the power of detection as well as freedom from interferences.
33
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
34
2
Spectrometric Instrumentation Atomic spectrometric methods of analysis essentially make use of equipment for spectral dispersion so as to isolate the signals of the elements to be determined and to make the full selectivity of the methodology available. In optical atomic spectrometry, this involves the use of dispersive as well as of non-dispersive spectrometers. The radiation from the spectrochemical radiation sources or the radiation which has passed through the atom reservoir is then imaged into an optical spectrometer. In the case of atomic spectrometry, when using a plasma as an ion source, mass spectrometric equipment is required so as to separate the ions of the dierent analytes according to their mass to charge ratio. In both cases suitable data acquisition and data treatment systems need to be provided with the instruments as well. When designing instruments for atomic spectrometry the central aim is to realize fully the ®gures of merit of the methods. They include the power of detection and its relationship to the precision, the freedom from spectral interferences causing systematic errors and the price/performance ratio, these being the driving forces in the improvement of spectrochemical methods (Fig. 7).
2.1
Figures of merit of an analytical method
In practice atomic spectrometric methods are relative methods and need to be calibrated. Calibration
The calibration function describes the functional relationship between the analytical signals, whether they be absorbances, absolute as well as relative radiation intensities or ion currents. The simplest form of the calibration function can be written as: Y acb
98
In the calibration this functional relationship is established from m measurements
2.1 Figures of merit of an analytical method Fig. 7. Qualities of an analytical method.
of the signals Yi obtained for samples with a known concentration c i . The coecients a and b are then determined by a least square linear regression procedure, which ensures that the deviations between the measured signals and the signals calculated from the regression equation are minimal. This is the case for a
mSci Yi ÿ Sc i SYi =mSc i2 ÿ
Sc i 2
99
and b
SYi ÿ aSc i =m
100
Often the inverse function of the calibration function is used, namely: c a0 Y b0
101
This is known as the analytical evaluation function. Calibration curves are often not linear, which necessitates use of a polynome of a higher degree to describe the calibration function. The latter may have the form: Y a0 a1 c a2 c 2
102
a0 ; a1 ; a2 ; . . . are determined by multivariate regression. Normally, a polynome of the 2nd degree is sucient to describe the calibration function over a large concentration range. A segmented calibration curve can also be used when the calibration function is not linear. Precision
When performing a series of measurements of analytical signals as they are obtained for a well-de®ned concentration of the analyte, there is a statistical uncer-
35
36
2 Spectrometric Instrumentation
tainty that stems from ¯uctuations in the analytical system. This gives rise to a statistical error. This determines the precision of an analytical method. The precision achievable is an important ®gure of merit of an analytical procedure and the statistical evaluation of analytical data has been treated extensively in classical textbooks (see e.g. Refs. [30±32]). The precision of a measurement procedure is expressed in terms of the standard deviation as: s
p S
Ym ÿ Yi 2 =
n ÿ 1
103
provided the distribution of the individual values can be described by a Gaussian function and, accordingly, is a normal one. Ym S
Yi =n is the mean value, Yi is an individual measurement value, n the number of individual measurements and s is an estimate for the standard deviation. The standard deviations for values resulting from dierent series of measurements are the result of a propagation of errors. Thus the coecients of variation
s 2 for a and b in the linear regression [Eqs. (99±100)] can be calculated as: s
a 2 m sY2 =mSci2 ÿ
Sci 2
104
s
b 2 sY2 Sci2 =mSci2 ÿ
Sci 2
105
and the con®dence levels for a and b are: Da Gt
P; f s
b and Db Gt
P; f s
b
106
where t is the value of the Student distribution (see e.g. Ref. [33]). For the precision of an analytical method not only the repeatability of single measurements but also the errors in the calibration procedure should be included. This is a complex problem, because according to the nature of the signal noise, the precision obtainable may vary considerably with the concentration level. In the case of a calibration the scattering of the measured Yi values around the calibration curve is given by the standard deviation of the regression: s
Y 2 S1m
Yi ÿ yi 2 =
m ÿ 2
107
yi are the signal values obtained for a standard sample with concentration c from the regression equation. The latter is calculated by a least square procedure from the pairs
Yi ; ci , where Yi are the measured values for ci; m is the total number of measurements taken with the standard samples and F m ÿ 2 is the number of degrees of freedom and: S
Yi ÿ yi 2 SYi2 ÿ bSYi ÿ aSxi Yi
108
Accordingly, the standard deviation for the concentration of the analytical sample
cX , which includes the error on the calibration, in the case of a calibration with
2.1 Figures of merit of an analytical method
samples, for which it is accepted that the errors in their concentration are low as compared with the scattering on a limited number of signals, can be calculated through a propagation of the errors as: sr
c ln 10 s
log c ln 10 b s
Y
109
b is the slope of the calibration curve. The magnitude of the concentrations of the analytical sample with respect to those of the calibration samples used also has to be considered, which can be included in Eq. (109) as: s
cX t cX ln 10 b s
Y p 1=m 1=nX
log cX ÿ log c 2 =S
log ci ÿ log c 2
110
t is the value from the Student table for a given probability and number of degrees of freedom, cX is the concentration determined for the unknown sample, b is the slope of the calibration curve (sensitivity), s
Y is the standard deviation of the regression, m is the number of replicates for all calibration samples, nX is the number of replicates for the unknown sample and log c is the mean for all calibration samples used [32]. Both Eqs. (109) and (110) are only valid for a limited concentration range, which is known as the linear dynamic range. This range is limited at the upper end by physical phenomena such as the detector over¯ow and at the lower end by the limit of determination. This limit is typical for a given analytical procedure and is the lowest concentration at which a determination can still be performed with a certain precision. When using calibration by standard additions, the sample matrix is present both in the unknown as well as in the calibration samples. Therefore, one can eectively correct for all interferences stemming from changes in the sensitivity resulting from in¯uences of matrix constituents on the signal generation. Provided one makes a single addition of an amount of analyte, which is free of errors, the statistical error for the analysis can be derived by a propagation of errors. When X is a function of n variables, for each of which N replicate measurements are made, the estimate of the standard deviation s
X can be calculated from the individual standard deviations for the n variables [s
k with k: 1; 2; 3; . . .] and the estimates Xk as: s 2
X
n X k1
qX=qXk sk 2
111
For: Y1 b c and after an addition cA for which:
112
37
38
2 Spectrometric Instrumentation
Y2 b
c cA
113
c cA =
Y2 ÿ Y1 Y1
114
A propagation of error according to Eq. (111) delivers: s 2
c fq=qY1 cA Y1 =
Y2 ÿ Y1 g 2 s
Y1 2 fq=qY2 cA Y1 =
Y2 ÿ Y1 g2 s
Y2 2
115
or s 2
c cA2 f1
Y2 ÿ Y1 Y1 =
Y2 ÿ Y1 2 g s
Y1 2 fY1 1=
Y2 ÿ Y1 2 g 2 s
Y2 2
116
and as cA c
Y2 ÿ Y1 =Y1 : s 2
c
c=Y1 Y2 =
Y2 ÿ Y1 s
Y1 2 c 1=
Y2 ÿ Y1 s
Y2 2
117
or: sr2
c Y22 =
Y2 ÿ Y1 2 sr2
Y1 ÿ sr2
Y2
118
Normally, there is also an uncertainty in the concentrations of the calibration samples used both in the case of calibration with synthetic calibration samples as well as in calibration by standard additions. In both cases these errors can also be included in the calculation of the analytical error [34]. When weighing an amount of sample M there is a weighing error sM and when dissolving this amount of sample in a volume V, there is an additional uncertainty sV in the volume to be taken into account. The resulting relative standard deviation can be calculated as: sr; c
p
sM =M 2
sV =V 2
119
and when further diluting to a volume V 0 : c 0 c
V=V 0 and the standard deviation for the sample is: sr; c 0
p
sc =c 2
sV =V 2
sV 0 =V 0 2
120
When calibrating with one synthetic calibration sample
c S giving a signal Ys , the concentration of the unknown sample cX can be calculated as: cX YX =YS c S
121
2.1 Figures of merit of an analytical method
and: s 2
cX
qcX =qYX 2 s 2
IX
qcX =qYS 2 s 2
YS
qcX =qc S 2 s 2
c S
122
with: qcX =qYX q=qYX
YX =YS c S CS =YS
123
qcX =qYS q=qYS
YX =YS c S ÿ
YX c S =YS2
124
qcX =qc S q=qc S
YX =YS c S YX =YS
125
and: s 2
cX c S =YS 2 s 2
YX ÿ
YX c S =Ys2 2 s 2
YS YX =YS 2 s 2
c S
126
Also in the case of calibration by standard additions, the uncertainty in the amount added in the case of one standard addition can be taken into account. When cX Yx =
YXA ÿ YX cA
127
s 2
cX
qcX =qYX 2 s 2
YX
qcX =qYXA 2 s 2
YXA
qcX =qcA 2 s 2
cA
128
with: qcX =qYX q=qYX YX =
YXA ÿ YX cA YXA =
YXA ÿ YX 2 cA
129
qcX =qYXA q=qYXA YX =
YXA ÿ YX cA ÿYX =
YXA ÿ YX 2 cA
130
qcX =qcA q=qcA YX =
YXA ÿ YX cA YX =
YXA ÿ YX
131
and: s 2
cX YXA =
YXA ÿ YX 2 cA 2 s 2
YX ÿYX =
YXA ÿ YX 2 cA 2 s 2
YXA YX =
YXA ÿ YX 2 s 2
cA
132
Noise
The precision which can be obtained with a certain measurement system depends on the noise in the system. The latter may have dierent causes and accordingly, dierent types of noise can be distinguished as follows (for a detailed treatment, see Ref. [35]).
39
40
2 Spectrometric Instrumentation
. ``Fundamental noise'' or ``random noise'': this type of noise is statistically dis-
. .
tributed and its amplitude as a function of the frequency can be written as a sum of many sinusoidal functions. This type of noise is related to the corpuscular nature of matter or to the quantization of radiation, respectively, and cannot be completely eliminated. ``Non-fundamental noise'' also called ``¯icker noise'' or ``excess noise''. Here the sign or the magnitude can correlate with well-de®ned phenomena. ``Interference noise'', resulting in contributions at well-de®ned frequencies and mostly stemming from components in the system.
The last two types can often be eliminated by appropriate ®ltering. In a so-called noise spectrum (Fig. 8) the noise amplitude is plotted as a function of the frequency [see Ref. [36]). Here one distinguishes between ``white noise'',
Fig. 8. Noise power spectra commonly found in chemical instrumentation. (A): ``white'' noise; (B): flicker
1= f noise and (C): interference noise. (Reprinted with permission from Ref. [36].)
2.1 Figures of merit of an analytical method Fig. 9. Noise power spectra for ICP-OES using a Fasseltype torch. (a): 0±500 Hz and (b): 0±24 Hz; sample: 10 000 mg/mL Al solution; power: 1.8 kW, outer gas: 18 L/min Ar; nebulizer pressure: 3 bar Ar; sample uptake rate: 1 mL/min. (Reprinted with permission from Ref. [37].)
which spreads over all frequencies and is almost always ``fundamental'' in origin, whereas for the ``1=f noise'' the amplitude decreases with the frequency and it has non-fundamental causes. Also discrete noise bands with well-de®ned causes may occur. They may stem from the source and be caused by the gas ¯ow dynamics or contributions from the vacuum line etc. In the case of an inductively coupled plasma for example (Fig. 9) [37], there is a 1=f noise contribution stemming from large droplets which vigorously and in some way unpredictably evaporate in the plasma. Furthermore, there is a white noise contribution spread out over all frequencies stemming from the nebulization. There is also a noise contribution in the form of a broad noise band at ca. 300 Hz. This stems from the pulsing of the plasma which is related to the gas ¯ow dynamics. These pulsations can be visualized by high-speed photography [38]. It was also found, for example, that in the case of slurry nebulization, the contribution of the 1=f noise may increase. This can be shown by the 1=f noise increase obtained when using a direct powder introduction (Fig. 10). The frequency noise component depending on gas ¯ow was found to depend both on dierent gas ¯ows as well as on the power level. This is understandable, as both parameters in¯uence the size and hence the expansion and shrinkage of the plasma and their frequencies. In the case of a toroidal microwave discharge obtained in a microwave plasma torch, the frequency dependent noise stemming from the gas ¯ow dynamics could also be found at a frequency of 150±250 Hz (Fig. 11) [39]. When using hydride
41
42
2 Spectrometric Instrumentation Fig. 10. Noise power spectra and precision in ICP-OES for the introduction of (a): a 10 000 mg/mL Al solution, (b): a 10 000 mg/mL Al suspension and (c): the dry powder (Al2 O3 , Sumitomo, Japan). Fassel-type torch with outer gas flow: 18.1 L/min Ar; (a), power: 1.75 kW, nebulizer pressure: 3 bar Ar, sample uptake rate: 1 mL/min; (b), power: 1.7 kW, nebulizer pressure: 3 bar Ar, sample uptake rate: 1 mL/min; (c), power: 1.8 kW, aerosol gas flow: 1.3 L/min. (Reprinted with permission from Ref. [37].)
generation only a gaseous mixture enters the MIP, which might be the reason why the 1=f noise contribution is then almost absent. For glow discharge atomic spectrometry, the sputtering processes and analyte introduction and interaction with the plasma are so smooth that both 1=f noise and white noise are virtually absent and the noise spectra are con®ned to some frequency dependent noise components (Fig. 12) [40], which either stem from the power line (50 or 60 Hz) or may also be introduced by the vacuum line (10±20 Hz). The latter component can be identi®ed, as it decreases when throttling the main vacuum line. In plasma mass spectrometry noise power spectra were also found to re¯ect the
2.1 Figures of merit of an analytical method
Fig. 11. Noise power spectra of the He I 501.56 nm line collected at 1, 3 and 5 mm above the top of the torch for a 220 W He MPT plasma as source for OES. Gas flow of the He support gas: 2.2 L/min; gas flow of He
carrier: 0.2 L/min. (a): low-frequency noise spectra (0±50 Hz), (b): high-frequency noise spectra (0±1000 Hz). (Reprinted with permission from Ref. [39].)
gas ¯ows dynamics (Fig. 13) [41]. However, in ICP-MS they also were found to be severely in¯uenced by the cooling of the spray chamber. This is understandable as this will in¯uence the droplet evaporation during aerosol transport as well as the concentration of vapor present in the aerosol. In particular, it was found that the level of the white noise follows the spray chamber temperature.
43
44
2 Spectrometric Instrumentation Amplitude noise spectra for a matrix line in glow discharge source (GDS) atomic emission spectrometry. Steel standard sample 218A (Research Institute CKD, Czech Republic); i: 50 mA; argon pressure: 600 Pa; burning voltage: 900 V; 0.35 m McPherson monochromator; line: Fe I 371.9 nm. (a): Without needle valve between the vacumm pump and the GDS, (b): with needle valve between the pump and the GDS. (Reprinted with permission from Ref. [40].) Fig. 12.
Accordingly, noise spectra are a strong diagnostic tool to trace the sources of noise, and to study the limitations stemming from dierent sources (source stability, atom reservoir stability, detector used, etc.). For instance, it will be important to see if the noise of the detector is predominant, as this type of noise can be described by Poisson statistics where: s2 n
133
n is the number of events per second and s its standard deviation. Alternatively, it might be that the background noise of the source is much more important or that ``¯icker noise'' or ``frequency-dependant noise'' are predominant. In the last case overtones will often occur. Signal-to-noise ratios
The signal-to-noise ratios, which can be measured, depend on all aspects that in¯uence the signal magnitude and also on the magnitude and the nature of the
2.1 Figures of merit of an analytical method
Fig. 13. Noise power spectra in ICP-MS for ArO (56 dalton). Meinhard nebulizer, power: 1350 W, outer gas: 14 L/min, intermediate gas: 1.4 L/min, nebulizer gas: 1.0 L/min, sample uptake rate: 1.05 mL/min, doubly distilled water. (Reprinted with permission from Ref. [41].)
noise. Depending on the nature of the noise several measures can be taken so as to improve signal-to-noise ratios in a system (for a discussion in the case of atomic ¯uorescence spectrometry, see Ref. [42]). Provided both the source (primary source and/or atom reservoir, nebulizer, etc.) as well as the detector contribute to the noise, the signal-to-noise ratio
S=N in the case of a Poisson distribution for all types of noise is then given by: q 2 s2 S=N S=
sDC S
134
sDC is the noise of the dark current of the detector (shot noise) and sS is the noise of the source (radiation or ion source or in the case of absorption measurements the primary source). In the case of absorption measurements, e.g., three cases can be considered.
. A continuous source and a continuously measuring detector: here the signal is s proportional to the mean radiant density of the source Bave and the measurement s time t0 , the shot noise Ns (being sS2 ) is proportional to the signal Bave and t0 , and
45
46
2 Spectrometric Instrumentation
the background noise of the atom reservoir is proportional to the background b signal Bave and: q s s Bb S=N K Bave =
Bave ave
135
. A pulsed source and a continuously measuring detector: here the signal is again proportional to the mean radiant density of the source, but when the pulse width is tp , the pulse frequency is f and the pulse heigth is Bps , it can be written as b Bps f tp and the background noise is still given by Bave . Accordingly: q b
S=N 0 K Bps f tP =
BPS f tP Bave
.
136
When the background noise is predominant, the improvement in signal-to-noise b ratio is BPs f tP =Bave , whereas the improvement in the case where the shot s noise dominates is given by
BpS f tP =Bave 1=2 A pulsed source and a synchronously pulsed detector: here the signal again is proportional to BPs , tP and f, however, the background level also depends on tP and f and: q b
S=N 00 K BPS f tP = f tP
BPS Bave
137
or
S=N 00 K BPS
f tP 1=2 =
q b
BPS Bave
138
and the improvement as compared with S=N or
S=N 0 can be considerable b S b when the background noise predominates (Bave > Bave and Bave > BPS ). Then the ÿ2 ÿ4 improvement is f tP . As f tP can be 10 to 10 and with pulsed lasers and so-called boxcar integrators large improvements of even 10ÿ8 in signal-to-noise ratios are possible. Power of detection
According to Kaiser [43], the limit of detection of an analytical procedure is the concentration at which the analytical signal can still be distinguished from a noise level with a speci®c degree of uncertainty. In the case of a 99.86% uncertainty and provided the signal ¯uctuations of the limiting noise source can be described by a normal distribution, the lowest detectable net signal YL is three times the relevant standard deviation: YL 3s
139
In atomic spectrometry the net signal is determined from the dierence between a
2.1 Figures of merit of an analytical method
brutto signal, including analyte and background contributions, and a background signal, a propagation of error has to be applied as: s
q 2 2
sSignal sbackground
140
and provided the total signal and the p background have almost the same absolute standard deviation, a factor of only 2 must be introduced. For photoelectric measurements s
I, the standard deviation of the measured signals contains several contributions and: s 2
I sP2 sD2 sf2 sA2
141
Here,ps the noise of the photoelectrons. When the photon ¯ux is n, P represents sP A
n sD is the dark current noise of the photomultiplier and is proportional to the dark current itself. sf is the ¯icker noise of the source and is proportional to the signal and sA is the ampli®er noise resulting from electronic components. The last contribution can usually be neglected, whereas sf is low for very stable sources (e.g., glow discharges) or can be compensated for by simultaneous line and background measurements. As sD A ID one should use detectors with low dark current, then the photon noise of the source limits the power of detection. In many cases it is not the background signal from the source or the measurement system but blank contributions that limit the power of detection, the limiting standard deviation is often the standard deviation of the blank measurements and this value must be included in Eq. (139) [44]. From the calibration function the detection limit then is obtained as: p cL a 0
3 2 s
142
The detection limit thus is closely related to the signal-to-background ratio and the signal-to-noise ratio. It is the concentration for which the signal-to-background ratio p equals 3 2 times the relative p standard deviation of the background or at which the signal-to-noise ratio is 3 2. The signal-to-noise ratio itself is related to the types of noise occurring in the analytical system. From the knowledge of the limiting noise sources well-established measures can be taken during signal acquisition to improve the signal-to-noise, as discussed before, and accordingly, also the power of detection of a system. Provided the ¯uctuations of the relevant background or blank and the ¯uctuations near this p signal level are not identical, the detection limit cannot be calculated using the 3 2s criterion but must be de®ned considering the error of the ®rst and the error of the second kind [45] (Fig. 14). Here, the error of the ®rst kind stems from the scattering of the blank or background values around a mean value, whereas the error of the second kind follows the con®dence intervals of the calibration curve, which are a function of the concentration. In particular, when the background signal stems mainly from the detector dark current whereas the ¯uc-
47
48
2 Spectrometric Instrumentation
Fig. 14. Calculation of the detection limit (xD ) taking into account the error of the first
a and the error of the second
b kind. (Reprinted with permission from Ref. [45].)
tuations of the signals for the calibration samples are related to the source and the signal generation, the error of the ®rst and the error of the second kind may dier. A similar diculty may arise when the distribution of the background measurement values and/or the standard signals is not normal. This can be tested by the w 2 and the ``Skewedness and excess'' (S & E) test [46]. In the w 2 test [47] the test value is:
w2
k X
wi ÿ n yi 2 =
n yi i1
143
with wi the number of measurements in a class i normalized to 1 as Swi 1 and yi the theoretical frequency in a class i (in tables for a normal distribution) again normalized to 1. The distribution must thus be investigated to establish whether the n measurements comply with a normal distribution with mean m m and standard deviation s s calculated from the measurements. The theoretical normal distribution for m and s can be calculated from: p p P
u 1=
2 p s expfÿ
x ÿ m= 2 sg dx
144
2.1 Figures of merit of an analytical method
Which theoretical frequency is found for the dierent classes is thus calculated. Every term
wi ÿ n yi 2 =
n yi is quadratic in a normally distributed value. The sum of k squares of normally distributed values is w 2 distributed with k degrees of freedom (k is the number of classes). As m and s are calculated from the data and Syi 1, three degrees of freedom have to be subtracted. The resulting test value can be compared with tabulated values for a given con®dence (e.g., 95%) and number of degrees of freedom and this can be done for the values or for a given function of them. When the latter is a logarithmic one for example, it can be controlled if the distribution of the data is normal or logarithmically normal. Whereas the w 2 test only gives information on the distribution of measured values over dierent classes the S & E test gives further information on whether the extreme values are not only systematically too high or too low and whether or not there is an excess of values around the mean. For a number of measured values one determines the parameter Q [46]: Q S
xi ÿ x 3 f i =
s 3 N
145
Here Q is a measure for the skewedness, x is the mean, xi is a measured value, f i is the number of measured values of a magnitude xi , s is the standard deviation and N the total number of measurements. One can further determine: e S
xi ÿ x 4 f i =
s 4 N ÿ 3
146
For a normal distribution both values should simultaneously be zero. With a ®nite number of N: p S f N
N ÿ 1=
N ÿ 2g Q
147
and E
N ÿ 1=
N ÿ 2
N ÿ 3
N 1e 6
148
In the case of a normal distribution the values of S and E should also have normal distributions with relative standard deviations: p sS G 6N
N ÿ 1=
N ÿ 2
N 1
N 3
149
and: q sE G 24N
N ÿ 1 2 =
N ÿ 3
N ÿ 2
N 3
N 5
150
If 66% of the experimental S and E values lie within the GsS or GsE limits it can be controlled. If the values scatter according to a normal distribution.
49
50
2 Spectrometric Instrumentation Fig. 15. Limit of determination and limit of detection.
Limit of determination
The limit of determination indirectly relates to the limit of detection. It is the concentration level from where a determination can be performed with a preset precision. The de®nition can be understood from the con®dence lines at each side of the calibration curves (Fig. 15), which diverge both at lower concentrations, as a result of sample inhomogeneities or noise magnitude, as well as at large concentrations, as a result of deviations from linearity in the calibration or source instabilities. When for a sample with a given concentration a number of signal measurements n are obtained with a standard deviation sS , the probability P of ®nding the value between the limits ms G uxsS will be the ratio of the integral under the whole population from ÿy < x < y to the integral within the limits m ÿ u sS < x < m u sS . On the contrary, for a single value, P will be the probability within which limits it deviates from the true value mS as a result of statistical ¯uctuations. For the mean mS of n measurements: p p mS ÿ u
P sS =
n < x < mS u
P sS =
n
151 p The analytical value then diers with a probability P by less than Gu
P sS =
n from the value m and this is the con®dence interval. It is the interval belonging to the mean value m in which with a preset probability one can assume a certain analytical error. This interval changes with the concentration and this determines the con®dence limits. The limit of determination is often pragmatically de®ned as being found at a concentration being 5±6 times the detection limit. Indeed, it is at this concentration level that the full precision of the analytical method is likely to be realized. Limit of guarantee of purity
The guarantee of purity
cG is the highest concentration that can be present in a sample, without being able to obtain an analytical signal, which can be dieren-
2.2 Optical spectrometers
tiated with a given probability from the limiting noise level. When excluding errors from sample heterogeneity, cG is given by: cG 2cL
152
This analytical ®gure of merit is of special importance especially in the control of high-purity substances and materials for microelectronics, and also in food analysis and other disciplines. In atomic spectrometry, the analytical signals measured often include contributions from non-spectrally resolved features stemming from constituents other than the analyte (e.g. matrix constituents). These contributions are known as spectral interferences. They can be corrected for by subtracting their contributions to the signal, which can be calculated from the magnitude of the interference and the concentration of the interferent. A special type of spectral interference is that which in¯uences the background signal on which the analyte signals are superimposed. For this type of interference a number of corrections are known. The degree of freedom from interferences is an important ®gure of merit for an analytical method.
2.2
Optical spectrometers
In optical atomic spectrometry the radiation emitted by the radiation source or the radiation which comes from the primary source and has passed through the atom reservoir has to be lead into a spectrometer. In order to make optimum use of the source, the radiation should be lead as complete as possible into the spectrometer. The amount of radiation passing through an optical system is expressed by its optical conductance. Its geometrical value is given by: G0
A
B
2
cos a1 cos a2 dA dB=a12
2 A
A B=a12
153
154
dA and dB are surface elements of the entrance and the exit apertures, a12 is the distance between them. a1 and a2 are the angles between the normals of the aperture planes and the radiation. When n is the refractive index of the medium, the optical conductance is given by: G G0 n 2
155
The radiant ¯ux through an optical system is given by: ftBG
156
t is the transmittance determined by re¯ection or absorption losses at the dierent
51
52
2 Spectrometric Instrumentation
optical elements and B is the radiant density of the source (in W/m 2 sr). For an optimal optical illumination of a spectrometer, the dispersive element, which serves to provide the spectrum, should be fully illuminated so as to obtain full resolution. However, no radiation should bypass the dispersive element, as this would cause stray radiation. Furthermore the optical conductance at every point of the optical system should be maximum. 2.2.1
Optical systems Illumination of the spectrometer
The type of illumination system with which these conditions can be full®lled, will depend on the dimensions of the source and the detector, the homogeneity of the source, the need to ®ll the detector homogeneously with radiation, the distance between the source and the entrance aperture of the spectrometer and on the focal length of the spectrometer. In conventional systems lenses as well as imaging mirrors are used. In the case of lenses, the lens material is important. Here glass lenses can only be used at wavelengths above 330 nm. In the case of quartz, radiation with wavelengths down to 165 nm can still be transmitted. However, as a result of the absorption of shortwavelength radiation by air, evacuation or purging of the illumination system and the spectrometer with nitrogen or argon is required when lines with wavelengths below 190 nm are measured. At wavelengths below 160 nm, evacuation and the use of MgF2 or LiF optics is required. With lenses mainly three illumination systems are of use. Imaging on the entrance collimator. A lens is placed immediately in front of the entrance slit and should image the relevant part of the radiation source on the entrance collimator (Fig. 16A). This has the advantage that the entrance slit is homogeneously illuminated, however, stray-radiation may easily occur inside the spectrometer. The distance between the source and the entrance slit (a) is given by the magni®cation required, as:
x=W a=f k
157
x is the width (or diameter) of the source, W is the width of the entrance collimator and f k its focal length. The f-number of the lens than is given by: 1=f 1=a 1=f k
158
. Example When a grating spectrometer with a focal length of 1.2 m and collimator dimensions of 55 55 mm 2 must be illuminated for a wavelength of p200 nm with a radiation source of 7 mm in diameter the enlargement for 55 2 77:8 mm is
2.2 Optical spectrometers
Fig. 16. Illumination of the optical spectrometer with lenses. (A): Imaging on the entrance collimator, (B): illumination with intermediate image, (C): imaging on the entrance slit.
b 77:9=7 and the distance d between the entrance slit of the spectrometer and the source should be b f k =d or d 108 mm, with f k the focal length. According to Eq. (158): 1=f 1=1200 1=108 and therefore a lens with a focal length of 99 mm at 200 nm is required. It must be considered that the focal length of a lens depends on the wavelength as: f
l1 =f
l2 n1 =n2
159
n1 and n2 are the refractive indices of the lens material at the respective wavelengths. For quartz, e.g., a factor of 0.833 has to be applied when passing from the Na 583 nm D-line to 200 nm. Illumination with an intermediate image. Here a ®eld lens is used to produce an intermediate image on a diaphragm. To illuminate the collimator mirror fully, the appropriate zone can be selected with the aid of a lens placed immediately in front of the exit slit. A third lens is used to illuminate the entrance slit homogeneously (Fig. 16B). The magni®cation is then divided over all 3 lenses, thus chromatic aberrations
53
54
2 Spectrometric Instrumentation
are minimized, but the set-up is highly in¯exible. The distances between the three lenses must be chosen so as to achieve the respective magni®cations. Accordingly:
x=I a1 =a2
160
D=sh a2 =a3
161
and I=W a3 =f k
162
I is the diameter of the intermediate image, D the diameter of the ®eld lens, sh the entrance slit height and a1 ; a2 and a3 are the distances between the respective lenses. Furthermore: a1 a2 a3 A
163
is usually ®xed because of the construction of the system, and 1=f 1 1=a1 1=a2
164
1=f 2 1=a2 1=a3
165
1=f 3 1=a3 1=f k
166
give the f-numbers of the respective lenses. One parameter (e.g. the width of the intermediate image) can be freely selected as x, W, sh , A and f k are ®xed.
. Example When a radiation source with a diameter of 4 mm is to be coupled with a 1 m monochromator with an entrance collimator of 50 mm width, an entrance slit heigth of 20 mm and a total distance between the radiation source and the entrance slit of 1 m, for an intermediate image I of 10 mm Eq. (162) gives: a3 200 mm. Eq. (166) gives f 3 . According to Eq. (160): 2:5 a1 a2 and from Eq. (163) 3:5 a1 800 mm, so that: a1 230 mm and a2 590 mm; f 1 is given by Eq. (164) and from Eq. (161): D 50 mm. Image on the entrance slit. With the aid of one lens this is also possible. Here the structure of the source appears on the entrance collimator and on the detector. This allows spatially-resolved line intensity measurements to be made when a detector with two-dimenional resolution such as a photographic emulsion or an array detector (see Section 2.2.2) are employed. This type of imaging is often used for diagnostic studies. Quartz ®ber optics. These have been found to be very useful for radiation transmission. Small lenses are used so as to respect the opening angle of the ®ber. This depends on the refractive index of the material, which because of optical transmission
2.2 Optical spectrometers
Fig. 17. Use of optical fibers for the illumination of an optical spectrometer. a: Opening angle of the fiber; d: lens diameter, D: diameter of the fiber, x: dimension of the zone in the radiation source to be selected, a1 : distance between lens and radiation source, a2 : distance between fiber entrance and lens.
reasons is usually quartz, and is often of the order of 30±40 . A typical illumination of a spectrometer with an optical ®ber (Fig. 17) uses a lens (diameter d) for imaging the source on the ®ber. Fibers or ®ber beams with a diameter of D 600 mm are often used. Then the magni®cation
x=D as well as the entrance angle tga
d=2=a2 and the lens formula determine the f number of the lens and the diameter d. At the exit of the ®ber, a lens is used so as to allow the radiation to enter the spectral apparatus, without causing stray radiation. With quartz ®bers it is easier to lead the optical emission into the spectral apparatus, however, it should be mentioned that the transmittance decreases seriously below 220 nm. This may give rise to detector noise limitations for analytical lines at lower wavelengths. As an alternative to lenses and optical ®bers so-called light pipes which use total re¯ection of radiation may also be used. Spectrometers
Spectral apparatus are used to produce the spectrum or to isolate narrow spectral ranges, without further spectrum deconvolution. In dispersive spectral apparatus, the spectrum is produced with a prism or a diraction grating. In non-dispersive spectral apparatus, spectral areas are isolated from a radiation beam, without any further spatial deconvolution, by re¯ection, interference or absorption in an interferometer or a ®lter monochromator, respectively. The ®lter is now only of use in ¯ame emission spectrometry. A dispersive spectral apparatus contains an entrance collimator, a dispersive element and an exit collimator (for an example discussion, see Refs. [48, 49]).
55
56
2 Spectrometric Instrumentation
With an entrance collimator a quasi parallel beam is produced from the radiation coming through the entrance aperture, which has a width se and a heigth he . The entrance collimator has a focal length f k and a width W. The diraction slit width
s0 and the diraction slit heigth
h0 are the half-widths given by: s0 l f =W
and h0 l f =h
167
The entrance aperture dimensions should not be made smaller than s0 and h0 , as then diraction will limit the resolution that can be obtained. The value of f =W is a measure of the amount of radiation energy entering the spectral apparatus. The exit collimator images the monochromatized radiation leaving the dispersive element on the exit slit. Here we have a series of monochromatic images of the entrance slit. In the case of one exit slit in which one line after another can be isolated by turning the dispersive element, we have a monochromator. In a polychromator the dispersive element is ®xed and there are many exit slits placed at locations where monochromatic images of the entrance slit for the lines of interest are obtained. They are often on a curved surface with a radius of curvature R (Rowland circle). Here simultaneous measurements of several lines and accordingly simultaneous multielement determinations are possible. In the case of a spectrographic camera, the lines are focussed in a plane or on a slightly curved surface, where a detector with two-dimensional resolution can be placed. With such a detector (photoplate, diode array detector, etc.), part of the spectrum over a certain wavelength range as well as the intensities of the signals, eventually at several locations in the source, can be recorded simultaneously. The energy per unit of surface on the detector is given by the irradiance: E f cos a=A
168
a is the angle between the surface of the radiation detector (A) and the incident radiation. Virtually the only dispersive elements now used are diraction gratings. Prisms are used only as predispersers. Distinction can be made between plane and concave gratings, of which the latter have imaging qualities. One also has to distinguish between mechanically ruled and holographically ruled gratings. As a result of the pro®le of the grooves, they have a fairly uniform radiant output over a large spectral area. Mechanically ruled gratings always have a so-called blaze angle and accordingly a blaze wavelength where the radiant energy delivered is at a maximum. In modern spectrometers re¯ection gratings are usually used. As a result of interference, a parallel beam of white radiation incident at an angle f1 with the grating normal is dispersed and at an angle f2 , radiation of wavelength l is diffracted (Fig. 18), according to the Bragg equation: sin f1 sin f2 ml=a
169
m is the order, f2 is the angle the diraction beam of wavelength l makes with the grating normal and a 1=nG is the grating constant, where nG is the number of
2.2 Optical spectrometers Fig. 18. Spectral dispersion at a diffraction grating. a: Grating constant, b: Blaze angle, f1 : angle of incident radiation, f2 : angle of diffracted radiation with wavelength l.
grooves per mm. When B is the width of the grating, the total number of grooves, N B nG , determines the theoretical resolving power R0 as: R0 B n G m
170
The angular dispersion can be obtained by dierentiation of Eq. (169) with respect to l: df2 =dl m=
a cos f2
171
The angular dispersion and the theoretical resolving power R0 are related as: df2 =dl R0 =Ds0 , where Ds0 is the width of the monochromatic beam where it exits the dispersive element. The reciprocal linear dispersion is given as: dx=dl
dx=df2
df2 =dl
f k =cos y 0
m=a cos f2
172
where y 0 is the angle between the plane of the detector and the direction of the sorting beam. The spectral slit width of the spectrometer is given by: Dl
dl=dx se
173
where se is the entrance slit width. The form and the depth of the grooves determine the intensity distribution in the spectrum. The angle for the maximum intensity is the so-called ``blaze'' angle b. The blaze wavelength for the order m can be calculated as: sin b lB m=2a cos
e=2 where
f1 ÿ f2 e.
174
57
58
2 Spectrometric Instrumentation Fig. 19. Important optical mountings for optical spectrometers with a plane (A: Ebert, B: Czerny± Turner) and a concave (C: Seya±Namioka, D: Paschen±Runge) grating. m: Mirror, se : entrance slit, sa : exit slit, gP : plane grating, gc : concave grating.
In a stigmatic spectral apparatus the height and the width of the slit are imaged in the same plane. In an autocollimation spectral apparatus the rays before and after the spectral dispersion pass through the same optical elements. With a polychromator all optical elements are usually ®xed, whereas in monochromators the desired wavelength is normally brought into the exit slit by turning the grating around its axis. With a plane grating several mountings (Fig. 19) can be used. In the Czerny± Turner mounting, two spherical mirrors with slightly dierent focal lengths are
2.2 Optical spectrometers
positioned at sligthly dierent angles. In this way spherical aberration can be adequately corrected for. This mounting is used for high-luminosity monochromators with fairly short focal lengths and highly dispersive gratings (more than 1800 grooves/mm). With the Ebert mounting, there is only one mirror serving as the entrance and the exit collimator. Accordingly, aberrations occur and the use of curved slits is necessary, so as to be able to realize the highest resolution and optical conductance. The Fastie±Ebert mounting is used for large spectrographs, where the entrance collimator and the exit collimator plane lie above each other. The spectrum is focussed in a plane and the f =W value can be as low as 1/30, which is very low and may require the use of a cylindrical lens to increase the irradiance on the detector. Photoplates, ®lms and photo multipliers are used as detectors. Normally, gratings are used at low orders
m < 4 and they have a small grating constant (1/3600 mm < a < 1=300 mm). The dierent orders can be separated by using special photomultipliers. For instance, with a solar-blind photomultiplier only radiation with a wavelength below 330 nm can be detected. This allows separation of the 1st order radiation at 400 nm from the 2nd order radiation at 200 nm. This can, for example, be applied in polychromators to double the practical resolution. Apart from gratings used at low orders, the so-called echelle gratings can be used. Their groove density is low (a up to 1/100 mm) but therefore order numbers of up to 70 can be used [50]. Here, the orders overlap and must be separated by using a second dispersive element (e.g. a prism) either with its axis parallel to the one of the echelle grating or in so-called crossed-dispersion mode. In the latter case the spectrum occurs as a number of dots with a height equalling that of the entrance slit. The optical conductance is given by: f
Bl t s h W H cos b=f 2
175
where Bl is the spectral radiance, t is the transmittance of the optics, W is the width and H the height of the grating. Echelle spectrometers often use an Ebert mounting and allow a high practical resolution to be obtained at low focal length
f . Therefore, they are thermally very stable and are used both with monochromators as well as polychromators. As two-dimensionally resolving array detector technology [vidicons, charge coupled devices (CCD) now also with image intensi®ers] with high spatial resolution and high sensitivity as well as signal-to-noise ratios, have become available, such Echelle spectrometers have become very useful. They allow the whole spectrum to be collected on one detector of limited dimensions (e.g. 1 in 1 in) and enable ¯exibility in selecting any analytical wavelength. The practical resolution then is ®nally limited by the pixel dimensions (e.g. 1000 pixels in one direction) and blooming as well as cross talk. In modern instrumentation a separate spectrometer is often built-in to zoom into remote spectral areas, as required, for example, for the alkali sensitive lines (Fig. 20). Mountings with a concave grating are also used. Here the radius of the grating determines the so-called Rowland circle. In the direction of dispersion the spectral lines are focussed on the Rowland circle and are monochromatic images of the
59
60
2 Spectrometric Instrumentation Fig. 20. Echelle mounting in optical set-up for CCD-ICPOES (Optima)(PerkinElmer Inc.).
entrance slit. In the Paschen±Runge mounting, the grating, the entrance slit and all the exit slits are ®xed on the Rowland circle. This mounting is most often used in large simultaneous polychromators with photoelectric detection. The number of detectors used here is often more than 30 and they are usually used in several orders. In this mounting, and when using conventional photomultipliers behind an exit slit, background acquisition is possible in a sequential way by slewing the entrance slit. This can be done very reproducibly by positioning the entrance slit in a stepping-motor driven unit drawn by a spring against a plate. This may provide for dynamic background correction as well as for sequential atomic emission spectrometry when working with time-stable sources. As the price of individual CCDs has come down, the whole relevant Rowland circle can be covered with CCDs. By positioning them below and above the plane of the exiting radiation and providing mirrors placed at 45 , coverage of the whole spectrum, without missing any wavelength areas, becomes possible and a fully simultaneous spectrometer with data acquisition possibilities for analytical lines and background measurements at any wavelength can be realized (Fig. 21). Such
Fig. 21. Optical set-up for
CCD-ICP-OES (CIROS) (Spectro Analytical Instruments GmbH).
2.2 Optical spectrometers
instruments are now commercially available, so the capacity of a photographically recording spectrograph for obtaining simultaneous information has now been realized in atomic spectrometry with however, the ease, precision and time-resolution of a photoelectrical system. In the Eagle mounting, the grating can be turned around its axis and moved along the direction of the radiation at the same time. The center and the focal plane change when switching to other wavelengths. This mounting is mechanically easy to construct, the astigmatism is low and the system can be easily housed in a narrow tank. It is often used in vacuum monochromators. The directions of the entering and the dispersed radiation beam in the Seya±Namioka mounting are constant and the angle between them is 70 30 0 . The entrance and exit slit as well as the grating axis are ®xed and the wavelength selection at the exit slit is performed by rotating the grating. At large angles a serious defocussing occurs and the aberrations are larger than in the case of the Eagle mounting. The Seya±Namioka system is also often used in vacuum monochromators. With most sequential spectrometers switching from one line to another by turning the grating is usually fairly quick. From Eq. (169) it can be calculated that, as the spectral lines have widths of 1±3 pm, the angle selection must be performed very accurately for wavelength selection. Indeed, for a grating with a 1=2400 mm and a line width of 1 pm the grating must be positioned with an accuracy of 10ÿ4 degrees. Therefore, one can scan the pro®le of the line or realize random access by using angle encoders. Computer-controlled stepper motors are used for turning the grating. Their angular resolution is often above 1000 steps per turn. In the case of a socalled sinebar drive, the number of steps performed by the motor is directly proportional to the wavelength displacement. Here a ball-bearing runs along while being pressed against a block moving on the spindle driven by the stepper motor. A further system uses a Paschen±Runge spectrometer with equidistant exit slits and a detector moving along the focal plane. Fine adjustment of the lines is then done by computer-controlled displacement of the entrance slit. 2.2.2
Radiation detectors
Photographic emulsions and photoelectric detection devices can be used as detectors for electromagnetic radiation between 150 and 800 nm. Among the photoelectric devices, photomultipliers are the most important but new solid state devices have become a useful alternative. Photographic emulsions
These were frequently used for analytical spectrography. They allow the whole spectrum to be recorded simultaneously and, accordingly, for multielement survey analysis they have a high information capacity. However, their processing is long, the precision obtainable is low and they do not allow on-line data processing. Ac-
61
62
2 Spectrometric Instrumentation
cordingly, the quantitative treatment of photographically recorded spectra remained only of limited importance for analytical use and the interest in photographic detection is now limited to research where documentation of the whole spectrum is required. Therefore, it will be treated only brie¯y here. When a photographic emulsion is exposed to radiation a blackening is produced, which is a function of the radiant energy accumulated during the exposure time. It is given by: S log
1=t ÿlog t log f0 =f
176
t is the transmission, f0 is the ¯ux through a non-irradiated part of the emulsion and f is the ¯ux obtained for an exposed part of the emulsion, when a white light beam is sent through the emulsion in a so-called densitometer. The emulsion characteristic for a selected illumination time gives the relationship between the logarithm of the intensity (Y log I ) and the blackening and has an S-shape (Fig. 22). Originally, only the linear part of the characteristic was used for quantitative work. However, soon transformations were described, which also allowed this characteristic to be linearized at low blackenings. The P-transformation has been
Fig. 22. Emulsion characteristic (S) and
transformation functions (P, W ).
2.2 Optical spectrometers
described by Kaiser [51] as: P g
Y ÿ Y0 S ÿ k D
177
where D log1=
1 ÿ t. The Seidel function is given by: W log
10 S ÿ 1
178
and thus: P
1 ÿ kS kW
179
Y0 is the inertia of the emulsion and the constants k and g describe the properties of the emulsion but they also depend on the densitometer. Y0 in most cases is not known as DP or DY values are usually used. g tan a and is the contrast of the emulsion. The emulsion can be calibrated with the aid of a step ®lter which is placed in front of the spectrometer. When the ratio of the intensities passing through two sectors of the ®lter is given by: DYm log
Im
1 =Im
2
180
the transformation equation becomes: DS g DYm k DD
181
where DD is the independent variable and DS the function. A linear regression for a number of line pairs gives k as slope and g DYm as intersection with the x-axis [52]. Y log I values for a given blackening then can be calculated as: Y k=g ÿ log
1 ÿ t ÿ 1=g log t Y0
182
and: DY ÿk=g DD 1=g DS
183
Errors on the intensity measurement stem from the graininess of the emulsion. According to Siedentopf [53], the standard deviation for measurements of small blackenings is given by: s
S
p 0:5
a=F S
184
a is the surface of a grain and F the surface of the densitometer slit (e.g. 10 mm 1 mm).
63
64
2 Spectrometric Instrumentation
. Example: For the Kodak SA 1 emulsion at a photometrically measured surface of 10 mm 1 mm and S A 0:2 the standard deviation sS 0:003, from which it can be calculated according to Eq. (184) that the grain diameter is about 1 mm. From s
S one can calculate the relative standard deviation of the intensities sr
I according to the rule of the propagation of errors as: sr
I ln 10 s
S 1=gf1 k=
10 S ÿ 1g
185
Typical values for g are 1±2.5 and for k 0.5±2. Non-sensitized emulsions can be used in the wavelength range 220±450 nm. At longer wavelengths green- or redsensitized emulsions can be used. They have a large g and accordingly a small dynamic range. At UV or VUV wavelengths the gelatine layer absorbs, so emulsions with less gelatine have to be used, which are fairly insensitive, or the radiation must be transformed to longer wavelengths with the aid of a scintillator, such as sodium salicylate. Whereas in optical atomic spectrometry photographic emulsions are now rarely used, they are made use of in mass spectrometry with spark sources in particular. Here they are very useful for survey semi-quantitative analyses, and they still are of great importance for routine high-purity metals analysis and semiconductor material analysis. For the precise and quick measurement of radiant intensities but also of ion currents photoelectric techniques are used almost exclusively. The measurements can easily be automated. The detectors used are photomultipliers, electron multipliers, photodiode array detectors, camera systems and other solid state detectors. Photomultipliers
These are most commonly used for precise measurements of radiant intensities. The photomultiplier has a photocathode and a series of so-called dynodes. Its properties and their theoretical basis have been well treated in Ref. [54]. The radiation releases photoelectrons from the photocathode as a result of the Compton eect. After acceleration these photoelectrons impact on the dynodes so that a large number of secondary electrons are produced, which after successive action on a number of dynodes leads to a measurable photocurrent at the anode
Ia . For a photon ¯ux Nf passing through the exit slit of the spectrometer and impacting on the photocathode, the ¯ux of photoelectrons produced
Ne is: Ne Nf Q:E:
l
186
Q:E: is the quantum eciency (up to 20%). The cathode current Ic is given by: Ic Ne e
e 1:9 10ÿ19 A s and:
187
2.2 Optical spectrometers
Ia Ic y
V
188
y
V is the ampli®cation factor (up to 10 6 ). The dark current Id is the Ia measured when no radiation enters the photomultiplier (about 1 nA). Q:E:, y
V and Id are characteristics of a certain type of photomultiplier. The photocurrents are lead to a measurement system. Firstly, Ia is fed to a preampli®er containing a high-ohmic resistor. The voltage produced is fed to an integrator. Each measurement channel mostly contains a preampli®er and an integrator. By selection of the resistors and capacitors, a high dynamic measurement range can be realized. When Ia accumulates in the capacitor with capacitance C (in Faradays) during an integration time t the potential obtained is: U Ia t=C
189
This voltage is digitized in an A/D convertor. The dynamic range of measurement is then ®nally limited at the lower side by the dark current of the photomultiplier and at the high side by saturation at the last dynode. The linear dynamic range of a photomultiplier is usually more than 5 decades of intensity. In the case of a photomultiplier the relative standard deviation of the measured intensities sr
I is determined by the noise of the photoelectrons and the dark current electrons. The quantum eciency and the ampli®cation at the dynodes are also important. Approximately: sr
I A 1:3
Ne; d ÿ1=2
190
where Ne; d is the total number of photoelectrons and dark current electrons in the measurement time t. The dark current should be so low that the lowest intensity signal to be measured (e.g. the spectral background intensity) produces a photoelectron ¯ux near which the dark current is negligible. Then the noise of the photoelectrons is predominant. Photomultipliers for the spectral range between 160 and 500 nm usually have an S20 photocathode, where Q:E:
l A 5±25%, the ampli®cation factors are 3±5 at dynode voltages between 50 and 100 V and there are 9±15 dynodes mounted in head-on or side-on geometry. For selected types the dark current may be below 100 photoelectrons/s (Ia < 10ÿ10 A). Red-sensitive photomultipliers have a bialkali photocathode. Their dark current is higher, but can be decreased by cooling. For wavelengths below 160 nm photomultipliers with MgF2 or LiF windows are used. So-called ``solar-blind'' photomultipliers are only sensitive for short-wavelength radiation (e.g. below 330 nm).
. Example: With an ICP atomic emission spectrometer, for the spectral background intensity at 250 nm a signal of 104 digits is obtained with a 10 V A/D convertor and 12 bit resolution (4096), thus representing a voltage of 0.254 V. When a capacitor with 1:11 10ÿ8 F and a measurement time of 10 s are used, this corresponds to a
65
66
2 Spectrometric Instrumentation
photomultiplier anode current of 2:8 10ÿ10 A. For the EMI 9789 QB photomultiplier with a voltage of 780 V, the ampli®cation is 1:11 10 5 , which means that the photocathode current is 2:5 10ÿ15 A or 1:5 10 4 photoelectrons/s. As the quantum eciency is 17% this represents a photon ¯ux at the photomultiplier of 8:7 10 4 photons/s. When the dark current of the photomultiplier is 0.1 nA, this means that the ¯ux of the dark current electrons is below 1000 electrons/s and that even with the spectral background at this wavelength, the noise of the photoelectrons is still dominant. Solid-state detectors
In optical atomic spectrometry solid-state detectors have become more and more important. They are multichannel detectors and include vidicon, SIT-vidicon, photodiode array detectors and image dissector tubes [55]. Furthermore, charge-coupled devices (CCD) and charge injection devices (CID) [56] have now been introduced. Photodiode arrays (PDA) may consist of matrices of up to 512 and even 1024 individual diodes (e.g. Reticon) which individually have widths of 10 mm and heights of up to 20 mm. The charge induced by the photoelectric eects gives rise to photocurrents which are sequentially fed into a preampli®er and a multichannel analyzer. Therefore, they are rapid sequential devices. They have the advantage that memory eects due to the incidence of high radiant densities on individual diodes (lag) as well as cross talk between individual diodes (blooming) are low. Their sensitivity in the UV and the VUV spectral range is low as compared with photomultipliers. Nevertheless, at present they are of interest for atomic emission spectrometry in particular as they can be coupled with microchannel plates giving an additional ampli®cation. This plate can be considered as being built up from a semiconductor plate in which parallel channels of low diameter at a high density have been provided. When a high potential is applied between both sides of the plate each tube acts as a single electron multiplier (Fig. 23). The signal-to-noise ratios
Fig. 23. Principle of the microchannel-plate/diode-array camera.
2.2 Optical spectrometers
Fig. 24. Early SIT vidicon system. (Courtesy of SSR Instruments, Inc.)
of photodiode arrays can be considerably improved by cooling them with the aid of a Peltier element or liquid nitrogen. Commercial instrumentation is now available (e.g. EG&G OMA 4; Spectroscopy Instruments Inc.) and has replaced silicon intensi®ed target (SIT) vidicon detectors for most applications. In vidicons the charge is integrated into the individual pixels and the simultaneously collected information is read out sequentially with the aid of an electron beam. When using an electrostatic beam intensi®er (Fig. 24), in a number of cases these devices are quite sensitive and also permit two-dimensionally resolved measurements. However, they suer from blooming. This means that in pixels near places receiving a high radiance a charge decay and thus a signal is also induced. Furthermore, lag is present, through which an irradiated pixel has some hysteresis eects, deteriorating subsequent signal acquisition. Photodiode array detectors have been used successfully, for example, in a segmented echelle spectrometer (Plasmarray, Leco Co) [57]. Here, spectral segments around the analytical lines are sorted by primary masks subsequent to spectral dispersion with an echelle grating and are detected after a second dispersion on a diode array. Accordingly, more than ten analytical lines and their spectral background contributions can be measured simultaneously. An example of this can be shown in the case of a glow discharge lamp (see Fig. 25) [58], however, it should be mentioned that restrictions arise due to detector noise limitations at low wavelengths. Image dissector tubes (Fig. 26) make use of an entrance aperture behind the photocathode, by which the photoelectrons stemming from dierent locations of the photocathode can be scanned and measured after ampli®cation in the dynode train, as in a conventional photomultiplier. Although used in combination with an echelle spectrometer with crossed dispersion for ¯exible rapid sequential analyses [59], these systems have not had any commercial breakthrough. This might be due to the limited cathode dimensions but also to stability problems. Charge transfer devices (CTD) [60] are solid-state multichannel detectors and integrate signal information as light strikes them, just as a photographic ®lm does. An individual detector in a CTD array consists of several conducting electrodes overlaying an insulating layer that forms a series of metal oxide semiconductor (MOS) capacitors. The insulator separates the electrodes from a doped silica region used for photogenerated charge storage. In this n-doped silicon, the major current carrier is the electron, and the minor carrier is the hole. When the electrodes are negatively charged with respect to the silicon, a charge inversion region is created
67
68
2 Spectrometric Instrumentation
Fig. 25. Spectral-segmented spectrum of a glow discharge [58]. Grimm-type glow discharge lamp with floating restrictor (8 mm diameter): 50 mA, 3.5 torr, 2 kV [42]; Plasmaarray spectrometer (LECO); steel sample: 217 A (Research Institute, CKD, Prague, Czech Republic).
under the electrodes. This region is an energetically favorable location for mobile holes to reside in. The promotion of an electron into the semiconductor conduction band, such as by the absorption of a photon, creates a mobile hole that migrates to and is collected in the inversion region. The amount of charge generated in a CTD detector is measured either by moving the charge from the detector element, where it is collected, to a charge-sensing ampli®er, or by moving it within the detector element and measuring the voltage change induced by this movement. These two modes of charge sensing are employed by the charge-coupled device (CCD) and the charge injection device (CID), respectively. In the CCD, the charge from each detector element is shifted in sequence to an ampli®er located at the end of a linear array or the corner of a two-dimensional array of detector elements. Charge is transformed to the on-chip ampli®er by sequentially passing the charge packets
Fig. 26. Image dissector tube. (a): Photocathode, (b): accelerating electrode (mesh), (c): deflection coils, (d): aperture, (e): electron multiplier.
2.2 Optical spectrometers Fig. 27. Charge-coupled device (CCD) detector. Layout of a typical three-phase CCD. Photogenerated charge is shifted in parallel to the serial register. The charge in the serial register is then shifted to the amplifier. (Reprinted with permission from Ref. [63].)
through a row of detectors. The organization of a two-dimensional, three-phase CCD array is illustrated in Fig. 27. Columns are read in parallel. With each reading, all of the charge in the imaging array is shifted towards the serial register by one row, whereas charge from the row adjacent to the serial register is transferred into the serial register. Once this done, the charge packets are shifted sequentially to the on-chip ampli®er. Modern CCDs are capable of thousands of transfers with almost immeasurable charge transfer losses. By their ability to transfer the photogenerated signal from the photoactive element to an on-chip ampli®er, a very high signal-tonoise ratio can be obtained, as compared with other detectors (see Fig. 28) [60]. Whereas in photodiode arrays the limiting noise is read noise, this is not the case here with CTDs. With CIDs the charge information can be read out nondestructively further increasing the signal-to-noise ratios. With CCDs binning can be applied, which is the process of summing the charge contained in multiple elements on the detector before sensing the total charge [61]. CTDs which can be used from the soft x-ray region to the IR are now commercially available and the signal-tonoise ratios achievable have become as high as with photomultipliers. The dynamic range has risen up to 7 orders of magnitude in the case of a CID [62].
Fig. 28. Calculated signal-tonoise ratios for different detectors at a photon flux of 10 photons/s for observation times ranging from 1 to 100 s at a wavelength of 600 nm. (Reprinted with permission from Ref. [60].)
69
70
2 Spectrometric Instrumentation Tab. 3. Evaluation of some new radiation detectors for atomic emission spectrometry.
Detector
Dimensions
Spectral range
Sensitivity versus photomultiplier
Ref.
Vidicon
12:5 12:5 mm 2 bidimens
with scintillator to 200 nm
55
Diode array
up to 25 mm up to 1024 monodimens 12.5 mm monodimens up to 60 mm bidimens up to 13 2 mm 2 bidimens
particularly for >350 nm
poorer particularly at 10 3 similar to PM dynamics >10 3
55
Diode array MCP Dissector tube CCD
200±800 nm 200±800 nm 200±800 nm
55, 64
59 56
These two-dimensional detectors [63] are ideally suited for coupling with an echelle spectrometer, which is state of the art in modern spectrometers for ICP atomic emission spectrometry as well as for atomic absorption spectrometers. As for CCDs the sensitivity is high and along with the signal-to-noise ratios achievable, they have become real alternatives to photomultipliers for optical atomic spectrometry (Table 3) and will replace them more and more. 2.2.3
Non-dispersive spectrometers
Filter monochromators are now used almost only for ¯ame photometry. They make use of interference ®lters, which may have a fairly low spectral bandpass (less than a few nm). However, it is also possible to use such ®lters for dynamic measurements of line and background intensities, and for transient signals, as occur in gas chromatography. The use of oscillating ®lters has been described, where the wavelength bandpass is slightly shifted by inclining them towards the radiation beam [65]. Non-dispersive multiplex spectrometers include Hadamard transformation spectrometers and Fourier transform spectrometers and are particularly useful for the case of very stable sources. In both cases the information, such as intensities at various wavelengths, is coded by a multiplex system, so that it can be recorded with a conventional detector. A suitable transformation is then used to reconstruct the wavelength dependence of the information. In Hadamard spectrometry use is made of a codation of the spectrum produced by recombining the information with the aid of a slit mask which is moved along the spectrum [66]. Fourier transform spectrometry [67] makes use of a Michelson interferometer (Fig. 29) to produce the interferogram. With the aid of a beam splitter the radiation is split into two parts each of which is directed to a mirror. When shifting the
2.2 Optical spectrometers Fig. 29. Principle of Fourier transform emission spectrometry.
mirror in one of the side arms, the interference for each wavelength is given by: I
x B
s1 cos
2psx
191
x is the optical path dierence, s is the wavenumber of the radiation (cmÿ1 ), I is the intensity measured with the detector, and B is the radiation density of the source. The frequency of the signal at the detector for a radiation with a frequency n is a function of the velocity v by which the optical path dierence is changed: f sv
192
Accordingly rapid changes in optical frequencies are transformed to lower frequencies, where they can be measured with the normal detector systems. For a wavelength of 400 nm or a frequency of 25000 sÿ1 and a scanning velocity of 0.08 cm/s the frequency becomes 2 kHz. Polychromatic radiation from a source accordingly gives an interferogram where the intensity of each point is the sum of all values resulting from Eq. (191). The central part contains the information at low resolution and the ends contain information at high resolution. A detector with a high dynamic range is thus required and this especially applies for line-rich atomic emission spectra with lines of widely diering intensities. Furthermore, the source must be very stable during the recording of the interferogram. The resolution obtained depends on the recording time, the spectral bandwidth and the number of repetitive scans. By applying a Fourier transform of the signal for each point of the interferogram, one can write:
y
y I
x B
s ds B
s cos
2psx ds
193 ÿy
C
ÿy
y ÿy
B
s cos
2psx ds
194
71
72
2 Spectrometric Instrumentation
C is a constant term and must be subtracted before the transformation and one ®nally obtains: B
s
y
I
x cos
2psx dx
ÿy
195
In the transformation the physical units are inverted. When the interferogram is expressed in optical path dierence units (cm), the spectrum is obtained in wavenumbers (cmÿ1 ) and when the interferogram is expressed in time units (s) the spectrum is in frequency units (sÿ1 ). Apart from sine and cosine functions, box-car and triangular, etc. functions are also known, for which the Fourier transformation can be calculated. When applying the Fourier transformation over the whole area Gy, the arm of the interferometer also would have to be moved from ÿy to y. When making a displacement over a distance of GL only, the interferogram has to be multiplied by a block function, which has the value of 1 between and ÿL and the value 0 outside. L then in¯uences the resolution that can be obtained. By digitizing the interferogram, a quick Fourier transform becomes possible provided a powerful computer is used. In the case of line-rich spectra this is possible with a limited but highly accurate displacement of the sidearm in a number of steps. In addition, repetitive scanning is applied so as to obtain a suciently intense image, during which a reference laser is used to ensure that the positioning of the mirror is reproducible. The technique has been well known in infrared spectroscopy for some time and now can be used for spectra in the visible and UV regions also, as suciently precise mechanical displacement systems and highcapacity computers have become available. Fourier transform atomic emission spectrometry is attractive for sources with a low radiance but a high stability. Furthermore, the wavelength calibration can be very accurate and the highest resolution can be realized. This makes the technique very attractive for line wavelength and line-shape measurements for diagnostics as well as for the development of line tables. However, only in the case of low-noise sources is it possible to obtain reasonable signal-to-noise ratios.
2.3
Mass spectrometers
As in the sources used in optical atomic spectrometry a considerable ionization takes place, they are also of use as ion sources for mass spectrometry. Although an overall treatment of instrumentation for mass spectrometry is given in other textbooks [68], the most common types of mass spectrometers will be brie¯y outlined here. In particular, the new types of elemental mass spectrometry sources have to be considered, namely the glow discharges and the inductively and eventually the microwave plasmas. In contrast with classical high voltage spark mass spectrometry (for a review see Ref. [69]) or thermionic mass spectrometry (see e.g. Ref. [70]), the plasma sources mentioned are operated at a pressure which is considerably
2.3 Mass spectrometers
higher (around 0.1 mbar up to atmospheric pressure) than the pressure in the mass spectrometer itself (10ÿ5 mbar). Consequently, an interface with the appropriate apertures has to be provided, along with high displacement vacuum pumps for bridging the pressure dierence between the source and the spectrometer itself. In commercial instrumentation fairly low-cost quadrupole mass spectrometers and also expensive double-focussing sector ®eld mass spectrometers are usually used (for a survey of mass spectrometers for analytical use, see Ref. [71]) and today new types of mass spectrometers such as time-of-¯ight mass spectrometers are being ®nd utilized in plasma atomic spectrometry. 2.3.1
Types of mass spectrometers
In all mass spectrometers used for atomic spectrometry a vacuum of 10ÿ5 mbar or better is maintained so as to avoid further ion formation from residual gas species or collisions of analyte ions with these species, by which the background signals would be increased. To this aim, turbomolecular pumps are used at present. They are preferred over diusion pumps as their maintenance is easier and back¯ow of oils does not occur. They are backed up by rotation pumps. To monitor the vacuum in the mass spectrometer a Penning gauge is generally used, whereas the vacuum in the sampling region, which is at the 0.1±1 mbar level, is often controlled with the aid of a capacitance pressure gauge. Quadrupole instruments
The spectrometer consists mainly of 4 equidistant and parallelly mounted rods (diameter 10±12 mm) between which a dc voltage and in addition a high-frequency ®eld (frequency: up to 1 MHz) or U V cos
ot is provided (Fig. 30). The dc voltage at the quadrupole should be slightly below the energy of the ions entering (usually below 30 eV). By varying the electric alternating ®eld, ions with dierent m=z can leave the analyzer sequentially and scanning becomes possible. A mathe-
Fig. 30. Principle of a quadrupole mass spectrometer.
73
74
2 Spectrometric Instrumentation
matical treatment of the processes taking place is possible with the aid of the Mathieu equations: A 8 e U=m r02 o 2
196
and Q 4 e V0 =m r02 o 2
197
with U the dc potential applied, m the mass of the ion, o=2p is the frequency of the alternating current, V0 the amplitude of the alternating current potential, r0 the radius of the quadrupole ®eld and e the elementary charge. Plotting A and Q on a diagram there is a zone for stable oscillation for ions within a narrow m=z range. The ions enter the instrument through the so-called skimmer and in addition have to pass the ion lenses and eventually a mask, which prevents the UV radiation from entering the spectrometer (``beam-stop''). By changing the voltages at the ion lenses, at the beam-stop and eventually at the quadrupole rods, the transmission of the spectrometer and the mass resolution for ions of a given mass and energy can be optimized. As these parameters all are interdependent, the optimization of the mass spectrometer is complex. By changing the quadrupole ®eld, the transmission for a certain ion changes. Accordingly, ions of a certain mass can be ®ltered out by manual setting of the ®eld or the spectrum can be scanned. The scanning velocity is normally limited to up to 30 000 dalton/s as a result of the high-frequency and dc components used in the spectrometer. Thus a mass interval of 0±300 dalton can be scanned in about 30 ms. The mass resolution achievable is mainly determined by the quality of the ®eld and in quadrupoles used for plasma mass spectrometry in the best case it is about 1 dalton. The line pro®les to a ®rst approximation are triangular, but side-wings also occur. Together with the isotopic abundances they determine the magnitude of the spectral interferences. Quadrupole mass spectrometers are rapid sequential measurement systems. Accordingly, the precision obtainable and also the ®gures of merit in the case of isotopic dilution techniques for calibration are limited by noise occurring in the source. Therefore, the price of quadrupole instruments is reasonable and also their transmission is high. This is important because it allows their use in plasma mass spectrometry, where the source itself can be kept at earth potential. This is a particular advantage in the case of inductively coupled plasmas. Magnetic mass analyzers
The oldest type of mass analyzers used, going back to the early 1920s, are magnetic mass analyzers. They consist of a curved ¯ight tube located between the poles of an electromagnet, so that the ®eld is perpendicular to the ¯ight direction of the ions. The ions ®rst pass through an electrical ®eld and then enter a magnetic ®eld where mass selection takes place. The electrical ®eld is used to extract the ions by a force:
2.3 Mass spectrometers
F zU
198
where z is the charge and U the voltage (up to 10 kV). The ions then enter a magnetic ®eld B and start to follow a curved trajectory on which they are held by the centrifugal force so that: m v 2 =r B z v
199
where r is the radius of the curved trajectory and v the velocity of the particle. Then: v B z r=m
200
As z U 1=2 m v 2 p v
2 z U=m
201
202
and by combining with Eq. (200): B z r=m
p
2 z U=m
or m=z B 2 r 2 =2U
203
By sweeping the ®eld B it is possible to scan the mass spectrum
m=z at an exit slit. As ions with dierent masses are de¯ected towards dierent locations in the focal plane simultaneous detection of ions with dierent masses is thus possible with these spectrometers, provided multichannel detection is applied. As the energy spread of the ion beam limits the resolving power of a de¯ecting magnetic sector mass analyzer, energy focusing must be applied. This can be done by providing an electrostatic analyzer in the ion optics. This electrostatic analyzer consists of two curved plates with a voltage of the order of 1.5±1 kV applied between them. For a mean radius r and a gap d operated with a voltage E, a singly charged ion with a mass m and a velocity v will be focussed through the exit slit only when: m v 2 =z E r=d
204
with the radius of the circular path being de®ned by the kinetic energy as: r
2 d=E z 0:5
mv 2
205
When bringing the magnetic sector and the electrostatic sector together, a double-
75
76
2 Spectrometric Instrumentation
focusing instrument is obtained, where ions are selected by mass only and where the ion energy of the particles does not play a role. The use of sector ®eld mass spectrometers requires the ion source to be brought to a high positive potential relative to the mass spectrometer entrance or the whole mass spectrometer to be at a high negative voltage with respect to the ion source. The low transmission of the complex arrangement accordingly is compensated for by a higher ion extraction yield. With sector ®eld instruments the resolution can easily be improved to below 0.1 dalton, enabling a number of spectral interferences in plasma mass spectrometry to be eliminated. The pro®les of the lines may be ¯at-topped but can be adjusted by optimizing with the aid of the widths of the source slit and the collector slit. The scanning speed cannot be increased in an unlimited manner, as hysteresis of the magnet may hamper this. When the signals are transient and multielement detection is required, there are certainly limitations here. Advanced types of mass spectrometers such as ion traps, ion cyclotron resonance and speci®cally time-of-¯ight mass spectrometers have also been considered for use in plasma mass spectrometry. Time-of-flight mass spectrometer
This spectrometer is based on the detection of ions produced from a single temporally well-de®ned event with high time resolution. The principle is well known and has long been applied in secondary ion mass spectrometry [72] and in laser micro mass spectrometry [73]. The basic formula for mass resolution is [71]: m=Dm 1=2
t=Dt The velocity of the ions is given by: p v 1:4 10 4
U=m
205
206
where m is the mass and U the accelerating voltage. Inserting 50 dalton for m and 5000 V for U would result in v 1:4 10 5 m/s. Hence for a ¯ight tube having a length l of 1.4 m one obtains, neglecting the time required to accelerate the ions: t l=v 10ÿ5 s. With modern instruments a time spread Dt of 5 ns and even lower can be obtained. Thus according to Eq. (205) a resolution of 2000 is feasable. The resolution can be increased further by using a re¯ectron, doubling the ¯ight time (Fig. 31). With this type of instrumentation high transmission can be achieved and high mass resolution can also be realized. In addition to ICP, mass spectrometry [74, 75] with the MIP [76] and glow discharges [77±79] has been developed. In the latter, packages of ions can be extracted on- or o-axis by using a pulsed ®eld and it is even possible to de¯ect the abundant argon or helium ®ller gas ions eciently to clean the spectrum, by applying a repeller pulse (Fig. 32). The technique has been shown to be particularly useful for transient signals, where a fair dynamic range and power of detection are achievable (see Ref. [80]). This also applies in the case of conventional ion detectors, such as dynode electron multipliers (Fig. 33).
2.3 Mass spectrometers
Fig. 31. On-axis ICP-time-of-flight mass spectrometer (Renaissance). (Courtesy of Leco Corp.)
Two-stage acceleration region in an ICP-TOF instrument. w: Beam width, v: beam velocity, R: repeller plate, f: repetition frequency, G1: extraction region entrance grid, Fig. 32.
d: packet length, a1: extraction region, a2: acceleration region, G2 flight-tube entrance grid, S: steering plates, D: deflection plates. (Reprinted with permission from Ref. [74].)
77
78
2 Spectrometric Instrumentation
Fig. 33. Continuous dynode electron multiplier. (Reprinted with permission from Ref. [81].)
Ion trap. In an ion trap [81] we have a tridimensional quadrupole mass spectrometer (Fig. 34), in which ions may be trapped or stored. This is achieved by applying a suitable rf voltage to a specially shaped electrode. Even during storage collisional dissociation as well as excitation may be achieved. The sensitiviy is excellent and a high resolution may also be achieved (in excess of 10 000 particularly when working with long acquisition times). This means one can also work with very weak ion beams [82]. The ®rst applications of an ion trap with an ICP have been published, both when using the ion trap behind a conventional quadrupole [83] as well as when coupling it through a suitable lens directly to the ICP [84].
Fig. 34. Ion trap detector analyzer, single stage configuration. (Reprinted with permission from Ref. [71].)
2.3 Mass spectrometers Fig. 35. Operating principles of ion cyclotron resonance mass spectrometry. (Reprinted with permission from Ref. [81].)
Ion cyclotron resonance analyzer
In this type of analyzer [85] the ions are trapped in circular orbits in a magnetic ®eld. To achieve this, three pairs of mutually perpendicular plates are placed in a strong magnetic ®eld. Externally produced ions may be injected into the cell, in which the time t for one revolution (Fig. 35) [81, 86] is given by: t
2 p rm =v
207
where v is the velocity and rm the radius of the orbit. Combination with Eq. (199) delivers: m=z B=2 p f
208
where f is the frequency. At a constant magnetic ®eld strength B, the m=z for the orbit depends on the frequency of the cyclotron. When applying an RF pulse of this frequency to two opposite plates, the respective ions will gain energy and when pulsing with a broad pulse distribution, many ions will come into resonance. This will lead to the respective signals in the receiving electrodes being a function of the ions in the cell. As the frequency distribution is determined by a Fourier transform of the induced currents, this method is known as Fourier transform mass spectrometry. In ion cyclotron resonance mass spectrometry a very high resolution can be achieved, as frequency distributions can be determined with a very high accuracy. Further-
79
80
2 Spectrometric Instrumentation Fig. 36. Marge Mattauch± Herzog system. (Reprinted with permission of Ref. [71].)
more, the method has a multiplex advantage and therefore allows for true simultaneous detection of ions over a wide mass range. Simultaneous mass spectrometer
This can be realized by using Mattauch±Herzog geometry, as is known from spark source mass spectrometry. Here the ions are conducted through an appropriate lens system into a magnetic analyzer and separated according to mass before reaching a linear focal plane (Fig. 36). A linear photodiode array with a microchannel plate can be used for the simultaneous and time-resolved detection of ions over the whole mass range. Multiple-collector magnetic mass analyzers
There are particularly useful when a highly accurate recording of ratios of ion currents for dierent isotopes is required. This is known from earlier work with thermal ionization used for isotopic analysis, and is important, for instance, in geochronology. An double-focusing instrument with a stigmatic focusing magnet and several individual collectors placed in the focal plane can be used to achieve such a system. Here the dispersion of the ions in the focal plane is larger than with the Mattauch±Herzog arrangement, allowing for the placement of several detectors. One also can use a beam-splitter in front of two quadrupole instruments [87]. As isotopic dilution becomes more and more important for tracer work as well as for highly accurate calibration, similar types of mass spectrometers may gain considerable importance. 2.3.2
Ion detection
For ion detection, several approaches are possible. From classical spark source mass spectrometry, the use of photographic plates is known, where very hard
2.3 Mass spectrometers
emulsions with low gelatine contents are required. The procedure for emulsion calibration is similar to the one described in optical emission spectrography, but dierent exposure times are normally used instead of the optical step ®lters. Provided an automated microdensitometer is used, mass spectrography is still a useful tool for survey analysis of solids down to the sub-mg/g level. However, for the precise work that is possible with the plasma sources, electron multipliers and pulse counting are usually used for ion detection. In elemental mass spectrometry the magnitudes of the signal to be detected may vary between 0.1 ion/s for the case of ultratrace analysis and 10 10 ions/s in the case of major elements, and depending on the ion source used. As the charge of an electron is 1:6 10ÿ19 Coulomb, these ion densities correspond to currents of between 10ÿ20 and 10ÿ9 A and for a such a range, several detectors need to be used. For low signals ( 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
88
3
Sample Introduction Devices In atomic spectrometry the sample must ®rst be brought into the form of an aerosol and atomized, before excitation involving emission, absorption or ¯uorescence and/or ionization can take place. Accordingly, the sample introduction is a very important step in all atomic spectrometric methods. Many approaches, which can be used for very dierent types of samples with respect to size, state of aggregate, stability, etc., have been investigated and treated extensively in textbooks (see e.g. Refs. [94±98]). Sample volatilization and signal generation can take place in a single source. However, sample volatilization can also take place in one source, and then to generate the signal the sample vapor or atom cloud is taken into a second source, usually with the aid of a carrier-gas ¯ow. The use of separate sources for sample volatilization and signal generation is known as the ``tandem source concept'', a term which was introduced by Borer and Hieftje [99]. This concept has the advantage that the best conditions for sample volatilization, with respect to source temperature and analyte residence time, are selected for the ®rst source and at the same time conditions can be selected in the second source for which, for example, the signal-to-background ratios and, accordingly, the power of detection are optimum or the matrix eects are lowest. This makes sense as these conditions may be totally dierent and using a single source for both sample volatilization and signal generation may lead to compromise conditions with losses in analytical performance. This has been demonstrated by the combination of spark ablation and plasma spectrometry for the direct analysis of electrically-conducting samples, when compared with conventional spark emission spectrometry. Both matrix eects from sample volatilization as well as from analyte excitation can be decreased as compared with spark emission spectrometry (see e.g. Ref. [100]) without losing the high power of detection of plasma atomic spectrometry. Similar eects are experienced when laser ablation is combined with the relatively cheap microwave plasma, where a matrix-independent calibration in many cases becomes feasible (see e.g. Ref. [101]). For sample volatilization of solids or dry solution residues both thermal evaporation and cathodic sputtering may be useful. The latter is particularly important in the case of low-pressure discharges, as here the ions are highly energized when they pass through the high-energy zones of the discharge and lose only a little en-
3 Sample Introduction Devices Techniques for sample introduction in plasma spectrometry [102]. Fig. 39.
ergy, because of the low number of collisions occurring in the low-pressure environment. Thermal volatilization takes place in sources where ohmic heat development is predominant, and this will be the case for sources where a large amount of electrical power is dissipated or high amounts of energy are absorbed. In the case of solutions, various nebulization devices are often used for the production of aerosols. For plasma sources at atmospheric pressure or ¯ames, the aerosol generation is frequently performed in a separate device. As shown schematically in Fig. 39, a number of sampling techniques for the analysis of solutions as well as for direct sampling of solids can be applied. The techniques used for plasma spectrometry [103] as well as sampling based on thermal evaporation and
89
90
3 Sample Introduction Devices
on cathodic sputtering, as performed in low-pressure discharges, will be treated in detail here.
3.1
Sample introduction by pneumatic nebulization
In the analysis of solutions, pneumatic nebulization and then the introduction of the aerosol into the source are already well known from the early work on ¯ame emission spectrometry. Pneumatic nebulizers must enable a ®ne aerosol to be produced, which leads to a high eciency, and the gas ¯ow used must be low enough to transport the droplets of the aerosol through the source with a low velocity. The production of ®ne droplets and the use of low gas ¯ows are essential to obtaining complete atomization in the source. Pneumatic nebulization of liquids is based on the viscous drag forces of a gas ¯ow passing over a liquid surface and entraining parts of the liquid, by which small independent droplets are produced. This may occur when the liquid is forced through a capillary tube and at the exit the gas then ¯ows concentrically around the tube or perpendicularly with respect to the liquid stream. A frit can also be used, which is continuously wetted and through which the gas passes. Nebulizers are mounted in a nebulization chamber, which has the function of separating o the larger particles and leading them into the waste. Impact surfaces are often provided, so as to fractionate the aerosol droplets further, by which the eciency of aerosol production increases. Whereas much research has been done on nebulizers with the aim of maximizing their eciency, much less has been done on spray chambers. They form a unit along with the nebulizer and the two can only be evaluated and optimized together. With respect to spray chambers when working with ¯ames, impact surfaces are often provided so as to fractionate the droplets formed by the nebulizer further and to trap the large droplets. According in plasma spectrometry the nebulization chamber according to Scott et al. [104] is used in the rule. However, by providing smooth sample run-o, the noise, for example in ICP spectrometry, can be much reduced, as shown in early work by Schutyser and Janssens, who investigated many types of spray chambers [105]. In the case of ICP mass spectrometry the amount of vapor formed from the liquid needs to be limited, as analyte is thus not brought into the plasma and this only increases risk of formation of molecular ions. To this aim it is very ecient to cool the nebulization chamber down. This leads on the one hand to a decrease in the background and, accordingly, an improvement in the signal-to-noise ratios (Table 4, [41]) and on the other hand to a decrease in the white noise level, as shown by the noise power spectra (Fig. 2). The spray chamber needs some time to ®ll up with the aerosol produced and some tailing of signals is seen, resultng from removal of the sample aerosol by the new incoming gas and aerosol. These build-up and wash-out times limit the speed of analysis and lead to a ¯attening of transient signals, when a vapor cloud passes through the spray chamber, as is the case in ¯ow injection analysis. In order to
3.1 Sample introduction by pneumatic nebulization Tab. 4. Influence of cooling of the spray chamber on the stability of ICP-MS (30 min short-term stability) with a Meinhard nebulizer and an aqueous standard solution; analyte concentration: 20 ng/mL of each element mentioned. (Reprinted with permission from Ref. [41].)
Element
RSD* without cooling
RSD a with cooling 10 C
9
3.03 2.18 2.17 1.94 2.49 3.07 3.08 3.52 3.66 3.38
2.46 2.75 2.31 1.84 2.26 1.29 1.28 1.38 1.61 1.97
Be Mg 59 Co 60 Ni 69 Ga 115 In 154 Sm 208 Pb 209 Bi 238 U 29
a
RSD (relative standard deviations) in % resulting from 10 replicate measurements.
minimize these dead times, small spray chambers with short build-up and rinsing times have been constructed by decreasing their volume and optimizing their form with respect to ideal aerosol droplet fractionation and very ecient surface renewal. This resulted e.g. in the cyclone spray chamber described by Wu and Hieftje [106]. They showed that with this cyclone spray chamber the wash-out times were short without sacri®cing analytical precision, as shown by the noise spectra, or deterioration in the power of detection. Indeed, for a 10 mg/mL solution of Mg in ICP-OES the wash-out time, de®ned as the time required to reach a signal below 1% of the analyte signal, is below 10 s and the signals are of a similar magnitude as with a Scott-type chamber. Systematic investigations are just beginning on the frequency and nature of pulsations in spray chambers of varying forms, based on the calculation of the ¯ow patterns on the one hand and measurements of droplet sizes and droplet velocities on the other [106a]. In the pneumatic nebulizers used in atomic spectrometry, the liquid ¯ow is usually of the order of 1 to a few mL/min and the full eciency of the nebulizer (a few %) can actually be realized at gas ¯ows of 2 L/min. However, even with gas ¯ows below 2 L/min, droplet diameters as low as about 10 mm and injection velocities below about 10 m/s are obtained. Pneumatic nebulization can be realized with a number of types of nebulizers. For ¯ame emission and atomic absorption as well as for plasma spectrometry, they include concentric nebulizers, so-called cross¯ow nebulizers, Babington nebulizers and fritted-disk nebulizers (Fig. 40) [95]. Concentric nebulizers
These have been known since the work of Gouy [107] in the middle of the 19th century, and in principle are they are self-aspirating. They can be made of metal, plastic or of glass. In atomic absorption spectrometry, PtaIr capillaries are often used so as to allow the aspiration of highly acidic solutions. In plasma atomic
91
92
3 Sample Introduction Devices Fig. 40. Pneumatic nebulizers for plasma spectrometry. A: concentric glass nebulizer, B: cross-flow nebulizer, C: Babington nebulizer, D: fritteddisk nebulizer. (Reprinted with permission from Ref. [95].)
spectrometry, the use of concentric glass nebulizers (Meinhard Associates) is typical [108]. They use a fairly low gas ¯ow (1±2 L/min) as compared with the nebulizers used in atomic absorption (up to 5 L/min) and are often self-aspirating. In both cases, however, the sample uptake rates are 1±2 mL/min and the aerosol eciencies 2±5%. The nebulizers work at pressures of 2±7 bar, which are speci®ed as are their sample uptake rates, being functions of their ori®ce dimensions (15±35 mm). Meinhard nebulizers can also be made of quartz, in which case even solutions containing dilute HF can be nebulized, without attacking the nebulizer. In the case of solutions with high salt contents, the danger for clogging and signal drift is high as compared with other nebulizers (Fig. 41) [109]. The maximum admissible salt content is 20±40 g/L but it varies considerably along with the type of salt involved. It is particularly high for easily hydrolyzable compounds such as aluminum salts and phosphates. Through long-term experiments, concentric glass nebulizers have also been shown to be sensitive to high concentrations of alkali and alkaline earth elements. The changes in nebulization eciency and sample uptake are connected with ionexchange phenomena at the glass wall, as mentioned in Ref. [109]. However, the in¯uence of such salts can be reduced considerably by wetting the nebulizer gas ¯ow. A type of concentric glass nebulizer which is operated at pressures of several tens of bars is known as a high eciency nebulizer (HEN) [110]. More recently micro-
3.1 Sample introduction by pneumatic nebulization
Fig. 41. Operating stability of different types of pneumatic nebulizers having high salt contents. (AA): concentric nebulizer as used in AAS work, solution: A16 g dissolved salts (mainly NaCl) per liter. (Reprinted with permission from Ref. [109].)
concentric nebulizers have also been described, for which the internal volume is very small and the liquid channel is connected directly to capillaries. One type that is available, the MCN-100, is very useful for coupling chromatography to plasma spectrometry, as the ¯ow used can be in the sub-mL/min range (Fig. 42) [111].
93
94
3 Sample Introduction Devices Fig. 42. Microconcentric nebulizer for ICP-OES/MS. (Reprinted with permission from Ref. [111].)
Cross-flow nebulizers
The cross-¯ow nebulizers are less sensitive to salt deposits at the nozzle. They are used in plasma spectrometry as well as in ¯ame work and require the use of a peristaltic pump for sample feeding. The pump must have a sucient number of rolls, so as to avoid pulsing, and even then may still introduce a low-frequency noise. However, dual tube pumps with a time delay between both channels, have been introduced, through which pulsing can be further minimized. In the cross¯ow nebulizer the sample solution is fed through a vertically mounted capillary and nebulized by a gas ¯ow entering through a horizontal capillary which ends close to the tip of the vertical one. The design of cross-¯ow nebulizers has been discussed previously by Kranz [112]. A cross-¯ow nebulizer for dc arc solution and AAS work was described by Valente and Schrenk [113]. For ICP work the nebulizer should have capillaries with diameters < 0.2 mm and a distance of 0.05±0.5 mm between the tips. The capillaries also can be made of metal, glass or plastic and they have similar characteristics to the concentric nebulizer. Types made of Ryton for example enable the aspiration of solutions containing higher concentrations of HF. Then aerosol gas ¯ows are around 1±2 L/min at 2±5 bar. This has been con®rmed by optimization measurements described by Fujishiro et al. [114]. They found that by varying the inside and the outside diameters of the capillary from 0.15 to 0.9 mm and from 0.5 to 1.5 mm, respectively, the pressure drop across the sample tube decreases from 150 to 50 mbar. By increasing the gas ¯ow from 0.75 to 1.75 L/min, the pressure drop across the sample tube was found to increase from 150 to 350 mbar. As shown in Fig. 43 [114], the variations in the drops in pressure considerably in¯uence the droplet size. Babington nebulizer
Here the sample solution is pumped through a wide tube into a groove, where the ®lm running o is nebulized by a gas ¯ow entering perpendicular to the ®lm of
3.1 Sample introduction by pneumatic nebulization
Fig. 43. Influence of pressure decrease across the sample tube of a cross-flow nebulizer on droplet size distribution. Pressure decrease: (): 95 mm H2 O, (o): 250 mm H2 O. (Reprinted with permission from Ref. [114].)
liquid. As the sample solution does not need to pass through a narrow capillary, solutions containing high salt concentrations and also slurries can be analyzed without the risk of clogging. This type of nebulizer, ®rst described by Babington in 1973 [115], was later used extensively used by Ebdon (see e.g. [116]) especially for slurry nebulization. Work with slurries requires that the slurries are ®rst nebulized and behave just as solutions with respect to the sample introduction into the aerosol. From electron probe micrographs of aerosol particles sampled on Nuclepore ®lters under isokinetic conditions, it was found that at nebulizer gas ¯ows of 3 L/min, being typical of plasma spectrometry but far below those for ¯ame atomic absorption, particles with a diameter of up to 15 mm can be found in the aerosol (Fig. 44) [117]. This would imply that powders with a grain size of up to about 15 mm could still be nebulized as could a solution. This, however, is not true as the mass distribution in the case of powders may be quite dierent in the slurry and in the aerosol, as shown for the case of SiC (Fig. 45) [118]. The nebulization limitations for the case of slurry nebulization thus must be investigated from case to case and leads to certain types of restrictions.
95
96
3 Sample Introduction Devices
Fig. 44. Scanning electron microscopy images of Al2 O3 particles sampled from a 1% suspension itself and from the aerosol produced from the suspension. Al2 O3 (VAW 143±85, Vereinigte Aluminium Werke, Bonn)
sieved fractions. GMK nebulizer with an impactor 3 mm from the orifice. Aerosol gas flow rate: 1 L/min, pumping rate: 2 mL/min. (Reprinted with permission from Ref. [117].)
Fritted-disk nebulizers
Nebulizers such as these have high nebulization eciencies and a low sample consumption (down to 0.1±0.5 mL/min). In particular, they have been shown to be useful for plasma spectrometry in the case of organic solutions or for coupling with liquid chromatography [119]. However, in comparison with the other types of nebulizers they suer from memory eects. Indeed, the rinsing times are much higher than the 10±20 s as required for the other pneumatic nebulizers. In the case of the so-called thimble nebulizer, as described by Liu et al. [120] (Fig. 46), these disadvantages have been eliminated, mean droplet sizes were reported to be of the 2 mm level, which is lower by a factor of 5 than with the HEN mentioned previously. Direct injection nebulizers
Direct injection nebulizers, as they were ®rst described by Kimberley et al. [121], are used without a nebulization chamber. In this way sample losses are low as
3.1 Sample introduction by pneumatic nebulization
Fig. 45. Particle size and mass distribution of a suspended SiC powder and the aerosol generated with a GMK nebulizer [118].
are the dead times. This can only be achieved when the aerosol droplets are ®ne enough to avoid sample deposition or incomplete evaporation in the source. High nebulization eciencies can then also be obtained, as shown by the comparative measurements reported by McLean et al. [122]. These features can only be realized when the drop in pressure at the top of the nebulizer is high or when the nebulizer is operated at high pressure. Oscillating nebulizer
A special type of pneumatic nebulizer is the oscillating nebulizer, described by Wang et al. [123] (see Fig. 47) [124]. Here the droplet distribution is narrow, the
97
98
3 Sample Introduction Devices Fig. 46. Demountable thimble frit nebulizer and its spray chamber (not drawn to scale). (Reprinted with permission from Ref. [120].)
droplet size is low, e.g. as compared with the concentric nebulizer and the eciency is also high. With pneumatic nebulizers the samples can be taken up continuously. Then after a few seconds of aspiration a continuous signal is obtained and it can be integrated for a certain time. However, small sample aliquots can also be injected directly into the nebulizer [125±128] or into a continuous ¯ow of liquid. The latter
Fig. 47. Oscillating capillary nebulizer with single pass spray chamber. (Reprinted with permission from Ref. [124].)
3.1 Sample introduction by pneumatic nebulization
is known as ¯ow injection analysis (FIA) and has been treated extensively (for the principles, see Ref. [129]). The signals in both cases are transient. They rise sharply and there is some tailing, resulting from generation of the analyte aerosol cloud at the front end and dilution of the aerosol at the signal end, respectively. The signals can last a few seconds up to eventually some 10 s and with sample volumes of 10± 20 mL similar signal intensities to continuous nebulization can often be obtained. These techniques enable analysis of small samples (down to around 10 mL), lower the salt deposits on nebulizers and maximize the absolute power of detection. Also useful in work with the introduction of small-volume samples, denoted as discrete sampling, the use of small cyclone chambers as well as of direct injection nebulizers has to be mentioned [121]. FIA has gained further importance as automation is easy and on-line sample preparation becomes possible. This is speci®cally due to the use of on-line digestion of slurries at high temperatures, as described e.g. by Haiber and Berndt [130], as well as by on-line preconcentration and matrix removal by coupling with column chromatography. In the case of dissolved Al2 O3 ceramic powders, matrix removal is easily possible by ®xation of the dithiocarbamate complexes of the impurity metals on a solid phase while aluminum is not retained. The traces can then be eluted from the column and e.g. directly determined by ICP mass spectrometry (Fig. 48) [131]. The other solution lies in retaining the matrix on the column, as e.g. is possible in the case of zirconium by the formation of thenoyltri¯uoroacetonates [132], which also could be ®xed on an RP
Fig. 48. ICP-MS coupled on-line to liquid chromatography using hydraulic high-pressure nebulization. (1): HPLC pump, (2): sample loop, (3): chromatographic column, (4):
HHPN nebulization chamber, (5): desolvation system with heating and two-stage cooling, (6): ICP torch, (7): drain. (Reprinted with permission from Ref. [131].)
99
100
3 Sample Introduction Devices
18 solid phase, in amounts of up to a few mg [133]. Further examples of on-line matrix removal are given in the literature (see e.g. Refs. [134±136]). The optimization of pneumatic nebulizers is aimed in particular at selecting the working conditions that give the optimum droplet size and ecieny. The so-called Sauter diameter d 0, i.e. the diameter for which the volume to surface ratio equals that of the complete aerosol, is given by the Nukuyama±Tanasawa equation (see Ref. [137]): 00
d 0 C=vG
s=r 0:5 C 0 fh=
s r 0:5 gC 1000
Q L =Q G C
000
218
vG is the gas velocity, Q G the gas ¯ow, Q L the liquid ¯ow, h the viscosity, r the density, s the surface tension of the liquid and C, C 0 and C 00 are constants. When the nebulizer gas ¯ow increases d 0 becomes smaller (through vG and Q G ) and the eciency of the sample introduction will increase and as a result the signals also do. However, as more gas is passed through the source, it will be cooled and, moreover, the residence time of the droplets will decrease and thus the atomization, excitation and ionization will also decrease. These facts will counteract any increase in the signals as a result of the improved sampling. Accordingly, the highest signal intensity and thus the greatest power of detection will be achieved at a compromise gas ¯ow. The physical properties of the liquid sample (s; h and r) also in¯uence d 0 , the eciency and the analytical signals. Dierences in the physical properties of the liquid samples to be analyzed thus lead to so-called nebulization eects. h has a large in¯uence. It in¯uences d 0 through the second term in Eq. (218). However, in the case of free sample aspiration it also in¯uences d 0 through Q L , which is given by the Poiseuille law as: Q L
p RL4 =
8 h l DP
219
RL is the diameter and l the length of the capillary. DP is the pressure dierence and is given by: DP DPU ÿ g h r ÿ f
s
220
DPU is the dierence between the pressure inside and outside the nebulizer, h is the height and f
s a correction factor. In order to minimize the nebulization eects one can utilize forced feeding with a peristaltic pump, a result of which is that Q L is no longer a function of h. One also can operate the nebulizer at a high gas ¯ow, by which Eq. (218) reduces to: d 0 C=vG
s=r 0:5
221
when Q G =Q L becomes >5000. However, the gas ¯ow may then cool the source down through which incomplete atomization or signal depressions or enhancements may lead to even stronger interferences. Droplet size distributions of aerosols can be determined by stray-light measure-
3.1 Sample introduction by pneumatic nebulization Fig. 49. Particle size distributions of pneumatic nebulizers for ICP. MAK, high-pressure cross-flow nebulizer; CGN, concentric glass nebulizer; JAB, Jarrell±Ash Babington nebulizer; JAC, Jarrell±.Ash fixed cross-flow nebulizer (Reprinted with permission from Ref. [139].)
ments (see e.g. Refs. in [122, 137]). Data obtained from sampling with cascade impactors could be in¯uenced by evaporation of the droplets on their way through the device, especially as isokinetic sampling may require the application of reduced pressure [138]. Various nebulizers have been used to provide data, e.g. Olsen and Strasheim (see Ref. [139]). When comparing the dierent nebulizers used in early work on ICP-OES (Fig. 49), they found that the high pressure cross-¯ow nebulizer has the highest SNR values, the lowest particle size and the narrowest droplet size distribution. The SNR values found in ICP-OES, for example, seemed to correlate with the fraction of the aerosol present as droplets with diameters below 8 mm. Particle size distributions depend on the nebulizer and its working conditions, on the liquid nebulized and also on the nebulization chamber. These relationships have been studied in depth, also for organic solutions (see e.g. Refs. [140, 141]). In these cases, an aerosol with a lower mean particle size is often produced with a fairly low gas ¯ow pneumatic nebulizer as compared with aqueous solutions and thus resulting in a higher eciency and power of detection. Jet-impact nebulization
Apart from pneumatic nebulization through gas entrainment other types of pneumatic nebulization have been realized. In so-called jet-impact nebulization, use is made of the impact of a high pressure jet of liquid on a surface [142].
101
102
3 Sample Introduction Devices
Fig. 50. Hydraulic high pressure nebulization system. (a): Solvent filter (20 mm), (b): solution reservoir (water, organic solvents), (c): highpressure pump, (d): solvent feed valve/sample loop, (e): plate with concentric holes;
(f ): high-pressure nebulization nozzle (10±30 mm) with integrated protection filter, (g): drain, (h): impact bead and (i): gas mixing chamber. (Courtesy of H. Berndt)
High-pressure nebulization
This has been described by Berndt [143], and makes use of the expansion of a liquid jet under high pressure at a nozzle (Fig. 50). Here much higher nebulization eciencies can be obtained even for solutions with high salt solutions and for oils. This is due to the small droplet size, which is in the 1±10 mm size [144]. The nebulization eciency can be improved further by heating the capillary, so that a mixture of vapor and aerosol is released at the nozzle [145]. In ICP-OES the detection limits can be decreased by an order of magnitude as compared with pneumatic nebulization, provided a heated spray chamber is used. Such systems are ideally suited for speciation work as they allow easy coupling of HPLC and atomic spectrometry [146]. A further advantage lies in the fact that the aerosol properties are fairly independent of the physical properties of the liquid. High-pressure nebulization is very promising for the coupling of on-line pre-enrichement or separations by high performance liquid chromatography (HPLC) with atomic spectrometry. In the case of plasma spectrometry, however, it is usually necessary to desolvate the aerosol produced, as high water loadings cool the plasma too much and decrease the analytical signals and, further, may also lead to increased matrix eects. Desolvation can be achieved by heating the aerosol leading to solvent evaporation, as previously applied in atomic absorption work by Syti [147] (Fig. 51). Other new ways of cooling in aerosol desolvation are possible by the use of Peltier elements. Also, IR heating may be used to heat up the aerosol. Another new method of
3.2 Ultrasonic nebulization
Fig. 51. Desolvation unit for ICP atomic spectrometry.
desolvation lies in the use of membranes, such as those made of Na®on. Here water molecules can diuse through, while salt aeosol particles as well as droplets are retained (Fig. 52). The principle and features as demonstrated by Yang et al. [148] may be particularly useful for various types of sample introduction used in plasma spectrometry.
3.2
Ultrasonic nebulization
By the interaction of suciently energetic acoustic waves of a suitable frequency with the surface of a liquid, geysers are formed by which an aerosol is produced.
Aerosol drying by using a single tube NafionR membrane desolvation stage. (Reprinted with permission from Ref. [148].) Fig. 52.
103
104
3 Sample Introduction Devices
Fig. 53. Ultrasonic nebulization. (A): Discontinuous system (reprinted with permission from Ref. [149]), (B): system with liquid flowing over the transducer (reprinted with permission from Ref. [150]).
The diameter of the aerosol droplets produced depends on the frequency and on the physical properties of the liquid sample. In the case of water for example, aerosol droplets formed at a frequency of 1 MHz have a diameter of about 4 mm. The energy can be focussed with the aid of a liquid lens onto the surface of the sample solution or the liquid sample can be pumped continuously over the transducer, which must then be cooled eciently (Fig. 53A and B). Ultrasonic nebulization has two considerable advantages over pneumatic nebulization. At ®rst the aerosol particles have a lower diameter and a fairly narrow particle size distribution as compared with pneumatic nebulization (1 as a result of the commencement of self-reversal. Particularly in trace analysis by AES the intensity of the spectral background I U can be used as reference signal and then in Eq. (272), c R should be replaced by c U. The calibration constant c U is the concentration at which the line and background intensities are equal (background equivalent concentration ± BEC value) and can be calculated as: c U
I X =I U 1=cÿ1
273
The ratio
I X =I U and the calibration constants c U for a given element and line depend on the radiation source and the working conditions selected but also on the spectral apparatus [338, 339]:
cU cU
dB U =dlDl L s eff R L p p=
ln 16 B0 s0 R 0 cU; y
A1
A2
q 4
s a =s eff 4 1 (274)
cU; y describes the in¯uence of the radiation source,
dB U =dl is the spectral radiation density for the background intensity, B0 is the radiant density for an analytical line at the concentration c 1 and Dl L is the physical width of the analysis line. The second term
A1 describes the in¯uence of the spectral apparatus. s eff
s 2e s02 s 2L s 2z 1=2
275
s L is the spectral slit width corresponding to the physical width of the spectral line
Dl L and R L l=Dl L , the resolution required to resolve the line width. s z is the slit width corresponding to the optical aberrations in the spectral apparatus and is de®ned as: s z Dl z dx=dl
276
Dl z can be determined by measuring the practical resolution of the monochromator from the deconvolution of the two components of the Hg 313.1 nm line doublet and subtraction of the contributions of the diraction slit width (s0 l f =W ) and the entrance slit width (s e ). The contribution of the natural width of the Hg lines can be neglected, as it is very low in the case of a hollow cathode lamp. In the case of monochromators s z is predominant, but in the case of a very high resolution it becomes less signi®cant and A1 ! 1. The factor A2 describes the in¯uence of the pro®le of the analysis line and the in¯uence of the eective measurement slit. In the case of photoelectric measurements the latter is the exit slit width of the spectrometer. This contribution is only
5.1 Principles
relevant when s a =s eff < 2, which is virtually only the case with photographic measurements where the line pro®le is scanned with a very narrow densitometer slit. For photoelectric measurements A2 s a =s eff as the exit slit width must be larger than the eective line width because of the thermal and mechanical stability of the system. With the aid of these equations, detection limits and signal to background ratios that have been obtained with two dierent spectrometers can be compared. For instance for a 0.9 m Czerny±Turner monochromator, with R 0 216 000 (grating width: 90 mm and grating constant: 1/2400 mm), s e 12 mm (^s e 4), s a 15 mm (^s a 5), practical resolution: 0.007 nm (R z 75 000); and for a 0.5 m Ebert monochromator: R 0 134 600 (grating width: 60 mm and grating constant: 1/2442.4 mm), s e 4 mm (^s e 1:5), s a 6 mm (^s a 2:3), practical resolution: 0.01 nm (R z 35 000), the ratio of the constants c U and the detection limits for an analysis line at 300 nm and with a line width of Dl L G 3 pm in the case of the same excitation source and both spectral apparatus should be: c L 0.9-m/c L 0.5-m 0:88 [180]. Detection limit
This is a very important ®gure of merit for an analytical method. As discussed in Section 2.1 in general, for an atomic emission spectrometric method it can be written as: c L c U
I X =I U
277
I X =I U is the smallest line-to-background ratio that can still be measured. When the de®nition is based on a 99.86% probability, it is given by 3 times the relative standard deviation of the background intensities. In the case of photographic or diode array registration the intensities of the line and the spectral background are recorded simultaneously, and: s s 2
I X I U s 2
I U 1=2
278
As the intensities of the line and the spectral background are nearly equal in the vicinity of the detection limit, their standard deviations will be equal and the relevant standard deviation reduces to: p s s
I U 2
279
There is some dierence in the standard deviations when line and background intensities are measured simultaneously, as in the case of photographic detection or with modern CCD spectrometers, or when they are measured sequentially as done in slew scanning systems. Indeed, in the ®rst case the ¯uctuations of line intensities and background intensities for a considerable part are correlated, especially at low signal levels and thus partially cancel. This may lead to a considerable gain in power of detection.
199
200
5 Atomic Emission Spectrometry
The value of the standard deviation can be measured directly, but in the case of a photographic measurement it can be calculated from the standard deviation of the blackenings with the aid of Eq. (184). This shows that for trace analyses, emulsions with a high g must be used. Accordingly, standard deviations s r
I U of about 0.005 can be obtained. This assumes a smooth background and these values re¯ect only the graininess of the emulsion and do not include contributions from source ¯uctuations. In the case of photoelectric measurements with a photomultiplier, s is the absolute standard deviation of the background intensity, measured at a wavelength adjacent to the spectral line or at a blank sample at the wavelength of the analytical line. It may consist of several contributions as: s 2
I U s 2P s 2D s 2A s 2V
280
s P is the photon noise of the source or the noise of the photoelectrons and p s P A n, when n is the number of photoelectrons considered; s D is the dark current noise of the photomultiplier and s D A I D , when I D is the dark current; s A is the ¯icker noise of the radiation source and s A A f, when f is the radiant density; and s V is the ampli®er noise. s V in most cases can be neglected as compared with other sources of noise. In the case of very constant sources s A is small. Owing to the proportionality between s D and I D it is advisable to choose a photomultiplier with a low dark current, so as to avoid detector noise limitations. Here limitations may remain when spectral lines at lower wavelengths are used. Indeed, then the spectral background intensities are usually low, so that the detector noise becomes predominant. This may speci®cally become the case when an Echelle spectrometer is used for the measurements. Owing to slit height limitations the photon ¯ux onto the detector then becomes very small. In many cases this is balanced by the fact that Echelle spectrometers have a shorter focal length, because of which the overall radiant ¯ux density in the focal plane is higher. Also in the case of CCD detector technology, limitations through the detector noise may occur. In particular, readout noise here may be limitating. This is because the detector is read many times so as to collect the charge accumulated during the numerous subsequent measurement cycles that are required to realize a suciently large linear dynamic range. In most other cases, the photon noise will be limiting for the level of detection. When signals include a blank contribution, this will also have ¯uctuations, which limit the power of detection in practice, the in¯uence of which can easily be shown. If I X is the lowest measurable signal without blank contribution, I X0 the lowest analytical signal including a considerable blank value, I U the background signal and I Bl the blank signal, I U ,
I X I U ,
I Bl I U and
I X0 I Bl I U can be measured and: q
I X0 I Bl I U ÿ
I Bl I U 3 s 2
I X0 I Bl I U s 2
I Bl I U
281 or p
I X0 I U ÿ I U 3 s 2
I X I U s 2
I U
282
5.1 Principles
This leads to:
I X0 =I U
I X =I U
1 I Bl =I U s r
I Bl I U =s r
I U
283
and the detection limit becomes [128]: c L0 c U f
I X =I U
1 I Bl =I U s r
I Bl I U =s r
I U g or c L s r
I Bl I U =s r
I U
1 I Bl =I U
284
As blank contributions themselves also scatter, which cannot be eliminated by blank subtraction, and as blank values scatter independently of the analytical signals, the presence of blank contributions always deteriorates the limit of detection achievable. Accordingly, to obtain the best limit of detection in AES the standard deviations of the background, the blank contributions and the c U values should be minimized. The in¯uence of the standard deviations of the line and background intensities are partly correlated. This has the consequence that the standard deviations of the net signals could be decreased when line and background intensities are measured simultaneously, as is possible with photographic emulsions and with the new array detectors that have simultaneous measurement capabilities at a large number of adjacent wavelengths. However, this is only true when the spectral background from the source and not the pixel-to-pixel noise is limitating. The c U values are a function of the radiation source, the elements and the lines (re¯ected in cU; y ) and also of the spectrometer (through A1 and A2 ). In order to keep A1 and A2 as close to one as possible, a spectral apparatus with high resolving power (R 0 high), a low entrance slit (s e 1:5s0 ) should be used and take s a s eff . On the other hand, it should be con®rmed that the spectral background intensities can still be measured. Indeed, the background intensities obtained are proportional to the entrance slit width and thus detector noise limitations could be encounterd when the slit widths become too narrow. Also thermal and mechanical stabilities become limitating in the case of narrow slit widths, by which, particularly in systems with on-peak integration instead of slew-scan procedures, the long-term stability can no longer be guaranteed. Apart from the high power of detection, also the realization of the highest analytical accuracy is very important. This relates to the freedom of interferences. Whereas the interferences stemming from in¯uences of the sample constituents on the sample introduction or on the volatilization, ionization and excitation in the radiation source dier widely from one source to another, most sources emit linerich spectra and thus the risks for spectral interferences in AES are high. In the wavelength range 200±400 nm, as an example, only for arc and spark sources have more than 200 000 spectral lines yet been identi®ed with respect to wavelength and element in the classical MIT Tables. Consequently, spectral interferences are much more severe than in AAS or AFS work.
201
202
5 Atomic Emission Spectrometry
Therefore, in AES it is also advisable to use high resolution spectrometers so as to minimize the risks of spectral interferences. A practical resolving power of 40 000, which guarantees that spectral lines at wavelengths of 300 nm and wavelength dierences of 0.0075 nm can still be spectrally resolved, is advisable. This is particularly the case when trace determinations must be performed in matrices emitting line-rich spectra. The nature of spectral interferences, may however, be dierent.
. Direct .
. .
coincidence of lines: when the wavelength dierence is less than the spectral bandwidth of the spectral apparatus. Line wings: spectral lines of matrix elements may have a width extending to up to several pm, as is known e.g. from the calcium lines in water samples from the early work on ICP-OES [341]. Here an estimation of the spectral interference for analytical lines situated on the wings has to be made by weighting the intensity contributions at well-de®ned wavelength dierences away from the line center. Contributions from band emission spectra, especially in the region of intensive band systems (CN at 370±385 nm, N 2 at 390±400 nm, NH at 340±350 nm, OH at 310±320 nm). Also here weighted corrections, as in the case of matrix line wings may be necessary. Stray radiation contributions usually have a more even spectral dependence and in a number of cases can be corrected for by measuring the stray radiation intensity on one side of the analytical line only.
Knowledge of the atomic spectra is also very important so as to be able to select interference-free analysis lines for a given element in a well-de®ned matrix at a certain concentration level. To do this, wavelength atlases or spectral cards for the dierent sources can be used, as they have been published for arcs and sparks, glow discharges and inductively coupled plasma atomic emission spectrometry (see earlier). In the case of ICP-OES, for example, an atlas with spectral scans around a large number of prominent analytical lines [329] is available, as well as tables with normalized intensities and critical concentrations for atomic emission spectrometers with dierent spectral bandwidths for a large number of these measured ICP line intensities, and also for intensities calculated from arc and spark tables [334]. The problem of the selection of interference-free lines in any case is much more complex than in AAS or AFS work.
5.2
Atomic emission spectrometers
As all elements present in the radiation source emit their spectrum at the same time, from the principles of AES it is clear that it is a multielement method and is very suitable for the determination of many elements under the same working conditions. Apart from simultaneous determinations, so-called sequential analyses can also be carried out, provided the analytical signals are constant. Sequential and
5.2 Atomic emission spectrometers
simultaneous multielement spectrometers both have their own possibilities and limitations. Sequential spectrometers
In most cases these include Czerny±Turner or Ebert monochromators with which the lines can be rapidly selected and measured one after another. Owing to the linerich emission spectra the focal lengths are often up to 1 m and a grating with a width of up to 100 mm and a constant of 1/2400 mm is used. These instruments require very accurate presetting of the wavelength, which is dicult on a randomaccess basis. Indeed, the halfwidth of a spectral line corresponds to a grating angle of 10ÿ4 degrees. Therefore, work can be performed in a scanning mode and with stepwise integration at dierent locations across the line pro®le or advanced techniques such as optical encoders and direct grating drives can be used. Systems employing ®xed optics with a multislit exit mask, a moveable photomultiplier and ®ne wavelength adjustment with the aid of computer-controlled displacement of the entrance slit are also used. Sequential spectrometers are ¯exible as any elemental line and any background wavelength within the spectral range (usually from 200 to up to 600 nm) can be measured, however, they are slow when many elements must be determined. They are of particular use in the case of stable radiation sources such as plasma sources at atmopheric pressure or glow discharges operated under steady state sputtering conditions. Spectral background intensity measurements can be accomplished by slew-scan procedures for the grating or also for the entrance slit. However, the spectral background and the line intensities in the case of monochannel instruments are not measured simultaneously, which is actually possible with array detectors. An interesting device for rapid scanning of spectra (the whole analytical range from 165 to 750 nm in 2 min) with a high resolution (7 pm line widths measured) and with individual integration times of 2.5 ms is the IMAGE system [93]. The high dynamic range is realized through rapid changes in the photomultiplier voltage while keeping the currents within a narrow range. The precision of wavelength access is so good that spectrum subtraction readily becomes possible as shown in the case of ICP-OES for the determination of Zr in a solution containing Fe (Fig. 90). Simultaneous spectrometers
Today these virtually always make use of photoelectric radiation detection, whereas earlier, spectrographs with photoplate recording were in wide use, especially for qualitative and semi-quantitative analyses. Simultaneous photoelectric spectrometers usually have a Paschen±Runge optical mounting. They have many exit slits and are particularly suitable for rapid determinations of many elements in a constant and well-understood sample matrix and the achievable precision is high. However, the analysis program is ®xed and accordingly simultaneous spectrometers are not suitable for the analysis of random samples. Owing to the stability require-
203
204
5 Atomic Emission Spectrometry Fig. 90. ICP atomic emission spectra for a 10 g/L Zr solution with and without 2 mg/L Fe in the vicinity of the Fe II 234.349 nm (a) and Fe II 239.562 nm (b) lines obtained with the IMAGE approach. (Reprinted with permission from Ref. [342].)
ments larger exit slits are often used, through which thermal drifts can be overcome and lower spectral resolution is obtained. In applications of trace analysis in particular, background correction is often required. This can be achieved by computer-controlled displacement of the entrance slit or by a rotation of a quartz refractor plate behind the entrance slit (Fig. 91). Here slew-scan procedures also enable scans from dierent samples to be superimposed, such as solutions of real samples, the dissolution reagents required and water e.g. in ICP-OES. In the case of copper a positive error could be attributed to a coincidence with a molecular band, probably from the BaO species (Fig. 92) [343]. As a result of the decrease in the price of CCDs, providing up to 20 CCDs is now feasable, covering the whole spectral range from the VUV to the visible, even including the alkali element lines with the aid of a supplementary grating, as shown in Fig. 93 for a commercially available system. Such a system makes line identi®cation very easy, as has already been shown by earlier work where a SIT vidicon was coupled to a conventional Czerny±Turner monochromator. By spectrum comparisons, analytical lines of the rare earths in a complex dissolved mineralogical sample in ICP-OES could easily be found. This is done by comparing back with stored spectra from pure analyte solutions, and at the same time background correction becomes easy as every wavelength is digitally accessible, and through appropriate software even the simulation of background spectra and their subtraction is possible. When the whole analytical range becomes accessible multiline calibration becomes easy, also in post-measurement evaluation when the spectra are saved. This
5.2 Atomic emission spectrometers Fig. 91. Background correction with a computercontrolled simultaneous ICPOES spectrometer.
may be very useful for improving or controlling the analytical accuracy limitations resulting from spectral interferences. A further advantage of a multi-CCD system is the linear dispersion and the onedimensional spectral dispersion. This both facilitates comparison of recorded and saved spectra and enables a high radiation throughput as the whole height of the
Fig. 92. Spectral interferences in ICP-OES as recorded in wavelength scans obtained by displacement of the entrance slit. 3 kW Ar-N2 ICP, Baird 1000 1 m spec-trometer, sample: 0.5534 mg of airborne dust sample HBW 2 per mL (±), 19 mg/mL of HBO3 , 19 mg/mL of HF, 4.9 N HNO3 (± ±); the same after the decomposition (± . ± . ); water ( ), line: Cu I 324.754 nm. (Reprinted with permission from Ref. [343].)
205
206
5 Atomic Emission Spectrometry
Fig. 93. ICP-OES spectrometer CIROS±Paschen±Runge circle with detector alignement. (Courtesy of Spectro Analytical Instruments.)
entrance slit can be used. However, there certainly are line interference problems, as with 20 CCDs of 1024 pixels each the spectral resolution for a 1 m Paschen± Runge instrument with a grating constant of 1/2400 mm may become too low in a number of cases. Echelle spectrometers
In addition, Echelle spectrometers are often used [50]. By combination of an ordersorter and an Echelle grating either in parallel or in crossed-dispersion mode, high practical resolution (up to 300 000) can be realized with an instrument of fairly low focal length (down to 0.5 m) (Fig. 94). Therefore, the stability as well as the luminosity are high. By using an exit slit mask with a high number of preadjusted slits, highly ¯exible and rapid multielement determinations are possible. As high-quality solid-state detectors are now available (Table 9) relatively compact high-resolution electronic spectrometers can be realized [344]. According to the grating equation, the wavelength dierence Dl
k between two subsequent orders with the same diraction angle b, which also gives the spectral length of order k is
5.2 Atomic emission spectrometers
Fig. 94. Echelle spectrograph with internal order separation in a tetrahedral mounting. (1): Entrance slits, (2): spherical collimator mirror, (3): prism, (4): Echelle grating, (5): spherical camera mirror, (6): focal plane (A).
Part of an iron spectrum recorded with an Echelle spectrograph with collimator and camera focal length of 500 mm, original size of the part: 20 15 mm 2 (B). (Reprinted with permission from Ref. [344].)
given by: Dl
k l
k ÿ l
k 1 l
k=k
285
where l
k is the central wavelength of the order k. To achieve the splitting of the spectral range of interest into as many sub-spectra as are necessary, the number k has to be according to Eq. 285 and the grating equation: dgl
286
For a groove distance e.g. of d 13:33 mm (75 grooves per mm) the wavelength range between 190 and 850 nm is split into approximately 100 orders (k 125±28), which superimpose for a 68 within a diraction angle ranging from b min 56:5 to b max 63:8 . Another specialty of the Echelle grating relates to the angular dispersion, normally given by: db=dl k=
d cos b
287
and for the case of autocollimation of an Echelle grating (a b Y B , the Blaze angle: db=dl 2 tan Y B
288
207
Type
CCD CCD CCD CCD CCD CCD CID
Detector
Kodak KAF-1000
Kodak KAF-6300
EEV 05-30
SITeCCD-10245B
Reticon RA2000J
DALSA1A-D9-5120
CIDTEC CID 38 PPR
1024 1024
24:6 24:6 3027 2048
27:65 18:43 1242 1152
27:5 25:9 1024 1024
24:6 24:6 2048 2048
27:6 27:6 5120 5120
61:44 61:44 512 512
14:34 14:34
24:0 24:0
28:0 28:0
12:0 12:0
13:5 13:5
24:0 24:0
22:5 22:5
9:0 9:0
Number of pixels (area/mm 2 )
Pixel area/mm 2
40±60%, 200±1100 nm
30%
50%, 200±1050 nm
70%, 190±1050 nm
45%, 400±1080 nm
45%, 400±1080 nm
35%, 400±1080 nm
Max. quantum (10% limits)
Tab. 9. Performance of selected charge-coupled devices. (Reprinted with permission from Ref. [344].)
1 500 000
130 000
100 000
400 000
500 000
85 000
500 000
Full well capacity/eC
100 g/L) and has found considerable use e.g. for water analysis [372], and especially for the analysis of seawater, brines and even oils. The multielement capacity is limited, as the optimal excitation zones for most elements dier slightly, and thus as a result of the small observation zones considerable losses in the power of detection occur when using compromise conditions. Owing to the high resolution of the Echelle spectrometer, DCP-AES is also very useful for the analysis of materials with line-rich emission spectra, such as highalloy steels, tungsten, molybdenum and other refractory metals. It has also entered into speciation work, where applications such as the determination of metalloproteins by coupling DCP-OES with chromatography have been described [373]. 5.5.1.2
5.5 Plasma source AES
5.5.2
Inductively coupled plasma AES
Inductively coupled plasmas (ICPs) in their present form go back to the work of Reed [374], who used these sources for crystal growth in the 1960s. They were introduced into spectrochemical analysis, by Green®eld et al. in 1964 [375] and by Wendt and Fassel [376], both of whom used them in their present form as sources for atomic emission spectrometry. 5.5.2.1 The inductively coupled plasma
In this type of plasma the energy of a high-frequency generator (in the frequency range of around 6±100 MHz) is transferred to a gas ¯ow at atmospheric pressure (mostly argon) in a quartz tube system with the aid of a working coil (Fig. 98). The electrons take up energy and collide with atoms, by which a plasma with a temperature of up to 6000 K is formed. At a suitable gas ¯ow, geometry of the torch and frequency, a ring-shaped toroidal plasma is formed, where the sample aerosol passes centrally through the hot plasma. The burner consists of three concentric quartz tubes. The aerosol is led with its carrier gas through the central tube. Between the outer and the intermediate tube a gas ¯ow is introduced tangentially. It takes up the high-frequency energy but also prevents the torch from melting. A torch of the type according to Fassel has a diameter of ca. 18 mm and can be operated at 0.6±2 kW and with 10±20 L/min of argon. In so-called mini-torches gas consumption can be down to about 6 L/min, through the use of a special gas inlet
Fig. 98. ICP-optical emission spectrometer. (a) gas supply, (b) sampler, (c) peristallic pump, (d) nebulizer, (e) nebulization chamber, (f ) hf generator, (g) ICP torch, (h) spectral apparatus, data acquisition, data processing.
219
220
5 Atomic Emission Spectrometry
or by further reduction of the torch dimensions [377]. It has been reported that ICPs can even be operated at powers of down to 200 W and with an argon consumption as low as 1 L/min [378]. However, this plasma is no longer a robust source that is capable of taking up wet aerosols of any aqueous sample solution [379]. Green®eld used a larger torch (outer diameter of up to 25 mm), in which an ICP can be operated at higher power and where also argon, oxygen or even air can be used as the outer gas ¯ows (up to 40 L/min). Here an intermediate argon ¯ow (about 8 L/min) is used. With the Fassel torch an intermediate gas ¯ow (about 1 L/ min) is useful when analyzing solutions of high salt contents or organic solutions, so as to prevent salt or carbon deposition, respectively. The operation of a plasma based on air has also been tried speci®cally for environmental monitoring purposes [380, 381]. The plasma can only be succesfully ignited using argon and then slowly changing the gas from argon to air or nitrogen. These discharges are operated at 1.5±2 kW, and with respect to smoothness of the spectral background and power of detection are much less powerful than the argon ICP. As in addition they are dicult to ignite and to keep stable, they have not achieved a breakthrough as yet. The ICP is almost in local thermal equilibrium. Indeed, the excitation temperatures (from atomic line intensity ratios) are about 6000 K [382] and the rotation temperatures (from the rotation lines in the OH bands) are 4000±6000 K (see Refs. [383, 384]). From the broadening of the Hû-line, an electron number density of 10 16 cmÿ3 is obtained, whereas from the intensity ratio of an ion and an atom line of the same element the electron number density found is 10 14 cmÿ3 . It has also been reported that measured line intensity ratios of ion to atom lines are higher by a factor of 100 than those calculated for a temperature of 6000 K and the electron number density found is 10 16 cmÿ3 . This indicates the existence of over-ionization. This can be understood from the excitation processes taking place. They include the following.
. Excitation through electrons or processes involving radiation:
.
Ar e ! Ar , Ar m (metastable levels: E @ 11:7 eV), and Ar (E i 15:7 eV) Ar hn ! Ar (radiation trapping) M e ! M , M e (electron impact) M e ! M hn (radiative recombination) Excitation through argon species: Ar M ! Ar M Ar m M ! Ar M (Penning eect) Ar M ! Ar M (charge transfer)
Accordingly, an over-population of the argon metastable levels would explain both the over-ionization as well as the high electron number density in the ICP. Indeed, it could be accepted that argon metastables act both as ionizing species and at the same time are easily ionized [385]. This could explain the fairly low interferences caused by easily ionized elements and the fact that ion lines are excited very e-
5.5 Plasma source AES
ciently, despite the fact that their norm temperatures are much higher than the plasma temperatures. However, the discrepancies shown are not encountered to such an extent when the temperatures from the Saha equation are used in the calculations [386]. Nevertheless, it can be accepted that a number of processes are predominant in well-de®ned zones of the plasma, as indicated by spatially-resolved measurements of various plasma parameters (see e.g. Ref. [387]). New detailed knowledge can be achieved by using Thomson scattering, with which both electron number densities and electron temperatures can be determined irrespective of whether the plasma is in local thermal equilibrium or not, and this can be achieved with high spatial resolution. Several investigations on the contribution of dierent excitation mechanisms, the in¯uence of easily ionized elements on the processes in the dierent zones of the plasma and the in¯uence of working parameters have been published by Hieftje and coworkers [388±390]. 5.5.2.2 Instrumentation
ICP generators are R-L-C-circuits producing a high-frequency current at high power (Fig. 98). The end-stage can be a triode, but nowadays transistor technology can also be used. The hf energy can be led to the remote plasma stand with the aid of a coaxial cable. This requires the use of a matching unit so as to tune the plasma optimally before and after ignition with the generator. However, in other systems the working coil can be provided as an integral part of the oscillator. Generators with a ®xed frequency as well as generators where the frequency can change with the load are also used. In the ®rst case, a change in the load changes the ratio of the forward power and the power re¯ected to the generator and necessitates retuning, whereas in the second case a small frequency shift occurs. Both types of generators may perform equally well provided the power supply to the plasma, which usually amounts to about 50% of the energy dissipated in the generator, remains constant. The stability of the power and the constancy of the gas ¯ows are both essential in order to obtain the highest analytical precision. ICP generators with frequencies ranging from 27 to 100 MHz have been used and in general 27.12 and 40 MHz generators are now available. As a result of detailed investigations by Mermet and coworkers only slight dierences in the performance in terms of detection limits and matrix interferences have been measured (see e.g. Ref. [391]). The ICP can be viewed side-on as well as end-on for atomic emission spectrometry. Whereas in the beginning side-on observation only was used, end-on observation was later found to have a number of advantages: (i) the optical pathway is longer, which may remove detector noise limitations especially when working at low UV and VUV wavelengths; (ii) it is easy to realize a system that is transparent at VUV wavelengths, enabling determinations to be made of elements such as the halogens, phosphorus, sulfur but also many metals at lines free of spectral interferences and with high line-to-background intensity ratios [392]. Therefore, limitations with respect to the linear dynamic range and higher easily ionized element interferences as a consequence of the larger in¯uence of geometry changes of the plasma may occur [393].
221
222
5 Atomic Emission Spectrometry
In plasma spectrometry the sample must be brought as a ®ne aerosol into the plasma, which as a rule is achieved by pneumatic nebulization. The aerosol should enter the hot plasma zones with a low injection velocity so as to realize residence times in the ms range that are required for an ecient atomization and excitation. Therefore, the carrier gas ¯ow should not exceed 1±2 L/min. With advanced pneumatic nebulizers for ICP spectrometry an aerosol with a mean droplet diameter in the 5 mm range can be obtained with an eciency of a few % [139, 140] for a gas ¯ow of 1±2 L/min and a sample uptake rate of 1±2 mL/min. The concentric nebulizer made of glass (Meinhard nebulizer) is self-aspirating, the cross-¯ow nebulizer may have capillaries made of dierent materials and is pump-fed but is less sensitive to clogging in the case of solutions containing high salt contents. The Babington-type nebulizer can be used successfully for slurry nebulizations as was ®rst shown by Ebdon and Cave [116]. Later it was used in ICP-AES for many types of powder analyses including also ceramic powders [117]. The fritted disk nebulizer has not found wide use. Apart from those mentioned, new approaches for ICP-AES lie in the use of the jet-impact nebulizer [142], direct injection nebulization [121] and high-pressure nebulization [143] (see Section 3.1). The high eciency nebulizer [110] works at high pressure delivering a high nebulization eciency at very low sample uptake, which is ideally suited for coupling ICP-AES or mass spectrometry with chromatography for speciation purposes. High eciency is also obtained with the so-called oscillating nebulizer, as diagnosed when using dierent advanced aerosol study techniques, as reviewed by McLean et al. [122]. In plasma spectrometry the idea of micro total analysis systems (mTAS) incorporating sample preparation, trace±matrix separation and determination in one system has also become increasingly important. The miniaturization of all steps beginning with those of sample pretreatement allows it to work with the most minute amounts of sample as e.g. required in biomedical research, to use lowest amounts of reagent and to minimize chemical waste problems. To this aim the use of high-eciency nebulizers working with minute amounts of sample is very suitable. So far work has been done with capillary electrophoresis, which can easily be housed in microstructured systems on a chip, as shown e.g. by Bings et al. [394]. In pneumatic nebulization for ICP-AES continuous sample feeding requires a sample aspiration time of about 30 s, so as to attain a stationary signal, a measurement time of around 5±10 s and a rinsing time again of 30 s at minimum. However, discrete sampling is also possible with injection systems known from ¯ame AAS [125, 126] and by ¯ow injection. Work with sample aliquots of down to 10 mL then becomes possible, which e.g. is particularly useful when it comes to working with microsamples [140], or for the analysis of solutions containing high salt contents [395]. For ICP-AES both sequential and simultaneous as well as combined instruments are used. In sequential spectrometers special attention is given to the speed of the wavelength access and in simultaneous spectrometers to the provision of background correction facilities. In combined instruments a number of frequently used channels are ®xed and with a moving detector or an integrated monochromator
5.5 Plasma source AES
access to further lines is possible. Echelle instruments gained interest as they allow the highest resolution at low focal length to be realized and as they can be successfully combined with advanced detectors such as CCDs. Paschen±Runge spectrometers with a large number of CCDs covering the whole analytically usable spectral range, certainly for ICP-AES, allow full use to be made of the spectral information at all wavelengths simultaneously. 5.5.2.3 Analytical performance
Power of detection
The nebulizer gas ¯ow, the power, and the observation height are important parameters to be optimized so as to get the highest power of detection for an element (single element optimization) or a number of elements (compromise conditions). Such optimizations can be done by single-factor studies or Simplex optimization (see e.g. Refs. [396, 397]). The nebulizer gas in¯uences the aerosol droplet size and nebulizer eciency but also the plasma temperature. Accordingly, for each element and line there will be a fairly sharp optimum for the aerosol gas ¯ow where the line-to-background ratio is maximum. The power mainly determines the plasma volume and will be optimum for so-called soft lines (atom lines and ion lines of elements with low ionization potential) at rather low values (1±1.2 kW), whereas for hard lines (lines of elements with high excitation potential and highly energetic ion lines) as well as in the case of organic solutions the optimum value is higher (1.5±1.7 kW). The optimum observation height will be a compromise between the analyte number densities (highest in the lower zones) and the completeness of atomization and excitation (a few mm above the working coil). Only elements such as the alkali metals are measured in the tail ¯ame. Usually an observation window (e.g. 2 2 mm 2 ) that is not too small is selected so as to secure suciently high photon ¯uxes through the spectrometer in order to avoid detector noise limitations. As the analytical lines in ICP-AES have a physical width of 1 to up to 20 pm and the spectra are line rich the use of a high-resolution spectrometer is required. The detection limits for most elements are at the 0.1 (Mg, Ca, . . .) to 50 ng/mL level (As, Th, . . .). In particular, for elements which have refractory oxides (such as Cr, Nb, Ta, Be, rare earths, . . .), a fairly low ionization potential and sensitive ion lines, the detection limits are lower than in AAS. On the contrary, the detection limits for elements with high excitation potentials, such as Cd, Zn and Tl, are lower in AAS than in ICP-AES. The detection limits for B, P, As, . . . , which have more sensitive atom lines, are somewhat lower in DCP-AES than in ICP-AES. For P, S, N, O and F, the most sensitive lines are at VUV wavelengths and here with vacuum spectrometers and VUV transparent pathways and optics, values of 450 ng/mL (Br), 200 ng/mL (Cl), 55 ng/mL (I), 3.9 ng/mL (P) and 3.5 ng/mL (S) can be obtained [348]. The detection limits obtained by radial and axial observation were found to be up to one order of magnitude lower in the axial as compared with the radial viewing, as reported by Ivaldi and Tyson [393]. These results were obtained with an Echelle
223
224
5 Atomic Emission Spectrometry
spectrometer using CCD detection. The results were found to be partly due to better signal to background ratios but also to lower RSD values. The latter may be understood from the better stabilization of the plasma. Interferences
Most interferences in ICP-AES are additive interferences and relate to coincidences of spectral background structures with the elemental lines used. They are minimized by the use of high-resolution spectrometers and background correction procedures. Indeed, in the spectral wavelength range of 200±400 nm more than 200 000 atomic emission lines of the elements have been listed and in addition molecular bands also occur. Accordingly, coincidences will often occur within the spectral line widths themselves. In order to facilitate the selection of interferencefree spectral lines in ICP-AES, Boumans worked out a system where the interferences for about 900 prominent lines [334] are given for spectrometers with a certain spectral slit width in terms of critical concentration ratios. Further line tables for ICP-AES [398] and atlases [329] are available, although incomplete. In ICP-AES multiplicative interferences also occur. They may have several causes, namely:
. Nebulization eects: As discussed earlier, dierences in the physical properties of
.
the dierent sample and calibration solutions lead to variations in the aerosol droplet size and thus also in the eciency of the nebulizer and the sample introduction. This eect is strongest in the case of free sample aspiration and relatively low nebulizer gas ¯ow and can be minimized (see Section 3.1). Easily ionized elements have a complex in¯uence: Firstly they may cause a decrease in excitation temperature, as energy is consumed for ionization. Further, they may shift the ionization equilibrium for partially ionized elements as they in¯uence the electron number densities through their easy ionization. They may also cause changes in plasma volume as a result of ambipolar diusion or eventually quench exciting species such as metastables [399]. After correcting for the in¯uence of easily ionized elements on the spectral background, signal enhancements or suppressions are negligible up to concentrations of 500 mg/mL of Na, K, Ca or Mg. At larger dierences in concentration matrix-matching of the standards should be applied.
Ivaldi and Tyson [393] reported that interferences caused by NaCl were found to be nearly twice as high in the case of axial viewing than in radial viewing and also atom to ion line intensity ratios were in¯uenced more severely in the case of axial viewing. Analytical precision
As the sample introduction into the plasma is very stable, RSDs in ICP-AES are about 1% and the limiting noise is proportional to this, including ¯icker noise and nebulizer noise.
5.5 Plasma source AES
However, as shown in the noise spectra (Fig. 9), there is a frequency dependent component at about 200 Hz, this being due to the gas ¯ow dynamics, as has been mentioned by Belchamber and Horlick [400]. The ¯icker noise present increases when particles are entering the ICP that can no longer be completely evaporated. This is e.g. the case when blowing powders into the ICP. At wavelengths below 250 nm, however, detector noise limitations may occur especially with high-resolution spectrometers. For integration times exceeding the 1 s level, the full precision is normally achieved. In the case of CCD detection, readout noise may also become important. With the aid of an internal standard, ¯uctuations in sample delivery in particular and up to a certain extent changes in excitation conditions can also be compensated for. To appreciate the full bene®t of internal standardization, the intensities of the analytical line and the internal standard line, and in trace determinations eventually also of the spectral background near the analytical line, should be measured simultaneously, which can be easily done in the case of a CCD spectrometer. The full analytical precision of ICP-AES in fact can be thus obtained. This can actually be demonstrated by the determination of the main components, as is required in stoichiometric determinations of advanced ceramics. After optimizing the sample dissolution procedure for the case of silicon±boron±carbo±nitride, which can be made speeded up by microwave-assisted decomposition with acids [401] silicon at the 60% level could be determined with a sequential spectrometer with an absolute error of analysis (standard deviation inluding replicated analyses as well as decompositions) of 1% or better. In the case of high-temperature superconductors, the precision at the 0.4±3% (m:m) level for PbaBiaSraCaaCuaO superconductors can be even higher, which is understandable from the less stringent dissolution conditions required and the thermochemical properties of the related elements, which are also less prone to volatilization losses and contamination from the laboratory environment. With the aid of analyses with an ICP-CCD based spectrometer, formulae can be determined as: Pb0:404G0:001 Bi1:795G0:007 Sr1:95G0:02 Ca2:11G0:02 Cu3:04G0:02O10:80:1 [402]. In many cases this results in relative errors of 1% and lower. This is possible by simultaneous measurements of line and background intensities, leaving more room for improvements by internal standardization. This, however, was found to be very critical with respect to line pair selection. It can thus be seen that ICP-AES could now replace spectrophotometric and titrimetric methods completely with respect to the precision achievable, also when considering the main components, as in addition it has been shown to be suitable for advanced working materials [403]. The principles of the selection of line pairs in the case of ICP-AES have been described well and in detail by Fassel et al. [404]. For an intensity ratio of two atomic emission lines from two dierent elements: Iqp
Xo =Isr
Ro fn
Xo A qp nqp gq Q Ro =n
Ro Asr nsr gs Q
Ro g fexp
Es ÿ Eq =kT g
294
Fluctuations in number densities of both elements X (analyte) and R (reference
225
226
5 Atomic Emission Spectrometry
element) are cancelled. In order to keep the in¯uence of ¯uctuations in the excitation temperature low, lines with similar excitation energies should be used and also their partition functions should be of the same order of magnitude, rendering the choice dicult. Two atom or two ion lines should also be selected and in order to correct for changes in sample delivery, the ionization has to be taken into account: n
X =n
R n
X K
Xo!X
n 2e K
Ro!R n e K
Ro!R K
R!R = n
R K
Ro!R
n 2e K
Xo!X n e K
Xo!X K
X!X2
295 with: K
Xo!X n
X n e =n
2 p m e k T 3=2 =
h 3 g 2 Q
X =Q
Xo expÿEi
Xo =
k T
296 K
X!X2 n
X2 n e =n f
2 p m e k T 3=2 =
h 3 g 2 Q
X2 =Q
X expÿEi
X =
k T
297
and similarly for R. These equations show that the analyte and reference elements should have ionization energies of the same order of magnitude. Changes occurring similarly for analyte and reference element number densities can be fairly well compensated for by selecting Iqp =Isr as the analytical signal, however, this applies less for changes in the temperatures. Departures from LTE in the ICP should be taken into account and also that analytical lines are selected for which the norm temperatures are far above the plasma temperatures. This is still more stringently controlled in the case of multielement determinations, where the selection of dierent reference elements for a number of analytes may actually be required. The possibilities of simultaneous background correction in the case of a twochannel spectrometer have been described in early ICP papers by Meyers and Tracy [405]. In the case of brass alloys with internal standardization, RSDs below 0.1% can be obtained with simultaneous ICP-AES spectrometers for copper and zinc, where ICP-AES comes near to the precision achievable with x-ray ¯uorescence spectrometry for metal analysis [406]. Multielement capacity
The multielement capacity of both sequential and simultaneous ICP-AES is high as is the linear dynamic range (up to 5 orders of magnitude). It is restricted on the lower side by the limits of determination and at the higher concentrations by the
5.5 Plasma source AES
beginning of self-reversal. This only starts at analyte concentrations of about 20 000 mg/mL, which is due to the fact that analyte atoms hardly occur in the cooler plasma zones. The linear range for axial viewing and radial viewing of the plasma may dier somewhat. For axial viewing as the optical pathway is longer, the selfabsorption and risks of self-reversal increase. The latter in particular is more likely when the analyte has to pass through a longer but cooler zone. This can be avoided both by skimming the ICP as well as by the use of a gas ¯ow entering perpendicular to the ICP axis and blowing the analyte away from the cooler zones. These dierences in linear dynamic range were also reported by Ivaldi and Tyson [393]. Special techniques
The analytical performance of ICP-AES can be considerably improved by using alternative techniques for sample introduction. Organic solutions Solutions such as in MIBK, xylene or kerosene can also be nebulized pneumatically and determinations in these solvents can be performed directly with ICP-AES (see e.g. Ref. [407]). The addition of oxygen or the use of higher power is helpful so as to avoid carbon deposits. It has also been suggested that the use of burners with sharp edges and improved ¯ow dynamics are very helpful at preventing carbon deposits [408]. Oils can be analyzed after dilution 1:10 with xylene (c L : 0.5±1 mg/g). Although xylene has been used as a diluent, however, MIBK and other ketones may be more problematic especially at a lower powers of operation. Heavy metals can be extracted from waste waters as APDTC complexes with organic solvents (such as MIBK, xylene or CCl 4 ) and thus separated from the accompanying easily ionized elements that cause interferences [407]. Advanced pneumatic nebulizers There has been increasing interest in these nebulizers for ICP-AES. Maximum dissolved solids nebulizers [111] are usually cross-¯ow nebulizers operated at a high pressure. They are often made from PTFE, including the capillaries, and thus can be used for work with HF containing solutions. In this case a nebulization chamber made of PTFE and an ICP torch with an internal tube made of Al2 O3 ceramics or BN should be used. High-eciency nebulizers are also cross-¯ow nebulizers, usually operated at very high pressures and are available from several manufacturers for ICP work. Fritted disk nebulizers have not made a great breakthrough for practical ICP work. More applications can be found for the Babington nebulizers when it comes to slurry nebulization for ®ne powders, based again on Ebdon's work. The LeÂgeÁre nebulizer and the grid nebulizer are other types of Babington nebulizers that have been developed and have been shown to oer good eciencies at relatively low gas ¯ows. Systems delivering monodisperse aerosols are very worthwhile for the study of evaporation processes and because of their relationship to the power of detection and interference eects in ICP-AES. Towards this aim, the MDMI (monodisperse dried microparticulate injector) has been used by Olesik (see e.g. Ref. [409]). It is a sys-
227
228
5 Atomic Emission Spectrometry
tem that combines a droplet generator and an oven, reproducibly generating monodisperse droplets of controllable size and velocity on demand. Ultrasonic nebulization This has been applied since the early work on ICP-AES [151]. Both nebulizer types where the sample liquid ¯ows over the nebulizer transducer crystal and types where the ultrasonic radiation (at 1 MHz frequency) is focussed through a membrane on the standing sample solution have been used. When applying aerosol desolvation the power of detection of ICP-AES can be improved by a factor of 10 by using ultrasonic nebulization. This speci®cally applies to elements such as As, Cd and Pb, which are of environmental interest. However, because of the limitations discussed in Section 3.2, the approach is of particular use in the case of dilute analytes such as in water analysis [150]. Additional ®ne detailed development, however, is regularly carried out, as with ICP-AES the process is crucial for elements such as Cd, As and Pb for which threshold values in fresh water samples can just still be measured reliably with this type of sample introduction. Such a development is the microultrasonic nebulizer (mUSN) operated with argon carrier gas, as described by Tarr et al. [410]. High pressure nebulization As has been described by Berndt, this has been shown to provide both a gain in power of detection as well as good short- and long-term stability in ICP-AES. It is interesting to see the gain in power of detection when the process is performed at high temperature [145]. A further development is its on-line coupling with sample decomposition of suspended material, which is important for water and for biological samples [130]. Thermospray
Since their introduction in the 1980s thermospray systems have been shown to increase the aerosol generation eciencies and accordingly the power of detection. In ICP-AES they have been further developed into systems where a gain in the power of detection and high accuracy can be achieved for real samples as compared with conventional pneumatic nebulization [411]. It should be kept in mind that both thermospray nebulization and high-pressure nebulization [143] successfully allow the analyte introduction eciency to be increased and thus also the power of detection, however, again only when aerosol desolvation is applied. They are especially interesting for speciation by on-line coupling of ICP-AES and HPLC, as shown later. Electrothermal evaporation and direct sample insertion This procedures allow the sampling eciency to increase so much that work with microsamples becomes possible and at the same time the power of detection for the concentration can be improved further. ETV-ICP-AES has been applied with graphite cups [189], graphite furnaces [186] but also with tungsten ®laments [412]. It has been applied not only with dry solution residues but also with dry suspension residues. The latter approach has been developed extensively by Krivan and coworkers and applied to the analysis of refractory powders (e.g. AlN and Al2 O3 [413]). In the case of Al2 O3 powders, it was shown by Wende [34] that volatilization aids were very helpful at increasing the
5.5 Plasma source AES
Volatilization of dry solution aliquots from a graphite boat evaporation system coupled to a simultaneous ICP optical emission spectrometer at different heating currents [34]. Fig. 99.
number of elements that can be volatilized from the matrix. It was also possible to avoid volatilization of the matrix in a number of cases. The use of AgCl as a thermochemical reagent, even in the case of relatively coarse powders weighed into a sampling boat, resulted in a fairly good volatilization of Ca, Fe, Na and Si from the matrix, however, volatilization occurs according to the boiling points of the analytes (Fig. 99). Direct sample insertion allows the direct analysis of used oils [189] and of microamounts of sediments [193] as well as the determination of volatile elements in refractory matrices [191]. Diculties lie in the calibration and in the signal acquisition. The latter necessitates a simultaneous and time-resolved measurement of the transient signals for the line and background intensities to be made in trace analysis. This may become easier when applying CCD detection with the appropriate software. Also in the case of direct sample insertion the use of thermochemical reagents has been found to be useful [414]. An extremely sensitive technique lies in its combination with laser ablation [415]. A very high degree of sample introduction eciency is reached, so the power of detection may be the highest possible in plasma atomic emission spectrometry. Hydride generation As compared with pneumatic nebulization, hydride generation with ¯ow cells [154] allows the detection limits for As, Se, etc. to be decreased by 1.5 orders of magnitude down to the ng/mL level. Thus a hydride generator is a useful accessory to ICP-AES for water analysis. However, the limitations due to possible systematic errors from heavy metals and from analytes being in the non-mineralized form remain. Working at high acid concentrations, the use of complexing agents to mask the heavy metals causing the interferences (tartrate in the case of copper for example) and mineralization by treatment with sulfuric acid until fuming to decompose organoarsenicals is helpful. Electrochemical hydride generation avoids blank contributions from the high amounts of acids that have to be used as well as from the
229
230
5 Atomic Emission Spectrometry
sodium borohydride and the NaOH required to stabilize its solutions. This was ®rst proposed by Liu et al. [416] for AAS work and was optimized by Hueber and Winefordner [417]. Comparative studies of conventional and electrochemical hydride generation with ICP-AES have also been performed and showed that the detection limit of 0.8 ng/mL for conventional hydride generation can certainly be improved by electrochemical hydride generation. Gas and liquid chromatography Coupled with ICP-AES these are useful approaches for speciation work, as shown by the early work of Cox et al. [418] for Cr III /Cr VI speciation, where Al2 O3 ®lled columns and elution with acids and NaOH were used. Anion exchangers could also be used. New impetus has arisen from the interfacing of thermospray and high-pressure nebulization to ICP-AES. ICP-AES remains of interest for the element-speci®c detection in the speciation of silicon compounds and is a good alternative to ICP-MS, where considerable spectral interferences hamper the power of detection that can be obtained. Direct solids sampling This is of special interest, despite the fact that ICP-AES is mainly of use for the analysis of liquids. In a number of cases it enables the same precision and accuracy to be obtained as in work with solutions but without the need for time-consuming sample dissolution involving analyte dilution. For compact solids arc and spark ablation are a viable approach for metals [100, 214]. Aerosols with particle sizes at the mm level [216] and detection limits at the mg/g level are obtained [213]. Owing to the separate ablation and excitation stages, matrix in¯uences are particularly low, as shown for aluminum [100] and for steel samples [213]. In the ®rst case, only for supereutectic silicon concentrations were matrix eects obtained (Fig. 100). For low-alloyed samples straight calibration curves are obtained and in the case of high-alloyed steels, even samples with widely dierent Cr or Ni contents are on the same calibration curves, which, are in fact slightly curved. Laser ablation can be used with similar ®gures of merit [419], also for electrically non-conducting samples and with lateral resolution (up to around 10 mm). The aerosol produced with a Nd:YAG laser has a particle size in the mm range. With steel samples, no selective volatilization was found to occur, which contrasted sharply with the case of brass. However, in the latter case spark ablation was also found to result in a high selective volatilization [225]. A very stable material ablation was possible with the so-called LINA (laser induced argon spark) spark system [297], where through a rotation of the focussing lens the plasma is moved over the sample surface. With ICP-AES, detection limits in steels were found to be in the mg/g range, even when measuring the spectral background with a high-purity steel sample (Table 12) and, as shown for a standard reference sample, fairly good analytical accuracy was obtained. It was found that by using the intensity of an Fe line as the internal standard signal the precision could be signi®cantly improved. Applications ICP-AES is now a routine analytical method, which is of speci®c use when a large number of elements have to be determined in many samples, whether they are 5.5.2.4
5.5 Plasma source AES
The analysis of Al and AlSi alloys by spark ablation coupled with ICP-OES. 3 kW argon-nitrogen ICP, Spectrovac 1000 and 1 kV spark at 25 Hz. (Reprinted with permission from Ref. [100].)
Fig. 100.
Tab. 12. Detection limits obtained with LINA spark coupled with sequential ICP-AES [217].
Sample matrix
Element/line (nm)
Concentration range ( %)
Limit of detection ( %)
Low-alloyed steel
Ni II 231.604 Cr II 267.716 Mo II 281.615 Mn II 293.306 Si I 212.412 Fe II 259.940 Ni II 231.604 Sn II 189.989 Pb II 220.353 Fe II 259.940 Cu I 324.754 Si I 212.412
0.59±4.45 0.102±2.95 0.09±1.00 0.047±0.95 0.033±1.46 0.022±0.73 0.009±0.36 0.19±0.92 0.49±3.01 0.0079±0.98 0.021±2.8 0.20±9.22
0.002 0.004 0.007 0.002 0.006 0.004 0.004 0.018 0.095 0.004 0.035 0.024
Brass
Aluminum
231
232
5 Atomic Emission Spectrometry
solutions or dissolved solids. Therefore, in many cases it is complementary to AAS as the power of detection of furnace AAS cannot be rivaled for most elements. ICPAES is of special interest for the analysis of geological samples, environmental analysis, clinical analysis, food analysis, metal analysis, chemicals analysis and in the certi®cation of reference materials. In the analysis of geological samples, sample dissolution with respect to nebulization eects but also to ionization interferences is important [399]. The analysis of rare earth minerals without [420] and with removal of the matrix has been reported. The analysis of environmentally-relevant samples is a major ®eld of application. Based on the work of Garbarino and Taylor [421], a method has been proposed by the US EPA (Environmental Protection Agency) [422] and later by DIN [423] for waste water analysis. The latter, standardized procedure describes the sample decomposition, the analytical range for 22 elements and frequent interferences of ICP-AES in waste water analysis. For the analysis of natural waters, hydride generation [424], preconcentration based on liquid±liquid extraction of the dithiocarbamate complexes [425], adsorption of trace elements onto activated carbon [426] and also co-precipitation [e.g. with In(OH)2 ] [427], etc. have been reported and special emphasis has been given to speciation (as given in the Refs. in [428]) and on-line preconcentration [134]. In clinical analysis, Ca, Fe, Cu, Mg, Na and K can be determined directly even in microsized serum samples [128]. Also electrothermal evaporation and direct sample insertion are particularly useful from this point of view. Generally, however, the power of detection of ICP-AES is too low for most analytical problems encountered in the life sciences. Metals analysis in particular should be mentioned in the ®eld of industrial products. Apart from solution analysis also direct metal analysis by spark ablation is very useful [100]. Through the availablity of CCD spectrometers the analytical precision that can be achieved is so high, that even for major components errors as low as or even lower than in classical chemical methods can be achieved. This is documented by the analysis results for stoichiometric determinations of hightemperature superconductors of the YaBaaCuaO type (Fig. 101) [402]. It also applies for ceramics of the SiaBaNaC type, where sample decomposition by various methods including microwave-assisted dissolution, dissolution by fusion and decomposition with acids at high-pressure combined with sequential ICP-AES successfully allowed the stoichiometry to be controlled [401]. In addition, the analysis of refractory powders by slurry nebulization for Al2 O3 [429], SiC [430] and even for ZrO2 [132] powder samples or analyses subsequent to well-investigated sample decomposition methods, also making use of microwave-assisted dissolution, [431] are very powerful methods. 5.5.3
Low-power high-frequency plasmas
High-frequency discharges at low power and with capacitive power coupling, known as the stabilized capacitively coupled plasma (SCP), as described by Gross et al. [432]
5.5 Plasma source AES
Fig. 101. Precision achievable in the analysis of hightemperature superconducting powders by sequential and simultaneous ICP-OES [402].
for element-speci®c detection in chromatography may, in addition to ICP-AES, be a useful alternative especially when advanced sample introduction techniques are used. The plasma is sustained in a water-cooled quartz tube with an internal diameter of 1 mm and is powered through two annular electrodes around the capillary and within a distance of about 10 mm. The plasma has a length of about 20 mm and can be operated with helium or argon at a power of 100 W at maximum and a frequency of 27.12 MHz. The plasma itself is very well established between the electrodes and is fairly robust. For sample introduction ¯ow-cell hydride generation can be used without the need to remove the excess of hydrogen from the reaction gases, giving a detection limit of 40 ng/mL for As. Also the discontinuous hydride generation technique using NaBH4 pellets can be used and then an absolute detection limit for As of 5 ng is obtained. These values are better by a factor of 2 than with a surfatron MIP as the atomic emission source but by a factor of 5 higher than in ICP-AES. Alternatively pneumatic nebulization can be applied provided a desolvation of the aerosol is performed and the ®gures of merit are then similar to those of an MIP according to Beenakker (detection limits in the 1±10 ng/mL range) [433]. With electrothermal evaporation from a tungsten ®lament and quartz ®ber optics, detection limits are at the 50±100 pg level for many elements, except for Fe which is subjected to spectral interferences from tungsten lines. This was established from the use of dierent working gases and especially from experiments with the addition of H2 to the argon. In the case of coupling with graphite furnace atomization Cu, Mg and Fe can be determined in serum samples without dilution for Fe and Cu and with a 1:100 fold dilution for Mg [434].
233
234
5 Atomic Emission Spectrometry
Fig. 102. Set-up for pneumatically operated direct insertion of dry solution residues into an SCP as a source for atomic emission spectrometry. (Reprinted with permission from Ref. [435].)
When working with the SCP, the plasma is so well established between the electrodes, that it is possible to insert a sample into the plasma, without moving it away. This opens up the possibility of drying sample aliquots on a quartz rod and then etching them from the rod by inserting it into the plasma (Fig. 102) [435]. Both with Ar and with He, for Pb, Mg, Cd and Cu detection limits in the 100±600 pg range are obtained for 20 mL samples, however, with considerable interferences from easily ionized elements. 5.5.4
Microwave plasmas
Microwave plasmas are operated at 1±5 GHz. They are produced in a magnetron. The electrons emitted by a glowing cathode are led through a cavity with a series of radially arranged resonance chambers to which an ultra-hf (UHF) ®eld is applied. Strong UHF currents are produced by resonance, which can be coupled out by the anode and transported by a coaxial cable. The microwave current is led to a waveguide. Only when the waveguide has dimensions below certain critical values can microwaves be transported. The transport eciency can be regulated with the aid of moveable walls and screws. A standing wave is produced in a cavity and the microwave energy is coupled into the resonator by a loop or with an antenna. Both l=2, l=4 cavities, etc. have been proposed [436] and allow tuning with respect to a favorable ratio of forward to re¯ected power. Microwave energy can also be transported through strip lines of suitable dimensions formed by metal vapor deposition or galvanically deposited on electrically non-conducting substrates. These strip lines should be protected against heating through re¯ected power losses by cooling
5.5 Plasma source AES
with a fan, a cooling-water circuit or a Peltier cooling element. They can have dimensions of below 1 mm in width, several 0.1 mm in thickness and varying length (up to 10 cm) and are very useful for attemps to miniaturize microwave resonant structures. Capacitively coupled microwave plasma
This type of single-electrode discharge goes back to the work of Mavrodineanu and Hughes [437]. Here the microwaves are led to a pin-shaped electrode and the surrounding burner housing acts as the counter electrode. A bush-form of discharge burns at the top of the electrode and the sample aerosol enters concentrically with this pin (Fig. 103). The plasma, which can be operated with 2±4 L/min argon, helium or nitrogen (at 400±800 W), is not in LTE (T ex 5000±8000 K and T g 3000±5700 K, n e 10 14 ) (see the radially and axially resolved measurements of Bings et al. [438]) and it is a current-carrying plasma. As the sample aerosol enters with a fairly low eciency into a hot zone where the background emission also will be high, the detection limits especially for elements with thermally stable oxides will be poor (see e.g. Ref. [439]), for example, Na 589 nm: 0.05; Be 234.9 nm: 0.06; Mg 285.2 nm: 0.04; Al 396.1 nm: 0.16 mg/mL. In addition, alkali elements cause high matrix eects; e.g. for 500 mg/mL they amount to 100% in the case of Mg. Therefore, spectrochemical buers are often used, such as CsCl. Because of the possibility of working with He, the source is also useful for the speciation of widely diverse elements including P, S and the halogens, as shown by Hanamura et al. [440]. A considerable advantage of the CMP is that it can also easily be operated with air only. Even in the case of organic solutions, such as diluted oils, trace determinations are then possible [441]. The high stability of an air operated CMP also allows optical emission spectrometry to be used with a CMP for real-time air quality monitoring. For environmentally relevant elements such as Fe, Cu, Pb and Cd, detection limits in air are obtained (Table 13), which are, for example, below the thresholds set by legislation in Germany [442]. It was also found that high water loadings, as are often present after the washing out of stack gases, or CO2 , as is always present in stack gases of waste incineration, signi®cantly in¯uence but do not drastically deteriorate the detection limits. It also should be possible to decrease the dimensions of a CMP considerably without deteriorating the analytical performance. In addition, wall stabilization of the plasma by additional gas ¯ows will certainly improve the robustness of the CMP still further and improve both power of detection as well as decrease the sensitivity towards variations in easily ionized element concentrations. Microwave induced plasmas
These electrodeless discharges (MIPs) are operated in a noble gas at fairly low power (mostly below 200 W). They have become important as sources for atomic
235
236
5 Atomic Emission Spectrometry
Fig. 103. Capacitively coupled microwave plasma (A) [442] and microwave induced plasma discharges (B, C). (B): MIP in a TM010 resonator according to Beenakker; (a): cylindrical wall, (b): fixed bottom, (c): removable lid, (d): discharge tube, (e): holder, (f ): coupling loop, (g): PTFE insulator, (h):
microwave connector, (i, j): tuning screws, (k): fine tuning stub, (l): holder (reprinted with permission from Ref. [442, 443] and related literature). (C): Surfatron type MIP; original version: 20 mm < lI < 250 mm, D 1 : 11 mm and D 2 : 40 mm, cut-off frequency : 2 GHz (reprinted with permission from Ref. [453]).
5.5 Plasma source AES Tab. 13. Detection limits (mg/m 3 ) obtained with OES using an air CMP with different water loading [442]. Maximally allowed concentrations of metals in flue gases in Germany : S Cd, Tl, Hg 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
254
6
Plasma Mass Spectrometry The plasma sources developed for atomic emission spectrometry have also been shown to be very suitable ion sources for mass spectrometry. This is particularly true for electrical discharges at pressures in the 1±5 mbar range and sources at atmospheric pressure since the powerful vacuum systems became available, with which the pressure dierence between the mass spectrometer (of the order of 10ÿ5 mbar) and the source can be bridged. Elemental mass spectrometry, however, goes back to the use of high-vacuum arcs and sparks, with which ultratrace and survey analyses of metal samples could be performed (for an overview, see e.g. Refs. in [68, 69]). Spark source mass spectrography with high-resolution sector ®eld mass spectrometers, is still very useful for a survey characterization of electrically-conducting solids down to the ng/g level. The spectra can be recorded on photographic plates, which are a permanent document and at least enable semi-quantitative analyses to be made. At the ng/g level this approach is suitable for the quality control of materials required in microelectronics. The technique has become very useful since computer-controlled densitometers have been available, which automatically record the blackenings of the elemental lines on the photoplates and convert them into logarithmic intensities, these being proportional to the logarithmic concentrations. Spark source mass spectrometry requires expensive sector-®eld mass spectrometers and despite the possibility of automated read-out of the spectra, highly skilled laboratory personnel for data evaluation are also required. This situation arises from the fact that high-vacuum sparks are very powerful sources producing vast amounts of multiply charged ions making the spectra line rich. The analytical precision achievable in spark mass spectrometry is low, however, special precautions such as the use of rotating electrodes have been found to be helpful. Spark source mass spectrometry has been used for the analysis of compact metals, from which electrodes could be made by turning-o or by ®xing drillings in electrode holders. In the case of metals of low melting point, such as gallium, liquid nitrogen cooled electrodes of the gallium were used. For electrically conducting as well as non-conducting powders electrodes were pressed from mixtures with gold powder, which is ductile and leads to electrodes that are electrically and heat conductive. The method is very powerful for dry solution residues as a result of its sensitivity and has been used for impurity detection in liquid aliquots obtained from trace± matrix separations in the analysis of high-purity materials.
6.1 ICP mass spectrometry
In the course of the late 1970s new mass spectrometric methods, which made use of the plasma sources known from optical atomic spectrometry came into use. They will be treated in detail and consist in particular of ICP mass spectrometry (ICP-MS) and glow discharge mass spectrometry (GD-MS), which have contributed to a considerable portion of the progress that has been made in elemental analysis as compared with spark source mass spectrometry.
6.1
ICP mass spectrometry
Plasma mass spectrometry developed on the basis of the experience gathered with the extraction of ions from ¯ames, as was done for diagnostic purposes in the mid1960s [493]. Douglas and French [494], towards the end of the 1970s, used an atmospheric pressure microwave plasma, into which they introduced dry aerosols, as an ion source for mass spectrometry and reported very low detection limits as compared with other methods of solution analysis. In 1980 Houk et al. [495] ®rst showed the analytical possibilities of ICP-MS and described the sampling of ions through a water-cooled aperture (of 0.6±1 mm) into a vacuum chamber with a pressure of a few mbar, maintained by a powerful roughing pump. From this chamber a second sampling takes place, through a second aperture of the same dimensions, into the high vacuum of a quadrupole mass spectrometer. This development became possible as a result of the improvement in stability and the spectral resolution of 1 dalton that is achievable with quadrupole mass ®lters, as well as by the further development of vacuum technology. As the method has the same excellent possibilities for sample introduction as ICP-AES but in addition enables much lower detection limits to be achieved, covers more elements and allows the determination of isotope concentrations, ICP-MS rapidly developed into an established analytical method [496]. 6.1.1
Instrumentation
In ICP-MS (Fig. 112) the ions formed in the ICP are extracted with the aid of a conical water-cooled sampler into the ®rst vacuum stage where a pressure of a few mbar is maintained. A supersonic beam is formed and a number of collision processes take place as well as an adiabatic expansion. A fraction is sampled from this beam through the conical skimmer placed a few cm away from the sampler. Behind the skimmer, ion lenses focus the ion beam now entering a vacuum of 10ÿ5 . This was originally done with the aid of oil diusion pumps or cryopumps, respectively, but very quickly all manufacturers switched to turbomolecular pumps backed by roughing pumps. The sampler and the skimmer are usually made of stainless steel and are both conically shaped with dierent cone angles. The sampler can also be made of copper, which has a better heat conductivity. In the case of HF containing solutions,
255
256
6 Plasma Mass Spectrometry
Fig. 112. Instrumentation for low-resolution ICP-MS. (a): nebulizer; (b): sampler; (c): ion optics; (d): quadrupole; (e): electronics; (f ): detector; (g): rf-generator; (h): roughing pump; (i): turbomolecular pump: (j): quadrupole rf generator.
platinum samplers can also be used. This is particularly worthwhile for the analysis of geological samples subsequent to wet chemical dissolution and removal of the silicates. The distance between the sampler and skimmer is critical with respect to the maximum power of detection and minimal ionization interferences. This also applies to the power transmission to the rf coil, where considerable dierences were found for coils powered centrally and coils powered at one of the ends [497]. The processes in the intermediate stage together with their in¯uence on the ion trajectories in the interface and also behind the second aperture (skimmer) are very important for the transmission of ions and for related matrix interferences, this being the topic of fundamental diagnostic studies (see e.g. Refs. in [496]). Vaughan and Horlick [498] used the commercial program MacSimion to model the einzel lens and the Bessel box input lens system of an inductively coupled plasma mass spectrometer and compared the results obtained for dierent lens voltage settings with experimental data. Today's commercially available instrumentation fall into three groups, namely quadrupole based instruments, high-resolution instruments and time-of-¯ight systems. For each system dierent ion extraction conditions must be selected. In a quadrupole, as a result of the high transmission, no high extraction voltage is required, which is totally dierent for a sector ®eld instrument. Here either the spectrometer or the plasma must be at a high potential. With quadrupoles a mass resolution of 1 dalton over the whole spectral range means the method suers from isobaric interferences especially at masses below 80 dalton. In the case of sector ®eld instruments, the resolution may be up to 5000 while in commercial instruments sucient transmission can still be maintained. As the spectral background intensities are low, detection limits are lower, not only as a result of the absence of a number of interferences but also because of better signal-to-background ratios [499]. For time-of-¯ight ICP-MS, both othogonal as well as linear end-on arrangements are used. The instruments available use a re¯ectron [74].
6.1 ICP mass spectrometry
Duty cycles are still fairly low as ion packages in a time scale of ms are sampled but spectrum development and acquisition still takes orders of magnitude longer. Despite high throughput, the power of detection has as yet not been fully exploited. Dual-channel mass spectrometers, where the ion beam is split inside the mass spectrometer [87] are very useful for isotope ratio work. However, they still use dierent detectors. Mattauch±Herzog arrangements with a ¯at focal plane have also been introduced for plasma mass spectrometry work. They are ideally suited for work with array detectors, which are certainly very useful for isotope ratio work, as within a narrow mass range isotope signals are actually measured simultaneously, together with their respective background signals [500]. Such detectors are available as microchannel plates coupled through a phosphor to photodiode array detectors. 6.1.2
Analytical features
The analytical features of ICP-MS are related to the production of ions in analytical ICPs as well as to the highly sensitive ion detection and the nature of the mass spectra themselves. On the basis of the Saha±Eggert equation, the degrees of ionization for the chemical elements can be calculated for a plasma temperature of 6000 K, an electron number density of 10 15 and partition functions according to Ref. [5] (Table 15) [501]. They range from over 99% for sodium to 10ÿ5 for chlorine. The argon ICP thus has good prospects for the determination of metals but for non-metals has a number of limitations as a result of these considerations. ICP mass spectra
In low-resolution ICP-MS the resolution in the spectra at best is 1 dalton and cluster ions are found particularly in the lower mass range signals, which in a number of cases cause spectral interferences with analyte ions. Cluster ions may be formed from dierent types of compounds present, namely: Tab. 15. Degree of ionization of some elements in an analytical ICP
(temperature: 6000 K, ne : 10 15 , electronic partition functions according from Ref. [5]. (Reprinted with permission from Ref. [501].) Element
Ionization energy (eV)
Degree of ionization ( %)
Ba Cl Fe Cu Na S Y
5.21 13.01 7.87 7.724 5.138 10.357 6.51
9.386 1:7 10ÿ5 0.231 0.637 0.996 9:7 10ÿ3 0.867
257
258
6 Plasma Mass Spectrometry
. solvents and acids: H , OH , H2 O , . . . , NO , NO 2 , . . . , Cl , . . . , SO , SO2 , SO3 H (when residual H2 SO4 is present in the measurement solution);
. radicals of gases in the surrounding atmosphere: O2 , CO , CO2 , N2 , NH , NO , . . . ;
. reaction products of the above mentioned species with argon: ArO , ArOH , . . . , ArCl , Ar 2,.... These cluster ions give strong signals in the mass range up to 80 dalton and may hamper the determination of the light elements as a result of spectral interferences. At higher masses, the singly charged ions of the heavier elements are found, as are doubly charged ions of the light elements. They occur particularly for elements with low ionization energies. In the mass spectra signals for compounds of the analyte with various other species also occur, such as MO , MCl , MOH , MOH 2 , . . . . These ions may be formed by the dissociation of nitrates, sulfates or phosphates in the plasma or they may result from reactions between analyte ions and solvent residuals or oxygen in the plasma or eventually in the interface between the plasma and the mass spectrometer inlet. In early work it was recognized that the selection of the sampling zone in the plasma and the aerosol carrier gas ¯ow may strongly in¯uence the occurrence of cluster ions and doubly charged analyte ions [502, 503]. This is clearly demonstrated by the results for lanthanum, where even doubly and triply charged ions may be expected in the argon plasma. This necessitates speci®c optimization of the power and especially of the carrier-gas ¯ow, which in¯uence both temperature and residence times of the sampled substance in the plasma. This is documented e.g. for the intensities of the signal for the ArO and the Cu ions in Fig. 113 [504]. Signals of the dierent isotopes are obtained for each element. Their intensity ratios correspond to the isotopic abundances in the sample. Use can be made of this fact for calibration by isotopic dilution with stable isotopes and in tracer experiments. The isotopic abundances of the elements, however, are also useful for the recognition of spectral interferences. The spectra in ICP-MS are less line rich than in ICP-AES. With quadrupole mass spectrometers, however, the resolving power of the spectrometers is also lower than in ICP atomic emission spectrometers. Accordingly, cluster ions will often cause spectral interferences and appropriate correction procedures must be worked out. With high-resolution sector ®eld instruments (Fig. 114) which have a resolution of 5000 or more, a number of these interferences will not occur. This is clearly shown by calculations of the minimum resolution that should be obtained so as to avoid the most frequent interferences occurring in plasma mass spectrometry with argon inductively coupled plasmas (Table 16) [499]. The fact that prominent spectral interferences in the analysis of real samples such as dissolved ceramic powders can be avoided is re¯ected by the signal scans for Cr and Fe, signals commonly interfered with by cluster ion signals, as seen in Fig. 115 [504]. In ICP-MS the background intensities are usually low. They are mainly due to the dark current of the detector and to signals produced by ions scattered inside the
6.1 ICP mass spectrometry
Fig. 113. Maximum values of ion energies for two ion species (A) and mass spectra in the mass regio 42±82 dalton in ICP-MS (B) when sampling ions at different gas flow rates for a
1 mg/mL solution of Cu at radial distance 0 mm and 1.5 kW, standard solution. (Reprinted with permission from Ref. [497].)
259
260
6 Plasma Mass Spectrometry Fig. 114. High-resolution ICP mass spectrometer. (Reprinted with permission from Ref. [504].)
mass spectrometer as a result of collisions with residual gas species or re¯ection. In order to shield for the direct UV radiation, a beam stop is often supplied in the mass spectrometer. Accordingly, in ICP-MS the ICP only contributes slightly to the background, which contrasts with ICP-AES where the background continuum arises from interactions between electrons and ions in the plasma, from molecular bands, from wings of broad matrix lines and from stray radiation. Therefore, calibration by standard additions is often very successful in ICP-MS, as the systematic errors originate more from signal enhancements or depressions than from spectral background interferences. Tab. 16. Minimally required resolving power (R) in order to avoid spectral interferences in ICP-MS (extracted from Ref. [499]).
Isotope
Polyatomic ion
R required
27
C15 N, 13 C14 N C14 N1 H 40 12 Ar C 35 16 1 Cl O H 39 16 K O 37 18 Cl O 40 16 Ar O 40 Ca16 O 44 Ca16 O 23 Na37Cl 40 23 Ar Na 31 16 P O2 32 16 S O2 40 35 Ar Cl 40 37 Ar Cl
1454 918 2375 1671 2670 2034 2502 2479 3057 2409 2790 1851 1951 7775 9181
Al V 52 Cr 51
55
Mn
56
Fe
60
Ni
63
Cu
64
Zn As 77 Se 75
12
12
6.1 ICP mass spectrometry Fig. 115. High-resolution ICP mass spectra from elements which suffer from Cl-induced spectral interferences in 0.4 M HCl: (a): 20 ng/mL of V, resolution 5000, (b): 20 ng/mL of Cr, resolution 5800 and (c): 100 ng/mL of As, resolution 7500. (Reprinted with permission from Ref. [504].)
Optimization
For the optimization of ICP-MS with respect to the highest power of detection, minimal spectral interferences and signal enhancements or depressions, as well as highest precision, the most important parameters are the power of the ICP, its gas ¯ows (especially the nebulizer gas), the burner geometry and the position of the sampler as well as the ion optics parameters. These parameters determine the ion yield and the transmission and accordingly also the intensities of the analyte and interference signals. At increasing nebulizer gas ¯ow the droplet size decreases (see Section 3.1), and thus the analyte introduction eciency goes up, however, this is at the expense of the residence time in the plasma, the plasma temperature
261
262
6 Plasma Mass Spectrometry
Fig. 116. Dependence of Ni , Pd and Pt signals on injector gas flow rate for a range of forward powers and sampling depths of 15, 20 and 25 mm. (Reprinted with permission from Ref. [505].)
and thus also of the ionization, as shown by optimization studies (Fig. 116) [505]. The optimization of the carrier gas ¯ow, the ion sampling location and the power is more critical in ICP-MS than in ICP-AES. Indeed, in the latter case, the observation window is often of the order of 2 2 mm 2 and thus considerably larger than in ICP-MS (below 1 mm 2 ). This leads to sharper optimization maxima in ICP-MS than in ICP-AES and accordingly larger dierences between results for single element optimization and results at compromise conditions, respectively. Furthermore, changes of the nebulizer gas ¯ow also in¯uence the formation and the breakdown of cluster ions, requiring optimization with respect to minimum spectral interferences as well. The carrier gas ¯ows in¯uence the ion energies, as shown for 63 Cu and the ArO in Ref. [497] and also the geometry of the aerosol channel. Normally the aerosol gas ¯ow is between 0.5 and 1.5 L/min. It must be optimized together with the power, which in¯uences the plasma volume and therewith the kinetics of the dierent processes taking place, and the position of the sampler. By changing the voltages at the dierent ion lenses the transmission for a given ion can be optimized, enabling the optimization of its detection limit and the minimization of interferences. In multielement determinations a compromise must always be made. For the easily ionized elements working at so-called cool plasma conditions has been shown to be very successful. From the calculation of the degrees of ionization
6.1 ICP mass spectrometry
it can be seen that here mainly singly ionized species are formed instead of multiply ionized species (see detection limits listed in Ref. [499]. For Na, K, Ca and Al detection limits in the sub-ppt range then become achievable. Power of detection
In order to obtain the optimum power of detection, the analyte density in the plasma, the ionization and the ion transmission must be maximized. The power will thus be between 0.6 and 2 kW and the sampler at about 10±15 mm above the tip of the injector. The detection limits, obtained at single element optimum conditions dier considerably from those at compromise conditions, but are still signi®cantly lower than in ICP-AES (Table 17). In general it can be said that for the heavy elements, which have complex atomic term schemes and accordingly very
Tab. 17. Detection limits in ICP-OES and ICP-MS.
Element
Ag Al Au B Cd Ce Co Cr Fe Ge Hg La Li Mg Mn Ni Pb Se Te Th Ti U W Zn a
Detection limit ICP-OES (ng/mL) [329]
ICP-MS (ng/mL) [506]
HR-ICP-MS (ng/L) [499]
7 20 20 5 3 50 6 6 5 50 20 10 80 0.1 1 10 40 70 40 60 4 250 30 2
0.03 0.2 0.06 0.04 0.06 0.05 0.01 0.3 Ð 0.02 0.02 0.05 0.1 0.7 0.1 Ð 0.05 0.8 0.09 0.02 Ð 0.03 0.05 0.2
0.4a 0.4a 0.8 5.4 0.5a Ð 0.14a 0.24a 0.9a 0.7 Ð Ð 0.012a 0.2a 0.14a 0.3a 0.12a Ð Ð Ð 0.4 Ð 0.15 0.2a
Cold plasma conditions.
263
264
6 Plasma Mass Spectrometry
line-rich spectra but with low intensity lines in ICP-AES, the detection limits in ICP-MS are much lower than in ICP-AES. For most elements the detection limits are very similar, except for the elements for which spectral interferences are a limiting factor. This applies to As ( 75 As with 40Ar 35 Cl ), Se ( 80 Se with 40Ar 40Ar ) and Fe ( 56 Fe with 40Ar 16 O ). The detection limits for elements with high ionization potential are usually very poor because of the limited formation of positive ions. They may, however, be lower when they are detected as negative ions (cL for Cl : 5 and for Clÿ : 1 ng/mL). The acids present in the measurement solution and the material of which the sampler is made (Ni, Cu, . . .) may in¯uence the occurrence of spectral interferences considerably and accordingly the detection limits for a number of elements. This is particularly important when analyzing solids subsequent to sample dissolution by treatment with acids. Here the measurement solution should hopefully not contain chlorine, phosphate or sulfate ions. If they do, it is advantageous to remove them by precipitation or fuming o and taking up the analytes in dilute nitric acid. The detection limits for ICP-MS in the case of solids thus suer from the necessary sample dilution. In this case the sample concentration in the measurement solution is often limited to 500 mg/mL, as for solutions of Al2 O3 or SiC powders [507]. This is due to the risk of sample depositions blocking the sampler and contrasts with ICP-AES where for these materials sample concentrations up to 5 or even 50 mg/mL are possible. For elements such as B, Mg, Fe, etc. the power of detection of ICP-AES with respect to the solid samples is accordingly higher than in ICP-MS. The detection limits also depend to a great extent on the type of mass spectrometer used. The values for quadrupole mass spectrometers are by orders of magnitude higher than in the case of high-resolution mass spectrometry. This applies particularly for strongly interfered elements such as Fe [499]. With time-of¯ight ICP-MS, the detection limits in the case of the light elements are up to one order of magnitude better than in quadrupole based ICP-MS. However, for the heavy elements they are up to a factor of ®ve worse [508]. Precision and memory effects
The constancy of the nebulizer gas ¯ow is of prime importance for the precision achievable in ICP-MS. After stabilizing the nebulizer gas ¯ows, relative standard deviations can be below 1%. They can be improved still further by internal standardization [509]. The tolerable salt concentrations (1±5 g/L) are much lower than in ICP-AES, because of the risks of sampler clogging and depend on the respective salts. The memory eects may become limiting and in the case of a high matrix load rinsing times of 1±2 min are required. A considerable portion of the noise originates from the nebulizer and the noise level can generally be decreased by cooling the spray chamber. The latter is shown to decrease the white noise somewhat as well as some 1= f contributions and also the standard deviations for the blanks [41]. This leads directly to an improvement in the detection limits.
6.1 ICP mass spectrometry
Interferences
Signal enhancements and depressions resulting from the matrix relate to the nebulization, in¯uences on the ionization in the ICP and to changes in the plasma geometry as well as in the ion energies. Changes of the nebulizer gas ¯ow in¯uence the nebulization eects (see Section 3.1), however, they also lead to changes of the aerosol channel geometry and the plasma temperature and hence also the interferences of easily ionized elements [510]. Space charge eects have also been shown to play an important role in the suppression of signals as a result of alkali metals [511]. Up to a certain amount these eects can be adequately corrected for by using an internal standard, for which the selection is only possible empirically. However, the guidelines are that elements with similar masses and ionization potentials should be chosen. Thus, it is often advisable to select more internal standards to cover elements in the whole analytical mass range, such as Sc for the low mass elements and Rh for the heavier elements. The use of internal standards was shown in early work to be very successful for serum samples [509]. The matrix eects can be eliminated to such a degree that calibration with aqueous solutions in the case of diluted serum samples is possible. In mass spectrometry signals are obtained for each isotope present. With the low mass resolution of quadrupole mass spectrometers (A1 dalton), this leads to a number of isobaric interferences, which can be corrected for with appropriate software. This type of interference depends only slightly on the working conditions, which is not the case for spectral interferences resulting from doubly charged ions, background species or cluster ions. The background species at low masses [512] cause considerable spectral interferences e.g. for 28 Si (with 14 N 14 N ), 31 P (with 16 16 N OH ), 80 Se (with 40Ar 40Ar ). Species such as 40Ar 16 O not only interfere with 56 Fe , but due to all isotopic combinations and any hydrides present, also with other transition metal isotopes ( 52 Cr, 53 Cr, 54 Cr, 54 Fe, 55 Mn, 56 Fe, 57 Fe, 58 Ni, 58 Fe and 59 Co). In the case of HCl, additional interferences for further transitional metal isotopes will result from Cl , ClO , ClN and ArCl species. Under ®xed working conditions these interferences do not change very much with the matrix composition and can be corrected for by subtraction. However, they limit the power of detection. The interactions of signals from doubly charged ions and cluster ions change considerably with the power, the sampler position and the nebulizer gas ¯ow. These interferences are particularly important for elements which have relatively low ionization potentials or that form thermally stable oxides (e.g. Ba, Sr, Mg, Ti). This has been shown by measurements of the signals of singly charged (M ), and doubly charged (M 2 ) ions as well as of metal oxides (MO ) and hydroxides (MOH ), e.g. in the case of Ba [503, 505]. Considerable progress with respect to spectral interferences in the case of quadrupole mass spectrometry has been possible through the use of collision- and reaction-induced dissociation of cluster species. This work goes back to the research of Douglas and French [513] and Barinaga et al. [514]. For collision-induced dissociation, it is advantageous to provide for hexapole or octapole arrangements between the skimmer and the mass spectrometer itself, and to ®ll these arrangements with the
265
266
6 Plasma Mass Spectrometry
appropriate gases. For collision-induced dissociation, helium as well as xenon, have been successfully used and for reaction-induced dissociation, e.g., hydrogen, nitrogen, oxygen and ammonia, as shown for the case of H2 in Ref. [515], and it has been used in the determination of S [516], B, Li, etc. A reduction of the interferences resulting both from spectral overlap and ionization can be realized in a number of cases by removal of acids and matrix residuals, as shown by on-line removal of these species by separations on solid phases, which has now been tested in the case of exchanges made at modi®ed nebulizer surfaces [517]. Isotopic dilution analysis
Dilution with stable isotopes oers the possibility of performing tracer experiments but also of circumventing systematic errors. The principle [518] can be applied for every element which has at least two stable or longlife isotopes. For its application the analyte with a known isotopic composition but which diers from that of the sample is added to a known amount of sample, and mixed thoroughly. The isotopic abundance ratio R then is given by: R NS h S
1 NA hA
1=NS h S
2 NA hA
2
298
NS is the number of atoms of the element to be determined in the sample and NA the number of atoms in the amount of standard added. h S and hA are the abundances of the isotope (1) and (2) in sample and standard added. Thus, NS or the absolute mass GS is given by: GS GA hA
1 ÿ R hA
2=R h S
2 ÿ h S
1
299
R results from the signals of the isotopes, h S
1 and h S
2 as a rule are the natural abundances. GA is the absolute amount of standard added, hA
1 and hA
2 are known from the isotopic composition. Isotope dilution in ICP-MS has been applied in studies on Pb (see e.g. Ref. [519]). Also tracer experiments for Fe in biological systems (see e.g. [520]) have been described. The precision achievable in the determination of isotope ratios for abundances which dier by a factor of less than 10 is in the lower percent range. For the determination of isotope ratios, the precision of TOF-ICP-MS has been studied in a preliminary comparison with other mass spectrometer systems [521]. Typical isotope ratio precisions of 0.05% were obtained, thus overtaking sector ®eld mass spectrometry with sequential detection, for which values of 0.1±0.3% for 63 Cu/ 65 Cu in Antarctic snow samples have been reported [522]. Similar results were obtained by Becker et al. [523] for Mg and Ca in biological samples (0.4± 0.5%). In principle, the features of TOF-ICP-MS may be superior to those of sequential sector ®eld or quadrupole mass spectrometry, however, true parallel detection of the signals as is possible with multicollector systems may be the de®ni-
6.1 ICP mass spectrometry
tive solution, as shown by Hirata et al. [524]. Here the use of detectors which allow true parallel measurement of the signals within the relevant mass range, just as the CCDs do for optical atomic spectrometry, may be the ultimate solution and bring about the ®nal breakthrough for ICP-MS isotope ratio measurements as is required in isotope dilution mass spectrometry. Alternative methods for sample introduction
Apart from continuous pneumatic nebulization, all sample introduction techniques known from ICP-AES have been used and are of use for ICP-MS. Similar to ICP-AES, the analysis of organic solutions is somewhat more dicult [525]. The addition of oxygen was found to be helpful when nebulizing euents from HPLC containing organic eluents such as acetonitrile. This was useful when using ICP-MS for on-line detection in speciation as well as in trace±matrix separations. Here, however, it is useful to use desolvation, even in the case of low consumption, high eciency nebulizers, such as the HEN or DIHEN. This can be done eciently with membrane desolvation using Na®on membranes [148]. The use of ultrasonic nebulization just as in ICP-AES allows the sampling eciency to be increased. The high water loading of the plasma has to be avoided, as is possible with desolvation, not only to limit the cooling of the ICP but also to keep the formation of cluster ions and the related spectral interferences low. These are complex, as for example, a change in water loading also in¯uences the pressure in the intermediate stage [526]. The use of high-pressure nebulization in ICP-MS has similar advantages and is suitable for coupling ICP-MS with HPLC. The set-up that is of use in speciation work is the same as the one used for on-line trace±matrix separations [527]. With the formation of volatile hydrides, the detection limits for elements such as As, Se and Sb can also be improved. As shown in Ref. [528], improvements for Pb were also obtained. They are due to improved analyte sampling eciency but also to the decrease in cluster ion formation resulting from the introduction of a water-free analyte, which also applies to the cold vapor technique in the case of Hg. Electrothermal vaporization (ETV) in addition to its features for the analysis of microsamples, in ICP-MS has the additional advantage of introducing a dry analyte vapor into the plasma. Hence, it has been found to be useful for elements for which the detection limits are high as a result of spectral interferences with cluster ions. In the case of 56 Fe, which is subject to interference by 40ArO , Park et al. [529] showed that the detection limit could be improved considerably by ETV. For similar reasons the direct insertion of samples in ICP-MS leads to the highest absolute power of detection (detection limits in the pg range and lower [530, 531]). Transient signals arise from electrothermal vaporization, where accordingly the number of elements that can be determined for one vaporization event is very limited, unless drastic decreases in analytical precision and power of detection are accepted. In this respect TOF-ICP-MS oers some advantages. Here simultaneous measurements of dierent isotopes are possible during one evaporation event
267
268
6 Plasma Mass Spectrometry
without losses in analytical performance. It is even possible to monitor the separation of the volatilization of interferents from the evaporation of the analytes in realtime, which is very helpful for removing both spectral and ionization interferences, as shown by Mahoney et al. [80]. Despite ICP-MS being mainly a method for the analysis of liquids and solids subsequent to dissolution, techniques for direct solids sampling have also been used. They are required particularly when the samples are dicult to bring into solution or in addition are electrically non-conducting and thus dicult to be analyzed with glow discharge or spark techniques. For the case of powders, such as coal, slurry nebulization with a Babington nebulizer can be applied in ICP-MS as well [532]. ETV oers good possibilities not only for powders but even for granulates also. With a novel furnace into which graphite boats can be introduced and where halogenated carbons can be used as volatilization aids, dierent volatile halogenide forming elements can be successfully evaporated from Al2 O3 . When applying slurry sampling under continuous ultrasonic treatment of the slurry, accurate results for powdered biological substances [533], and also for WC [534] and Al2 O3 powders down to the sub-mg/g level [535] can be obtained. For the direct analysis of metals spark ablation can be applied and the detection limits are in the ng/g range, as shown for steels [536]. When analyzing metals [537] as well as electrically non-conducting samples, laser ablation combined with ICP-MS is very useful. A Nd:YAG laser with a repetition rate of 1±10 pulses/s and an energy of around 0.1 J has been used. For ceramic materials such as SiC, the ablated sample amounts are of the order of 1 ng and the detection limits down to the 0.1 mg/g level. When using lasers with dierent wavelengths, material ablation as well as the minimum crater diameters achievable may well vary, both of which also vary with the gas atmosphere (argon or helium), as studied by Guenther and Heinrich [538]. Despite the availability of advanced lasers for certain sample types a number of questions remain. For the analysis of CuaZn alloys, the calibration behaves nonlinearly, which can be explained by a change in the mass ablation with the composition, and also when using dierent lasers with dierent pulse lengths and wavelengths [539]. Laser ablation is now so controllable that, in the case of multicollector ICP-MS, isotopic analyses for individual grains of minerals can be performed with a precision in the isotopic ratio of better than 0.02% (2s level) [524]. Laser ablation can also be used successfully for samples which are dicult to analyze directly by any other method, e.g. plastics and glass samples. The use of individual laser spikes will be especially interesting in the case of TOF-ICP-MS, as then it will be possible to perform real multielement determinations from the material cloud generated by a single laser shot. 6.1.3
Applications
ICP-MS is especially promising for the areas where ICP-AES is applied but where further improvement in the power of detection is required. This is the case in trace
6.1 ICP mass spectrometry
analysis for geological samples and speci®cally for hydrogeological samples as well as for trace determinations in metals, in biology and medicine as well as in environmental analysis. Geological samples
In the case of geological samples ICP-MS is of interest where multielement determinations are required and ICP-AES cannot be used because of the lack of power of detection or spectral interferences. This applies to the determination of low concentrations of the rare earth elements [506, 540]. Clogging and corrosion of the sampler may be critical and requires rinsing and working with solutions having concentrations below 0.1%. Hydrogeological samples, as described by Garbarino and Taylor [541], can be analyzed very accurately for Ni, Cu, Sr, Cd, Ba, Tl and Pb by isotope dilution ICP-MS. For a series of trace elements ETV was found to be useful so as to reduce spectral interferences. This applies particularly to volatile elements such as Tl [542] but in the case of a metal ®lament vaporizer also to Pt, Pd, Ru and Ir [543]. For the investigation of inclusions in minerals, laser ablation ICPMS is very powerful. It can even be used for the analysis of liquid inclusions in minerals, which provide important information for geologists [544]. Metals and ceramics
Trace determinations down to the sub-mg/g level are possible in metals and ceramics as the analyte concentrations may be up to 5 g/L. ICP-MS therefore is really an alternative to ICP-AES for the analysis of metals with line-rich spectra. This has been shown in the case of high-temperature alloys [545]. However, matrix interferences also ®nally limit the power of detection and matrix removal is useful to make further improvements to the power of detection and calibration. All classical principles of trace±matrix separation are very helpful in this respect. In the case of Al2 O3 powders subsequent to sample dissolution, the Al can be separated o by ®xing the APDC complexes of a number of analytes on a solid phase and by releasing them into the ICP after matrix removal [131]. For SiC the acid residues remaining after decomposition of the powders with fuming sulfuric acid can be removed by fuming o and taking up the residue with dilute nitric acid, by which the detection limits can often be improved by one order of magnitude [546]. Using high-resolution ICP-MS the detection limits for Al2 O3 ceramics are found to be of the same order of magnitude as in the case of quadrupole ICP-MS coupled on-line to matrix removal. In the latter case, however, Cr could not be determined down to low concentrations, as a result of the interference of the 52 Cr signal with the ArC signal [504]. For the direct analysis of steels by ICP-MS, Jiang and Houk [547] used arc ablation and reported detection limits at the 0.1 mg/g level and calibration curves being linear to concentrations of 0.1%. Arrowsmith and coworkers showed that both in metals as well as in ceramic samples direct analyses could be performed with
269
270
6 Plasma Mass Spectrometry
laser ablation coupled to ICP-MS [537, 548, 549]. Laser ablation ICP-MS is now commercially available and of great use for survey analysis of solids and also for inclusions. Biological and medical samples
For biological and medical samples, ICP-MS has facilitated a considerable enlargement of the series of elements that can be determined directly and, thus, is of great importance for speciation and bioavailability studies. Normal levels of a number of trace constituents in clinical samples have been determined by ICP-MS [550]. In urine good agreement of ICP-MS results with those of other techniques was obtained for elements with masses beyond 81 (Pb, Cd, Hg and Tl). Deviations were found for As, Fe and Se, which could be partially eliminated by precipitation of the chlorides from the measurement solution. With ICP-MS Pb can be determined in blood [551] and the bioavailability of Zn has been studied [552]. For the analysis of small samples, as shown for blood analysis [553], or for the analysis of samples with high salt contents, ¯ow injection can be applied. By coupling ICP-MS with chromatographic techniques, metals bound to dierent protein fractions can also be determined separately [554]. In addition, metabolic studies can be performed by isotopic dilution work, which is very promising for medical applications. For applications in the life sciences, where limited sample volumes or fairly complex mixtures are the norm, chromatographic techniques enabling high chromatographic resolution, such as capillary zone electrophoresis, coupled on-line to ICPMS are very powerful. Applications such as the determination of organoselenium compounds and of metalloproteins in serum [556] or the separation of the six relevant arsenic compounds in water at their 1±2 mg/L level [557] should be mentioned. The use of low-consumption high-eciency nebulizers, such as the direct injection high eciency nebulizer [558] are of great value in these applications. Environmental work
ICP-MS is very promising in the area of environmental studies. Many elements can be determined directly in drinking water. In waste water analysis sample decomposition by treatment with HNO3 aH2 O2 is often required and the most frequent isobaric interferences have been described [559]. For seawater analysis, the salt contents makes sample pretreatment necessary, which can be done by chelate extraction. Beauchemin et al. [560] obtained a preconcentration of a factor of 50 by sorption of the trace elements onto an SiO2 column treated with 8-hydroxyquinoline and determined Ni, Cu, Zn, Mo, Cd, Pb and U in seawater. In river water Na, Mg, K, Ca, Al, V, Cr, Mn, Cu, Zn, Sr, Mo, Sb, Ba and U could be determined directly and Co, Ni, Cd and Pb after the above mentioned preconcentration procedure. For As, preconcentration by evaporation of the sample was sucient. Isotope dilution delivers the highest accuracy [561] and the procedure has been applied to
6.1 ICP mass spectrometry
the characterization of a standard reference sample [560]. ICP-MS, subsequent to sample decomposition with HNO3 aH2 O2 , has also been used for trace determinations in marine sediments [562] and for the trace characterization of marine biological samples [563, 564]. Owing to the extremely high power of detection ICPMS can be used to determine very low background concentrations in unpolluted areas. With high-resolution ICP-MS and a microconcentric high-eciency nebulizer Rh, Pd and Pt can be detemined in ice samples with detection limits down to 0.02, 0.08 and 0.008 pg/g, respectively [565] as can the actinides in environmental samples [566]. ICP-MS coupled with chromatography has become very important for speciation of environmentally relevant elements. For the speciation of chromium in waste water samples from the galvanic industry, Andrle et al. [567] ®xed the CrIII and CrVI species present through a reaction with dithiocarbamates. In this way, stable complexes were formed, which could then be separated by chromatography. The preparation of these complexes can be performed in a ¯ow system which includes a thermostated reactor and ®xation of the complexes on a solid phase. The separation can be performed by on-line coupling with reversed-phase chromatography and detection by ICP-MS using hydraulic high-pressure nebulization, as introduced for atomic absorption spectrometry by Berndt [143]. Calibration of the procedure is performed by standard additions so as to correct for any shifts in the CrIII aCrVI equilibrium during the complexation reaction. The chromatograms, in the case of a waste water sample, showed that multielement speciation is certainly possible, as elements other than Cr may be present in dierent valence states and accordingly also form dierent complexes with dithiocarbamates. Other methods of speciation of chromium in water samples lie in the use of anion exchange resins, which were shown by Barnowski et al. [568] to retain both the CrIII and the CrVI species. This approach has the advantage that chloride ions which could possibly be present are retained and accordingly do not cause spectral interferences, e.g. when determining preconcentrated iron species. Gas chromatography coupled with ICP-MS enables the determination of volatile organometal compounds, such as the organolead compounds, to be performed, with a high power of detection. In other cases a derivatization has to be performed, e.g. ethylation with tetraethylborate or a Grignard reaction (for a discussion see Ref. [452]). Suitable coupling systems have been described combining GC and quadrupole or sector ®eld ICP-MS (see e.g. Prange and Jantzen [569] and Heisterkamp et al. [570]), where the transfer line is heated. A drawback is the risk of skewing the gas chromatography peaks when using these mass spectrometers in the multielement mode. As with gas chromatography separations can be accelerated still further using multi-capillary gas chromatography [571]; skewing of peaks then becomes an even greater risk. The related systematic errors can be eliminated by using TOF-ICP-MS, as shown by Leach et al. [572]. For GC use of a large and bulky ICP is also no longer required and savings in investment and operating costs can be made by switching to microwave plasma sources.
271
272
6 Plasma Mass Spectrometry
6.1.4
Outlook
Further trends in mass spectrometry with discharges at atmospheric pressure lie in the use of alternative plasmas, in the progress of developments to mass spectrometric equipment and in the improvement of sample introduction. Without loss in analytical performance ICPs can be operated at lower gas ¯ows and power consumption especially at higher frequencies (up to 100 MHz) (see e.g. Ref. [573]). The addition of molecular gases such as N2 has been thoroughly investigated by Lam and Horlick [574], who found a decrease in the formation of cluster ions in a number of cases. Helium ICPs have been investigated by Chan and Montaser [575] and these could be very useful for the determination of the halogens, as shown in Ref. [576]. MIPs have also been used as sources for mass spectrometry, as e.g. described by Satzger et al. [577]. Whereas sampling capabilities are practically limited to coupling with gas chromatography or electrothermal atomization, MIPs at higher power may be useful alternative sources, as they can also be operated in nitrogen [578]. This has been shown for a high-power nitrogen discharge by Okamoto [579], where a number of interferences known from argon plasmas could be avoided. Moreover, the operating costs of such a system are lower and for speci®c applications, such as process analysis, it might be of interest. Other plasmas at atmospheric pressure, such as the FAPES (furnace atomic plasma emission source) developed by Blades [580] have been used as ion sources for mass spectrometry. With FAPES detection limits in the fg range can be obtained, as microsamples can be analyzed with virtually no transport losses (Fig. 117) [581]. However, further investigations on interferences certainly still need to be made. By using high-resolution systems instead of quadrupole based instruments spectral interferences can be eliminated in a number of cases. To this aim, sector ®eld instruments, provided they become cheaper, and time-of-¯ight instruments, especially for the case of transient signals, begin to ®nd uses [75]. As on-line coupling of high resolving separation systems with ICP-MS becomes of more and more use and multielement speciation is requested not only in volatile but also in liquid samples, plasma-TOF-MS becomes of greater interest. This is shown by the ®rst results obtained with capillary zone electrophoresis coupled with TOF-ICP-MS by Bings et al. [582]. Advanced types of mass spectrometers, well-known from work in organic mass spectrometry, such as ion cyclotron resonance mass spectrometry [583] and ion traps may become of use. Both constitute approaches for achieving extremely high spectral resolution and, in the case of soft plasmas, also work with clusters. With ICP Fourier transform ion cyclotron resonance mass spectrometry a resolving power of up to 88 000 has been realized and detection limits at the submg/L level found. A microwave plasma torch can be operated very stably with argon as well as with helium and can be used as an ion source for time-of-¯ight mass spectrometry. Such a system, as described by Pack et al. [76], is very useful for element-speci®c detec-
6.1 ICP mass spectrometry
Fig. 117. FAPIMS (furnace atomization plasma ionization MS) workhead and sampling interface, (1): x, y, z translation; (2): viewing/ sampling port; (3): ICC furnace; (4): rf center electrode; (5): type N rf connector; (6): brass collet; (7): plasma gas (He) inlet; (8):
photodiode temperature sensor; (9): to roughing pump; (10): MACOR support rings; (11): modified sampler cone; (12): standard skimmer cone. (Reprinted with permission from Ref. [581].)
tion in gas chromatography, as peak skewing is absent. It is easy to bring the GC column up immediately behind the plasma and to heat the complete transfer line by resistance heating of a copper wire wound around the capillary (Fig. 118). In this way there is no chance of deterioration of the chromatographic resolution and, as helium can be used as the working gas, the detection limits for halogens are very low. When recording the chromatograms of the halogenated hydrocarbons, carbon and chlorine can be monitored simultaneously (Fig. 119). The stability of the plasma is obvious from the ¯at signal for chlorine at the elution time for methanol. The detection limits for chlorinated compounds were shown to be in the fg range and are considerably lower than in a low-pressure ICP and a quadrupole mass spectrometer. This may partly be due to the small size of the microwave discharge as compared with the ICP, resulting in a lower analyte dilution. The coupling of time-of-¯ight mass spectrometry with a helium microwave plasma torch and gas chromatography may accordingly become a real alternative to the microwave in-
273
274
6 Plasma Mass Spectrometry
Fig. 118. GC-MPT-TOFMS experimental set-up. Capillary column extends from the GC oven to the tip of the MPT, where a plasma is formed. The plasma is sampled through a 0.5 mm orifice into the TOFMS for mass analysis. (Reprinted with permission from Ref. [76].)
duced plasma atomic emission detector (MIP-AED) and also to ICP-MS coupled with gas chromatography for the determination of organometallic or halogenated compounds. In the ®eld of sample introduction, hyphenated systems and devices allowing online preconcentration, automated sample introduction or speciation will become more and more important. Summarizing, the potential of ICP-MS lies in the fact that analyses can now be performed with the ¯exible sampling of an ICP, with true multielement capacity and a high power of detection. Manufacturers
. Agilent Technologies, Palo Alto, CA, USA . Finnigan MAT, Bremen, Germany . GBC, Australia . Hitachi, Tokyo, Japan
6.2 Glow discharge mass spectrometry Fig. 119. Isotope-specific chromatograms of halogenated hydrocarbons (chlorobutane to chlorohexane) in methanol obtained by GCMPT-TOFMS. Twin boxcar averagers are used for data collection. (a): Signal from 12 C, (b): signal from 35 Cl. (Reprinted with permission from Ref. [76].)
. Leco, St Joseph, MI, USA . Micromass Ltd., UK . PerkinElmer Sciex, Norwalk, CT, USA . Spectro Analytical Instruments, Kleve, Germany . Thermo Jarrell Ash, Waltham, USA . Varian Ass., Mulgrave, Australia . VG Elemental, Sussex, UK 6.2
Glow discharge mass spectrometry
Glow discharges [584], known from their use as radiation sources for atomic emission spectrometry have also became recognized as powerful ion sources for mass spectrometry. This development started with spark source mass spectrometry, where continuous eorts were made to arrive at more stable sources with the added advantage that the matrix dependency of the analyte signals would be lower than in spark sources [69]. In glow discharge mass spectrometry the analyte is volatilized by sputtering and the ions are produced in the negative glow of the discharge as a result mainly of collisions of the ®rst kind with electrons and Penning ionization. Subsequently, an ion beam is extracted which in its composition is representative of the sample. Between the glow discharge and the spectrometer a reduction in gas pressure is required. As the glow discharge is operated at a pressure of around 1 mbar, a two-
275
276
6 Plasma Mass Spectrometry
stage vacuum system is required. A cathodic extraction can be done at the cathode plume, by taking the sample as the cathode and drilling a hole in it. The aperture should reach the hot plasma center. In most cases the extraction of the ions, however, is done anodically. As a result of impurities in the ®ller gas and the complexity of the processes taking place, not only do analyte and ®ller gas ions occur in the mass spectra but also many other ions. Spectral interferences therefore can occur and for various reasons:
. isobaric interferences of isotopes of dierent elements, e.g. 40Ar with 40 Ca or 92
Zr with
92
Mo ;
. interferences of analyte ion signals with doubly charged ions, e.g. 56 Fe 2 with Si (this type of interference, however, is much rarer than in earlier work with high vacuum sparks); interferences of analyte ions with cluster ions formed from analyte and gas species, e.g. 40Ar 16 O with 56 Fe ; interferences by signals from residual gas impurities, e.g. 14 N 16 O 1 H with 31 P forming, for instance, metal argides. 28
. .
Mass spectrometric measurements contributed to clarifying the mechanisms of reactions with reactive species such as air and and water vapor. Indeed, the presence of water vapor may seriously in¯uence ionization and excitation in the glow discharge plasma, thus leading to a decrease in analytical signals such that they are barely observable [585]. The major consequence would be an oxidation of the sample surface, quenching of argon metastables, low sputtering, and loss of analyte atoms due to enhanced gas-phase reactions. Therefore, the presence of water vapor should be limited by using high-purity gases, optimized vacuum systems and getters as appropriate. Cryogenic cooling may also be helpful. In a subsequent study, water was pulse injected into the discharge, to clarify the behavior of various cathodic materials (Cu, Fe, and Ti) [586]). Depending on the reactivity of these metals, oxides were formed during the pulsed injection and shortly afterwards. The eects for Ti were found to be larger than for Cu and Fe. General models of GD-MS, as presented by Vieth and Huneke [587], include electron ionization three-body recombination and compare fairly well with experimental measurements of singly- and multiply-charged Ar ions. Penning ionization was found to predominate when the ion species had ionization energies below the energy of the Ar metastables. It can be shown with an electrostatic probe how GDMS parameters are in¯uenced by the voltage of an auxiliary electrode inserted into the plasma [588]. The plasma potential and the ion energies seem to follow the bias potentials well whereas the electron temperatures behave in a more complex fashion. The formation and use of signals for doubly charged analyte ions in GD-MS, when the singly charged ions suer from interferences, has been treated by Goodner et al. [589]. A study of Ar, He, Kr, Ne and N2 , from the point of view of spectral interferences and sputtering, showed that Ar gave the best sputtering while with respect to signals, memory eects and cost, Kr was the worst [590]. The
6.2 Glow discharge mass spectrometry
dependence of the ion current, ion intensity and energy distribution on power, carrier gas pressure and sampling distance in the case of a magnetron rf-GD for MS was investigated for a borosilicate glass cathode, and ion intensities were found to depend strongly on pressure and distance [591]. The possibility of separating the interfering signals depends on the instrument resolution. In the case of quadrupoles the mass resolution at best is unity and these interferences can only be corrected for by mathematical procedures. However, even with high-resolution instruments having a resolution of 5000 many inteferences remain, especially those with hydrides in the mass range 80±120 Da. 6.2.1
Instrumentation
Glow discharges operated at pressures in the 0.1±1 mbar range can be coupled to mass spectrometers by using an aperture between both chambers with a size of about 1 mm. Much work has been done with dc discharges. The Grimm-type source with ¯at cathodes is usually used. The glow discharge may, however, also use pin-shaped samples which can be introduced into the source without even admitting any air. Harrison and Bentz [584] accordingly coupled discharges with a pin cathode as well as hollow cathode plumes to a high-resolution sector ®eld mass spectrometer. This geometry was found in 1984 in the early glow discharge mass spectrometers (VG 9000) available through VG Instruments (Fig. 120A). A double-focussing spectrometer in so-called reversed Nier±Johnson geometry is used, with which a spectral resolution of about 10 4 can be obtained. The mass range up to 280 Da can be covered by using an acceleration voltage of 8 kV. A Faraday-cup and a so-called Daly±Multiplier are used as detectors with the possibility of continuous switching. The vacuum of 10ÿ6 mbar is maintained with the aid of oil diusion or turbomolecular pumps. To lower the background, the source housing can be cooled down to act as a cryo-pump. However, with further developments quadrupole mass spectrometers have also been used to detect the ions produced in a glow discharge. No high voltage is required to realize a suciently high ion extraction yield and rapid scans can easily be performed, which enables quasi-simultaneous multielement measurements to be made. An electrostatic analyzer is often used in front of the spectrometer as an energy ®lter. Jakubowski et al. [592] described a combination of a Grimm-type glow discharge with a quadrupole mass spectrometer (Fig. 120B) and studied the basic in¯uence of the working and construction parameters. Glow discharge mass spectrometers using quadrupoles are now commercially available. A special dc GD for combination with a sector-®eld instrument has been designed for the analysis of high-purity Si [593]. The Si ion yield was found to increase with gas pressure, probably as a consequence of the enhancement in the Ar metastable population. The signals for the Si 2 decrease perhaps as a result of an increase in dissociative collisions. Detection limits for elements such as Al, As, B, Cl, Fe, P and U are at a level of 6 10 10 ±6 10 13 atoms/cm 3 .
277
278
6 Plasma Mass Spectrometry Fig. 120. Glow discharge mass spectrometry. A: Sector field based instrument (a): glow discharge source; (b): magnetic sector field; (c): electrostatic energy analyzer; (d): Daly detector; (e): Faraday cup (VG 9000, courtesy of VG Instru-ments) and B: a quadrupole based low resolution glow discharge mass spectrometer (a): source; (b): ion optics; (c): quadrupole mass filter; (d): differential pumping system; (e): detector. Similar to described in Ref. [611].
A modi®ed version of the Grimm-type GDL has been described by Shao and Horlich [594]. The source has a ¯oating restrictor and is designed so as to replace an ICP torch in an ICP-MS. Therefore, the anode is slightly positive with respect to the earthed skimmer interface of the MS system. The simultaneous analysis of an unknown sample and a reference material was carried out by means of a system based on two pulsed GD sources housed within the same tube [595]. Optimization of the relative position of the two cathodes was achieved by evaluating the signals produced in GD-MS when using the same specimen for each of them. The introduction of rf-powered glow discharges in MS has led to a large broadening of the applicability of the method [596]. In initial studies it was shown that the bias potential is directly proportional to rf power and inversely proportional to the discharge gas pressure. When coupled to a high-resolution double-focussing mass spectrometer [597], problems of rf shielding, grounding and coupling to the accelerating potential were faced. To this end a source capable of handling both pin and ¯at samples was built. It operated stably at 13.56 MHz, 10±50 W and 0.1±1 mbar. Ion currents up to 10ÿ9 A and a mass resolution of 8000 could be obtained.
6.2 Glow discharge mass spectrometry
Depth pro®ling is also possible as a result of the radial ¯ow of support gas onto the sample surface. A successful extraction and focussing of ions necessitated parametric studies with respect to ion kinetic energies. The analytical ®gures of merit of rf-GD-MS also with respect to depth pro®ling have been evaluated by Marcus et al. [598]. Two current versions of the lamp were considered, namely one for ¯at and one for pin samples. An rf-GD-TOF mass spectrometer has been described by Myers et al. [599]. The ion optics which focus ions toward the entrance of the TOF instrument are the same as those used in the original ICP-TOF-MS. By means of pin-shaped brass samples of varying lengths, the sample±skimmer distance in the GD-TOF-MS instruments was optimized at 4 mm, while discharge gas pressure and power provided the best results in the 50±60 W and 0.3 mbar range, respectively. The application of small negative potentials to the skimmer cone (extraction ori®ce) was found to improve signals only slightly. However, higher negative potentials reduced both signals and resolving power. The skimmer potential was also found to aect the ®nal kinetic energies of the ions before their extraction in the TOF. At 0.3 mbar all ions extracted for mass analysis were found to have approximately the same kinetic energy and detection limits were stated to be at a level of 1 mg/g. Work has also been carried out with TOF-MS using pulsed GD sources. The eect of the distribution of the kinetic energy was lowered by resorting to a twostage acceleration ®eld by using a dc microsecond-pulsed GD as described by Hang et al. [600]. Here the sample was placed on the tip of a direct insertion probe. The con®guration permitted the use of current intensities as high as 500 mA with 200 ns pulses being applied. During the active cycle of pulsed operation a larger net signal (by up to 2 orders of magnitude) can be produced than could be generated through a comparable power level in the dc mode. Furthermore, the fast pulsed discharge operation may permit a diagnostic evaluation of the plasma processes and give improved analytical performance when used with TOF-MS. An rf-planar-magnetron GD has also been coupled with TOF-MS. For rapid sample changing without venting the mass spectrometer, a sliding PTFE seal was placed at the interface. The seal in turn holds a Macor ring, shields it from the plasma and supports the sample [601]. Detection limits for conducting and insulating materials were of the order of 0.1 and 10 mg/g for B and Mg in Macor, and Bi, Cr, Mn, Ni, and Pb in aluminum, respectively. The source±spectrometer combination still needs improvements at the interface with respect to the extraction location for analyte ions, the scattered ion noise and the extraction repetition rate. When applying a hollow cathode (HC) discharge as the ion source, two groups of ions (respectively with high and low kinetic energy) were detected in the plasma of a reversed HC source combined with an MS [602]. The former group of ions are produced in the negative glow region, while the latter are formed in the extraction ori®ce of the cathode base. Discrimination between these two groups by setting the acceleration voltage so as to separate the high-energy group (analyte atoms) from the low-energy group (Ar carrier gas and cluster ions) could lead to improved analytical performance. Hollow cathodes have also been used for the direct analysis of liquids. As re-
279
280
6 Plasma Mass Spectrometry
ported by You et al. [603] a particle beam liquid chromatography (LC)-MS device can be interfaced with a heated HC unit. The high eciency of the thermal nebulization system ensures that analyte particles from the aqueous solution can be transported into the heated HC cell for subsequent vaporization and excitation as well as ionization. The discharge characteristics were investigated in the case of Ar and He as support gas, the latter giving less line-rich emission spectra. To explore in detail the behavior of the system, signals for Cs and Na were studied as a function of the discharge pressure, current, solvent ¯ow rate, tip temperature, etc. Optimum results were obtained at low ¯ow rates (even down to 0.2 mL/min) and temperatures of 200±220 C, respectively. The system can be used successfully as an ion source for mass spectrometry. The transport of analytes was found to be enhanced by the addition of HCl to the solutions due to an enhancement in analyte size favored by this reagent, as revealed by scanning electron microscopy. Typically, the particle size was in the 2±8 mm range with transport eciencies of 4±18%, as found for Cu, Fe and Na test solutions [604]. Gas sampling GD was ®rst described by McLuckey et al. as a soft ionization source [490]. It was shown to facilitate the ionization of gaseous compounds while giving both elemental as well as molecular information, as shown for the case of AsH3 produced by hydride generation (Fig. 121) [488] and can be used successfully for the ionization of any atomic as well as molecular vapor. By alternating from soft to harsher conditions, it is possible to change from atomic to molecular information. This can be illustrated by the spectra obtained from ferrocene vapor [605]. McLuckey proposed the use of glow discharges coupled with quadrupole ion-trap MS for the determination of high-explosive substances in the vapor phase. The gas sampling GD system was found to be very eective at forming negative ions, e.g. from mono-, di- and trinitrophenols, mono-, di- and trinitrotoluenes, S, and others. Also low pressure ICPs and MIPs, often used as ion sources (see e.g. Creed et al. [606]) are very similar in their performance to glow discharges, except for them being electrodeless discharges. They have shown particular merits for elementspeci®c detection in chromatography, where detection limits down to the pg/s level
Fig. 121. Mass spectrum obtained from argon gas sampling glow discharge coupled with hydride generation for the determination of As. Solution: 10 mg/mL As; argon pressure: 0.2 Torr; voltage: 760 V; current: 20 mA; quadrupole mass spectrometer (Balzers), vacuum in 3rd stage: 2 10ÿ5 Torr, signal detection: secondary electron multiplier. (Reprinted with permission from Ref. [488].)
6.2 Glow discharge mass spectrometry
Fig. 122. Dielectric barrier (db) discharge microstrip plasma in set-up for diode laser AAS. (DL1, DL2): diode lasers, (BS): beam splitter, (M): mirror, (G): grating, (PD1, PD2): photodiodes. (Reprinted with permission from Ref. [608].)
can be obtained. These values, however, were found to be surpassed for halogens by microwave discharges at atmospheric pressure coupled with TOF-MS. Glow discharges have also now been miniaturized. In the discharge in a microstructured system, described by Eijkel et al. [607], molecular emission is obtained and the system can be used successfully to detect down to 10ÿ14 g/s methane with a linear response over two decades. A barrier-layer discharge for use in diode laser atomic spectrometry has also been described recently (Fig. 122) [608]. 6.2.2
Analytical performance
With ¯at as well as pin form cathodes the discharge is usually operated at about 1 mbar, a gas ¯ow rate of a few mL/min and the current is about 2±5 mA with a burning voltage of 500±1500 V. Normally the power is stabilized and the pressure, the current, the voltage or the power are taken as optimization parameters. All electrically-conductive samples and even semi-conductors can be analyzed directly. After a preburn phase the discharge can be stable for many hours. Detection limits down to the ng/g range are obtained for most of the elements. Mass spectrometric methods are relative methods and need to be calibrated with known reference samples. To this aim so-called relative sensitivity factors are used. They are de®ned as:
281
282
6 Plasma Mass Spectrometry
RSF xel =x ref c ref =cel
300
where xel and x ref are the isotope signals for the element to be determined and a reference element, respectively, and c is their concentration. When matrix eects are absent the factors are 1. In GD-MS the RSFs are found to be much more closer to 1 as compared for instance with spark source mass spectrometry [609] and secondary ion mass spectrometry. Particularly in quadrupole systems spectral interferences may limit the power of detection as well as the analytical accuracy. This is shown by a comparison of a high- and a low-resolution scan of a mass spectrum (Fig. 123) [609, 610]. In the ®rst scan (A), the signals of 56 Fe 2 , 28 Si , 12 C 16 O and 2 N 2 are clearly separated, which is not, however the case in the quadrupole mass spectra. In the latter scan, interferences can be recognized from comparisons of the isotopic abundances. Also physical means such as the use of neon as an alternative discharge gas [611] can be used to overcome the spectral interference problem. Similar approaches can be followed with all types of glow discharges discussed under Section 6.2.1. 6.2.3
Analytical applications
The use and topical applications of GD-MS as sources for MS has been well covered in a review by Caroli et al. [612]. Bulk concentrations
GD-MS is of use for the direct determination of major, minor and trace bulk concentrations in electrically-conductive and semi-conductor solids down to concentrations of 1 ng/g. Because of the limited ablation (10±1000 nm/s), the analysis times especially when samples are inhomogeneous are long. It has been applied speci®cally to the characterization of materials such as Al, Cd, Ga, In, Si, Te, GaAs and CdTe. A comparative study of spark-ablation ICP-MS and GD-MS in the case of steel has been reported by Jakubowski and coworkers [536, 613]. The RSFs for a number of trace elements and the measurement precision are very similar in both cases. Steel analysis by GD-MS bene®ts from the addition of 1% of H2 to the Ar discharge gas [614], the explanation for which is certainly complex. For certi®ed reference steels, including superalloys, reliable analysis results can be obtained. The determination of Mo, Nb and Zr in steels by GD-MS was found to be aected by the formation of multiply-charged cluster ions (metal argides) [615]. A correction based on the assumption that the rate of formation of the singly-charged argide is the same for all analytes and coincident with that of FeAr was used. The capabilities of low resolution GD-MS were shown by the example of steel analysis [616], where detection limits were down to 1 ng/g and up to 30 elements could be determined.
6.2 Glow discharge mass spectrometry
Fig. 123. Glow discharge mass spectrum at high resolution (A) (reprinted with permission from Ref. [610]) and at low resolution (B) (reprinted with permission from Ref. [592]).
Rare earth elements have been determined in metallic Gd, La, Nd, Pt and Tb by GD-MS. Surface contamination can be removed by a 10 min predischarge and careful assessment of polyatomics interferences performed. At the 1 mg/g level RSDs are up to 40%. Also rf-GD-MS has been applied to the analysis of metals and its ®gures of merit discussed [599]. Both for high-purity substances [617] and semiconductor-grade materials [618] up to 16 elements can be determined in 5N and 6N type samples. At high resolu-
283
284
6 Plasma Mass Spectrometry
tion the determination of Ca and K is also possible [611]. Mykytiuk [619] showed that GD-MS can now eectively replace spark source mass spectrometry for a number of applications with excellent detection limits (often at the ng/g level) and with greater independence from the matrix composition. Interferences are still limiting in a number of cases, but Oksenoid et al. [620] showed that a substantial improvement by several orders of magnitude in analyte-to-interfering Ar gas intensity could be obtained by optimizing the lamp geometry, the working voltages and the gas ¯ow. This allowed elements such as Br, K and Rb to be determined down to 1 mg/g, 1 mg/g and 10 ng/g, respectively. Routine bulk trace determinations in high-purity metals often revealed dierences of one order of magnitude [621], which was attributed to sample inhomogeneity. Vieth and Huneke [609] determined 56 elements in six alloys and found good agreement between the RSFs and those calculated on the basis of a semi-empirical model accounting for diusion, ionization recombination and ion-extraction phenomena. High-purity and alloyed Al can be analyzed with a presputtering time of 30 min at a voltage higher than 1000 V and a current of 5.5 mA. For the actual analysis 1000 V and 2 mA were found to be suitable [622]. An investigation of background spectral interferences revealed little in¯uence of gaseous, aqueous and solution species, although the formation of metal argides and multiply charged Ar species could not be disregarded in the evaluation of mass spectra [623]. For Al the detection limits of the rf-GD were found to be higher than those stated for commercially available high-resolution GD mass spectrometers, namely, in mg/g: 0.61 for Mn, 0.53 for Ni, 0.19 for Pb, 0.17 for Cr and 0.15 for Bi [601]. In Al samples Th and U concentrations of 20 ng/g can still be determined [617]. Limits of detection in the ng/g range and below with precisions of between 7 and 30% were obtained in the analysis of Al- and Co-based alloys by GD-MS [624], where the importance of ultraclean working conditions was stressed. Saito et al. [625] determined B in highpurity Mo down to 0.1 ng/g. For the elements C, N and O in steel at concentrations of 20 mg/g and below the lamp must be cooled to reduce the background species intensities, evacuated over a period of 20 min to allow the determination of N, presputtering should be used for the determination of C and the Ar puri®ed with a ZrO catalyst to improve the quanti®cation of O. When using cooling in the case of high-purity Ga, the detection limits were below the blank values of chemical analysis [626]. This also seemed to be useful in the analysis of YBa2 Cu3 Ox and other high-temperature superconductors with a pin GD [627]. For metal alloys isotope ratios in solid samples can be determined for U and Os by GD-MS [628]. To avoid interferences Kr was selected as the discharge gas, but the results were still seriously aected by multiply-charged ions of Ca, Kr and Si. Isotopic measurements of Pd charged electrolytically with protium and deuterium were performed with a precision of better than 0.07%, however, only with a highresolution instrument could interferences be prevented [629]. Also cooling of the discharge cell was required so as to eliminate hydride interferences from adsorbed H2 O. Trace elements were determined in Pt powder by GD-MS after ascertaining their RSFs [630]. The interference of PtH could not be disregarded in the case of
6.2 Glow discharge mass spectrometry
Au and Ir, requiring cryocooling to eliminate it. Accuracy and precision were in the range 10±15 and 5±10%, respectively, while the power of detection turned out to be better by 1±2 orders of magnitude than in ICP atomic emission spectrometry. High-purity Ti has been analyzed by GD-MS to ascertain the Sc content [631]. The interference of 50 Ti 40Ar 2 limited the power of detection to 25 ng/g. Magnetron GD plasmas have been used for high-precision analysis of pure Cu, Mg, Al, Zn-based and Mn alloys and showed potential for high-precision analyses [632]. Copper-based alloys have been analyzed by quadrupole GD-MS using a system with optimized ori®ce-to-cathode distance, energy analyzer setting and bias voltage and trace elements were determined down to the mg/g level with a precision of 2±5%. Hutton and Raith [633] reported accuracies and precision similar to those of conventional arc and spark analysis. In 5N and 6N gallium Ca and K can be determined at high resolution with detection limits below the blanks of chemical analysis [634]. For compact SiC samples the analysis results were in good agreement with those of neutron activation analysis [618]. Powder samples
Electrically non-conducting powder samples can be analyzed after mixing with a metal powder such as Cu, Ag or Au and briquetting, as is known from experience with GD-OES. However, degassing and oxygen release may necessitate long preburn times and introduce instabilities. Also blanks arising from the metal powder may be considerable, e.g. for Pb in the case of copper powder. The compacted pellet approach for powders was used by De Gendt et al. [635], who investigated the in¯uence of the host matrix type (Ag, Cu), cooling and cathode shape (pins, disks) for some Fe ore certi®ed reference materials. While binder and sample matrix had little eect on experimental RSFs, cooling and sample geometry played a signi®cant role. Careful optimizing and controlling the working parameters allows accuracy and precision of the measurements of better than 10% to be obtained. The use of a secondary cathode for GD-MS of non-conducting samples such as solid glasses or sintered iron ores has been reported [636]. It was found that dierent sample types require ad hoc optimization of the measurement parameters, e.g. by changing the ratio between the sample and the secondary cathode signals. Furthermore, the purity of the secondary cathode may in¯uence blanks and detection limits and the electrical resistivity of the sample is important too. Iron meteorites have been analyzed for major (Co, Fe, Ni) as well as for minor and trace elements (up to 53 elements with detection limits down to the ng/g level) by GD-MS [637]. The isotopic composition was also shown to enable some information on the origin of the outer shell to be obtained and anomalies for C and N concentrations could be seen. Teng et al. [638] examined some factors aecting trace determinations in soils, in particular red clay and forest soil. Satisfactory results were obtained by re-sorting
285
286
6 Plasma Mass Spectrometry
the RSFs from soils CRMs. In particular the in¯uence of sample oxygen content and host binder identity were examined and GD-MS was found to be less prone to matrix eects than laser ablation ICP-MS. Betti et al. [639] used GD-MS for the determination of trace isotopes in soil, sediments and vegetation by blending with a conductive host matrix, namely Ag and using the secondary cathode technique so as to achieve a stable discharge. In this way detection limits in the pg/g range could be obtained for the radioisotopes 137 Cs, 239 Pu, 241 Pu, 90 Sr and 232 Th using an integration time of 1 h and a mass resolution of 100. Both total elemental concentrations and organometallic species could be determined in soil by means of GD-MS and GC-MS [640]. Suspended particulate matter can be analyzed after depositing it on high-purity In [641]. With detection limits at the sub-mg/g level data for 53 elements in a 10 mg sample were obtained, which proved to agree by better than a factor of 2 for 34 elements with results from other techniques. Atmospheric particulate matter was subjected to analysis by GD-MS by Schelles and Van Grieken [642]. Air was pumped through a single-ori®ce impactor stage in which the aerosol was collected on a metal support, which was then used as the cathode in the GD unit. The maturity of GD-MS as a technique capable of routinely providing complete chemical analyses at the ultratrace level for insulating solid materials has clearly been demonstrated by quanti®cation of a full range of elements (from Li to U) in coal and coal ¯y ash [643]. The samples were mixed with high-purity Ag powder as the binder and pressed into a pin shaped pellet by means of a polypropylene mold. Critical steps in the determination process were the inhomogeneous distribution of elements within and among the ¯y ash particles and the purity of the binder. The presence of highly volatile compounds hindered the application to bituminous coal. The analysis of Al2 O3 powders was possible after mixing with copper powder, but the presputtering was found to be critical. Also both argon and neon can be used as working gases, allowing detection limits down to the mg/g level to be obtained and optimization towards freedom from interferences [644]. Compact non-conductive samples
The ability of rf-GD sources to sputter or atomize compact non-conductive samples directly has been investigated intensively. The eects of cryogenic cooling, power, pressure and distance between sample and exit ori®ce have been investigated to improve the performance of rf-GD-MS in the analysis of oxides [645]. The ®rst parameter was found to be crucial to removing gaseous interfering species. After careful optimization a precision of 5% could be obtained and the RSFs were 0.5±3 depending on the matrix, allowing semiquantitative analyses only. The rf-GD-MS analysis of glass and ceramic samples has also been reported [646]. By adjusting the ion transfer optics of the double-focussing mass spectrometer the ratio of analyte-to-background contaminant ion intensities can be optimized. Macor, a fairly insoluble non-conducting glass ceramic used as an insulating material capable of withstanding high voltages, can be analyzed directly [647] and O, Si, Mg, Al and B determined.
6.2 Glow discharge mass spectrometry
Pigmented polymer coatings on steel and PTFE as well as PTFE-based copolymers can be analyzed by rf-GD-MS in order to ®ngerprint them [648]. Advantages compared with SIMS and x-ray photoelectron spectroscopy are that the method is fast and does not require dissolution of the sample, and thermal volatilization processes do not appear to take place. A tandem system consisting of pulsed dye laser ablation and ionization in a GD source for MS have been described by Barshick and Harrison [649]. The role played by the working gas (Ar, He, Ne) on redeposition of sputtered material has also been clari®ed. Removal of interfering species in GD-MS is possible through the use of getters such as Ag, C, Ta, Ti and W [650]. This approach has been applied successfully in the determination of rare earths so as to avoid oxide ion formation. Depth profiles
These can be recorded just as in GD-OES, however, with a much higher resolution so that here thin multilayers can still be characterized [613] (Fig. 124]. Also the conversion of the intensity scale into a concentration scale is easier and the sensitivity higher. By the use of dierent working gases (see e.g. Ref. [651]) the scope of the method can be continuously adapted compared to other methods for depthpro®ling. In the depth-pro®ling analysis of steel by GD-MS, initial degassing caused serious interference problems independent of whether fast or slow erosion rates were adopted. Thermal degassing under vacuum conditions in the ion source before igniting the discharge has been shown to be helpful. In frictional brass-coated steels quanti®cation of continuously varying concentrations of Cu, Fe and Zn can be performed based on a linear combination of the RSFs and/or sputter rates for both Fe and brass [652]. Technical layers in a CraNi system with a thickness of 30 nm to 10 mm can be characterized with respect to composition and thickness with GD-MS using a penetration rate of up to 0.1 mm/s and a depth resolution of 10 nm [613]. Dry solution residues
Analysis of dry solution residues can also be performed successfully with GD-MS. This has been shown with the aid of hollow cathodes, with which Sr, Ba and Pb can be determined [653] and where in 10±20 mL up to 70 elements can be determined with detection limits in the pg range [654]. With a Grimm-type discharge solution residues can be formed on the cathode by evaporating microvolumes of analyte solutions having low total salt contents. With noble metals such as Pt and Ir [655], use can be made of cementation to ®x the analyte onto a copper target, so that it is preconcentrated and can be sputtered reproducibly. This has been found to be a way of ®xing pg amounts of Ir on a copper plate prior to analysis with GD-MS. A GD-MS method has been developed for the analysis of microliter volumes of aqueous solutions which permitted the long-term acquisition of data [656]. Samples were either adsorbed on pin-shaped electrodes prepared by pressing high-
287
288
6 Plasma Mass Spectrometry Fig. 124. Intensity versus time
profile obtained in GD-MS analysis of a CraNi multilayer system on Si substrate: (a): discharge current and (b): single-ion monitoring profiles of A: Si, B: Cr and C: Ni (I), and concentration versus depth profile calculated from measurements (II). (Reprinted with permission from Ref. [613]).
purity Ag powder or by preparing a slurry of Ag powder and the solution subsequently dried and pressed into a pin. In both instances homogeneity and particle size of the individual materials are critical. Tests were carried out with the NIST SRMs 3171 and 3172 multielement solutions of Al, Ba, Be, Cd, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, Se, Sr and Zn. The average relative error was about 14% (2±30%) indicating the need for further work. For the determination of Pt in urine of patients treated with Cisplatin, the residues were dissolved in water and aliquots dried on the tips of carbon electrodes allowing the determination of both Pb and Pt down to the ng/g level. It was thus possible to conclude that Pb was displaced from body compartments and mobilized by administering Pt [657].
6.2 Glow discharge mass spectrometry
Acid digestion is recommended for the analysis of waste oils from of vehicles. A few mL of the resulting aqueous leachates can then be pipetted onto Ag powder and the slurries dried and pressed into polyethylene slugs to produce pins that can eventually be submitted to GD-MS analysis for the assay of their Pb contents [658]. Determinations can be performed by isotope dilution and concentrations as low as 3 mg/g Pb determined with a precision of better than 5%. GD-MS has also been used for the analysis of crude oils [659] and Cr, Cu, Fe. Mg, Na, Ni, Pb, Si, Sn and Ti can be determined in NIST SRMs, SPEX organometallic standard oils and re®ned oil composites. The method performs very well for limited amounts of sample, but the polyatomic interferences are a drawback. Gases and vapors
Glow discharges are also very useful to excite and ionize dry vapors as generated by electrothermal vaporization or hydride generation. In the ®rst case, Guzowski et al. [605] showed that by electrothermal vaporization of dry solution aliquots in a graphite cup atomizer, vapor sampling was possible and particularly in the case of TOF-MS allowed reliable determinations to be made. The system is very useful for studying vaporization eects, as signals of dierent elements can actually be monitored simultaneously and this also applies for isotope signals. It also was found that by alternating hard and soft conditions, even molecular information from bromoform or ferrocene can be obtained. Hydride generation can also be used for sample introduction into a helium, argon or neon discharge and depending on the pressure and current more molecular ions or atomic species are found. The technique has excellent sensitivity and a fairly good dynamic range and precision [660]. In speciation, glow discharges are excellent detectors for GC work as shown earlier. In addition to the low power and pressure ICPs they can be used successfully for element-speci®c detection for gas chromatography. An rf-GD-MS system has been used as a detector for GC by Olson et al. [661]. The set-up should consist of a temperature-controlled transfer line of stainless steel from the exit of the GC to the inlet of the GD source. The system has been tested with tetraethyl-Pb, tetraethyl-Sn and tetrabutyl-Sn and provided useful structural information for the identi®cation of these compounds through the observation of fragment peaks; the detection limits were down to 1 pg. Manufacturers
Glow discharge (GD) mass spectrometers with high (h) and low (l) resolution are available from several manufacturers.
. Extrel, Pittsburgh, PA, USA: GD(l) . Turner Scienti®c, Finnigan, Bremen, Germany: ICP GD(l) . VG Instruments, Fisons, Cheshire, UK: GD(h), GD(l)
289
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
290
7
Atomic Fluorescence Spectrometry In atomic ¯uorescence spectrometry the analyte is brought into an atom reservoir (¯ame, plasma, glow discharge, furnace) and excited by absorbing monochromatic radiation emitted by a primary source. This can be a continuous source (xenon lamp) or a line source (hollow cathode lamp, electrodeless discharge lamp or a tuned laser). Subsequently, the ¯uorescence radiation emitted, which may be of the same wavelength (resonance ¯uorescence) or of a longer wavelength (non-resonance ¯uorescence) is measured with or without being spectrally resolved.
7.1
Principles
A simpli®ed model for the ¯uorescence process can be drawn up for two levels [662]. When a two-level system is considered (Fig. 125) and excitation is expected to occur only as a result of absorption of radiation with radiant density rn, without any contributions from collision processes to the excitation, the population of the excited level
n2 can be given by: dn2 =dt n1 B12 rn ÿ n2 A 21 k 21 B21 rn
301
where B21 is the transition probability, n the frequency and n1 is the population of the lower level. A 21 is the transition probability for spontaneous emission, B21 the transition coecient for stimulated emission and k 21 is the coecient for decay by collisions. The exciting radiation is supposed to be ``white'', this means a constant radiation density over the absorption pro®le. When the statistical weights of n1 and n2 are equal, B21 B12 and when in addition n T n2 n1 , this being the condition for a closed two-level system in thermal equilibrium, Eq. (301) can be solved as: n2 n T B rn t r 1 ÿ exp
ÿt=t r because n2 0 when t 0.
302
7.1 Principles Transitions in a two-level system for atomic fluorescence spectrometry. (Reprinted with permission from Ref. [662].)
Fig. 125.
t r
2 B rn A 21 k 21 ÿ1
303
and is referred to as the response time of the pumping process. In the case of primary sources with low intensity and atom reservoirs at low pressure, e.g. glow discharges in which quenching is low: n2
t n T B rn tsp 1 ÿ exp
ÿt=tsp
304
or in equilibrium: n2
teq n T B rn tsp
305
When the primary source is of a low intensity and quenching occurs in the atom reservoir, as is the case in an atmospheric pressure furnace, ¯ame or plasma: n2
t n T B rn t 1 ÿ exp
ÿt=t
306
with t
A 21 k 21 ÿ1
307
being a longer response time. When the absorption of radiation increases up to a certain value A 21 and k 21 become negligible. Then n2 n T =2 and becomes independent of the radiant density of the exciting radiation and a state of saturation is reached. This situation can be realized when lasers are used as primary sources. The radiaton density required to obtain saturation can be calculated as: rus
A 21 k 21 =2B
308
291
292
7 Atomic Fluorescence Spectrometry
this being the radiant density with which an equilibrium population at the point of half saturation is obtained. Apart from ¯uorescence with two levels, cases where more than two levels are involved in the process are also known. The radiating level may often be somewhat below the level to which pumping occurs and the transition between the two levels can be radiationless and take place through heat losses. Also the ¯uorescence transition sometimes does not end at the ground level. With two levels resonance ¯uorescence occurs, where stray radiation from the exciting radiation limits the power of detection. In the case of three-level systems there is non-resonant ¯uorescence, where this limitation does not apply. However, here the radiant densities are much lower. Therefore, populating the excited level will only be suciently successful when using very intensive primary sources such as tunable lasers. In atomic ¯uorescence work, the radiant density of the primary source can vary with:
. time: pulsed or continuous sources; . frequency: when the source has a certain spectral pro®le, i.e. white or line radiation with a certain line width;
. space: the source can be spatially inhomogeneous, through which suitable optics for the sampling of the ¯uorescence radiation is required. The absorption pro®le of the atoms in the atom reservoir is a function of the different line broadening mechanisms and contibutions to the physical line width. When self-absorption is negligible, the intensity of the ¯uorescence radiation is given by: BF n2 hu 0
1=4p A 21 n2 hu 0
1=4p
1=tsp
309
Where l is the thickness of the ¯uorescence layer. By combination with Eq. (304), it becomes obvious that:
. at primary sources with low intensities the ¯uorescence increases linearly with the radiant density of the primary source;
. when the radiant density of the primary source is sucient to obtain saturation (e.g. a tunable laser) the radiation density of the ¯uorescence radiation cannot be increased further by increasing the intensity of the primary source. Equilibrium equations for three-level systems can be set up just as for two-level systems. The models must be corrected because of the following.
. The absorption of the exciting radiation is only proportional to the density of analyte atoms and the depth of the atom reservoir in the ®rst approximation. As the intensity decreases with the depth because of absorption, a correction factor
7.2 Instrumentation
must be used and the primary intensity is given by: rn
nf1 ÿ expÿk
n Lg dn
310
. The ¯uorescence radiation itself is also only proportional to the atom number densities and the pathlength through the atom reservoir in a ®rst approximation because of self-absorption. Both of these conditions depend on the set-up and also on the fact that a continuous source or a line source is used. As especially in the case of non-resonant AFS the ¯uorescence intensities are low, special precautions have to be taken for signal acquisition. Indeed, the signalto-noise ratios can be improved considerably by using pulsed signals and phasesensitive ampli®cation, as is easily done e.g. in the case of laser sources (see Section 2.1).
7.2
Instrumentation
A system for AFS consisits of a primary source, an atom reservoir and a detection system for measuring the ¯uorescence radiation. In some cases the detection can actually be performed without the need for spectral dispersion, e.g. by using ®lters only. Primary sources
As primary sources, continuous sources such as a tungsten halogenide or a deuterium lamp can be used. They have the advantage that multielement determinations are possible. However, because of the low radiant densities saturation is not obtained and the power of detection is not fully exploited. With line sources such as hollow cathode sources and electrodeless discharge lamps much higher radiances can be obtained. Even ICPs into which a concentrated solution is introduced can be used as a primary source, through which multielement determinations become possible. With laser sources in the continuous as well as in the pulsed mode saturation can be obtained. In most cases dye lasers are used. They have a dye (mostly a substituted coumarine) as the active medium. This liquid is continuously pumped through a cuvette and has a broad wavelength band (ca. 100 nm), within which a wavelength can be selected with the aid of a grating or prism. By using several dyes the whole wavelenth range down to 440 nm can be covered. The laser is often pumped by a second laser (e.g. an excimer laser) and can be operated in a pulsed mode. As its energy output is very high, it is possible to apply frequency doubling and Raman shifting, with which the accessible wavelength range can be decreased down to the 200 nm range. Accordingly, with some ¯exibility saturation can be
293
294
7 Atomic Fluorescence Spectrometry
realized for almost any term. In the case of resonant transitions ( f 0:1±1) the energy required therefore is 10±100 mW, whereas for a non-resonant transition
f 10ÿ3 it is up to 1 W. Tunable dye lasers allow frequencies of up to around 10 Hz to be achieved and the pulse length can be down to 2±10 ns. Fluorescence volumes
Chemical ¯ames are very useful as ¯uorescence volumes, because of their good sample uptake capabilities and temporal stability. As the temperature is limited (3000 K for the C2 H2 aN2 O ¯ame) problems arise for elements with thermally stable oxides, as is known from experience with ¯ame AAS. These problems do not occur when an ICP is used as the atom reservoir [663]. In the ICP the measurements should be made in the higher part of the tail ¯ame as otherwise the ionized and excited analyte fraction is too high. To keep the formation of oxides and the sideways diusion of analyte low, a burner with an extended outer tube should be used for AFS work. In a graphite furnace the atomization eciency is high and due to the limited expansion of the vapor cloud, the residence time of the analyte atoms is high as well. Accordingly, graphite furnace atomization laser AFS has extremely low absolute detection limits. Discharges under reduced pressure can also be used as the ¯uorescence volume. They have the advantage that quenching of the ¯uorescent levels, which may limit the power of detection but can also cause matrix eects, is lower [664]. Elements with low intensity ¯uorescence lines (e.g. Eu, Tm and Y) have been determined in aqueous solutions by depositing and drying nanoliter amounts of sample on the Ni cathode of a miniature GD source used as the atom reservoir [665]. The atomic cloud thus formed was excited by a Cu-vapor laser-pumped dye laser to detect ¯uorescence directly. Absolute detection limits of 2 fg for Eu, 0.08 fg for Tm and 1.2 pg for Y were achieved and the total time for analysis from sample probing to data acquisition did not exceed 5 min. Furthermore, laser-produced plasmas may also be used as the ¯uorescence volume [226], which has been shown to be particularly useful for direct trace determinations in electrically non-conducting solids. Detection
The detection of the ¯uorescence radiation diers in resonant and in non-resonant AFS. In the ®rst case, the radiation is measured in a direction perpendicular to that of the incident exciting radiation. However the system will suer from stray radiation and emission of the ¯ame. The latter can be eliminated by using pulsed primary sources and phase-sensitive detection. In the case of non-resonant ¯uorescence, stray radiation problems are not encountered, although the ¯uorescence intensities are lower, which necessitates the use of lasers as primary sources and spectral apparatus that will isolate the ¯uorescence radiation. A set-up for laser excited AFS (Fig. 126) may make use of a pulsed dye laser pumped by an excimer laser. The selection of the excitation line is then done by the choice of the dye and
7.3 Analytical performance Fig. 126. Experimental set-up for laser induced atomic fluorescence work. (a): Flame, ICP, etc., (b): dye laser; (c): pumping laser, (d): photomultiplier, (e): monochromator, (f ): boxcar integrator, (g): data treatment and display.
the tuning of the laser. Pulses with a length of 2±10 ns occur at a frequency of up to 25 Hz. A mirror can be placed behind the source so that the exciting radiation passes through the ¯uorescence volume twice and to lead the largest possible fraction of the ¯uorescence radiation into the monochromator. The photomultiplier signal is then ampli®ed and measured with a box-car integrator of which a ns window is synchronized with the pumping laser.
7.3
Analytical performance Detection limits
The detection limits in AFS are low particularly for elements with high excitation energies, such as Cd, Zn, As, Pb, Se and Tl. In ¯ame AFS, elements which form thermally stable oxides such as Al, Mg, Nb, Ta, Zr and the rare earths are hampered by insucient atomization. This is not the case when an ICP is used as the ¯uorescence volume. Here the detection limits for laser excitation and non-resonant ¯uorescence are lower than in ICP-AES (Table 18) [663]. ICP-AFS can be performed for both atomic and ionic states [664]. Using a graphite furnace as the ¯uorescence volume and laser excitation, the detection limits are very low; e.g. Ag: 3 10ÿ3 , Co: 2 10ÿ3 , Pb: 2 10ÿ3 ng/mL [665, 666]. Very sensitive determinations of Mo can be performed by dry solution residue analysis with laser-excited AFS in a hollow cathode discharge as the atomizer, as shown by Grazhulene et al. [667]. Bolshov et al. [668] showed that very low levels of lead in Antartic ice samples could be determined by laser-excited AFS using dry solution residue analysis with graphite furnace atomization. When performing laser-excited AFS at a laser plume, it would appear to be useful to produce the laser plasma at pressures below atmospheric pressure, as then the ablation depends only slightly on the matrix [226, 669].
295
296
7 Atomic Fluorescence Spectrometry Tab. 18. Detection limits in laser atomic fluorescence spectrometry.
Element
Detection limit (ng/mL) Laser AFS ICP [663]
Ag Al B Ba Co Cu Eu Fe Ga Ir Mn Mo Na Pb Pt Si Sn Ti Tl V Y Zr
Laser AFS furnace [665] 0.003
0.4 4 0.7 0.002 0.002 10 0.003 1 0.2 0.006 5 0.02 0.000025 4 1 3 1 7 3 0.6 3
Spectral interferences
In atomic ¯uorescence spectrometry spectral interferences are low as the ¯uorescence spectra are not line rich. Linear dynamic range
The linear dynamic range of AFS is similar to that of OES or even larger. Indeed there is no background limitation nor is there a serious limitation because of selfreversal, both of which limit the linear dynamic range in OES at the lower end and at the higher end, respectively. It may well be that the exciting radiation becomes less intensive when penetrating into the ¯uorescence volume and also that ¯uorescence radiation is absorbed by ground-state atoms before leaving the atom reservoir. These limitations, however, are less stringent than self-reversal in OES. The application of laser-excited AFS in the analytical laboratory up to now has been seriously hampered by the complexity and cost of tunable lasers, which may change with the availability of less expensive but powerful and also tunable diode lasers, which can be operated in the complete analytically important wavelength range.
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
297
8
Laser Enhanced Ionization Spectrometry Laser enhanced ionization spectrometry (LEI) [670] is based on the optogalvanic eect, which has been known since the early work on gas discharges. It was found that the current in a discharge under reduced pressure increases when irradiating with monochromatic radiation of a wavelength which corresponds to a transition in the term diagram of the discharge gas. As an analytical method LEI is based on a measurement of the increase of the analyte ionization in an atom reservoir resulting from a selective excitation of atoms or molecules with the aid of laser radiation [669, 670].
8.1
Principles
Ionization of atoms X in an atom reservoir can result from collisions with energyrich atoms of another species M (Fig. 48) and the rate constant for the reaction k1i
M X M e X can be written as: k1i nX 8kT=
p m r 1=2 Q MX exp
ÿEi =kT
311
where nX is the concentration of analyte atoms, Q MX is the cross section for collisions between X and M, which are atoms of another element, k is Boltzmann's constant, Ei is the ionization energy and T is the temperature in the atom reservoir. When the atoms X are in an excited state E2 above the ground level, it can easily be seen that k1i will change into k2i exp
E2 =kT k1i
312
and the increase in ionization for a level that is 1 eV above the ground level in the case of a ¯ame with T 2500 K is e 5 150. The ionization equilibrium will change although the recombination rate remains the same. Moreover, states with
298
8 Laser Enhanced Ionization Spectrometry
high energies (so-called Rydberg states) have a longer lifetime (t A n 3 with n being the main quantum number) and therefore ionization through collisions or through ®eld ionization increases. However, the probability of ionization does not increase in an unlimited way with E2 in one-step processes as the transition probability for the transition 1 ! 2 decreases with E2 ÿ E1 . The increase in ionization also depends on the population of the level 2. In order to get a suciently high population in the excited level, to which the transition probability should be suciently high, a laser must be used. Indeed, for a closed system and two terms n1 and n2 , the equilibrium is characterized by the equation: B12 rn n1 B21 rn n2 A21 n2 k21 n2
313
where B21
g1 =g2 B12 and n T n1 n2 n i . B12 and B21 are the transition probabilities for absorption and stimulated emission, respectively, and n1 , n2 , n i and n T the analyte atom densities for the lowest state, the excited state, the ionized state and the total number densities. g1 and g2 are the statistical weights, A21 is the transition probability for spontaneous emission and k21 the coecient for collisional decay. Accordingly, n2 B12 rn
n T ÿ n i =
1 g1 =g2 B12 rn A21 k21
314
n2 will increase until A21 k21 becomes negligible and then n2 =n1 g2 =g1 . The required radiation densities are therefore about 10±100 mW for a resonant transition ( f 0:1±1) and 1 W/cm 3 for a non-resonant transition ( f 0:001). Furthermore, the use of a continuous wave laser allows n i to be maximum and the use of multistep excitation may be helpful especially when high-energy terms (Rydberg states) have to be populated. Indeed, when F is de®ned as: F n2 =
n i n2 n2 =
n T ÿ n i n2 =nt
315
ÿdn2 =dt k2i n2 k2i F
nt ÿ n i k2i F nt
316
Integration leads to: ln nt ÿk2i F t C
317
and as nt n T ÿ n i and at t 0 nt n T : n i n T 1 ÿ exp
ÿk2i F t
318
For an irradiation over the time Dt1 n i ! n T when: Dt1 g 1=
k2i F
319
Accordingly, the laser irradiation should last long enough therefore cw (continous
8.1 Principles
wave) lasers might be more appropriate than pulsed lasers, provided the radiant densities are suciently high. It could also be that it is more bene®cial to populate the Rydberg level in two steps, because it is easier to reach a level that is near to the ionization energy when very short wavelengths are not used. For an ionization potential of 7 eV, radiation with l < 220 nm is required, as for 1 eV 1:6 10ÿ16 erg, l is 1240 nm (because of E h c=l). In addition, the oscillator strength decreases with E2 ÿ E1 and in the case of two-step procedures the selectivity also increases. LEI as an analytical technique thus requires a tunable laser, an atom reservoir, a galvanic detection system and read-out electronics including a boxcar integrator. It was ®rst realized experimentally with a ¯ame as the atom reservoir using pneumatic nebulization for sample uptake [670]. A pulsed dye laser pumped with ¯ash lamps is used in most cases, as here the selection of dierent dyes, frequency doubling and Raman shifting, which is possible as a result of the high energies available, allows the whole spectral range down to 200 nm to be covered. However, diode lasers, particularly for multistep excitation, and dye lasers pumped with an excimer or a nitrogen laser can also be applied. The atom reservoir can be a ¯ame, furnace or also a plasma discharge at atmospheric and at reduced pressure. In the case of a ¯ame (Fig. 127), the sample is usually introduced by pneumatic nebulization. The laser enters the ¯ame a few mm above the burner head and a large collector electrode is situated a few mm above the laser beam location. This allows the ions formed to be extracted with a high eciency. The collector electrode should be large so as to minimize changes in the ®eld as a result of changes in electron number densities due to dierent concomitant concentrations. It should also be positioned inside the space containing the ions produced by LEI so as to minimize analyte signal losses. The measurement unit includes phase-sensitive ampli®cation to ensure high S/N ratios, but does not require any spectrometric equipment as only an ion current is being measured.
Fig. 127. Flame laser enhanced
ionization spectrometry [670, 671]. (a): Flashlamp/dye laser, (b): high voltage, (c): trigger photodiode, (d): preamplifier, (e): pulse amplifier, (f ): active filter, (g): boxcar averager, (h): chart recorder. (Reprinted with permission from Ref. [671]).
299
300
8 Laser Enhanced Ionization Spectrometry
8.2
Figures of merit
The sensitivity in LEI depends on the properties of the laser system, the terms populated, the atom reservoir and the detection system. It can be written as: S c XB b
Aexp
ÿEi =kT B12
A rnp
A
320
where c is a constant, XB is the number density of analyte atoms in the lower level according to Boltzmann, b is the atomization eciency in the ¯ame, rnp
A is the spectral radiation density per laser pulse, Ei is the ionization energy of the excited level and S is the sensitivity in nA/(ng/mL). The noise includes contributions from ¯icker noise, nebulizer noise and also detector noise, and when applying phasesensitive ampli®cation detection limits are in the sub-ng/mL range (Table 19) [670, 671]. They are lowest when using a furnace as an atom reservoir because here the highest atom number densities and the longest residence times are encountered. In the case of pulsed lasers the radiation densities enhance the sensitivities, whereas in the case of cw lasers the n i values increase as a result of improved excitation eciencies. In the so-called thermionic diode, a heated collector electrode is simply brought into the vicinity of the laser beam and isolated towards the vessel walls. No extraction voltage is applied and the ion current ¯owing as a result of the decreasing negative space charge around the collector, as a result of the ionization by the laser, Tab. 19. Comparison of limits of detection for cw and pulsed laser LEI. (Reprinted with permission from Ref. [671].)
Cw element/wavelength (nm) Ca 422.7 Ca 300.7 Ga 417.2 Ga 287.4 In 451.1 In 410.1 In 303.9 K 766.5 K 294.3 Li 460.3 Li 670.8 Na 589.0 Na 285.3 Rb 420.2 Rb 780.0 Rb 629.8 Sr 460.7
Pulsed limit of detection (ng/mL)
dynamic range (orders of magnitude)
1
5
60
>3
0.1 16
>4 4.5
limit of detection (ng/mL)
0.1 0.07
0.006 0.1
>4 1
20
>4 0.001
0.025
5 0.05
0.7 0.09
3 >5
0.4
>5
2
8.3 Analytical applications
is taken as the analytical signal. The whole system is usually operated at reduced pressure and allows detection of 10 2 atoms to be made (for Refs., see [672]). In LEI the matrix eects are low. This applies in particular to samples with complicated atomic spectra, as no spectral interferences are encounterd. Elements with low ionization energies may cause matrix eects. The latter could be due to a change in the ion current as a result of changes in the electron number density in the ¯ame. This can be compensated for by measuring alternately with and without laser irradiation. Ionization of the matrix by the laser may also lead to errors in the ion currents measured. These can be eliminated by modulation of the frequency of the exciting laser radiation around n, which corresponds to E2 ÿ E1 hn, where E2 is the intermediate level (Rydberg state) and E1 the ground level. Furthermore, the ion production as a result of the laser irradiation may in¯uence the collection ecieny at the electrode as the electric ®eld around the electrode changes. This eect can be minimized by using a large electrode and by collecting all of the ions produced. This sets limits on the pulsing frequency. Indeed, for a distance of 5 mm between the collector and the burner, a ®eld of 100 V/cm and a mobility of 20 cm 2 / V s, this being typical for a ¯ame in which T 2500 K, the highest allowed pulsing rate is 4000 Hz. Even so, for Na concentrations of 100 mg/mL matrix eects are still negligible. The linear dynamic range in ¯ame LEI is of the order of 3 decades. However, the multielement capacity is limited as monochromatic laser radiation is required. When using two lasers and applying two-photon spectroscopy, only those atoms that do not have a velocity component in the observation direction will undergo LEI. Then the absorption signals become very narrow (Doppler-free spectroscopy). This enhances the selectivity and the power of detection, however, it also makes isotope detection possible. Uranium isotopic ratios can thus be detected, similarly to with atomic ¯uorescence [673] or diode laser AAS. Thus for dedicated applications a real alternative to isotope ratio measurements with mass spectrometry is available.
8.3
Analytical applications
LEI has been applied successfully to the trace determination of Tl [674] for certi®cation purposes, and for combinations with laser evaporation and all other atomization techniques represents a powerful approach to detection. Laser photoionization and galvanic detection have been applied to hollow cathode dark space diagnostics [675]. Photoionization is produced to measure the dark space widths of linear ®eld distributions directly. A theoretical model has been developed and its predictions veri®ed with experimental ®ndings for a uranium hollow cathode discharge operated in neon or xenon. Variations in the ground-state densities of sputtered neutrals have also been measured. Apart from the galvanic detection of the ion currents, direct mass spectrometric detection of the ions can also be applied, as is the case with resonance ionization mass spectrometry (RIMS) [676]. In addition, ionization can be performed by multiphoton absorption, which requires very intense primary sources.
301
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
302
9
Sample Preparation for Atomic Spectrometry In a large number of atomic spectrometric methods sample preparation is required but depending on the method or sample introduction technique used, it can vary widely from a simple polishing step to a complete sample dissolution and subsequent analyte±matrix separation.
9.1
Sample preparation in direct compact sample analysis
In the direct analysis of metal samples, as in spark or x-ray ¯uorescence spectrometric analysis for process or product control, as treated in the pertinent literature (see e.g. Ref. [677]) treatment of the sample surface by turning-o or polishing with suitable emery paper may be adequate. However, in order to obtain the highest precision and accuracy, special procedures such as turning-o with a diamond in the case of glow discharge atomic spectrometry for bulk analysis [476] may be very helpful. New procedures such as laser ablation with a moving laser beam (LINA) no longer require any treatment of either metal or of ceramic or glass surfaces.
9.2
Grinding, sieving and compaction of powders
Grinding is of paramount importance since it allows homogenization in the case of direct solids analysis, and when applying sample dissolution it is often indispensable so as to increase the eectiveness of treatment with the aggressive reagents required for decomposition and dissolution. Sample grinding can be done in rotation mills using vessels and pestles made of dierent materials. Stainless steel or very hard materials such as tungsten carbide or agate are very useful. When grinding hard materials the dierence in hardness between the sample materials and the grinding tools should at be least 2 on the Mohs hardness scale. In the case
9.2 Grinding, sieving and compaction of powders
of SiC for example, which is produced as a solid material, grinding with WC introduces considerable amounts of W and with steel mills, Fe is found in the sample materials. When grinding SiC granulates in a high-purity mill made of SiC the abrasion of the mill could be up to a few % of the sample [546]. Grinding allows the mean particle size to be brought down to around 10 mm and in the case of wet grinding to around 5 mm. After grinding, particle sizes of powders can easily be determined with laser diraction of a slurry of the material, or by automated electron probe microanalysis after depositing a layer on Nuclepore ®lters and coating with C or Au to render the sample electrically conductive (see Fig. 62 for comparative measurements) [203]. In order to sieve samples, sets of sieves with increasing mesh number and made of dierent materials, including very inert ones, can be used. For the compaction of powders it is often necessary to mix the powder with a binder, to homogenize the mixture and to apply high pressure. Some powders such as Cu, Au, Ag and graphite powder may be mixed in a ratio of up to of 5:1 with the sample powder and electrically- as well as heat-conductive and even vacuumtight pellets can be obtained. To reduce the amount of sample material required, it is often sucient just to place the sample binder mixture in the middle of
Fig. 128. Preparation of electrically- and heatconductive as well as vacuum-tight pellets for the analysis of powders by direct solids sampling atomic spectrometric methods. (a) Side-on view, (b): top view (according to Ref. [210]).
303
304
9 Sample Preparation for Atomic Spectrometry
the pellet (Fig. 128) [210]. Pressures required to obtain stable pellets can be up to 80 T/cm 2 .
9.3
Sample dissolution
Sample dissolution can be performed with liquid or solid reagents at elevated temperatures causing a chemical reaction or melting of the mixture. The dissolution aids may introduce considerable contamination, as they are present in large excess. In addition, the vessels and also the laboratory atmosphere may constitute a source of contamination, whereas analyte losses particularly due to volatilization or adsorption onto the vessel walls may occur. Dissolution procedures for many types of materials can be found in the literature. They are used together with methodology for trace matrix separations based on classical chemical principles. 9.3.1
Wet chemical methods
Wet chemical dissolution is often performed with acids and the eectiveness increases with the oxidizing capacity. The latter depends on the acid (HCl < HNO3 < HClO4 < HClaHNO3 ) and the maximum temperature, depending on the boiling point of the acid (HCl < HNO3 < H2 SO4 < H3 PO4 ). As the maximum temperature increases with the pressure work is often carried out in PTFE vessels in steel autoclaves at high pressure (up to 200 atm). The acids that give very soluble salts, such as HNO3 or HClO4 (not for K) are advantageous. Acids such as HNO3 and HCl can be very eciently puri®ed by isothermic distillation. Quartz is preferred over glass as a vessel material. However, when HF is required for the treatment of silicon containing materials, vessels made of Pt or PTFE should be used. Whereas quartz and glass can be cleaned successfully by purging them with steam or HNO3 vapors, Pt is rinsed most eciently with dilute chlorine free HNO3 . In order to keep contamination risks low during chemical treatment in open systems, work in cleanrooms including the use of laminar-¯ow clean benches is advisable [678]. 9.3.2
Fusion procedures
Fusion with alkali ¯uxes (Na2 CO3 , K2 CO3 , Na2 B4 O7 , etc.) as well as acid ¯uxes (KHSO4 ) or more volatile salts such as NH4 HSO4 in an excess of often 10:1 is most eective for the dissolution of many refractory powders such as Al2 O3 or ZrO2 . However, these methods are hampered by the purity of the ¯uxes or abrasion of the vessels when it comes to applications for trace analysis. Recrystallization is a way forward, except for chemicals such as NH4 HSO4 , which can be prepared from pure chemicals (ClHSO3 and pure NH3 ) [312]. In the case of ZrO2 powders such ¯uxes
9.4 Flow injection analysis
are very useful, as the excess can be volatlized easily and the fusion can be made in quartz vessels at no more than 500 C. Special equipment for the parallel fusion of several samples including automatic pouring of the molten mixtures into water for dissolution or leaching are commercially available. 9.3.3
Microwave-assisted methods
Microwave-assisted treatment with acids, sometimes at pressures of up to 80 atm in vessels made of ®ber enforced plastic is very ecient for increasing the speed and eectiveness of dissolution procedures. This is related to the fact that the liquid mixture is heated directly and not by convection through the wall. Many procedures have been described in a monograph [679] and in the appropriate literature. Also procedures for slurries with ¯ow-through systems have been proposed [680] and used very eciently, e.g. for powdered biological materials. Systems enabling the dissolution of several samples to be carried out simultaneously and under rigid pressure and temperature control [681] are now commercially available. Resistance-heated systems have also been proposed for on-line dissolution as an alternative to microwave-heated systems, mainly for the treatment of slurries of biological samples [130]. 9.3.4
Combustion techniques
Combustion in an oxygen atmosphere of samples that have an organic matrix has long been used for sample decomposition. In the so-called Trace-O-mat [682] cellulose powder is often added to the sample so that a number of inorganic samples such as soils can even be decomposed. The decomposition products are then collected in a liquid nitrogen cooled trap. As a further development for inorganic refractory samples, ¯uorination has been proposed. This can be done with minute (about 10 mg) amounts of elemental ¯uorine in nickel reactors [683, 684]. The volatile reaction products can be determined on-line by IR spectrometry [685] or gas mass spectrometry [686]. In order to eliminate spectral interferences a controlled distillation of the condensed reaction products from a liquid nitrogen cooled system can be applied and this allows, for example, the main contaminants of Si3 N4 to be determined with reasonable precision.
9.4
Flow injection analysis
Trace matrix separations and concentration procedures are required in combined procedures, which are a prerequisite to characterizing reference materials, as then
305
306
9 Sample Preparation for Atomic Spectrometry
calibration with solutions containing only the elements to be determined can be applied. Such procedures are now used in ¯ow injection based systems.
9.5
Leaching sample preparation methods
The isolation of analytes by leaching, as is often done in soil analysis, is often dictated by the diculties encountered when dissolving samples completely. However, also for speciation work, where the analytes are labile such procedures are often the only way to obtain the analyte solutions (see e.g. Ref. [687] for the isolation of organomercury compounds from ®sh samples).
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
307
10
Comparison with Other Methods Atomic absorption, optical emission and atomic ¯uorescence as well as plasma mass spectrometry and new approaches such as laser enhanced ionization now represent strong tools for elemental analysis including speciation and are found in many analytical laboratories. Their power of detection, reliability in terms of systematic errors and their costs re¯ecting the economic aspects should be compared with those of other methods of analysis, when it comes to the development of strategies for solving analytical problems (Table 20).
10.1
Power of detection
AAS using furnaces remains a most sensitive method, certainly for easily volatilized elements with high excitation energies and this also applies to AFS using laser excitation. AAS will de®nitely bene®t from the development of diode lasers, as it is then possible to eliminate the use of spectrometers for spectral isolation. Moreover, the dynamic range problems could be decreased by measuring in the wings of the absorption lines and the background measurement also could be performed by tuning the laser. This necessitates, however, obtaining lasers with which the analytical wavelengths down to the VUV can be covered. LEI is another most powerful approach with respect to power of detection, especially when applying Doppler-free spectroscopy. However, these methods have limited multielement capacity. From this point of view, plasma atomic spectrometric methods are more powerful. ICPOES will remain an important tool for multielement determinations in liquid samples and its power of detection is between that of ¯ame and furnace AAS. Thus it is the appropriate method for tasks such as waste water analysis. As a result of the availability of new detectors enabling multiline calibration and actual simultaneous acquisition of line and spectral background intensities, ICP-OES is now capable of high-precision analyses, which previously could only be done with laborious wet chemical and spectrophotometric methods. ICP-MS has provided further progress particularly with respect to power of detection, provided the matrix loading of the sample solutions remains low. There may also be limitations in the low mass range (below mass 80), due to interfer-
308
10 Comparison with Other Methods Tab. 20. Figures of merit of analytical methods.
Method Atomic absorption Flame Furnace Atomic emission Dc arc Spark DE plasma ICP MIP Glow discharge Laser Atomic ¯uorescence (laser) LEI X-ray spectrometry XRF Electron microprobe PIXE Auger electron spectroscopy Mass spectrometry Spark ICP Glow discharge SIMS Laser Activation analysis Electrochemistry Spectrophotometry Spectro¯uorimetry Chromatography Gas Liquid
Detection limit
Matrix effect
Cost
(a)
Low; medium; very high. a TRXRF total re¯ection x-ray ¯uorescence.
ences by cluster ions. The use of sample introduction techniques generating dry aerosols has certainly improved matters, e.g. for Fe, As and Se. Further improvements are possible by the use of high-resolution mass spectrometry using sector ®eld instruments. However, lower cost instruments are required so as to make the technique aordable to a broader circle of users. Time-of-¯ight mass spectrometry will de®nitely help signi®cantly wherever transient signals are obtained and multielement determinations are required, such as in electrothermal evaporation, single shot laser ablation and gas chromatographic detection. Therefore, multicollector mass spectrometers eventually with multichannel detection are the ultimate solu-
10.2 Analytical accuracy
tion to precise isotope ratio determinations as are required in tracer or isotope dilution work. Whereas ICP atomic spectrometry is certainly a most useful method for multielement determinations in liquid samples at low concentration levels, glow discharges are very useful for the analysis of solids. Glow discharge atomic emission gained its merits speci®cally for depth pro®ling analysis and in addition is useful for bulk analysis together with standard methods such as arc and spark emission spectrometry. Glow discharge mass spectrometry was ®rst found to be of interest for the analysis of ultrapure metals, as required in microelectronics. However, quanti®cation is dicult and certi®ed samples are required for calibration. When applying dry solution residue analysis, GD-MS has extremely low absolute detection limits. This approach is very useful for many applications in the life sciences, as it combines a high power of detection with the possibility of isotope dilution. In mass spectrometry this approach oers new possibilities, as more classical thermionic techniques are only applicable to elements of which the species can be evaporated. The use of soft discharges is interesting, as molecular information can be obtained. With solids, soft laser ablation combined with time-of-¯ight mass spectrometric detection is also useful. Optical atomic and mass spectrometric methods can be used for the determination of the light elements, which is an advantage over x-ray spectrometric methods. Apart from this restriction, total re¯ection x-ray ¯uorescence provides a high power of detection, high accuracy and high multielement capacity, all with a minimum of sample preparation [688]. It should be considered together with developments such as work with polarized x-rays enabling much lower detection limits to be achieved, also with energy-dispersive systems and local analysis with focussed x-ray radiation, as well as with the use of soft x-rays facilitating speciation in a number of cases [689].
10.2
Analytical accuracy
The analytical accuracy of methods can only be discussed with regard to the complete analytical procedure applied. Therefore, it is necessary to optimize sample preparation procedures and trace±matrix separations speci®cally to the requirements of the analytical results in terms of accuracy, power of detection, precision, cost and the number of elements and increasingly of the species to be determined. However, the intrinsic sensitivity to matrix interferences of the dierent methods of determination remains important. In optical emission and in mass spectrometry, spectral interferences remain an important limitation to the analytical accuracy achievable. In atomic emission this applies particularly to the heavier elements as they have the more line rich atomic spectra. When these heavy metals are present as the matrix, as is often the case in metal analysis, the necessitity of matrix separations is obvious when trace analyses
309
310
10 Comparison with Other Methods
have to be performed. In order to overcome limitations by spectral interferences, high resolution Echelle spectrometers are ®nding more and more uses. They are compact and thus combine high resolution together with excellent stability and further enable multielement determinations to be made by using advanced detector technology, e.g. with CCDs. The latter are also very useful for classical Paschen± Runge spectrometers with many CCDs covering the whole analytical range. In mass spectrometry spectral interferences limit the accuracy in the low mass range in particular. Progress could be expected from the use of sector ®eld and time-of-¯ight mass spectrometry. However, in the ®rst case transmission and in the second case dynamic range problems must be given attention. Signal depressions and enhancements are a further main cause of interferences in ICP-MS. They can be succesfully taken care of by using standard additions, as in ICP-MS the spectral background is low. Furthermore, internal standardization may well allow compensatation to be made for easily ionized elements eects. This is more dicult in ICP-AES, where the spectral background especially in trace analysis is considerable and may be in¯uenced strongly by changes in the concentrations of easily ionized elements. In AAS and AFS, limitations to the analytical accuracy are mostly related to physical and chemical interferences and are due less to spectral interferences. In furnace AAS thermochemical processes limit the achievable accuracy and necessitate temperature programs to be carefully worked out in order to cope with errors arising from thermochemical eects. In AFS and also in LEI, it is necessary to control matrix in¯uences relating to quenching when analyzing real samples. Because of the necessity to characterize reference materials traceability in the measurements is very important. In atomic spectrometry background acquisition methods have improved so much that although it is not an absolute methodology, every step in the calibration and in the measurement processes can be extremely well characterized.
10.3
Economic aspects
The power of detection and the accuracy of analytical methods cannot be discussed without considering the expense arising from instruments and operating costs as well as from the laboratory personnel. Methods allowing multielement analyses to be performed and achieving a high throughput of samples are certainly advantageous for routine laboratories. In this respect, ICP spectrometric methods in particular oer good possibilities despite the high instrument costs and the high consumption of power and gases. Miniaturized spectrometers using new detector technology are both very powerful and at the same time much less expensive. Microplasmas are cheap to construct, as are the plasma generation and operation, and for well de®ned purposes such as element-speci®c detection in chromatography are already proving to be useful.
10.3 Economic aspects
In many cases, however, the costs arising from sample preparation will become decisive, which favors x-ray spectrometric methods, provided the earlier mentioned limitations are not encountered. Future progress will certainly depend on the availabilty of on-line sample treatment using, for example, ¯ow injection and eventually on-line sample dissolution as is possible in some cases with microwaveassisted heating. Also the realization of separations in miniaturized systems and with minute amounts of reagents is very promising. In each instance the question of which method should be selected will have to be discussed for each type of analytical task to be solved.
311
Analytical Atomic Spectrometry with Flames and Plasmas. Jose A. C. Broekaert Copyright > 2002 Wiley-VCH Verlag GmbH & Co. KGaA ISBNs: 3-527-30146-1 (Hardback); 3-527-60062-0 (Electronic)
312
Literature 1 Kirchhoff G. R. and Bunsen R.
2
3
4 5
6
7
8
9
10
(1860) Chemical analysis by spectrum-observation, Philos Mag 20: 89±98. Walsh A. (1955) The application of atomic absorption spectra to chemical analysis, Spectrochim Acta 7: 108±117. Grotrian W. (1928) Graphische Darstellung der Spektren von Atomen mit ein, zwei und drei Valenzelektronen, Springer, Berlin. Griem H. R. (1964) Plasma Spectroscopy, McGraw-Hill, New York. de Galan L., Smith R. and Winefordner J. D. (1968) The electronic partition functions of atoms and ions between 1500 K and 7000 K, Spectrochim Acta, Part B 23: 521±525. Pearce W. J. and Dickerman P. J. (eds) (1961) Symposium on Optical Spectrometric Measurements of High Temperatures, University of Chicago Press, Chicago. Boumans P. W. J. M. (1966) Theory of Spectrochemical Excitation, Hilger & Watts, London. Robinson D. and Lenn P. D. (1967) Plasma diagnostics by spectroscopic methods, Appl Opt 6: 983±1000. Kalnicki D. J., Kniseley R. N. and Fassel V. A. (1975) Inductively coupled plasma optical emission spectroscopy. Excitation temperatures experienced by analyte species, Spectrochim Acta, Part B 30: 511±525. Corliss C. H. and Bozman W. R. (1962) Experimental transition probabilities for spectral lines of 70 elements derived from the NBS tables of spectral line intensities. The wavelengths,
11 12
13
14
15
16
17
18
energy levels transition probability and oscillator strength of 25000 lines between 200 and 900 nm for 112 spectra of 70 elements, NBS monograph 53, Washington DC. UnsoÈld A. (1955) Physik der SternatmosphaÈren, Springer, Berlin. Junkes J. and Salpeter E. W. (1961) Linienbreiten in der photographischen Spektralphotometrie II. Das optische Linienpro®l, Ric Spettrosc 2: 255±483. Bruce C. F. and Hannaford P. (1971) On the width of atomic resonance lines from hollow cathode lamps, Spectrochim Acta, Part B 26: 207±235. Broekaert J. A. C., Leis F. and Laqua K. (1979) Application of an inductively coupled plasma to the emission spectroscopic determination of rare earths in mineralogical samples, Spectrochim Acta, Part B 34: 73±84. Lochte-Holtgreven W. (ed) (1968) Plasma Diagnostics, North-Holland Publ. Company, Amsterdam. Elwert G. (1952) Verallgemeinerte Ionisationsformel eines Plasmas, Z Naturforsch 7a: 703±711. Mermet J. M., Spectroscopic diagnostics: basic concepts, in Boumans P. W. J. M. (ed) (1987) Inductively coupled plasma emission spectroscopy, Part II, WileyInterscience, New York, 353±386. Diermeier R. and Krempl H. (1967) Thermische Anregungsfunktionen und Normtemperaturen von Atomund Ionenlinien in Zweikomponentenplasmen, Z Phys 200: 239±248.
Literature 19 Herzberg G. (1950) Molecular spectra
20
21
22
23
24
25
26
27
28
29
30
and molecular structure, 2nd edition, Van Nostrand Reinhold Ltd., New York. Pearce R. W. B. and Gaydon A. G. (1953) The identi®cation of molecular spectra, Chapman and Hall Ltd., London. Broekaert J. A. C. (1977) Determination of rotational temperatures in a transitional type hollow cathode, Bull Soc Chim Belges 86: 895±906. Raeymaekers B., Broekaert J. A. C. and Leis F. (1988) Radially resolved rotational temperatures in nitrogenargon, oxygen-argon, air-argon and argon ICPs, Spectrochim Acta, Part B 43: 941±949. Ishii I. and Montaser A. (1991) Radial excitation temperatures in argon±oxygen and argon±sair ICP, Spectrochim Acta, Part B 46: 1197± 1206. Ferreira N. P., Human H. G. C. and Butler L. R. P. (1980) Kinetic temperatures and electron densities in the plasma of a side view Grimm-type glow discharge, Spectrochim Acta, Part B 35: 287±95. Bockasten K. (1961) Transformation of observed radiances into radial distribution of a plasma, JOSA 51: 943±947. Hieftje G. M. (1992) Plasma diagnostic techniques for understanding and control, Spectrochim Acta, Part B 47: 3±25. È ber die Elwert G. (1952) U Ionisations- und Rekombinationsprozesse in einem Plasma und die Ionisationsformel der Sonnenkorona, Z Naturforsch 7a: 432±439. Broekaert J. A. C. (1987) Trends in optical spectrochemical trace analysis with plasma sources, Anal Chim Acta 196: 1±21. Borer M. and Hieftje G. M. (1991) Tandem sources for atomic emission spectrometry, Spectrochim Acta Rev 14: 463±486. Doerffel K. (1984) Statistik in der analytischen Chemie, Verlag Chemie, Weinheim.
31 Sachs L. (1969) Statistische Auswert-
ungsmethoden, Springer Verlag, Berlin. 32 Nalimov V. V. (1963) The application
33
34 35
36
37
38
39
40
41
42
of mathematical statistics to chemical analysis, Pergamon Press, Oxford. Miller J. C. and Miller J. N. (1989) Statistics for analytical chemistry, Ellis Horword Ltd. Publishers, Chichester. Wende M. C., unpublished work. Ingle Jr. J. D. and Crouch S. R. (1988) Spectrochemical analysis, Prentice-Hall International, Inc., Englewood Clis. Hieftje G. M. (1972) Signal-to-noise enhancement through instrumental techniques, Anal Chem 44: 81A±88A. Van Borm W. A. and Broekaert J. A. C. (1990) Noise characteristics in inductively coupled plasma optical emission spectrometry using slurry nebulization and direct powder introduction techniques, Anal Chem 62: 2527±2532. Winge R. K., Eckels D. E., DeKalb E. L. and Fassel V. A. (1988) Spatiotemporal characteristics of the inductively coupled plasma, J Anal At Spectrom 3: 849±855. Pereiro R., Min Wu., Broekaert J. A. C. and Hieftje G. M. (1994) Direct coupling of continuous hydride generation with microwave plasma torch atomic emission spectrometry for the determination of arsenic, antimony and tin, Spectrochim Acta, Part B 49: 59±73. Broekaert J. A. C., Brushwyler K. R., Monnig C. A. and Hieftje G. M. (1990) Fourier-transform-atomic emission spectrometry with a Grimmtype glow-discharge source, Spectrochim Acta, Part B 45: 769±778. Pollmann D., Pilger C., HergenroÈder R., Leis F., TschoÈpel P. and Broekaert J. A. C. (1994) Noise power spectra of inductively coupled plasma mass spectrometry using a cooled spray chamber, Spectrochim Acta, Part B 49: 683±690. Omenetto N. and Winefordner J. D. (1979) Atomic ¯uorescence spectrometry basic principles and applications, Prog Anal Spectrosc 2: 1±183.
313
314
Literature 43 Kaiser H. (1947) Die Berechnung der
44
45
46
47
48
49 50
51
52
53 54
55
56
57
58
Nachweisemp®ndlichkeit, Spectrochim Acta 3: 40±67. Kaiser H. and Specker H. (1956) Bewertung und Vergleich von Analysenverfahren, Z Anal Chem 149: 46±66. Hubaux A. and Vos G. (1970) Decision and detection limits for linear calibration curves, Anal Chem 42: 849±855. Behrends K. (1967) Folgen analytische Fehlerkurven eine Gauûverteilung? Fresenius' Z Anal Chem 235: 391±401. Himmelblau D. (1969) Process analysis by statistical methods, Wiley, New York. Grove E. L. (1971) Analytical Emission Spectrometry, Vol. 1, Parts I and II. Decker Inc., New York. Williams C. S. and Becklund O. A. (1972) Optics, Wiley, New York. Keliher P. N. and Wohlers C. C. (1976) Echelle grating spectrometers in analytical spectrometry, Anal Chem 48: 333A±340A. È ber SchwaÈrzKaiser H. (1948) U ungstransformationen, Spectrochim Acta 3: 159±190. Boumans P. W. J. M. (1971) O-line data processing in emission spectrography, XVI Coll Spectrosc Intern, Heidelberg, Preprints, Vol. II, 247±253. Siedentopf H. (1934), Z Phys 35: 454. Zworykin V. K. and Ramberg E. G. (1949) Photoelectricity and its application, Wiley, New York. Talmi Y. (1979) Multichannel image detectors, American Chemical Society Symposium Series 102. Sweedler J. V., Jalkian R. F. and Denton M. B. (1989) A linear charge coupled device detector system for spectroscopy, Appl Spectrosc 43: 953±961. Hieftje G. M. and Brushwyler K. R. (1991) Glow-discharge atomic emission spectrometry with a spectrally segmented photodiode-array spectrometer, Appl Spectrosc 45: 682±691. Broekaert J. A. C., Brushwyler K. R. and Hieftje G. M., unpublished work.
59 Danielson A., Lindblom P. and
60
61
62 63
64
65
66
67
68
69
70
SoÈdermann E. (1974) The IDES spectrometer, Chem Scr 6: 5±9. Sweedler J. V., Bilhorn R. B., Epperson P. M., Sims G. R. and Denton M. B. (1988) Highperformance charge transfer device detectors, Anal Chem 60: 282A±291A. Epperson P. M. and Denton M. B. (1989) Binning spectral images in a charge-coupled device, Anal Chem 61: 1513±1519. Bilhorn R. B. (1987) PhD Dissertation, University of Arizona. Bilhorn R. B., Epperson P. M., Sweedler J. V. and Denton M. B. (1987) Charge transfer device detectors for analytical optical spectroscopy ± operation and characteristics, Appl Spectrosc 41: 1114±1124. Furuta N., Brushwyler K. R. and Hieftje G. M. (1989) Flow-injection analysis utilizing a spectrally segmented photodiode array ICP emission spectrometer I. Microcolumn preconcentration for the determination of molybdenum, Spectrochim Acta, Part B 44 (1989) 349±358. Rosenkranz B., Breer C. B., Busher W., Bettmer J. and Cammann K. (1997) The plasma emission detector ± a suitable detector for speciation and sum parameter analysis, J Anal At Spectrom 12: 993±996. Treado P. J. and Morris M. D. (1989) A thousand points of light: the Hadamard transform in chemical analysis and instrumentation, Anal Chem 61: 723A±734A. Faires L. M. (1986) Fourier transforms for analytical atomic spectroscopy, Anal Chem 58: 1023A±1043A. Adams F., Gijbels R. and Van Grieken R. (eds) (1988) Inorganic mass spectrometry, Wiley, New York. Bacon J. R. and Ure A. (1984) Spark source mass spectrometry: recent developments and applications, Analyst 109: 1229±1254. Heumann K. G., Beer F. and Weiss H. (1983) Chlorid Spurenbestimmung
Literature
71
72
73
74
75
76
77
78
79
in Reinstchemikalien durch negative Thermionen-Massenspektrometrie, Mikrochim Acta I: 95±108. Brunnee C. (1987) The ideal mass analyzer: fact or ®ction, Int J Mass Spectrom Ion Processes 76: 125±237. Èdenauer F. G. Benninghoven A., Ru and Werner H. W. (1986) Secondary ion mass spectrometry, Wiley, New York. Denoyer E., Van Grieken R., Adams F. and Natusch D. F. S. (1982) Laser microprobe mass spectrometry I: basic principles and performance characteristics, Anal Chem 54: 26A± 41A. Mahoney P. P., Ray S. J. and Hieftje G. M. (1997) Time-of-¯ight mass spectrometry for elemental analysis, Appl Spectrosc 51: 16A±28A. Myers D. P., Li G., Yang P. and Hieftje G. M. (1994) An inductively coupled plasma time-of-¯ight mass spectrometer for elemental analysis. Part I: optimization and characterization, J Am Soc Mass Spectrom 5: 1008±1016. Pack B. W., Broekaert J. A. C., Guzowski J. P., Poehlmann J. and Hieftje G. M. (1998) Determination of halogenated hydrocarbons by helium microwave plasma torch timeof-¯ight mass spectrometry coupled to gas chromatography, Anal Chem 70: 3957±3963. Harrison W. W. and Hang W. (1996) Pulsed glow discharge time-of-¯ight mass spectrometry, J Anal At Spectrom 11: 835±840. Majidi V., Moser M., Lewis C., Hang W. and King F. L. (2000) Explicit speciation by microsecond pulsed glow discharge time-of-¯ight mass spectrometry: concurrent acquisition of structural, molecular and elemental information, J Anal At Spectrom 15: 19±25. Hang W., Yang P., Wang X., Yang C., Su Y. and Huang B. (1994) Microsecond pulsed glow discharge time-of-¯ight mass spectrometer, Rapid Commun Mass Spectrom 8: 590± 594.
80 Mahoney P. P., Ray S. J., Li G. and
81
82
83
84
85
86
87
Hieftje G. M. (1999) Preliminary investigation of electrothermal vaporization sample introduction for inductively coupled plasma time-of¯ight mass spectrometry, Anal Chem 71: 1378±1383. Turner P. J., Millis D. J., SchroÈder E., Lapitajs G., Jung G., Iacone L. A., Haydar D. A. and Montaser A. (1998) Instrumentation for low- and high-resolution ICPMS, in: Montaser A. (ed) Inductively coupled plasma mass spectrometry. Wiley-VCH, New York, 421±501. Stalford G. C., Kelley P. E., Syka J. E. P., Reynolds W. E. and Todd J. F. J. (1984) Recent improvements in and applications of advanced ion trap technology, Int J Mass Spectrom Ion Processes 60: 85. Koppenaal D. W., Barinaga C. J. and Smith M. R. (1994) Performance of an inductively coupled plasma source ion trap mass spectrometer, J Anal At Spectrom 9: 1053±1058. Eiden G. C., Barinaga C. J. and Koppenaal D. W. (1997) Bene®cial ion±molecule reactions in elemental mass spectrometry, Rapid Commun Mass Spectrom 11: 37±42. Wilkins C. L. and Gross M. L. (1981) Fourier transform mass spectrometry for analysis, Anal Chem 53: 1661A± 1676A. Milgram K. E., White F. M., Goodner K. L., Watson C. H., Clifford H., Koppenaal D. W., Barinaga C. J., Smith B. H., Winefordner J. D., Marshall A. G., Houk R. S. and Eyler J. R. (1997) High-resolution inductively coupled plasma Fourier transform ion cyclotron resonance mass spectrometer, Anal Chem 69: 3714±3721. Warren A. R., Allen L. A., Pang H. M., Houk R. S. and Janghorbani M. (1994) Simultaneous measurement of ion ratios by inductively coupled plasma mass spectrometry with a twin-quadrupole instrument, Appl Spectrosc 48: 1360±1366.
315
316
Literature 88 Gray A. L. and Date A. R. (1983) ICP
89
90
91
92
93
94
95
96
97
98
source mass spectrometry using continuous ¯ow ion extraction, Analyst 108: 1033±1050. Lichte F. E., Meier A. L. and Crock J. G. (1987) Determination of the rare-earth elements in geological materials by ICP-MS, Anal Chem 59: 1150±1157. Jeffreys H. and Jeffreys B. (1962) Methods of mathematical physics, statistics, Cambridge University Press, Cambridge, 488. Horlick G. and Shao Y. (1992) Inductively coupled plasma mass spectrometry for elemental analysis, in: Montaser A. and Golightly D. W. (eds) Inductively coupled plasmas in analytical atomic spectrometry, 2nd edition. VCH, New York. Vaughan M. A. and Horlick G. (1990) Ion trajectories through the input optics of an inductively coupled plasma mass spectrometer, Spectrochim Acta, Part B 45: 1301±1311. Merten D., Broekaert J. A. C. and LeMarchand A. (1999) Spectrum scanning in rapid sequential atomic emission spectrometry with the inductively coupled plasma, Spectrochim Acta, Part B 54: 1377±1382. Sneddon J. (ed) (1990) Sample introduction in atomic spectroscopy, Elsevier, Amsterdam. Broekaert J. A. C. and Boumans P. W. J. M. (1987) Sample introduction techniques in ICP-AES, in: Boumans P. W. J. M. (ed) Inductively coupled plasma emission spectroscopy, Vol. I. Wiley, New York, 296±357. Montaser A. and Golightly D. W. (eds) (1992) Inductively coupled plasmas in analytical atomic spectrometry, 2nd edition, VCH Publishers, Inc., Weinheim. Browner R. F. and Boorn A. W. (1984) Sample introduction, the Achilles heel of atomic spectrometry? Anal Chem 56: 786A±798A. Browner R. F. and Boorn A. W. (1984) Sample introduction techniques for atomic spectrometry, Anal Chem 56: 875A±888A.
99 Borer M. W. and Hieftje G. M.
100
101
102
103
104
105
106
106a
107
108
109
(1991) Tandem sources for atomic emission spectrometry, Spectrochim Acta Rev 14: 463±486. Aziz A., Broekaert J. A. C., Laqua K. and Leis F. (1984) A study of direct analysis of solid samples using spark ablation combined with excitation in an inductively coupled plasma, Spectrochim Acta, Part B 39: 1091±1103. È bbing J. and Niemax K. Ciocan A., U (1992) Analytical applications of the nmicrowave induced plasma used with laser ablation of solid samples, Spectrochim Acta, Part B 47: 611±617. Broekaert J. A. C. and ToÈlg G. (1987) Recent developments in atomic spectrometry methods for elemental trace determinations, Fresenius' J Anal Chem 326: 495±509. Broekaert J. A. C. (1987) Trends in optical spectrochemical trace analysis with plasma sources, Anal Chim Acta 196: 1±21. Scott R. H., Fassel V. A. and Kniseley R. N. (1974) Inductively coupled plasmas for optical emission spectroscopy, Anal Chem 46: 75±80. Schutyser P. and Janssens E. (1979) Evaluation of spray chambers for use in inductively coupled plasma atomic emission spectrometry, Spectrochim Acta, Part B 34: 443±449. Wu M. and Hieftje G. M. (1992) A new spray chamber for ICP spectrometry, Appl Spectrosc 46: 1912±1918. Schaldach G., Berger L. and Berndt H. (2000) Optimization of a spray chamber design using computer simulation and evolution strategies, 2000 Winter Conf on Plasma Spectrometry, Fort Lauderdale, Abstracts, p. 43. Gouy G. L. (1879) Photometric research on colored ¯ames, Ann Chim Phys 18: 5±101. Meinhard J. E. (1979) Pneumatic nebulizers, present and future, in: Barnes R. M. (ed): Applications of plasma emission spectroscopy. Heyden, London, 1±14. Wohlers C. C. (1977) Comparison of nebulizers for the inductively coupled plasma, ICP Inf Newsl 3: 37±51.
Literature 110 Liu H. and Montaser A. (1994)
111
112
113
114
115
116
117
118 119
120
Phase-Doppler diagnostic studies of primary and tertiary aerosols produced by a high-eciency nebulizer, Anal Chem 66: 3233±3242. Vanhaecke F., Van Holderbeke M., Moens L. and Dams R. (1996) Evaluation of a commercially available microconcentric nebulizer for inductively coupled plasma mass spectrometry, J Anal At Spectrom 11: 543±548. Kranz E. (1972) Untersuchungen uÈber die optimale Erzeugung und FoÈrderung von Aerosolen fuÈr spektrochemische Zwecken, Spectrochim Acta, Part B 27: 327. Valente S. E. and Schrenk W. G. (1970) The design and some emission characteristics of an economical dc arc plasmajet excitation source for solution analysis, Appl Spectrosc 24: 197. Fujishiro, Kubota M. and Ishida R. (1984) A study of designs of a cross ¯ow nebulizer for ICP atomic emission spectrometry, Spectrochim Acta, Part B 39: 617±620. Babington R. S. (1969) Method of atomizing liquid in a mono-dispersed spray, U.S. Patent 3,421,692. Ebdon L. and Cave M. R. (1982) A study of pneumatic nebulisation systems for ICP emission spectrometry, Analyst 107: 172±178. Raeymaekers B., Graule T., Broekaert J. A. C., Adams F. and TschoÈpel P. (1988) Characteristics of nebulized suspensions of refractory oxide powders used for the production of ceramics and their evaporation behaviour in an inductively coupled plasma, Spectrochim Acta, Part B 47: 923±940. Lathen C., unpublished work. Layman L. R. and Lichte F. E. (1982) Glass frit nebulizer for atomic spectrometry, Anal Chem 54: 638±642. Liu H., Clifford R. H., Dolan S. P. and Montaser A. (1996) Investigation of a high-eciency nebulizer and a thimble glass frit nebulizer for elemental analysis of biological materials by inductively coupled
121
122
123
124
125
126
127
128
129
plasma atomic emission spectrometry, Spectrochim Acta, Part B 51: 27±40. Kimberley E. L., Rice G. W. and Fassel V. A. (1984) Direct liquid injection sample introduction for ¯ow injection analysis and liquid chromatography with inductively coupled argon plasma spectrometric detection, Anal Chem 56: 289±292. McLean J. A., Minnich M. G., Lacone L. A., Liu H. and Montaser A. (1998) Nebulizer diagnostics: fundamental parameters, challenges, and techniques on the horizon, J Anal At Spectrom 13: 829±842. Wang L., May S. W., Browner R. F. and Pollock S. H. (1996) Low-¯ow interface for liquid chromatographyinductively coupled plasma mass spectrometry speciation using an oscillating capillary nebulizer, J Anal At Spectrom 11: 1137±1146. B'Hymer C., Sutton K. and Caruso J. A. (1998) Comparison of four nebulizer±spray chamber interfaces for the high-performance liquid chromatographic separation of arsenic compounds using inductively coupled plasma mass spectrometric detection, J Anal At Spectrom 13: 855±858. Sebastiani E., Ohls K. and Riemer G. (1973) Ergebnisse zur ZerstaÈubung dosierter LoÈsungsvolumina bei der AAS, Fresenius' Z Anal Chem 264: 105±109. Berndt H. and Slavin W. (1978) Automated trace analysis of small samples using the ``injection method'' of ¯ame atomic absorption spectrometry, At Absorpt Newsl 17: 109. Greenfield S. and Smith P. B. (1972) The determination of trace metals in microlitre samples by plasma torch emission, Anal Chim Acta 59: 341±348. Aziz A., Broekaert J. A. C. and Leis F. (1982) The optimization of an ICP injection technique and the application to the direct analysis of smallvolume serum samples, Spectrochim Acta, Part B 37: 381±389. Ruzicka J. and Hansen E. H. (1988) Flow injection analysis, Wiley, New York.
317
318
Literature 130 Haiber S. and Berndt H. (2000) A
131
132
133 134
135
136
137
137 138
139
novel temperature (360 C)/highpressure (30 Mpa) ¯ow system for online digestion applied to ICP spectrometry, Fresenius' J Anal Chem 368: 52±58. Pollmann D., Leis F., ToÈlg G., TschoÈpel P. and Broekaert J. A. C. (1994) Multielement trace determinations in Al2 O3 ceramic powders by inductively coupled plasma mass spectrometry with special reference to on-line preconcentration, Spectrochim Acta, Part B 49: 1251±1258. Lobinski R., Broekaert J. A. C., TschoÈpel P. and ToÈlg G. (1992) Inductively-coupled plasma atomic emission spectroscopic determination of trace impurities in ZrO2 -powder, Fresenius' J Anal Chem 342: 569±580. Vogt T., unpublished work. Hartenstein S. D., Ruzicka J. and Christian G. D. (1985) Sensitivity enhancements for ¯ow injection analysis ICP-AES using an on-line preconcentration ion-exchange column, Anal Chem 57: 21±25. Fang Z. (1993) Flow injection separation and preconcentration, VCH, Weinheim. Schramel P., Xu L., Knapp G. and Michaelis M. (1993) Multi-elemental analysis in biological samples by online preconcentration on 8-hydroxyquinoline±cellulose microcolumn coupled to simultaneous ICP-AES, Fresenius' J Anal Chem 345: 600±606. Nukuyama S. and Tanasawa Y. (1938, 1939) Experiments on the atomization of liquids in air stream, Trans Soc Mech Eng (Japan) 4: 86 and 5: 68. Allen T. (1981) Particle size measurement, Chapman & Hall, London. Van Borm W., Broekaert J. A. C., KlockenkaÈmper R., TschoÈpel P. and Adams F. C. (1991) Aerosol sizing and transport studies with slurry nebulization in inductively coupled plasma spectrometry, Spectrochim Acta, Part B 46: 1033±1049. Olsen S. D. and Strasheim A. (1983) Correlation of the analytical signal to the characterized nebulizer spray, Spectrochim Acta 38: 973±975.
140 Boorn A. W. and Browner R. (1982)
141
142
143
144
145
146
147
148
149
Eects of organic solvents in ICPAES, Anal Chem 54: 1402±1410. Kreuning G. and Maessen F. J. M. J. (1987) Eects of the solvent plasma load of various solvents on the excitation conditions in medium power ICP, Spectrochim Acta, Part B 42: 677±688. Doherty M. P. and Hieftje G. M. (1984) Jet-impact nebulization for sample introduction in ICP spectrometry, Appl Spectrosc 38: 405±411. Berndt H. (1988) High-pressure nebulization: a new way of sample introduction for atomic spectroscopy, Fresenius' J Anal Chem 331: 321±323. Posta J. and Berndt H. (1992) A high-performance ¯ow/hydraulic high pressure nebulization system (HPF/ HHPN) in ¯ame AAS for improved elemental trace determinations in highly concentrated salt solutions, Spectrochim Acta, Part B 47: 993±999. Berndt H. and Yanez J. (1996) Hightemperature hydraulic high-pressure nebulization: a recent nebulization principle for sample introduction, J Anal At Spectrom 11: 703±712. Posta J., Berndt H., Luo S. K. and Schaldach G. (1993) Highperformance ¯ow ¯ame atomic absorption spectrometry for automated on-line separation and determination of Cr(III)/Cr(VI) and preconcentration of Cr VI, Anal Chem 65: 2590±2595. Syti A. (1974) Developments in methods of sample injection and atomization in atomic spectrometry, CRC Crit Rev Anal Chem 4: 155± 225. Yang J., Conver T. S., Koropchak J. A. and Leighty D. A. (1996) Use of a multi-tube Na®on membrane dryer for desolvation with thermospray sample introduction to inductively coupled plasma-atomic emission spectrometry, Spectrochim Acta, Part B 51: 1491±1504. Wendt R. and Fassel V. E. (1965) Induction coupled plasma spectrometric excitation source, Anal Chem 37: 920±922.
Literature 150 Fassel V. A. and Bear B. R. (1986)
151
152
153
154
155
156
157
158
Ultrasonic nebulization of liquid samples for analytical ICP atomic spectroscopy, Spectrochim Acta, Part B 41: 1089±1113. Boumans P. W. J. M. and De Boer F. J. (1975) Studies of an ICP for OES. II Compromise conditions for simultaneous multielement analyses, Spectrochim Acta, Part B 30: 309±334. Barnett N. W., Chen L. S. and Kirkbright G. F. (1984) The rapid determination of arsenic by OES using a microwave induced plasma source and a miniature hydride generation device, Spectrochim Acta, Part B 39: 1141±1147. Manning D. C. (1971) A high sensitivity arsenic±selenium sampling system for atomic absorption spectroscopy, At Absorpt Newsl 10: 123. Thompson M., Pahlavanpour B., Walton S. J. and Kirkbright G. F. (1978) Simultaneous determination of trace concentrations of arsenic, antimony, bismuth, selenium and tellurium in aqueous solution by introduction of the gaseous hydrides into an ICP source for emission spectrometry, Analyst 103: 568±579. Broekaert J. A. C. and Leis F. (1980) Application of two dierent ICPhydride techniques to the determination of arsenic, Fresenius' Z Anal Chem 300: 22±27. Bulska E., TschoÈpel P., Broekaert J. A. C. and ToÈlg G. (1993) Dierent sample introduction systems for the simultaneous determination of As, Sb and Se by microwave-induced plasma atomic emission spectrometry, Anal Chim Acta 271: 171±181. Tao H. and Miyazaki A. (1991) Determination of germanium, arsenic, antimony, tin and mercury at trace levels by continuous hydride generation-helium microwave-induced plasma atomic emission spectrometry, Anal Sci 7: 55±59. Schickling C., Yang J. and Broekaert J. A. C. (1996) Optimization of electrochemical hydride generation coupled to microwaveinduced plasma atomic emission
159
160
161
162
163
164
165
166
spectrometry for the determination of arsenic and its use for the analysis of biological tissue samples, J Anal At Spectrom 11: 739±746. Stringer C. S. and Attrep Jr. M. (1979) Comparison of digestion methods for determination of organoarsenicals in wastewater, Anal Chem 51: 731±734. Welz B. and Melcher M. (1983) Investigations on atomisation mechanisms of volatile hydrideforming elements in a heated quartz cell, Analyst 108: 213±224. Vien S. H. and Fry R. C. (1988) Ultrasensitive simultaneous determination of arsenic, selenium, tin and antimony in aqueous solution by hydride generation gas chromatography with photoionisation detection, Anal Chem 60: 465±472. Zhang Li., Zhe-Ming N. and XiaoQuan S. (1989) In situ preconcentration of metallic hydrides in a graphite furnace coated with palladium, Spectrochim Acta, Part B 44: 339±346. Zhang Li., Zhe-Ming N. and XiaoQuan S. (1989) In situ concentration of metallic hydride in a graphite furnace coated with palladium ± determination of bismuth, germanium and tellurium, Spectrochim Acta, Part B 44: 751±758. Sturgeon R. E., Willie S. N., Sproule G. I., Robinson P. T. and Berman S. S., Sequestration of volatile element hydrides by platinum group elements for graphite furnace atomic absorption, Spectrochim Acta, Part B 44: 667±682. Haase O., Klare M., KrengelRothensee K. and Broekaert J. A. C. (1998) Evaluation of the determination of mercury at the trace and ultratrace levels in the presence of high concentrations of NaCl by ¯ow injection-cold vapour atomic absorption spectrometry using SnCl2 and NaBH4 , Analyst 123: 1219±1222. Nakahara T., Yamada S. and Wasa T. (1990) Continuous-¯ow determination of trace iodine by atmosphericpressure helium microwave-induced
319
320
Literature
167
168
169
170
171
172
173
174
plasma atomic emission spectrometry using generation of volatile iodine from iodide, Appl Spectrosc 44: 1673± 1678. Nakahara T., Mori T., Morimoto S. and Ishikawa H. (1995) Continuous¯ow determination of aqueous sulfur by atmospheric-pressure helium microwave-induced plasma atomic emission spectrometry with gas-phase sample introduction, Spectrochim Acta, Part B 50: 393±403. Rodriguez J., Pereiro R. and SanzMedel A. (1998) Glow discharge atomic emission spectrometry for the determination of chlorides and total organochlorine in water samples via on-line continuous generation of chlorine, J Anal At Spectrom 13: 911± 915. Sanz-Medel A., Valdes-Hevia y Temprano M. C., Bordel Garcia N. and Fernandez de la Campa M. R. (1995) Generation of cadmium atoms at room temperature using vesicles and its application to cadmium determination by cold vapor atomic spectrometry, Anal Chem 67: 2216± 2223. Preuû E. (1940) BeitraÈge zur spektralanalytischen Methodik. II Bestimmung von Zn, Cd, Hg, In, Tl, Ge, Sn, Pb, Sb und Bi durch fraktionnierte Destillation, Z Angew Mineral 3: 1. L'vov B. V. (1961) The analytical use of atomic absorption spectra, Spectrochim Acta 17: 761±770. Maûmann H. (1968) Vergleich von Atomabsorption und Atom¯uoreszenz in der GraphitkuÈvette, Spectrochim Acta 23: 215±226. L'vov B. V., Pelieva L. A. and Sharnopol'skii A. I. (1977) Decrease in the eect of the base during the atomic-absorption analysis of solutions in tube furnaces by vaporization of samples from a graphite substrate, Zh Prikl Spektrosk 27: 395±399. McNally J. and Holcombe J. A. (1987) Existence of microdroplets and dispersed atoms on the graphite
175
175a
176
177
178
179
180
181
182
surface in electrothermal atomizers, Anal Chem 59: 1105±1112. West T. S. and Williams X. K. (1969) Atomic absorption and ¯uorescence spectroscopy with a carbon ®lament atom reservoir. Part I. Construction and operation of atom reservoir, Anal Chim Acta 45: 27±41. Sychra V., Kolihova D., Vyskocilova O., Hlavac R. and È schel P. (1979) Electrothermal Pu atomization from metallic surfaces. Part 1. Design and performance of a tungsten-tube atomizer, Anal Chim Acta 105:263±270. Delves H. T. (1970) A micro-sampling method for the rapid determination of lead in blood by atomic-absorption spectrophotometry, Analyst 95: 431± 438. Berndt H. and Messerschmidt J. (1979) Eine Mikromethode der Flammen-Atomabsorptions- und Emissionsspektrometrie (Platinschlaufen-Methode) I. Die Grundlagen der Methode ± Anwendung auf die Bestimmung von Blei und Cadmium in Trinkwasser, Spectrochim Acta, Part B 34: 241±256. Brooks E. I. and Timmins K. J. (1985) Sample introduction device for use with a microwave-induced plasma, Analyst 110: 557±558. L'vov B. V. (1978) Electrothermal atomization ± the way toward absolute methods of atomic absorption analysis, Spectrochim Acta, Part B (1978) 153±193. Aziz A., Broekaert J. A. C. and Leis F. (1982) A contribution to the analysis of microamounts of biological samples using a combination of graphite furnace and microwave induced plasma atomic emission spectroscopy, Spectrochim Acta 37: 381±389. Miller-Ihli N. J. (1989) Automated ultrasonic mixing accessory for slurry sampling into a graphite furnace atomic absorption spectrometer, J Anal At Spectrom 4: 295±297. Grobenski Z., Lehmann R., Tamm R. and Welz B. (1982) Improvements
Literature
183
184
185
186
187
188
189
190
in graphite furnace atomic absorption microanalysis with solid sampling, Mikrochim Acta [Wien] I: 115±125. Volynsky A. B. (1998) Applications of graphite tubes modi®ed with highmelting carbides in electrothermal atomic absorption spectrometry I. General approach, Spectrochim Acta, Part B 53: 509±535. Volynsky A. B. (1998) Applications of graphite tubes modi®ed with highmelting carbides in electrothermal atomic absorption spectrometry II. Practical aspects, Spectrochim Acta, Part B 53: 1607±1645. Millard D. L., Shan H. C. and Kirkbright G. F. (1980) Optical emission spectrometry with an inductively coupled radiofrequency argon plasma source and sample introduction with a graphite rod electrothermal vaporization device. Part II. Matrix, inter-element and sample transport ects, Analyst 105: 502±508. Aziz A., Broekaert J. A. C. and Leis F. (1982) Analysis of microamounts of biological samples by evaporation in a graphite furnace and inductively coupled plasma atomic emission spectroscopy, Spectrochim Acta, Part B 37: 369±379. Broekaert J. A. C. and Leis F. (1985) An application of electrothermal evaporation using direct solids sampling coupled with microwave induced plasma optical emission spectroscopy to elemental determinations in biological matrices, Mikrochim Acta II: 261±272. Kantor T. (2000) Sample introduction with graphite furnace electrothermal vaporization into an inductively coupled plasma: eects of streaming conditions and gaseous phase additives, Spectrochim Acta, Part B 55: 431±448. Sommer D. and Ohls K. (1980) Direkte ProbeneinfuÈhrung in ein stabil brennendes ICP, Fresenius' Z Anal Chem 304: 97±103. Salin E. D. and Horlick G. (1979) Direct sample insertion device for ICP
191
192
193
194
195
196
197
198
emission spectrometry, Anal Chem 51: 2284±2286. Zaray Gy., Broekaert J. A. C. and Leis F. (1988) The use of direct sample insertion into a nitrogen± argon inductively coupled plasma for emission spectrometry I. Technique optimization and application to the analysis of aluminium oxide, Spectrochim Acta, Part B 43: 241±253. Karnassios V. and Horlick G. (1989) A computer-controlled direct sample insertion device for ICP-MS, Spectrochim Acta, Part B 44: 1345±1360. Blain L. and Salin E. D. (1992) Methodological solutions for the analysis of sediment samples by direct sample insertion ICP atomic emission, Spectrochim Acta, Part B 47: 205±217. Hauptkorn S. and Krivan V. (1994) Determination of silicon in boron nitride by slurry sampling electrothermal atomic absorption spectrometry, Spectrochim Acta, Part B 49: 221±228. GreÂgoire D. C., Maawali S. A. I. and Chakrabarti C. L. (1992) Use of Mg/ Pd chemical modi®ers for the determination of volatile elements by electrothermal vaporization ICP-MS: eect on mass transport eciency, Spectrochim Acta, Part B 47: 1123± 1133. Nickel H. and Broekaert J. A. C. (1999) Some topical applications of plasma atomic spectrochemical methods for the analysis of ceramic powders, Fresenius' J Anal Chem 363: 145±155. Lobinski R., Van Borm W., Broekaert J. A. C., TschoÈpel P. and ToÈlg G. (1992) Optimization of slurry nebulization inductively-coupled plasma atomic emission spectrometry for the analysis of ZrO2 -powder, Fresenius' J Anal Chem 342: 563±568. McKinnon P. W., Giess K. C. and Knight T. V. (1981) A clog-free nebulizer for use in inductively coupled plasma-atomic emission spectroscopy, in: Barnes R. M. (ed) Developments in atomic plasma spectrochemical analysis. Heyden, London, 287±301.
321
322
Literature 199 Barnes R. M. and Nikdel S. (1976)
200
201
202
203
204
205
206
Temperature and velocity pro®les and energy balances for an inductively coupled plasma discharge in nitrogen, J Appl Phys 47: 3929±3934. Hieftje G. M., Miller R. M., Pak Y. and Wittig E. P. (1987) Theoretical examination of solute particle vaporization in analytical atomic spectrometry, Anal Chem 59: 2861± 2872. Merten D., Heitland P. and Broekaert J. A. C. (1997) Modelling of the evaporation behaviour of particulate material for slurry nebulization inductively coupled plasma atomic emission spectrometry, Spectrochim Acta, Part B 52: 1905± 1922. Raeymaekers B., Van Espen P. and Adams F. (1984) The morphological characterization of particles by automated scanning electron microscopy, Mikrochim Acta II: 437±454. Broekaert J. A. C., Brandt R., Leis F., Pilger C., Pollmann D., TschoÈpel P. and ToÈlg G. (1994) Analysis of aluminium oxide and silicon carbide ceramic materials by inductively coupled plasma mass spectrometry, J Anal At Spectrom 9: 1063±1070. Docekal B. and Krivan V. (1992) Direct determination of impurities in powered silicon carbide by electrothermal atomic absorption spectrometry using the slurry sampling technique, J Anal At Spectrom 7: 521±528. È rerKrivan V., Barth P. and Schnu Patschan C. (1988) An electrothermal atomic absorption spectrometer using semiconductor diode lasers and a tungsten coil atomizer: design and ®rst applications, Anal Chem 70: 3625±3632. Lucic M. and Krivan V. (1999) Analysis of aluminium-based ceramic powders by electrothermal vaporization inductively coupled plasma atomic emission spectrometry using a tungsten coil and slurry sampling, Fresenius' J Anal Chem 363: 64±72.
207 Min H. and Xi-En S. (1989) Analysis
208
209
210
211
212
213
215
of ultra®ne ZrO2 for minor and trace elements by a slurry sampling inductively coupled plasma atomic emission spectrometric method, Spectrochim Acta, Part B 44: 957± 964. Obenauf R. H., Bostwick R., McCann M., McCormack J. D., Merriman R. and Salem D. (1994) Spex handbook of sample preparation and handling, Spex, Metuchen. Johnes J. L., Dahlquist R. L. and Hoyt H. E. (1971) A spectroscopic source with improved analytical properties and remote sampling capacity, Appl Spectrosc 25: 629±635. El Alfy S., Laqua K. and Maûmann H. (1973) Spektrochemische Analysen mit einer Glimmentladungslampe III. Entwicklung und Beschreibung einer Universalmethode zur Bestimmung von Hauptbestandteilen in elektrisch nichtleitenden pulverfoÈrmigen Substanzen, Fresenius' Z Anal Chem 263: 1±14. Human H. G. C., Scott R. H., Oakes A. R. and West C. D. (1976) The use of a spark as a sampling±nebulising device for solid samples in atomic absorption spectrometry, atomic ¯uorescence and ICP emission spectroscopy, Analyst 101: 265±271. Engel U., Kehden A., Voges E. and Broekaert J. A. C. (1999) Direct solid atomic emission spectrometric analysis of metal samples by an argon microwave plasma torch coupled to spark ablation, Spectrochim Acta, Part B 54: 1279±1289. Lemarchand A., Labarraque G., Masson P. and Broekaert J. A. C. (1987) Analysis of ferrous alloys by spark ablation coupled to inductively coupled plasma atomic emission spectrometry, J Anal At Spectrom 2: 481±484. Broekaert J. A. C., Leis F. and Laqua K. (1985) A contribution to the direct analysis of solid samples by spark erosion combined to ICP-OES, in: Sansoni B. (ed) Instrumentelle Multielementanalyse. Verlag Chemie, Weinheim.
Literature 216 Raeymaekers B., Van Espen P.,
217
218
219
220
221 222 223
224
225
226
227
Adams F. and Broekaert J. A. C. (1988) A characterization of sparkproduced aerosols by automated electron probe micro-analysis, Appl Spectrosc 42: 142±150. Kehden A., Flock J., Vogel W. and Broekaert J. A. C. (2000) Direct solids atomic spectrometric analysis of metal samples by laser induced argon spark ablation coupled to ICPOES, Appl Spectrosc, in press. Golloch A. and Siegmund D. (1997) Gliding spark spectroscopy ± rapid survey analysis of ¯ame retardants and other additives in polymers, Fresenius' J Anal Chem 358: 804±811. Barnes R. M. and Malmstadt H. V. (1974) Liquid-layer-on-solid-sample spark technique for emission spectrochemical analysis, Anal Chem 46: 66±72. Bings N., Alexi H. and Broekaert J. A. C., UniversitaÈt Dortmund, unpublished work. Omenetto N. (1979) Analytical laser spectroscopy, Wiley, New York. Moenke-Blankenburg L. (1989) Laser micro analysis, Wiley, New York. Laqua K., Analytical spectroscopy using laser atomizers. (1985) in: Martellucci S. and Chester A. N. (eds) Analytical laser spectroscopy, Plenum Publishing Corporation, New York, 159±182. Leis F., Sdorra W., Ko J. B. and Niemax K. (1989) Basic investigations for laser microanalysis: OES of laserproduced sample plumes, Mikrochim Acta [ Wien] II: 185±199. Gagean M. and Mermet J. M. (1998) Study of laser ablation of brass materials using inductively coupled plasma atomic emission spectrometric detection, Spectrochim Acta, Part B 53: 581±591. Sdorra W., Quentmeier A., Niemax K. (1989) Basic investigations for laser microanalysis. II. Laser-induced ¯uorescence in laser-produced sample plumes, Mikrochim Acta [ Wien] II: 201±218. Uebbing J., Brust J., Sdorra W.,
228
229
230
231
232
233
234
235
236
237
Leis F. and Niemax K. (1991) Reheating of a laser-produced plasma by a second laser, Appl Spectrosc 45: 1419±1423. Hemmerlin M. and Mermet J. M. (1996) Determination of elements in polymers by laser ablation ICP AES: eect of the laser beam wavelength, energy and masking on the ablation threshold and eciency, Spectrochim Acta, Part B 51: 579±589. Margetic V., Pakulev A., Stockhaus, Bolshov M., Niemax K. and HergenroÈder R. (2000) A comparison of nanosecond and femtosecond laser-induced plasma spectroscopy of brass samples, Spectrochim Acta, Part B 55: 1771± 1785. Kaminsky M. (1965) Atomic and ionic impact phenomena on metal surfaces, Springer Verlag, Berlin. Penning F. M. (1957) Electrical discharges in gases, Philips Technical Library, Eindhoven, p. 41. Boumans P. W. J. M. (1973) Sputtering in a glow discharge for spectrochemical analysis, Anal Chem 44: 1219±1228. Broekaert J. A. C. (1987) State-ofthe-art of glow discharge lamp spectrometry, J Anal At Spectrom 2: 537±542. Marcus R. K. (1996) Radiofrequency powered glow discharges: opportunities and challenges, J Anal At Spectrom 11: 821±828. Anderson G. S., Meyer W. N. and Wehner G. K. (1962) Sputtering of dielectrics by high-frequency ®elds, J Appl Phys 33: 2991±2992. Christopher S. J., Ye Y. and Marcus R. K. (1997) Ion kinetic energy distributions and their relationship to fundamental plasma parameters in a radio frequency glow discharge, Spectrochim Acta, Part B 52: 1627±1644. Laqua K. (1991) Spektralanalyse mit Glimmentladungen±GegenwaÈrtiger Stand und Zukunftsaussichten, in: Borsolorf R; Fresenius W., GuÈnzler H., Huber W., LuÈdenwald L., ToÈlg G. and Wisser H. (eds) Analytiker
323
324
Literature
238
239
240
241
242
243
244
245
246
Taschenbuch, Band 10, Springer, Berlin, 297±349. Broekaert J. A. C., Bricker T., Brushwyler K. R. and Hieftje G. M. (1992) Investigations of a jet-assisted glow discharge lamp for optical emission spectrometry, Spectrochim Acta, Part B 47: 131±142. Bogaerts A. and Gijbels R. (1997) Calculation of crater pro®les on a ¯at cathode in a direct current glow discharge, Spectrochim Acta, Part B 52: 765±777. Dogan M., Laqua K. and Maûmann H. (1972) Spectrochemische Analysen mit einer Glimmentladungslampe als Lichtquelle- I Elektrische Eigenschaften, Probenabbau und spektraler Charakter, Spectrochim Acta, Part B 26: 631±649. Marcus R. (ed) (1993) Glow discharge spectrometries, Plenum Press, New York. Van Straaten M., Vertes A. and Gijbels R. (1991) Sample erosion studies and modeling in a glow discharge ionization cell, Spectrochim Acta, Part B 46: 283±280. Bogaerts A. and Gijbels R. (1998) Modeling of direct current glow discharges and comparison with experiments: how good is the agreement? J Anal At Spectrom 13: 945±953. Christopher S. J., Hartenstein M. L., Marcus R. K., Belkin M. and Caruso J. A. (1998) Characterization of helium/argon working gas systems in a radiofrequency glow discharge atomic emission source. Part I: Optical emission, sputtering and electrical characteristics, Spectrochim Acta, Part B 53: 1181±1196. Heintz M. J., Mifflin K., Broekaert J. A. C. and Hieftje G. M. (1995) Investigations of a magnetically enhanced Grimm-type glow discharge source, Appl Spectrosc 49: 241±246. Heintz M. J., Broekaert J. A. C. and Hieftje G. M. (1997) Analytical characterization of a planar magnetron radio frequency glow-discharge source, Spectrochim Acta, Part B 52: 579±591.
247 Walsh A. (1955) The application of
248
249
250
251
252
253
254
255
256
atomic absorption spectra to chemical analysis, Spectrochim Acta 7: 108±117. Welz B. and Sperling M. (1999) Atomic absorption spectrometry, WileyVCH, Weinheim. Price W. J. (1979) Spectrochemical analysis by atomic absorption, Heyden, London. Ebdon L. (1982) An introduction to atomic absorption spectroscopy, Heyden, London. HergenroÈder R. and Niemax K. (1988) Laser atomic absorption spectroscopy applying semiconductor lasers, Spectrochim Acta, Part B 43: 1443±1449. Ng K. C., Ali A. H., Barber T. E. and Winefordner J. D. (1988) Multiple mode semiconductor diode laser as a spectral line source for graphite furnace atomic absorption spectroscopy, Anal Chem 62: 1893±1895. Koirthyohann R. S. and Pickett E. E. (1965) Background correction in long path atomic absorption spectrometry, Anal Chem 37: 601±603. Moulton G. P., O'Haver T. C. and Harnly J. M. (1990) Signal to noise ratios for continuous source atomic absorption spectrometry using a linear photodiode array to monitor sub-nanometre wavelength intervals, J Anal At Spectrom 5: 145±150. È tz M., BeckerHeitmann U., Schu Roû H. and Florek S. (1996) Measurements on the Zeemansplitting of analytical lines by means of a continuum source graphite furnace atomic absorption spectrometer with a linear charge coupled device array, Spectrochim Acta, Part B 51: 1095±1105. È tz M., BeckerHeitmann U., Schu Roû H. and Florek S. (1998) Spektroskopische Untersuchungen in der Graphitrohr-AAS mit Hilfe eines hochau¯oÈsenden KontinuumstrahlerSpektrometers, in: Vogt C., Wennrich R. and Werner G. (eds) CANAS 97, Colloquium Analytische Atomspektroskopie, UniversitaÈt Leipzig und Umweltforschungszentrum LeipzigHalle GmbH, Leipzig.
Literature Èrer257 Niemax K., Zybin A., Schnu
258
259
260
261
262
263
264
265
266
267
Patschan C. and Groll H. (1996) Semiconductor diode lasers in atomic spectrometry, Anal Chem 68: 351A± 356A. È rerNiemax K., Groll H. and Schnu Patschan C. (1993) Element analysis by diode laser spectroscopy, Spectrochim Acta Rev 15: 349±377. È rer-Patschan C. and Zybin A., Schu Niemax K. (1992) Simultaneous multi-element analysis in a commercial graphite furnace by diode laser induced ¯uorescence, Spectrochim Acta, Part B 47: 1519±1524. Silver J. A. (1992) Frequencymodulation spectroscopy for trace species detection: theory and comparison among experimental methods, Appl Opt 31: 707±717. Èrer-Patschan C., Groll H., Schnu Kuritsyn Yu. and Niemax K. (1994) Wavelength modulation diode laser atomic absorption spectrometry in analytical ¯ames, Spectrochim Acta, Part B 49: 1463±1472. Èwer Jakubowski N., Jepkens B., Stu D. and Berndt H. (1994) Speciation analysis of chromium by inductively coupled plasma mass spectrometry with hydraulic high pressure nebulization, J Anal At Spectrom 9: 193±198. Barshick C. M., Shaw R. W., Young J. P. and Ramsey J. M. (1994) Istopic analysis of uranium using glow discharge optogalvanic spectroscopy and diode lasers, Anal Chem 66: 4154±4158. Amos M. D. and Willis J. B. (1966) Use of high-temperature pre-mixed ¯ames in atomic absorption spectroscopy, Spectrochim Acta 22: 1325± 1343. Hermann R. and Alkemade C. T. J. (1983) Chemical analysis by ¯ame photometry, 2nd edition, Interscience Publishers, New York. Alkemade C. Th. J., Hollander Tj., Snelleman W. and Zeegers P. J. Th. (1982) Metal vapours in ¯ames, Pergamon Press, Oxford. Dean J. A. and Rains T. C. (eds) (1969) Flame emission and atomi
268
269
270
271
272
273
274
275
276
277
absorption spectrometry, Vol. 1. Marcel Dekker, New York. Rubeska I. and Musil J. (1979) Interferences in ¯ame spectroscopy: their elimination and control, Prog Anal Spectrosc 2: 309±353. Greenway G. M., Nelms S. M., Skhosana I. and Dolman S. J. L. (1996) A comparison of preconcentration reagents for ¯ow injection analysis ¯ame atomic spectrometry, in: McIntosk S. and Brindle I. (eds) Special Issue: Flow injection analysis, state of the art applied to atomic spectrometry, Spectrochim Acta, Part B 51: 1909±1915. Fang Z. I. and Welz B. (1989) Optimisation of experimental parameters for ¯ow injection ¯ame atomic absorption spectrometry, J Anal At Spectrom 4: 83±89. Slavin W. (1986) Flames, furnaces, plasmas, how do we choose? Anal Chem 58: 589A±597A. È tsch Frech W., Baxter D. C. and Hu B. (1986) Spatially isothermal graphite furnace for atomic absorption spectrometry using side-heated cuvettes with integrated contacts, Anal Chem 58: 1973±1977. Berndt H. and Schaldach G. (1988) Simple low-cost tungsten-coil atomiser for electrothermal atomic absorption spectrometry, J Anal At Spectrom 3: 709±713. L'vov B. V., Bayonov P. A. and Ryabchuk (1981) A macrokinetic theory of sample vaporization in electrothermal atomic absorption spectrometry, Spectrochim Acta, Part B 36: 397±425. Paveri-Fontana S. L., Tessari G. and Torsi G. C. (1974) Time-resolved distribution of atoms in ¯ameless atomic absorption spectrometry. A theoretical calculation, Anal Chem 46: 1032±1038. Slavin W., Manning D. C. and Carnrick G. R. (1981) The stabilized temperature platform furnace, At Spectrosc 2: 137±145. van den Broek W. M. G. T. and de Galan L. (1977) Supply and removal of sample vapor in graphite thermal atomizers, Anal Chem 49: 2176±2186.
325
326
Literature 278 Schlemmer G. and Welz B. (1986)
279
280
281
282
283
284
285
286
Palladium and magnesium nitrates, a more universal modi®er for graphite furnace atomic absorption spectrometry, Spectrochim Acta, Part B 41: 1157± 1166. Welz B., Curtius A. J., Schlemmer G., Ortner H. M. and Birzer W. (1986) Scanning electron microscopy studies on surfaces from electrothermal atomic absorption spectrometry-III. The lanthanum modi®er and the determination of phosphorus, Spectrochim Acta, Part B 41: 1175±1201. Krivan V. (1992) Application of radiotracers to methodological studies in atomic absorption spectrometry, J Anal At Spectrom 7: 155±164. Kantor T., Bezur L., Pungor E. and Winefordner J. D. (1983) Volatilization studies of magnesium compounds by a graphite furnace and ¯ame combined atomic absorption method. The use of a halogenating atmosphere, Spectrochim Acta, Part B 38: 581±607. L'vov B. V. (1990) Recent advances in absolute analysis by graphite furnace atomic absorption spectrometry, Spectrochim Acta, Part B 45: 633±655. Bulska E. and Ortner H. M. (2000) Intercalation compounds of graphite in atomic absorption spectrometry, Spectrochim Acta, Part B 55: 491±499. De Jonghe W. G. C. and Adams F. C. (1979) The determination of organic and inorganic lead compounds in urban air by atomic absorption spectrometry with electrothermal atomization, Anal Chim Acta 108: 21±30. Chakraborti D., De Jonghe W. R. A., Van Mol W. E., Adams F. C. and Van Cleuvenbergen R. J. A. (1984) Determination of ionic alkyllead compounds in water by gas chromatography/atomic absorption spectrometry, Anal Chem 56: 2692± 2697. Yamamoto M., Yasuda M. and Yamamoto Y. (1985) Hydride generation AAS coupled with ¯ow injection analysis, Anal Chem 57: 1382±1385.
È hn 287 Jackwerth E., Willmer P. G., Ho
288
289
290
291
292
293
294
R. and Berndt H. (1979) A simple accessory for the determination of mercury and the hydride-forming elements (As, Bi, Sb, Se and Te) using ¯ameless atomic absorption spectroscopy, At Absorpt Newsl 18: 66±68. Kaiser G., GoÈtz D., Schoch P. and ToÈlg G. (1975) Emissionsspektrometrische Elementbestimmung im Nano- und Picogramm-Bereich nach Ver¯uÈchtigung der Elemente in mit Mikrowellen induzierte Gasplasmen I. Extrem nachweisstarke Quecksilberbestimmung in waÈssrigen LoÈsungen, Luft, organischen und anorganischen Matrices, Talanta 22: 889±899. Piwonka J., Kaiser G. and ToÈlg G. (1985) Determination of selenium at ng/g- and pg/g-levels by hydride generation-AAS in biotic materials, Fresenius' Z Anal Chem 321: 225±234. Sturgeon R. E., Willie S. N. and Berman S. S. (1985) Hydride generation-atomic absorption determination of antimony in seawater with in-situ concentration in a graphite furnace, Anal Chem 57: 2311±2314. Welz B. and Melcher M. (1981) Mutual interactions of elements in the hydride technique in atomic absorption spectrometry, Anal Chim Acta 131: 17±25. Sanz-Medel A., del Rosario Fernandez de la Campa M., Gonzalez E. B. and FernandezSanchez M. L. (1999) Organised surfactant assemblies in analytical atomic spectrometry, Spectrochim Acta, Part B 54: 251±287. Berndt H. (1984) Probeneintragssystem mit Probenverbrennung oder Probenvorverdampfung fuÈr die direkte Feststoanalyse und fuÈr die LoÈsungsspektralanalyse, Spectrochim Acta, Part B 39: 1121±1128. Docekal B. and Krivan V. (1992) Direct determination of impurities in powdered silicon carbide by electrothermal atomic absorption spectrometry using the slurry sampling technique, J Anal At Spectrom 7: 521±528.
Literature 295 Hornung D. and Krivan V. (1999)
296
297
298
299
300
301
302
303
304
305
Solid sampling electrothermal atomic absorption spectrometry for analysis of high-purity tungsten trioxide and high-purity tungsten blue oxide, Spectrochim Acta, Part B 54: 1177± 1191. Gough D. S. (1976) Direct analysis of metals and alloys by atomic absorption spectrometry, Anal Chem 48: 1926±1930. Bernhard A. E. (1987) Atomic absorption spectrometry using sputtering atomization of solid samples, Spectroscopy 2: 118. Ohls K. (1987) Ein neuer Weg zum Einsatz der Atomabsorptionsspektrometrie mit Hilfe der direkten Atomisierung fester Proben, Fresenius' Z Anal Chem 327: 111±118. Kantor T., Polos L., Fodor P. and Pungor E. (1976) Atomic absorption spectrometry of laser-nebulized samples, Talanta 23: 585±586. Mizuike A. (1983) Enrichment techniques for inorganic trace analysis, Springer, Berlin. Zybin A. and Niemax K. (1997) GC analysis of chlorinated hydrocarbons in oil and chlorophenols in plant extracts applying element-selective diode laser plasma detection, Anal Chem 69: 755±757. È rer-Patschan Ch. Zybin A., Schnu and Niemax K. (1995) Wavelength modulation diode laser atomic absorption spectrometry in modulated low-pressure helium plasmas for element-selective detection in gas chromatography, J Anal At Spectrom, 10: 563±567. Koch J. and Niemax K. (1998) Characterization of an element selective GC-detector based on diode laser atomic absorption spectrometry, Spectrochim Acta, Part B 53: 71±79. Koch J., Miclea M. and Niemax K. (1999) Analysis of chlorine in polymers by laser sampling and diode-laser atomic absorption spectrometry, Spectrochim Acta, Part B 54: 1723±1735. Berndt H. and Sopczak D. (1987) Fehlererkennung und
306
307
308
309
310
311
312
313
314
Programmoptimierung in der ET-AAS durch zeitaufgeloÈste Signale. Direkte Elementspurenbestimmung in Harn: Cd, Co, Cr, Ni, Tl and Pb, Fresenius' Z Anal Chem 329: 18±26. De Loos-Vollebregt M. T. C. and De Galan L. (1985) Zeeman atomic absorption spectrometry, Prog Anal At Spectrom 8: 47±81. Broekaert J. A. C. (1982) Zeeman atomic absorption instrumentation, Spectrochim Acta, Part B 37: 65±69. Smith S. B. Jr. and Hieftje G. M. (1983) A new background-correction method for atomic absorption spectrometry, Appl Spectrosc 37: 419± 414. Yamamoto M., Murayama S., Ito M. and Yasuda M. (1980) Theoretical basis for multielement analysis by coherent forward scattering atomic spectroscopy, Spectrochim Acta, Part B 35: 43±50. Wirz P., Debus H., Hanle W. and Scharmann A. (1982) Synchronous multielement detection by forward scattering in a transverse magnetic ®eld (SYNFO), Spectrochim Acta, Part B 37: 1013±1020. Kingston H. M. and Jassie L. B. (1988) Introduction to microwave sample preparation: theory and practice, American Chemical Society, Washington DC. Kohl F., Jakubowski N., Brandt R., Pilger C. and Broekaert J. A. C. (1997) New strategies for trace analyses of ZrO2 , SiC and Al2 O3 ceramic powders, Fresenius' J Anal Chem 359: 317±325. TschoÈpel P. and ToÈlg G. (1982) Comments on the accuracy of analytical results in nanogram and picogram trace analysis of the elements, J Trace Microprobe Techn 1: 1±77. Berndt H. and Jackwerth E. (1979) Mechanisierte Mikromethode (``Injektionsmethode'') der FlammenSpektrometrie (AtomabsorptionAtomemission) fuÈr die Bestimmung der Serumelektrolyte und der Spurenelemente (Fe, Cu, Zn) ± Teil 1, J Clin Chem Clin Biochem 19: 71±76.
327
328
Literature 315 Alt F. and Massmann (1978)
316
317
318
319
320
321
322
323
324
Bestimmung von Blei in Blut mittels Atomabsorptionsspektrometrie, Spectrochim Acta, Part B 33: 337±342. Konvar U. K., Littlejohn D. and Halls D. J. (1990) Hot-injection procedures for the rapid analysis of biological samples by electrothermal atomic-absorption spectrometry, Talanta 37: 555±559. Burba P. and Willmer P. G. (1987) Multielement-preconcentration for atomic spectroscopy by sorption of dithiocarbamate-metal complexes (e.g. HMDC) on cellulose collectors, Fresenius' Z Anal Chem 329: 539±545. Jackwerth E., Mizuike A., Zolotov Y. A., Berndt H., HoÈhn R. and Kuzmin N. M. (1979) Separation and preconcentration of trace substances I. Preconcentration for ionorganic trace analysis, Pure Appl Chem 51: 1195±1211. Alt F., Jerono U., Messerschmidt J. and ToÈlg G. (1988) The determination of platinum in biotic and environmental materials, 1. mg/kg±g/kgrange, Mikrochim Acta 3: 299±304. Subramanian K. S. (1988) Determination of chromium(III) and chromium(VI) by ammonium pyrrolidinecarbodithioate-methyl isobutyl ketone furnace atomic absorption spectrometry, Anal Chem 60: 11±15. Dirkx W. M. R., Lobinski R. and Adams F. C. (1993) A comparative study of gas chromatography with atomic absorption and atomic emission detection for the speciation analysis of organotin, Anal Sci 9: 273±278. Oda C. E. and Ingle J. D. (1981) Speciation of mercury by cold vapor atomic absorption spectrometry with selective reduction, Anal Chem 53: 2305±2309. Schickling C. and Broekaert J. A. C. (1995) Determination of mercury species in gas condensates by on-line coupled high-performance liquid chromatography and cold-vapor atomic absorption spectrometry, Appl Organomet Chem 9: 29±36. Schermer S., Juricka L; Paumard J; Beiurohr E., Moitysik F. M. and
325
326
327
328
329
330 331
332
333
334
335
336
Broekaert J. A. C. Optimization of electrochemical hydride generation in a miniaturized ¯ow cell coupled to microwave-induced plasma, atomic emission spectrometry for the determination of selenium, Fresenius Journal of Analytical Chemistry, in press. Butcher D. J., Zybin A., Bolshov M. A. and Niemax K. (1999) Speciation of methylcyclopentadienyl manganese tricarbonyl by high performance liquid chromatography± diode laser atomic absorption spectrometry, Anal Chem 71: 5379±5385. Gaspar A. and Berndt H. (2000) Beam injection ¯ame furnace atomic absorption spectrometry: a new ¯ame method, Anal Chem 72: 240±246. Gerlach W. and Schweitzer E. (1930) Die chemische Emissionsspektralanalyse, Vol. 1. L. Voss, Leizig. Idzikowski A., Gadek S. and Mleczko W. (1973) Atlas of the arc spectra for 70 elements for the grating spectrograph, Monograph 3, Technical University of Wroclaw (Poland). Winge R. K., Fassel V., Peterson V. J. and Floyd M. A. (1985) ICP-AES ± An atlas of spectral information, Elsevier, Amsterdam. Harrison G. R. (1991) Wavelength tables, The MIT Press, Cambridge, MA. Saidel A. N., Prokofiev V. K. and Raiski S. M. (1955) Spektraltabellen, V E B Verlag Technik, Berlin. Gatterer A. and Junkes J. (1945, 1947, 1959) Atlas der Restlinien, 3 volumes, Specola Vaticana, Citta del Vaticano. Salpeter E. W. (1973) Spektren in der Glimmentladung von 150 bis 400 nm, 1st to 5th volume. Specola Vaticana, Citta del Vaticano. Boumans P. W. J. M. (1984) Line coincidence tables for ICP-OES, Volumes 1 and 2, Pergamon Press, Oxford. Broekaert J. A. C. (1983) Evaluation of a ¯exible ICP-spectrometer using a SIT as radiation detector, Abstract 23 CSI, Amsterdam, Spectrochim Acta, Part B 38: Supplement,44. Scheibe G. (1933) Chemische Spektralanalyse, in: BoÈttger W., Physikalische Methoden der analytischen Chemie, 1: 27.
Literature 337 Lomakin B. A. (1930) Quantitative
338
339
340
341
342
343
344
345
346
347
Spektralbestimmung von Wismut in Kupfer, Z Anorg Allg Chem 187: 75±90. Laqua K. Optische Emissionsspektrometrie (1980) in: Ullmann's EncyklopaÈdie der technischen Chemie, Vol. 5. Verlag Chemie, Weinheim, 441±500. Laqua K., Hagenah W. D. and WaÈchter H. (1967) Spektrochemische Spurenanalyse mit spektralaufgeloÈsten Analysenlinien, Fresenius' Z Anal Chem 221: 142±174. Aziz A., Broekaert J. A. C., Leis F. (1981) The optimization of an ICPinjection technique and the application to the direct analysis of small-volume serum samples, Spectrochim Acta, Part B 36: 251±260. Winge R. K., DeKalb E. L. and Fassel V. A. (1985) Comparative complexity of emission spectra from ICP, DC arc and spark excitation sources, Appl Spectrosc 39: 673±676. Merten D., Broekaert J. A. C. and LeMarchand A. (1997) Application of a rapid sequential inductively coupled plasma optical emission spectrometric method for the analysis with linerich emission spectra by dierent means of sample introduction, J Anal At Spectrom 12: 1387±1391. Broekaert J. A. C., Wopenka B. and Puxbaum H. (1982) Inductively coupled plasma optical emission spectrometry for the analysis of aerosol samples collected by cascade impactors, Anal Chem 54: 2174±2179. Becker-Ross H. and Florek S. V. (1997) Echelle spectrometers and charge-coupled devices, Spectrochim Acta, Part B 52: 1367±1375. Barnard T. W., Crockett H., Ivaldi J. C. and Lundberg P. L. (1993) Design and evaluation of an Echelle grating optical system for ICP-OES, Anal Chem 65: 1225±1230. Butler-Sobel C. (1982) Use of an extended torch with an ICP for the determination of nitrogen in aqueous solutions, Appl Spectrosc 36: 691±693. Broekaert J. A. C. and Zeeman P. B. (1984) Some experiments with an extended torch for the determination of nitrogen in aqueous solution with
348
349
350
351 352
353
354
355
356
357
358
the aid of ICP AES, Spectrochim Acta, Part B 39: 851±853. Heitland P. (2000) Verwendung von Spektrallinien in der ICP-OES im WellenlaÈngenbereich 125±190 nm, GIT Labor-Fachzeitschrift 847±850. Christian G. D. and Feldmann F. J. (1971) A comparison study of detection limits using ¯ame-emission spectroscopy with the nitrous oxide± acetylene ¯ame and atomic-absorption spectroscopy, Appl Spectrosc 25: 660± 663. Rosen B. (1970) International tables of selected constants 17, Pergamon Press, Oxford. Addink N. W. H. (1971) DC arc analysis, McMillan, London. Riemann M. Z. (1966) Der Kaskadebogen als Anregungsquelle fuÈr die Spektralanalyse von LoÈsungen, Z Anal Chem 215: 407±424. Todorovic M., Vukanovic V. and Georgijevic V. (1968) Optimal conditions for the use of a d.c. arc in an inhomogeneous magnetic ®eld, Spectrochim Acta, Part B 24: 571±577. Broekaert J. A. C. (1973) Calculation of detection limits in spectrographic analysis, Bull Soc Chim Belg 82: 561± 567. Radwan Z., Strzyzewska B. and Minczewski J. Spectrographic determination of trace amounts of rare earths I. Analysis of lanthanum and yttrium by the power sifting method. (1963), Appl Spectrosc 17: 1±5. Radwan Z., Strzyzewska B. and Minczewski J. (1963) Spectrographic determination of trace amounts of rare earths. II. Analysis of lanthanum, yttrium, europium, praesodymium neodymium and samarium by the carrier distillation method, Appl Spectrosc 17: 60±62. Radwan Z., Strzyzewska B. and Minczewski J. (1966) Spectrographic determination of trace amounts of rare earths. III. Determination of all rare earth elements by the carrier distillation method, Appl Spectrosc 20: 236±240. Avni R. (1978) in Grove E. L. (ed): Applied atomic spectroscopy, Vol. 1. Plenum Press, New York.
329
330
Literature 359 Spitsberg A. T. and Otis J. S. (2000)
360
361
362
363
364
365
366
367
368
369
370
371
Diculties with direct-current/arc optical emission spectrometric analysis of molybdenum powder for chromium, iron and nickel, Appl Spectrosc 11: 1707±1711. Slickers K. (1992) Die automatische Atom-Emissions-Spektralanalyse, BruÈhlsche UniversitaÈtsdruÈckerei, Gieûen. Koch K. H. (1984) BeitraÈge aus der Stahlindustrie zur angewandten Atomspektroskopie, Spectrochim Acta, Part B 39: 1067±1079. Leis F., Broekaert J. A. C. and Laqua K. (1987) Design and properties of a microwave boosted glow discharge lamp, Spectrochim Acta, Part B 42: 1169±1176. Jakubowski N. and Stuewer D. (1988) Comparison of Ar and Ne as working gases in analytical glow discharge mass spectrometry, Fresenius' Z Anal Chem 331: 145±149. Tohyama K., Ono J., Onodera M. and Saeki M. (1978) Research and Development in Japan. Okochi Memorial Foundation, 31±35. Coleman D. M., Sainz M. A. and Butler H. T. (1980) High frequency excitation of spark-sampled metal vapor, Anal Chem 52: 746±753. Korolev V. V. and Vainshtein E. E. (1959) Plasma generator as excitation source in spectroscopic analysis, Zh Anal Khim 14: 658±662. Margoshes M. and Scribner B. (1959) The plasma jet as a spectroscopic source, Spectrochim Acta 15: 138±145. Kranz E. (1964) in: Emissionsspektrokopie, Akademie-Verlag, Berlin, p. 160. Reednick J. (1979) A unique approach to atomic spectroscopy high energy plasma excitation and high resolution spectrometry, Am Lab May: 127±133. Decker R. J. (1980) Some analytical characteristics of a three electrode DC argon plasma source for optical emission spectrometry, Spectrochim Acta, Part B 35: 19±35. Leis F., Broekaert J. A. C. and Waechter H. (1989) A three-
372
373
374 375
376
377
378
379
380
381
electrode direct current plasma as compared to an inductively coupled argon plasma, Fresenius' Z Anal Chem 333: 2±5. Johnson G. W., Taylor H. E. and Skogerboe R. K. (1979) Determination of trace elements in natural waters by the DC argon plasma multielement atomic emission spectrometer (DCP-MAES) Technique, Spectrochim Acta, Part B 34: 197±212. Urasa I. T. and Ferede F. (1987) Use of direct current plasma as an element selective detector for simultaneous ion chromatographic determination of arsenic(III) and arsenic(V) in the presence of other common anions, Anal Chem 59: 1563±1568. Reed T. B. (1961) Induction-coupled plasma torch, J Appl Phys 32: 821±824. Greenfield S., Jones I. L. and Berry C. T. (1964) High-pressure plasmas as spectroscopic emission sources, Analyst 89: 713±720. Wendt R. H. and Fassel V. A. (1965) Induction coupled plasma spectrometric excitation source, Anal Chem 37: 920±922. Rezaaiyaan R., Hieftje G. M., Anderson H., Kaiser H. and Meddings B. (1982) Design and construction of a low-¯ow, low-power torch for ICP spectrometry, Appl Spectrosc 36: 626±631. Van Der Plas P. S. C. and De Galan L. (1984) A radiatively cooled torch for ICP-AES using 1 L minÿ1 of argon, Spectrochim Acta, Part B 39: 1161± 1169. Van Der Plas P. S. C., De Waaij A. C. and De Galan L. (1985) Analytical evaluation of an air-cooled 1 L minÿ1 argon ICP, Spectrochim Acta, Part B 4O: 1457±1466. Meyer G. A. and Thompson M. D. (1985) Determination of trace element detection limits in air and oxygen inductively coupled plasmas, Spectrochim Acta, Part B 40: 195±207. Bacri J., Gomes A. M., Fieni J. M., Thouzeau F. and Birolleau J. C. (1989) Transport coecient calculation in air±argon mixtures: explana-
Literature
382
383
384
385
386
387
388
389
tion of the improvement of detection limits of pollutants in air inductively coupled plasmas by argon addition, Spectrochim Acta, Part B 44: 887±895. Kalnicki D. J., Fassel V. A. and Kalnicky R. N. (1977) Electron temperatures and electron number densities experienced by analyte species in ICP with and without the presence of an easily ionized element, Appl Spectrosc 31: 137±150. Raeymaekers B., Broekaert J. A. C. and Leis F. (1988) Radially resolved rotational temperatures in nitrogen± argon, oxygen±argon, air±argon and argon ICPs, Spectrochim Acta 43: 941± 949. Ishii I. and Montaser A. (1991) A tutorial discussion on measurements of rotational temperature in ICP, Spectrochim Acta, Part B 46: 1197± 1206. Boumans P. W. J. M. and de Boer F. J. (1977) An experimental study of a 1-kW, 50-MHz RF ICP with pneumatic nebulizer and a discussion of experimental evidence for a nonthermal mechanism, Spectrochim Acta, Part B 32: 365±395. De Galan L. (1984) Some considerations on the excitation mechanism in the inductively coupled argon plasma, Spectrochim Acta, Part B 39: 537±550. Huang M., Hanselmann D. S., Yang P. Y. and Hieftje G. M. (1992) Isocontour maps of electron temperature, electron number density and gas kinetic temperature in the Ar inductively coupled plasma obtained by laser-light Thomson and Rayleigh scattering, Spectrochim Acta, Part B 47: 765±786. Huang M., Lehn S. A., Andrews E. J. and Hieftje G. M. (1997) Comparison of electron concentrations, electron temperatures and excitation temperatures in argon ICPs operated at 27 and 40 MHz, Spectrochim Acta, Part B 52: 1173±1193. Hanselman D. S., Sesi N. N., Huang M. and Hieftje G. M. (1994) The eect of sample matrix on electron density, electron temperature and gas
390
391
392
393
394
395
396
397
398
temperature in the argon inductively coupled plasma examined by Thomson and Rayleigh scattering, Spectrochim Acta, Part B 49: 495± 526. Hanselman D. S., Withnell R. and Hieftje G. M. (1991) Side-on photomultiplier gating system for Thomson scattering and laser-excited atomic ¯uorescence spectroscopy, Appl Spectrosc 45: 1553±1560. Michaud-Poussel E. and Mermet J. M. (1986) In¯uence of the generator frequency and the plasma gas inlet area on torch design in ICP-AES, Spectrochim Acta, Part B 41: 125±132. Brenner J. B., Zander A., Cole M. and Wiseman A. (1997) Comparison of axially and radially viewed inductively coupled plasmas for multielement analysis. Eect of sodium and calcium, J Anal At Spectrom 12: 897±906. Ivaldi J. C. and Tyson J. F. (1995) Performance evaluation of an axially viewed horizontal inductively coupled plasma for optical emission spectrometry, Spectrochim Acta, Part B 50: 1207±1226. Bings N. H., Wang C., Skinner C. D., Colyer C. L., Thibault P. and Harrison D. J. (1999) Micro¯uid devices connected to fused-silica capillaries with minimal dead volume, Anal Chem 71: 3292±3296. Broekaert J. A. C. and Leis F. (1979) An injection method for the sequential determination of boron and several metals in waste-water samples by inductively-coupled plasma atomic emission spectrometry, Anal Chim Acta 109: 73±83. Walters F. H., Parker L. R. Jr., Morgan S. L. and Deming S. N. (1991) Sequential simplex optimization, CRC Press, Inc., Boca Raton. Moore G. L., Humphries-Cuff P. J. and Watson A. E. (1984) Simplex optimization of a nitrogen-cooled argon inductively coupled plasma for multielement analysis, Spectrochim Acta, Part B 39: 915±929. Parsons M. L., Forster A. R. and Anderson D. (1980) An atlas of
331
332
Literature
399
400
401
402
403
404
405
406
spectral interferences in ICP spectroscopy, Plenum Press, New York. Broekaert J. A. C., Leis F. and Laqua K. (1979) Some aspects of matrix eects caused by sodium tetraborate in the analysis of rare earth minerals with the aid of inductively coupled plasma atomic emission spectrometry, Spectrochim Acta, Part B 34: 167±175. Belchamber R. M. and Horlick G. (1982) Noise power spectra of optical and acoustic emission signals from an ICP, Spectrochim Acta, Part B 37: 17± 27. Mann S., Geilenberg D., Broekaert J. A. C. and Jansen M. (1997) Digestion methods for advanced ceramic materials and subsequent determination of silicon and boron by inductively coupled plasma atomic emission spectrometry, J Anal At Spectrom 12: 975±979. Geilenberg D., Gerards M. and Broekaert J. A. C. (2000) Determination of the stoichiometric composition of high-temperature superconductors by ICP-OES for production control, Mikrochim Acta 133: 319±323. Kucharkowski R., Vogt C. and Marquardt D. (2000) Accurate and precise spectrochemical analysis of Bi-Pb-Sr-Ca-Cu-O high-temperature superconductor materials, Fresenius' J Anal Chem 366: 146±151. Barnett W. B., Fassel V. A. and Kniseley R. N. (1968) Theoretical principles of internal standardization in analytical emission spectroscopy, Spectrochim Acta 23: 643±664. Meyers S. A. and Tracy D. H. (1983) Improved performance using internal standardisation in ICP emission spectroscopy, Spectrochim Acta, Part B 38: 1227±1253. Broekaert J. A. C., KlockenkaÈmper R. and Ko J. B. (1983) Emissionsspektrometrische PraÈzisionsbestimmung der Hauptbestandteile von Cu/Zn-Legierungen mittels Glimmlampe und ICP, Fresenius' Z Anal Chem 316: 256±260.
407 Broekaert J. A. C., Leis F. and
408
409
410
411
412
413
414
Laqua K. (1981) The application of an argon/nitrogen inductively-coupled plasma to the analysis of organic solutions, Talanta 28: 745±752. Boumans P. W. J. M. and LuxSteiner M. Ch. (1982) Modi®cation and optimization of a 50 MHz inductively coupled argon plasma with special reference to analyses using organic solvents, Spectrochim Acta 37: 97±126. Olesik J. W. and Hobbs S. E. (1994) Monodisperse dried microparticulate injector: a new tool for studying fundamental processes in inductively coupled plasmas, Anal Chem 66: 3371±3378. Tarr M. A., Zhu G. and Browner R. F. (1993) Micro¯ow ultrasonic nebulizer for inductively coupled plasma atomic emission spectrometry, Anal Chem 65: 1689±1695. Vermeiren K., Vandecasteele C. and Dams R. (1990) Determination of trace amounts of cadmium, lead, copper and zinc in natural waters by ICP-AES with thermospray nebulization after enrichment on Chelex100, Analyst 115: 17±22. Dittrich K., Berndt H., Broekaert J. A. C., Schaldach G. and ToÈlg G. (1988) Comparative study of injection into a pneumatic nebuliser and tungsten coil electrothermal vaporisation for the determination of rare earth elements by inductively coupled plasma optical emission spectrometry, J Anal At Spectrom 3: 1105±1110. Lucic M. and Krivan V. (1999) Analysis of aluminium-based ceramic powders by electrothermal vaporization inductively coupled plasma atomic emission spectrometry using a tungsten coil and slurry sampling, Fresenius' J Anal Chem 363: 64±72. Karanassios V., Abdullah M. and Horlick G. (1990) The application of chemical modi®cation in direct sample insertion-inductively coupled plasma-atomic emission spectrometry, Spectrochim Acta, Part B 45: 119± 130.
Literature 415 Liu X. R. and Horlick G. (1994) In-
416
417
418
419
420
421
422
423
424
situ laser ablation sampling for inductively coupled plasma atomic emission spectrometry, Spectrochim Acta, Part B 50: 537±548. Liu Y., Wang X., Yuan D., Yang P., Huang B. and Zhuang Z. (1992) Flow-injection-electrochemical hydride generation technique for atomic absorption spectrometry, J Anal At Spectrom 7: 287±291. Hueber D. M. and Winefordner J. D. (1995) A ¯owing electrolytic hydride generator for continuous sample introduction in atomic spectrometry, Anal Chim Acta 316: 129±44. Cox A. G., Cook L. G. and McLeod C. W. (1985) Rapid sequential determination of chromium III/chromium VI by ¯ow injection analysis/ICP AES, Analyst 110: 331±333. Ishizuka T. and Uwamino Y. (1983) ICP emission spectrometry of solid samples by laser ablation, Spectrochim Acta, Part B 38: 519±527. Brenner I. B., Watson A. E., Steele T. W., Jones E. A. and Goncalves M. (1981) Application of an argon± nitrogen inductively coupled radiofrequency plasma (ICP) to the analysis of geological and related materials for their rare earth contents, Spectrochim Acta, Part B 36: 785±797. Garbarino J. R. and Taylor M. E. (1979) An inductively-coupled plasma atomic-emission spectrometric method for routine water quality testing, Appl Spectrosc 33: 220±226. U.S. Environmental Protection Agency, Inductively coupled plasmaatomic emission spectrometric method for trace element analysis of water and wastes, Washington, DC, 1979. DIN 38 406 (1987) Bestimmung der 24 Elemente Ag, Al, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, K, Mg, Mn, Mo, Na, Ni, P, Pb, Sb, Sr, Ti, V, Zn und Zr durch Atomemissionsspektrometrie mit induktiv gekoppeltem Plasma (ICP-AES)(E22), Beuthe Verlag, Berlin. Thompson M., Pahlavanpour B., Walton S. J. and Kirkbright G. F. (1978) Simultaneous determination of
425
426
427
428
429
430
431
432
trace concentrations of arsenic, antimony, bismuth, selenium and tellurium in aqueous solutions by introduction of the gaseous hydrides into an ICP source for emission spectrometry. Part II. Interference studies, Analyst 103: 705±713. Miyazaki A., Kimura A., Bansho K. and Umezaki Y. (1982) Simultaneous determination of heavy metals in waters by ICP-AES after extraction into diisobutyl ketone, Anal Chim Acta 144: 213±221. Berndt H., Harms U. and Sonneborn M. (1985) MultielementSpurenvoranreicherung aus WaÈssern an Aktivkohle zur Probenvorbereitung fuÈr die Atomspektroskopie (FlammenAAS, ICP/OES), Fresenius' Z Anal Chem 322: 329±333. Hiraide M., Ito T., Baba M., Kawaguchi H. and Mizuike A. (1980) Multielement preconcentration of trace heavy metals in water by preconcentration and ¯otation with indium hydroxide for ICP-AES, Anal Chem 52: 804±807. Broekaert J. A. C., Gucer S. and Adams F. (1990) Metal speciation in the environment, Springer Verlag, Berlin. Graule T., von Bohlen A., Broekaert J. A. C., Grallath E., KlockenkaÈmper R., TschoÈpel P. and ToÈlg G. (1989) Atomic emission and atomic absorption spectrometric analysis of high-purity powders for the production of ceramics, Fresenius' Z Anal Chem 335: 637±642. Docekal B., Broekaert J. A. C., Graule T., TschoÈpel P. and ToÈlg G. (1992) Determination of impurities in silicon carbide powders, Fresenius' J Anal Chem 342: 113±117. Merten D., Broekaert J. A. C., Brandt R. and Jakubowski N. (1999) Analysis of ZrO2 powders by microwave assisted digestion at high pressure and ICP atomic spectrometry, J Anal At Spectrom 14: 1093±1098. Gross R., Platzer B., Leitner E., Schalk A., Sinabell H., Zach H. and Knapp G. (1992) Atomic emission gas chromatographic detection ± chemical and spectral
333
334
Literature
433
434
435
436
437
438
439
440
interferences in the stabilized capacitive plasma (SCP), Spectrochim Acta, Part B 47: 95±106. Luge S., Broekaert J. A. C., Schalk A. and Zach H. (1995) The use of dierent sample introduction techniques in combination with the low power stabilized capacitive plasma (SCP) as a radiation source for atomic emission spectrometry, Spectrochim Acta, Part B 50: 441±452. Luge S. and Broekaert J. A. C. (1996) Use of a low power, high frequency stabilized capacitive plasma combined with graphite furnace vaporization for the atomic emission spectrometric analysis of serum samples, Anal Chim Acta 332: 193± 199. Herwig F. and Broekaert J. A. C. (2000) Optimization of a direct sample insertion into a stabilized capacitive plasma for optical emission spectrometry (SCP-OES), Mikrochim Acta 134: 51±56. Skogerboe R. K. and Coleman G. N. (1976) Microwave plasma emission spectrometry, Anal Chem 48: 611A± 622A. Mavrodineanu R. and Hughes R. C. (1963) Excitation in radiofrequency discharges, Spectrochim Acta 19: 1309±1307. Bings N. H., Olschewski M. and Broekaert J. A. C. (1997) Twodimensional spatially resolved excitation and rotational temperatures as well as electron number density measurements in capacitively coupled microwave plasmas using argon, nitrogen and air as working gases by spectroscopic methods, Spectrochim Acta, Part B 52: 1965±1981. Disam A., TschoÈpel P. and ToÈlg G. (1982) Emissionsspektrometrische Bestimmung von elementspuren in waÈssrigen LoÈsungen mit einem mantelgasstabilisierten kapazitiv gekoppelten Mikrowellenplasma (CMP), Fresenius' Z Anal Chem 310: 131±143. Hanamura S., Smith B. W. and Winefordner J. D. (1983) Speciation of inorganic and organometallic
441
442
443
444
445
446
447
448
compounds in solid biological samples by thermal vaporization and plasma emission spectrometry, Anal Chem 55: 2026±2032. Seelig M., Bings N. H. and Broekaert J. A. C. (1998) Use of a capacitively coupled microwave plasma (CMP) with Ar, N2 and air as working gases for atomic spectrometric elemental determinations in aqueous solutions and oils, Fresenius' J Anal Chem 360: 161±166. Seelig M. and Broekaert J. A. C. (2000) Investigations on the on-line determination of metals in air ¯ows by capacitively coupled microwave plasma atomic emission spectrometry (CMP-OES), Spectrochim Acta, Part B in press. Beenakker C. I. M. (1977) Evaluation of a microwave-induced plasma in helium at atmospheric pressure as an element-speci®c detector for gas chromatography, Spectrochim Acta, Part B 32: 173±178. Kollotzek D., TschoÈpel P. and ToÈlg G. (1982) LoÈsungsemissionsspektrometrie mit mikrowelleninduzierten Mehrfaden- und Hohlzylinderplasmen I.- Die radiale Verteilung von Signalund UntergrundintensitaÈten, Spectrochim Acta, Part B 37: 91±96. Heltai Gy., Broekaert J. A. C., Leis F. and ToÈlg G. (1990) Study of a toroidal argon and a cylindrical helium microwave induced plasma for analytical atomic emission spectrometry ± I. Con®gurations and spectroscopic properties, Spectrochim Acta, Part B 45: 301±311. Deutsch R. D. and Hieftje G. M. (1985) Development of a microwaveinduced nitrogen discharge at atmospheric pressure (MINDAP), Appl Spectrosc 39: 214±222. Wu M. and Carnahan J. W. (1990) Trace determination of cadmium, copper, bromine and chlorine into a helium microwave-induced plasma, Appl Spectrosc 44: 673±678. Moussanda P. S., Ranson P. and Mermet J. M. (1985) Spatially resolved spectroscopic diagnostics of an argon MIP produced by surface
Literature
449
450
451
452
453
454
455
456
457
wave propagation (Surfatron), Spectrochim Acta, Part B 40: 641±651. Kollotzek D., Oechsle D., Kaiser G., TschoÈpel P. and ToÈlg G. (1984) Application of a mixed-gas microwave induced plasma as an on-line element-speci®c detector in highperformance liquid chromatography, Fresenius' Z Anal Chem 318: 485±489. Kollotzek D., TschoÈpel P. and ToÈlg G. (1984) Three-®lament and toroidal microwave induced plasmas as radiation sources for emission spectrometric analysis of solutions and gaseous samples. II. Analytical performance, Spectrochim Acta, Part B 39: 625±636. Uden P. (1992) Element-speci®c chromatographic detection by atomic emission spectroscopy, American Chemical Society, Washington. Lobinski R. (1997) Elemental speciation and coupled techniques, Appl Spectrosc 51: 260A±278A. Hubert J., Moisan M. and Ricard A. (1979) A new microwave plasma at atmospheric pressure, Spectrochim Acta, Part B 33: 1±10. Richts U., Broekaert J. A. C., TschoÈpel P. and ToÈlg G. (1991) Comparative study of a Beenakker cavity and a surfatron in combination with electrothermal evaporation from a tungsten coil for microwave plasma optical emission spectrometry (MIPAES), Talanta 38: 863±869. Yang J., Schickling C., Broekaert J. A. C., TschoÈpel P. and ToÈlg G. (1995) Evaluation of continuous hydride generation combined with helium and argon microwave induced plasmas using a surfatron for atomic emission spectrometric determination of arsenic, antimony and selenium, Spectrochim Acta, Part B 50: 1351± 1363. Jin Q., Zhu C., Borer M. and Hieftje G. M. (1991) A microwave plasma torch assembly for atomic emission spectrometry, Spectrochim Acta, Part B 46: 417±430. Prokisch C., Bilgic A. M., Voges E., Broekaert J. A. C., Jonkers J., van Sande M. and van der Mullen J. A. M. (1999) Photographic plasma
458
459
460
461
462
463
464
images and electron number density as well as electron temperature mappings of a plasma sustained with a modi®ed argon microwave torch (MPT) measured by spatially resolved Thomson scattering, Spectrochim Acta, Part B 54: 1253±1266. Engel U., Prokisch C., Voges E., Hieftje G. M. and Broekaert J. A. C. (1998) Spatially resolved measurements and plasma tomography with respect to the rotational temperatures for a microwave plasma torch, J Anal At Spectrom 13: 955±961. Prokisch C. and Broekaert J. A. C. (1998) Elemental determinations in aqueous and acetonitrile containing solutions by atomic spectrometry using a microwave plasma torch, Spectrochim Acta, Part B 53: 1109± 1119. Pereiro R., Wu M., Broekaert J. A. C. and Hieftje G. M. (1994) Direct coupling of continuous hydride generation with microwave plasma torch atomic emission spectrometry for the determination of arsenic, antimony and tin, Spectrochim Acta, Part B 49: 59±73. Bilgic A. M., Prokisch C., Broekaert J. A. C. and Voges E. (1998) Design and modelling of a modi®ed 2.45 GHz coaxial plasma torch for atomic spectrometry, Spectrochim Acta, Part B 53: 773±777. Bilgic A. M., Engel U., Voges E., Èckelheim M. and Broekaert J. A. Ku C. (2000) A new low-power microwave plasma source using microstrip technology for atomic emission spectrometry, Plasma Sources Sci Technol 9: 1±4. Engel U., Bilgic A. M., Haase O., Voges E. and Broekaert J. A. C. (2000) A microwave-induced plasma based on microstrip technology and its use for the atomic emission spectrometric determination of mercury with the aid of the cold-vapor technique, Analytical Chemistry 72: 193±197. Bilgic A. M., Voges E., Engel U. and Broekaert J. A. C. (2000) A low-
335
336
Literature
465
466
467
468
469
470
471
472
473
power 2.45 GHz microwave induced helium plasma source at atmospheric pressure based on microstrip technology, J Anal At Spectrom 15: 579±580. Bogaerts A., Donko Z., Kutasi K., Bano G., Pinhao N. and Pinheiro M. (2000) Comparison of calculated and measured optical emission intensities in a direct current argon± copper glow discharge, Spectrochim Acta, Part B 55: 1465±1479. Bogaerts A. and Gijbels R. (2000) Description of the argon-excited levels in a radio-frequency and direct current glow discharge, Spectrochim Acta, Part B 55: 263±278. Broekaert J. A. C. (1979) Some observations on sample volatilisation in a hot-type hollow cathode, Spectrochim Acta, Part B 34: 11±17. Zil'bershtein Kh. I. (1977) Spectrochemical analysis of pure substances, Adam Hilger, Bristol, 173±227. Thelin B. (1981) The use of a high temperature hollow cathode lamp for the determination of trace elements in steels, nickel-base alloys and ferroalloys by emission spectrometry, Appl Spectrosc 35: 302±307. Schepers C. and Broekaert J. A. C. (2000) The use of a hollow cathode glow discharge (HCGD) as an atomic emission spectrometric element speci®c detector for chlorine and bromine in gas chromatography, J Anal At Spectrom 15: 51±65. Broekaert J. A. C. (1976) Application of hollow cathode excitation coupled to vidicon detection to the simultaneous multielement determination of toxic elements in airborne dust ± A unique sampling-analysis procedure for lead and cadmium, Bull Soc Chim Belg 85: 755±761. Dempster M. A. and Marcus R. K. (2000) Optimization of hollow cathode diameter for particle beam/hollow cathode glow discharge atomic emission spectrometry, Spectrochim Acta, Part B 55: 599±610. Caroli S. (1985) Improved hollow cathode lamps for atomic spectroscopy, Ellis Horwood, Chichester.
È dke 474 Falk H., Hoffmann E. and Lu
475
476
477
478
479
480
481
482
483
Ch. (1984) A comparison of furnace atomic nonthermal excitation spectrometry (FANES) with other atomic spectroscopic techniques, Spectrochim Acta, Part B 39: 283±294. Grimm W. (1968) Eine neue Glimmentladungslampe fuÈr die optische Emissionsspektralanalyse, Spectrochim Acta, Part B 23: 443±454. Ko J. B. (1984) New designs of glow discharge lamps for the analysis of metals by atomic emission spectrometry, Spectrochim Acta, Part B 39: 1405±1423. Steers E. B. M. (1997) Charge transfer excitation in glow discharge sources: the spectra of titanium and copper with neon, argon and krypton as the plasma gas, J Anal At Spectrom 12: 1033±1040. Steers E. B. M. and Leis F. (1991) Excitation of the spectra of neutral and singly ionized atoms in the Grimm-type discharge lamp with and without supplementary microwave excitation, Spectrochim Acta, Part B 46: 527±537. Leis F., Broekaert J. A. C. and Laqua K. (1987) Design and properties of a microwave boosted glow discharge lamp, Spectrochim Acta, Part B 42: 1169±1176. Ko J. B. and Laqua K. (1975) Vergleich verschiedener Anregungsverfahren bei der emissionsspektrometrischen Analyse von Aluminium-Legierungen, Proc XVIII Coll Spectrosc Intern, GAMS, Volume II, 549±556. Bengtson A. (1985) A contribution to the solution of the problem of quanti®cation in surface analysis work using glow discharge atomic emission spectroscopy, Spectrochim Acta, Part B 40: 631±639. Hoffmann V., Kurt R., KaÈmmer K., Thielsch R., Wirth Th. and Beck U. (1999) Interference phenomena at transparent layers in glow discharge optical emission spectrometry, Appl Spectrosc 53: 987±990. Winchester M. R., Lazik C. and Marcus R. K. (1991) Characterization of a radio frequency glow discharge
Literature
484
485
486
487
488
489
490
491
emission source, Spectrochim Acta, Part B 46: 483±499. Hoffmann V., Uhlemann H. J., PraÈûler F. and Wetzig K. (1996) New hardware for radio frequency powered glow discharge spectroscopies and its capabilities for analytical applications, Fresenius' J Anal Chem 355: 826±830. Belkin M., Caruso J. A., Christopher S. J. and Marcus R. K. (2000) Characterization of helium/ argon working gas systems in a radiofrequency glow discharge atomic emission source. Part II: Langmuir probe and emission intensity studies for Al, Cu and Macor samples, Spectrochim Acta, Part B 53: 1197± 1208. Lowe R. M. (1978) A modi®ed glowdischarge source for emission spectroscopy, Spectrochim Acta, Part B 31: 257±261. Ferreira N. P., Strauss J. A. and Human H. G. C. (1983) Developments in glow discharge emission spectrometry, Spectrochim Acta, Part B 38: 899±911. Broekaert J. A. C., Pereiro R., Starn T. K. and Hieftje G. M. (1993) A gas-sampling glow discharge coupled to hydride generation for the atomic spectrometric determination of arsenic, Spectrochim Acta, Part B 48: 1207±1220. Broekaert J. A. C., Starn T. K., Wright L. J. and Hieftje G. M. (1998) Studies of a helium-operated gas-sampling Grimm-type glow discharge for the atomic emission spectrometric determination of chlorine in halogenated hydrocarbon vapors, Spectrochim Acta, Part B 53: 1723±1735. McLuckey S. A., Glish K. G., Asano K. G. and Grant B. C. (1988) Atmospheric sampling glow discharge ionization source for the determination of trace organic compounds in ambient air, Anal Chem 60: 2220±2227. Pereiro R., Starn T. K. and Hieftje G. M. (1995) Gas-sampling glow discharge for optical emission spectrometry. Part II: Optimization
492
493
494
495
496
497
498
499
500
501
and evaluation for the determination of non-metals in gas-phase samples, Appl Spectrosc 49: 616±622. Margetic V., Pakulev A., Stockhaus A., Bolshov M., Niemax K. and HergenroÈder R. (2000) A comparison of nanosecond and femtosecond laser-induced plasma spectroscopy of brass samples, Spectrochim Acta, Part B 55: 1771±1785. Sugden T. M. in: Reed R. L. (ed) (1965) Mass spectrometry, Academic Press, New York, p. 347. Douglas D. J. and French J. B. (1981) Elemental analysis with a microwave-induced plasma quadrupole mass spectrometer system, Anal Chem 53: 37±41. Houk R. S., Fassel V. A., Flesh G. D., Svec H. J., Gray A. L. and Taylor Ch. E. (1980) Inductively coupled argon plasma as an ion source for mass spectrometric determination of trace elements, Anal Chem 52: 2283± 2289. Date A. R. and Gray A. L. (1989) Applications of ICP-MS, Blackie, Glasgow. Jakubowski N., Raeymaekers B. J., Broekaert J. A. C. and Stuewer D. (1989) Study of plasma potential eects in a 40 MHz inductively coupled plasma mass spectrometry system, Spectrochim Acta, Part B 44: 219±228. Vaughan M. A. and Horlick G. (1990) Ion trajectories through the input ion optics of an ICP-MS, Spectrochim Acta, Part B 45: 1301± 1311. Jakubowski N., Moens L. and Vanhaecke F. (1998) Sector ®eld mass spectrometers in ICP-MS, Spectrochim Acta, Part B 53: 1739± 1763. Solyom D. A., Burgoyne T. W. and Hieftje G. M. (1999) Plasma source sector ®eld mass spectrometry with array detection, J Anal At Spectrom 14: 1101±1110. Broekaert J. A. C. (1990) ICP Massenspektrometrie, in: GuÈnzler H., Borsdorf R., Fresenius W., Huber W., Kelker G., LuÈdenwald I., ToÈlg G. und
337
338
Literature
502
503
504
505
506
507
508
509
510
Wisser H. (eds) Analytiker Taschenbuch, Band 9. Springer-Verlag, Berlin, 127±163. Gray A. L., Houk R. S. and Williams J. G. (1987) Langmuir probe potential measurements in the plasma and their correlations with mass spectral characteristics in ICP-MS, J Anal At Spectrom 2: 13±20. Vaughan M. A. and Horlick G. (1986) Oxide, hydroxide and doubly charged analyte species in ICP mass spectrometry, Appl Spectrosc 40: 434± 445. Tittes W., Jakubowski N., Stuewer D., Toelg G. and Broekaert J. A. C. (1994) Reduction of some selected spectral interferences in inductively coupled plasma mass spectrometry, J Anal At Spectrom 9: 1015±1020. Vaughan M. A., Horlick G. and Tan S. H. (1987) Eect of the operating parameters on analyte signals in inductively coupled plasma mass spectrometry, J Anal At Spectrom 2: 765±772. Date A. R. and Gray A. L. (1985) Determination of trace elements in geological samples by inductively coupled plasma source mass spectrometry, Spectrochim Acta, Part B 40: 115±122. Broekaert J. A. C., Brandt R., Leis F., Pilger C., Pollmann D., Tschoepel P. and Toelg G. (1994) Analysis of aluminium oxide and silicon carbide ceramic materials by inductively coupled plasma mass spectrometry, J Anal At Spectrom 9: 1063±1070. Tian X., Emteborg H. and Adams F. C. (1999) Analytical performance of axial inductively coupled plasma time of ¯ight mass spectrometry (ICPTOFMS), J Anal At Spectrom 14: 1807±1814. Vandecasteele C., Nagels M., Vanhoe H. and Dams R. (1988) Suppression of analyte signal in ICPMS and the use of an internal standard, Anal Chim Acta 211: 91±98. Crain J. S., Houk R. S. and Smith F. G. (1988) Matrix interferences in ICP MS: some eects of skimmer ori®ce
511
512
513
514
515
516
517
518
519
520
diameter and ion lense voltages, Spectrochim Acta, Part B 43: 1355± 1364. Fraser M. M. and Beauchemin D. (2000) Eect of concomitant elements on the distribution of ions in inductively coupled plasma-mass spectroscopy. Part 1. Elemental ions, Spectrochim Acta, Part B 55: 1705± 1731. Tan S. H. and Horlick G. (1986) Background spectral features in ICP mass spectrometry, Appl Spectrosc 40: 445±460. Douglas D. J. and French J. B. (1992) Collisional focusing eects in radio frequency quadrupoles, J Am Soc Mass Spectrom 3: 398±408. Barinaga C. J., Eiden G. C., Alexander M. L. and Koppenaal D. W. (1996) Analytical atomic spectroscopy using ion trap devices, Fresenius' J Anal Chem 355: 487±493. Feldmann I., Jakubowski N. and Stuewer D. (1999) Application of a hexapole collision and reaction cell in ICP-MS. Part I: Instrumental aspects and operational optimization, Fresenius' J Anal Chem 365: 415±421. Mason P. R. D., Kaspers K. and van Bergen M. J. (1999) Determination of sulfur isotope ratios and concentrations in water samples using ICP-MS incorporating hexapole ion optics, J Anal At Spectrom 14: 1067±1074. Riepe H. G., Gomez M., Camara C. and Bettmer J. (2000) Modi®cation of capillaries coupled to micro-¯ow nebulizers: a new strategy for on-line interference removal in inductively coupled plasma mass spectrometry, J Mass Spectrom 35: 891±896. Heumann K. G. (1982) Isotope dilution mass spectrometry for microand trace element determination, Trends Anal Chem 1: 357±361. Dean J. R., Ebdon L. and Massey R. J. (1987) Selection of mode for the measurement of lead isotope ratios by ICP mass spectrometry and its application to milk powder analysis, J Anal At Spectrom 2: 369±374. Ting B. T. G. and Janghorbani M. (1987) Application of ICP-MS to
Literature
521
522
523
524
525
526
527
528
529
530
accurate isotopic analysis for human metabolic studies, Spectrochim Acta, Part B 42: 21±27. Vanhaecke F., Moens L., Dams R., Allen L. and Georgitis G. (1999) Evaluation of the isotope ratio performance of an axial time-of-¯ight ICP mass spectrometer, Anal Chem 71: 3297±3303. Vanhaecke F., Moens L. and Dams R. (1997) Applicability of highresolution ICP-mass spectrometry for isotope ratio measurements, Anal Chem 69: 268±273. Becker J. S. and Dietze H. J. (1998) Ultratrace and precise isotope analysis by double-focusing sector ®eld inductively coupled plasma mass spectrometry, J Anal At Spectrom 13: 1057±1063. Hirata T. and Yamaguchi T. (1999) Isotopic analysis of zirconium using enhanced sensitivity-laser ablationmultiple collector-inductively coupled plasma mass spectrometry, J Anal At Spectrom 14: 1455±1459. Hausler D. (1987) Trace element analysis of organic solutions using ICP mass spectrometry, Spectrochim Acta, Part B 42: 63±73. Zhu G. and Browner R. F. (1988) Study of the in¯uence of water vapor loading and interface pressure in ICPMS, J Anal At Spectrom 3: 781±790. Jakubowski N., Feldmann I., Èwer D. and Berndt H. (1992) Stu Hydraulic high pressure nebulization ± application of a new nebulization system for inductively coupled plasma mass spectrometry, Spectrochim Acta, Part B 47: 119±129. Wang X., Viczian M., Lasztity A. and Barnes R. M. (1988) Lead hydride generation for isotope analysis by ICP-MS, J Anal At Spectrom 3: 821±828. Park C. J., Van Loon J. C., Arrowsmith P. and French J. B. (1987) Sample analysis using plasma source mass spectrometry with electrothermal sample introduction, Anal Chem 59: 2191±2196. Boomer D. W., Powell M., Sing R. L. A. and Salin E. D. (1986)
531
532
533
534
535
536
537
538
Application of a wire loop direct sample-insertion device for ICP-MS, Anal Chem 58: 975±976. Hall G. E. M., Pelchat J. C., Boomer D. W. and Powell M. (1988) Relative merits of two methods of sample introduction in ICP-MS: electrothermal vaporisation and direct sample insertion, J Anal At Spectrom 3: 791±797. Williams J. G., Gray A. L., Norman P. and Ebdon L. (1987) Feasibility of solid sample introduction by slurry nebulisation for ICP-MS, J Anal At Spectrom 2: 469±472. Fonseca R. W., Miller-Ihli N. J., Sparks C., Holcombe J. A. and Shaver B. (1997) Eect of oxygen ashing on analyte transport eciency using ETV-ICP-MS, Appl Spectrosc 51: 1800±1806. Hauptkorn S., Krivan V., Gercken B. and Pavel J. (1997) Determination of trace impurities in high-purity quartz by electrothermal vaporization inductively coupled plasma mass spectrometry using the slurry sampling technique, J Anal At Spectrom 12: 421±428. Wende M. C. and Broekaert J. A. C. (2001) Investigations on the use of chemical modi®ers for the direct determination of trace impurities in Al2 O3 ceramic powders by slurry electrothermal evaporation coupled to inductively coupled plasma mass spectrometry (ETV-ICP-MS), Fresenius' J Anal Chem, 370: 513±520. Jakubowski N., Feldmann I., Sack È wer D. (1992) Analysis of B. and Stu conducting solids by ICP-MS with spark ablation, J Anal At Spectrom 7: 121±125. Arrowsmith P. (1987) Laser ablation of solids elemental analysis by ICP mass spectrometry, Anal Chem 59: 1437±1444. Guenther D. and Heinrich C. A. (1999) Comparison of the ablation behaviour of 266 nm Nd:YAG and 193 nm ArF excimer lasers for LA-ICP-MS analyses, J Anal At Spectrom 14: 1369±1474.
339
340
Literature 539 Borisov O. V., Mao X. L., Fernandez
540
541
542
543
544
545
546
547
548
A., Caetano M. and Russo R. E. (1999) Inductively coupled plasma mass spectrometric study of nonlinear calibration behavior during laser ablation of binary CuaZn alloys, Spectrochim Acta, Part B 54: 1351± 1365. Date A. R. and Hutchinson D. (1987) Determination of rare earth elements in geological samples by ICP source mass spectrometry, J Anal At Spectrom 2: 269±281. Garbarino J. R. and Taylor H. E. (1987) Stable isotope dilution analysis of hydrologic samples by ICP-MS, Anal Chem 59: 1568±1575. Park C. J. and Hall G. E. M. (1988) Analysis of geological materials by ICP-MS with sample introduction by electrothermal vaporisation, J Anal At Spectrom 3: 355±361. Gregoire D. C. (1988) Determination of platinum, ruthenium and iridium in geological materials by ICP-MS with sample introduction by electrothermal vaporization, J Anal At Spectrom 3: 309±314. Audetat A., Guenther D. and Heinrich Ch. A. (1998) Formation of a magmatic-hydrothermal ore deposit: insight with LA-ICP-MS analysis of liquid inclusions, Science 279: 2091± 2094. McLeod C. W., Date A. R. and Cheung Y. Y. (1986) Metal ions in ICP-mass spectrometric analysis of nickel-base alloys, Spectrochim Acta, Part B 41: 169±174. Pilger C., Leis F., TschoÈpel P., Broekaert J. A. C. and Toelg G. (1995) Analysis of silicon carbide powders with ICP-MS subsequent to sample dissolution without and with matrix, Fresenius' J Anal Chem 351: 110±116. Jiang S. J. and Houk R. S. (1986) Arc nebulisation for elemental analysis of conducting solids by ICP mass spectrometry, Anal Chem 58: 1739±1743. Arrowsmith P. and Hughes S. K. (1988) Entrainment and transport of laser ablated plumes for subsequent
549
550
551
552
553
554
556
557
558
elemental analysis, Appl Spectrosc 42: 1231±1239. Arrowsmith P. (1988) in: Barnes R. M. (ed) Abstracts of the 1988 Winter Conference on Plasma Spectrochemistry, San Diego, The University of Massachusetts, p. 54. Lyon T. D. B., Fell G. S., Hutton R. C. and Eaton A. N. (1988) Evaluation of ICAP MS for simultaneous multielement trace analysis in clinical chemistry, J Anal At Spectrom 3: 265± 272. Delves H. T. and Campbell M. J. (1988) Measurements of total lead concentrations and of lead isotope ratios in whole blood by use of ICP source MS, J Anal At Spectrom 3: 343±348. Serfass R. E., Thompson J. J. and Houk R. S. (1986) Isotope ratio determination by ICP MS for zinc bioavailability studies, Anal Chim Acta 188: 73±84. Dean J. R., Ebdon L., Crews H. M. and Massey R. C. (1988) Characteristics of ¯ow injection ICP MS for trace metal analysis, J Anal At Spectrom 3: 349±354. Dean J. R., Munro S., Ebdon L., Crews H. M. and Massey R. C. (1987) Studies of metalloprotein species by directly coupled high-performance liquid chromatography ICP-MS, J Anal At Spectrom 2: 607±610. Baker S. A. and Miller-Ihli N. J. (1999) Comparison of a cross-¯ow and microconcentric nebulizer for chemical speciation measurements using CZE-ICP-MS, Appl Spectrosc 53: 471±478. Van Holderbeke M., Zhao Y., Vanhaecke F., Moens L., Dams R. and Sandra P. (1999) Speciation of six arsenic compounds using capillary electrophoresis±inductively coupled plasma mass spectrometry, J Anal At Spectrom 14: 229±234. McLean J. A., Zhang H. and Montaser A. (1998) A direct injection high-eciency nebulizer for inductively coupled plasma mass spectrometry, Anal Chem 70: 1012±1020.
Literature 559 Herzog R. and Dietz F. (1987)
560
561
562
563
564
565
566
567
Anwendung der ICP MS in der Wasseranalytik, GewaÈsserschutz, Wasser, Abwasser, 92: 109±141. Beauchemin D., McLaren J. W., Mykytiuk A. P. and Berman S. S. (1988) Determination of trace metals in an open ocean water reference material by ICP-MS, J Anal At Spectrom 3: 305±308. McLaren J. W., Beauchemin D. and Berman S. S. (1987) Application of isotope dilution ICP mass spectrometry to the analysis of marine sediments, Anal Chem 59: 610±613. McLaren J. W., Beauchemin D. and Berman S. S. (1987) Determination of trace metals in marine sediments by ICP mass spectrometry, J Anal At Spectrom 2: 277±281. Ridout P. S., Jones H. R. and Williams J. G. (1988) Determination of trace elements in a marine reference material of lobster hepatopancreas (TORT-1) using ICPMS, Analyst 113: 1383±1386. Beauchemin D., McLaren J. W., Willie S. N. and Berman S. S. (1988) Determination of trace metals in marine biological reference materials by ICP MS, Anal Chem 60: 687±691. Barbante C., Cozzi G., Capodaglio G., Van de Velde K., Ferrari C., Veysseyse A., Boutron C. F., Scarponi G. and Cescon P. (1999) Determination of Rh, Pd, and Pt in polar and alpine snow and ice by double-focusing ICP-MS with microconcentric nebulization, Anal Chem 71: 4125±4133. Becker J. S., Dietze H. J., McLean J. A. and Montaser A. (1999) Ultratrace and isotope analysis of long-lived radionuclides by inductively coupled plasma quadrupole mass spectrometry using a direct injection higheciency nebulizer, Anal Chem 71: 3077±3084. Andrle C. M., Jakubowski N. and Broekaert J. A. C. (1997) Speciation of chromium using reversed phasehigh performance liquid chromatography coupled to dierent spectro-
568
569
570
571
572
573
574
575
metric detection methods, Spectrochim Acta, Part B 52: 189±200. Barnowski C., Jakubowski N., Stuewer D. and Broekaert J. A. C. (1997) Speciation of chromium by direct coupling of ion exchange chromatography with inductively coupled plasma mass spectrometry, J Anal At Spectrom 12: 1155±1161. Prange A. and Jantzen E. (1995) Determination of organometallic species by gas chromatography ICP MS, J Anal At Spectrom 10: 105±109. Heisterkamp M., De Smaele T., Candelone J. P., Moens L., Dams R. and Adams F. C. (1997) Inductively coupled plasma mass spectrometry hyphenated to capillary gas chromatography as a detection system for the speciation analysis of environmental samples, J Anal At Spectrom 12: 1077±1081. Rodriguez J., Mounicon S., Lobinski R., Sidelnikov V., Patrushev Y. and Yamanaku M. (1999) Species-selective analysis by microcolumn multicapillary gas chromatography with inductively coupled plasma mass spectrometric detection, Anal Chem 71: 4534±4543. Leach A. M., Heisterkamp M., Adams F. C. and Hieftje G. M. (2000) Gas chromatographyinductively coupled plasma time-of¯ight mass spectrometry for the speciation analysis of organometallic compounds, J Anal At Spectrom 15: 151±155. Michaud-Poussel E. and Mermet J. M. (1986) In¯uence of the torch frequency and the plasma-gas inlet area on torch design in ICP AES, Spectrochim Acta, Part B 41: 125± 132. Lam J. W. H. and Horlick G. (1990) A comparison of argon and mixed gas plasmas for ICP MS, Spectrochim Acta, Part B 45: 1313±1325. Chan S. K. and Montaser A. (1987) Characterization of an annular helium ICP generated in a low-gas¯ow torch, Spectrochim Acta, Part B 42: 591±597.
341
342
Literature 576 Nam S. H., Masamba W. R. L. and
577
578
579
580
581
582
583
584
Montaser A. (1993) Investigation of helium inductively coupled plasmamass spectrometry of metals and nonmetals in aqueous solutions, Anal Chem 65: 2784±2790. Satzger R. D., Fricke F. L., Brown P. G. and Caruso J. A. (1987) Detection of halogens as positive ions using a He microwave induced plasma as an ion source for mass spectrometry, Spectrochim Acta, Part B 42: 705±712. Wilson D. A., Vickers G. H. and Hieftje G. M. (1987) Use of microwave-induced nitrogen discharge at atmospheric pressure as an ion source for elemental mass spectrometry, Anal Chem 59: 1664± 1670. Okamoto Y. (1994) High-sensitivity microwave-induced plasma mass spectrometry for trace element analysis, J Anal At Spectrom 9: 745± 749. Blades M. W. (1994) Atmosphericpressure, radiofrequency, capacitively coupled helium plasmas, Spectrochim Acta, Part B 49: 47±57. Stewart I. I., Guevremont R. and Sturgeon R. E. (1999) The use of a sampler±skimmer interface for ion sampling in furnace atomization plasma ionization mass spectrometry, Anal Chem 71: 5146±5156. Bings N. H., Costa-Fernandez J. M., Guzowski Jr. J. P., Leach A. M. and Hieftje G. M. (2000) Time-of-¯ight mass spectrometry as a tool for speciation analysis, Spectrochim Acta, Part B 55: 767±778. Milgram K. E., White F. M., White M., Goodner K. L., Watson C. H., Koppenaal D. W., Barinaga C. W., Smith B. H., Winfordner J. G., Marshall A. G., Houk R. S. and Eyler J. R. (1997) High-resolution inductively coupled plasma Fourier transform ion cyclotron resonance mass spectrometry, Anal Chem 69: 3714±3721. Harrison W. W. and Bentz B. L. (1988) Glow discharge mass
585
586
587
588
589
590
591
592
593
spectrometry, Prog Anal Spectrosc 11: 53±110. Ratliff P. H. and Harrison W. W. (1994) The eects of water vapour in glow discharge mass spectrometry, Spectrochim Acta, Part B 49: 1747± 1757. Ratliff P. H. and Harrison W. W. (1995) Time-resolved studies of the eects of water vapor in glow discharge mass spectrometry, Appl Spectrosc 49: 863±871. Vieth W. and Huneke J. C. (1990) Studies on ion formation in a glow discharge mass spectrometry ion source, Spectrochim Acta, Part B 45: 941±949. Taylor W. S., Dulak J. G. and Ketkar S. N. (1990) Characterization of a glow discharge plasma as a function of sampling ori®ce potential, J Am Soc Mass Spectrom 1: 448. Goodner K. L., Eyler J. R., Barshick C. M. and Smith D. H. (1995) Elemental quanti®cation based on argides, dimers, and doubly charged glow discharge ions, Int J Mass Spectrom Ion Processes 146/147: 65± 73. Douglas Y. M., Duckworth C., Cable P. R. and Marcus R. K. (1994) Eects of target gas in collisioninduced dissociation using a double quadrupole mass spectrometer and radiofrequency glow discharge, J Am Soc Mass Spectrom 5: 845±851. Molle C., Wautelet D., Dauchot J. P. and Hecq M. (1995) Characterization of a magnetron radiofrequency glow discharge with a glass cathode using experimental design and mass spectrometry, J Anal At Spectrom 10: 1039±1045. È wer D. and ToÈlg Jakubowski N., Stu G. (1986) Improvement of ion source performance in glow discharge mass spectrometry, Int J Mass Spectrom Ion Processes 71: 183±197. Milton D. M. P., Hutton R. C. and Ronan G. A. (1992) Optimization of discharge parameters for the analysis of high-purity silicon wafers by magnetic sector glow discharge mass
Literature
594
595
596
597
598
599
600
601
602
spectrometry, Fresenius' J Anal Chem 343: 773. Shao Y. and Horlick G. (1991) Design and characterization of glow discharge devices as complementary sources for an ``ICP'' mass spectrometer, Spectrochim Acta, Part B 46: 165±174. Klingler J. A. and Harrison W. W. (1991) Glow discharge mass spectrometry using pulsed ual cathodes, Anal Chem 63: 2982±2984. Marcus R. K., Harville T. R., Mei Y. and Shick Jr. C. R. (1994) R.f. powered glow discharges elemental analysis across the solids spectrum, Anal Chem 66: 902A±911A. Saprykin A. I., Melchers F. G., Becker J. S. and Dietze H. J. (1995) Radio-frequency glow discharge ion source for high resolution mass spectrometry, Fresenius' J Anal Chem 353: 570±574. Marcus R. K. (1994) Radiofrequency powered glow discharges for emission and mass spectrometry: operating characteristics, ®gures of merit and future prospects, J Anal At Spectrom 9: 1029±1037. Myers D. P., Heintz M. J., Mahoney P. P., Li G. and Hieftje G. M. (1994) Characterization of a radio-frequency glow discharge/time-of-¯ight mass spectrometer, Appl Spectrosc 48: 1337± 1346. Hang W., Walden W. O. and Harrison W. W. (1996) Microsecond pulsed glow discharge as an analytical spectroscopic source, Anal Chem 68: 1148±1152. Heintz M. J. and Hieftje G. M. (1995) Eect of driving frequency on the operation of a radiofrequency glow discharge emission source, Spectrochim Acta, Part B 50: 1125± 1141. Deng R. C. and Williams P. (1994) Suppression of cluster ion interferences in glow discharge mass spectrometry by sampling high-energy ions from a reversed hollow cathode ion source, Anal Chem 66: 1890± 1896.
603 You J., Fanning J. C. and Marcus R.
604
605
606
607
608
609
610
611
612
K. (1994) Particle beam aqueous sample introduction for hollow cathode atomic emission spectroscopy, Anal Chem 66: 3916±3924. You J., Dempster M. A. and Marcus R. K. (1997) Studies of analyte particle transport in a particle beamhollow cathode atomic emission spectrometry system, J Anal At Spectrom 12: 807±815. Guzowski J. P. Jr., Broekaert J. A. C. and Hieftje G. M. (2000) Electrothermal vaporization for sample introduction into a gas sampling glow discharge time-of¯ight mass spectrometer, Spectrochim Acta, Part B 55: 1295±1314. Creed J. T., Davidson T. M., Shen W. and Caruso J. A. (1990) Lowpressure helium microwave-induced plasma mass spectrometry for the detection of halogenated gas chromatographic euents, J Anal At Spectrom 5: 109±113. Eijkel J. C. T., Stoeri H. and Manz A. (1999) A molecular emission detector on a chip employing a direct current microplasma, Anal Chem 71: 2600±2606. Miclea M., Kunze K., Musa G., Franzke J. and Niemax K. (2001) The dielectric barrier discharge ± a powerful microchip plasma for diode laser spectrometry, Spectrochim Acta, Part B 56: 37±43. Vieth W. and Huneke J. C. (1991) Relative sensitivity factors in glow discharge mass spectrometry, Spectrochim Acta, Part B 46: 137±154. Charalambous P. M. (1987) The application of a glow discharge mass spectrometer in the steel industry, Steel Res 5/87: 197±203. Èwer D. (1988) Jakubowski N. and Stu Comparison of Ar and Ne as working gases in analytical glow discharge mass spectrometry, Fresenius' Z Anal Chem 335: 680±686. Caroli S., Senofonte O. and Modesti G. (1999) Glow discharge atomic spectrometry, Adv At Spectrom 5: 173±234.
343
344
Literature 613 Jakubowski N. and Stuewer D.
614
615
616
617 618 619
620
621
622
623
(1992) Application of glow discharge mass spectrometry with low mass resolution for in-depth analysis of technical layers, J Anal At Spectrom 7: 951±958. Smithwick R. W., Lynch D. W. and Franklin J. C. (1993) Relative ion yields measured with a highresolution glow discharge mass spectrometer operated with an argon/ hydrogen mixture, J Am Soc Mass Spectrom 4: 178±285. Takahashi T. and Shimamura T. (1994) Determination of zirconium, niobium and molybdenum in steel by glow discharge mass spectrometry, Anal Chem 66: 3274±3280. Jakubowski N., Stuewer D. and Vieth W. (1987) Performance of a glow discharge mass spectrometer for simultaneous multielement analysis of steel, Anal Chem 59: 1825±1830. Technical Information GD012, VG Instruments, Manchester, U.K. Application Note 02.681, VG Instruments, Manchester, U.K. Mykytiuk A. P., Semeniuk P. and Berman S. (1990) Analysis of high purity metal and semiconductor materials by glow discharge mass spectrometry, Spectrochim Acta Rev 13: 1±10. Oksenoid K. G., Liebich V. and Pietsch G. (1996) Improvement of analyte to interferent line intensity ratio in direct current glow discharge mass spectrometry, Fresenius' Z Anal Chem 355: 863±865. Wilhartitz P., Ortner H. M., Krismer R. and Krabichler D. (1990) Classical analysis involving trace-matrix separation versus solid state mass spectrometry: a comparative study for the analysis of highpurity Mo, W and Cr, Mikrochim Acta 2: 259±271. Vassamillet L. F. (1989) Use of glow discharge mass spectrometers for quality-control of high-purity aluminium, J Anal At Spectrom 4: 451±455. Feng X. and Horlick G. (1994) Analysis of aluminium alloys using
624
625
626 627
628
629
630
631
632
inductively coupled plasma and glow discharge mass spectrometry, J Anal At Spectrom 9: 823±831. Venczago C. and Weigert M. (1994) Application of glow discharge mass spectrometry (GDMS) for the multielement trace and ultratrace analysis of sputtering targets, Fresenius' J Anal Chem 350: 303± 309. Saito M., Hirose F. and Okochi H. (1995) Comparative study of the determination of micro-amounts of boron in high-purity molybdenum by a spectrophotometric method using curcumin, spark source mass spectrometry and glow discharge mass spectrometry, Anal Sci 11: 695±697. Technical Information GD701, VG Instruments, Manchester, U.K. Tigwell M., Clark J., Shuttleworth S. and Bottomley M. (1992) Analysis and characterization of rare earths by plasma source mass spectrometry, Mater Chem Phys 31: 23±27. Ecker K. H. and Pritzkow W. (1994) Measurement of isotope ratios in solids by glow discharge mass spectrometry, Fresenius' J Anal Chem 349: 207±208. Donohue D. L. and Petek M. (1991) Isotopic measurements of palladium metal containing protium and deuterium by glow discharge mass spectrometry, Anal Chem 63: 740±744. Van Straeten M., Swenters K., Gijberls R., Verlinden J. and Adriaenssens E. (1994) Analysis of platinum powder by glow discharge mass spectrometry, J Anal At Spectrom 9: 1389±1397. Held A., Taylor P., Ingelbrecht C., De BieÁvre P., Broekaert J. A. C., Van Straeten M. and Gijbels R. (1995) Determination of the scandium content of high-purity titanium using inductively coupled plasma mass spectrometry and glow discharge mass spectrometry as part of its certi®cation as a reference material, J Anal At Spectrom 10: 849±852. Shi Z., Brewer S. and Sachs R. (1995) Application of a magnetron glow discharge to direct solid
Literature
633
634 635
636
637
638
639
640 641
642
sampling for mass spectrometry, Appl Spectrosc 49: 1232±1238. Hutton R. C. and Raith A. (1992) Analysis of pure metals using a quadrupole-based glow discharge mass spectrometer, J Anal At Spectrom 7: 623±627. VG Instruments, Technical Information GD 701, Manchester, U.K. De Gendt S., Schelles W., Van Èller V. (1995) Grieken R. and Mu Quantitative analysis of iron-rich and other oxide-based samples by means of glow discharge mass spectrometry, J Anal At Spectrom 10: 681±687. Schelles W., De Gendt S., Muller V. and Van Grieken R. (1995) Evaluation of secondary cathodes for glow discharge mass spectrometry analysis of dierent nonconducting sample types, Appl Spectrosc 49: 939± 944. Shimamura T., Takahashi T., Honda M. and Nagai H. (1993) Multi-element and isotopic analyses of iron meteorites using a glow discharge mass spectrometer, J Anal At Spectrom 8: 453±460. Teng J., Barshick C. M., Duckworth D. C., Morton S. J., Smith D. H. and King F. L. (1995) Factors in¯uencing the quantitative determination of trace elements in soils by glow discharge mass spectrometry, Appl Spectrosc 49: 1361± 1366. Betti M., Giamarelli S., Hiernaut T., Rasmussen G. and Koch L. (1996) Detection of trace radioisotopes in soil, sediment and vegetation by glow discharge mass spectrometry, Fresenius' J Anal Chem 355: 642±646. Barshick (1996) Rapid Commun Mass Spectrom 10: 341. Takahashi T. and Shimamura T. (1994) Determination of zirconium, niobium and molybdenum in steel by glow discharge mass spectrometry, Anal Chem 66: 3274±3280. Schelles W. and Van Grieken R. E. (1996) Direct current glow discharge mass spectrometric analysis of macor ceramic using a secondary cathode, Anal Chem 68: 3570±3574.
643 Luo F. C. H. and Huneke J. C. (1991)
644
645
646
647
648
649
650
651
652
Proc Intern Conf on elemental analysis of coal and its by-products, Barren River Resort, KY, September 9±11, p. 17. Woo J. C., Jakubowski N. and Stuwer D. (1993) Analysis of aluminium oxide powder by glow discharge mass spectrometry with low mass resolution, J Anal At Spectrom 8: 881±889. De Gendt S., Van Grieken R., Hang W. and Harrison W. W. (1995) Comparison between direct current and radiofrequency glow discharge mass spectrometry for the analysis of oxide-based samples, J Anal At Spectrom 10: 689±695. Saprykin A. I., Becker J. S. and Dietze H. J. (1996) Optimization of an rf-powered magnetron glow discharge for the trace analysis of glasses and ceramics, Fresenius' J Anal Chem 355: 831±835. Èller Schelles W., De Gendt S., Mu V. and Van Grieken R. (1995) Evaluation of secondary cathodes for glow discharge mass spectrometry analysis of dierent nonconducting sample types, Appl Spectrosc 49: 939± 944. Shick C. R. Jr., De Palma P. A. Jr. and Marcus R. K. (1996) Radiofrequency glow discharge mass spectrometry for the characterization of bulk polymers, Anal Chem 68: 2113±2121. Barshick C. M. and Harrison W. W. (1989) The laser as an analytical probe in glow discharge mass spectrometry, Mikrochim Acta 3: 169±177. Mei Y. and Harrison W. W. (1991) Getters and plasma reagents in glow discharge mass spectrometry, Spectrochim Acta, Part B 46: 175± 182. Hecq M., Hecq A. and Fontignies M. (1983) Analysis of thin metallic ®lms by glow discharge mass spectrometry, Anal Chim Acta 155: 191±198. Behn U., Gerbig F. A. and Albrecht H. (1994) Depth pro®ling of frictional brass coated steel samples by glow discharge mass spectrometry, Fresenius' J Anal Chem 349: 209±210.
345
346
Literature 653 Mattson W. A., Bentz B. L. and
654
655
656
657
658
659
660
661
Harrison W. W. (1976) Coaxial cathode ion source for solids mass spectrometry, Anal Chem 48: 489±491. Foss G. O., Svec H. J. and Conzemius R. J. (1983) The determination of trace elements in aqueous media without preconcentration using a cryogenic hollowcathode ion source, Anal Chim Acta 147: 151±162. Jakubowski N., Stuewer D. and Toelg G. (1991) Microchemical determination of platinum and iridium by glow discharge mass spectrometry, Spectrochim Acta, Part B 46: 155±163. Barshick C. M., Duckworth D. C. and Smith D. H. (1993) Analysis of solution residues by glow discharge mass spectrometry, J Am Soc Mass Spectrom 4: 47±53. Evetts I., Milton D. and Mason R. (1991) Trace element analysis in body ¯uids by glow discharge mass spectrometry: a study of lead mobilization by the drug cis-platin, Biol Mass Spectrom 20: 153±159. Barshick C. M., Smith D. H., Wade J. W. and Bayne C. K. (1994) Isotope dilution glow discharge mass spectrometry as applied to the analysis of lead in waste oil samples, J Anal At Spectrom 9: 83±87. Barshick C. M., Smith D. H., Hackney J. H., Cole B. A. and Wade J. W. (1994) Glow discharge mass spectrometric analysis of trace metals in petroleum, Anal Chem 66: 730± 734. Wetzel W., Broekaert J. A. C. and Hieftje G. M., Determination of arsenic by hydride generation coupled to time-of-¯ight mass spectrometry with a gas sampling glow discharge, spectrochimica Acta, Part B, submitted for publication. Olson L. K., Belkin M. and Caruso J. A. (1996) Radiofrequency glow discharge mass spectrometry for gas chromatographic detection: a new departure for elemental speciation studies, J Anal At Spectrom 11: 491± 496.
662 Omenetto N. and Winefordner J.
663
664
665
666
667
668
669
670
D. (1979) Atomic ¯uorescence spectrometry: basic principles and applications, Prog Anal At Spectrom 2: 1±183. Omenetto N., Human H. G. C., Cavalli P. and Rossi G. (1984) Laser excited atomic and ionic nonresonance ¯uorescence detection limits for several elements in an argon inductively coupled plasma, Spectrochim Acta, Part B 39: 115±117. Omenetto N., Smith B. W., Hart L. P., Cavalli P. and Rossi G. (1985) Laser induced double-resonance ¯uorescence in an inductively coupled plasma, Spectrochim Acta, Part B 40: 1411±1422. Bolshov M. A., Zybin A. V. and Smirenkina I. I. (1981) Atomic ¯uorescence spectrometry with laser sources, Spectrochim Acta, Part B 36: 1143±1152. Smith B. W., Farnsworth P. B., Cavalli P. and Omenetto N. (1990) Optimization of laser excited atomic ¯uorescence in a graphite furnace for the determination of thallium, Spectrochim Acta, Part B 45: 1369± 1373. Grazhulene S., Khvostikov V. and Sorokin M. (1991) The possibilities of glow discharge cathode sputtering for laser atomic-¯uorescence analysis of microelectronics materials, Spectrochim Acta, Part B 46: 459±465. Bolshov M. A., Rudniev S. N., Rudnieva A. A., Boutron C. F. and Hong S. (1997) Determination of heavy metals in polar snow and ice by laser excited atomic ¯uorescence spectrometry with electrothermal atomization in a graphite cup, Spectrochim Acta, Part B 52: 1535± 1544. Sdorra W. and Niemax K. (1990) Temporal and spatial distribution of analyte atoms and ions in microplasmas produced by laser ablation of solid samples, Spectrochim Acta, Part B 45: 917±926. Turk G. C., Travis J. C., DeVoe J. R. and O'Haver T. C. (1979) Laser enhanced ionization spectrometry in
Literature
671
672
673
674
675
676
677 678
679
680
681
analytical ¯ames, Anal Chem 51: 1890±1896. Havrilla G. J., Weeks S. J. and Travis J. C. (1982) Continuous wave excitation in laser enhanced ionization spectrometry, Anal Chem 54: 2566±2570. Niemax K. (1985) Spectroscopy using thermionic diode detectors, Appl Phys, Part B 38: 147±157. Smith B. W., Quentmeier A., Bolshov M. and Niemax K. (1999) Measurement of uranium isotope ratios in solid samples using laser ablation and diode laser-excited atomic ¯uorescence spectrometry, Spectrochim Acta, Part B 54: 943±958. Omenetto N., Berthoud Th., Cavalli P. and Rossi G. (1985) Analytical laser enhanced ionization studies of thallium in the air± acetylene ¯ame, Anal Chem 57: 1256± 1261. Babin F. J. and Gagne J. M. (1992) Hollow cathode discharge (HCD) dark space diagnostics with laser photoionisation and galvanic detection, Appl Phys, Part B 54: 35±45. Young J. P., Shaw R. W. and Smith D. H. (1989) Resonance ionization mass spectrometry, Anal Chem 61: 1271A±1279A. Koch K. H. (1999) Process analytical chemistry, Springer, Berlin. Tschoepel P. and Toelg G. (1982) Comments on the accuracy of analytical results in nanogram and picogram trace analysis of the elements, J Trace Microprobe Techn 1: 1±77. Kingston H. M. and Haswell S. J. (1997) Microwave enhanced chemistry ± fundamentals, sample preparation and applications, American Chemical Society, Washington DC. Pichler U., Haase A., Knapp G. and Michaelis M. (1999) Microwave enhanced ¯ow system for hightemperature digestion of resistant organic materials, Anal Chem 71: 4050. Zischka M., Kettisch P., Schalk A.
682
683
684
685
686
687
688
689
and Knapp G. (1998) Closed microwave-assisted wet digestion with simultaneous control of pressure and temperature in all vessels, Fresenius' J Anal Chem 361: 90±95. Knapp G., Raptis S. E., Kaiser G., Toelg G., Schramel P. and Schreiber B. (1981) A partially mechanized system for the combustion of organic samples in a stream of oxygen with quantitative recovery of the trace elements, Fresenius' Z Anal Chem 308: 97±103. Jacob E. (1989) Inorganic multielemental analysis by ¯uorine volatilization systems (FV/FTIR and FV/MS), Fresenius' Z Anal Chem 333: 761±762. Jacob E. (1989) A novel micro elemental analyzer for the determination of non-metals and some metals, Fresenius' Z Anal Chem 334: 641±642. Kaiser G., Mayer A., Friess M., Riedel R., Harris M., Jacob E. and Toelg G. (1995) Critical comparison of ICP-OES, XRF and ¯uorine volatilization±FT-IR spectrometry for the reliable determination of the silicon main constituent in ceramic materials, Fresenius' J Anal Chem 352: 318±326. Kipphardt H., Garten R. P. H., Jacob E., Broekaert J. A. C. and Toelg G. (1997) Mass spectrometric analysis of ceramics after decomposition with elemental ¯uorine, Mikrochim Acta 125: 101±105. Gerbersmann C., Heisterkamp M., Adams F. C. and Broekaert J. A. C. (1997) Two methods for the speciation analysis of mercury in ®sh involving microwave-assisted digestion and gas chromatographyatomic emission spectrometry, Anal Chim Acta 350: 273±285. KlockenkaÈmper R. (1997) Totalre¯ection x-ray ¯uorescence analysis, Wiley, New York. Pappert E., Flock J. and Broekaert J. A. C. (1999) Speciation of chromium in solid materials with the aid of soft-x-ray spectrometry, Spectrochim Acta, Part B 54: 299±310.
347